K Number
K123647
Manufacturer
Date Cleared
2013-02-14

(79 days)

Product Code
Regulation Number
870.5800
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CircuFlow 5208 Sequential Compression Device is a compression device based on sequential pneumatic compression technique which is intended for the treatment of the following conditions:

  • Lymphedema
  • Venous stasis ulcers
  • Venous insufficiency
  • Peripheral edema
    The device is intended for home and hospital use.
Device Description

The CircuFlow 5208 is a digitally controlled sequential pneumatic compression device designed to apply compression to a limb. The CircuFlow 5208 enables different treatment pressures and treatment times that should be used according to physician prescription. When activated, air flows into garments chambers, the pump provides gradient pressurization to the chambers (sequential inflation of distal to proximal, with distal chambers inflated to a greater pressure than the proximal ones). After each chamber is inflated, the pressure is held constant until all chambers are inflated, in order to prevent reverse gradient flow. Once all chambers are inflated, they are then all released simultaneously, and the cycle repeats. Pressure within each chamber can be programmed and individual chambers can be skipped or the pressure decreased in the case of a wound.

AI/ML Overview

The provided document describes a 510(k) summary for the CircuFlow 5208 Sequential Compression Device. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving efficacy through clinical trials with specific acceptance criteria related to a disease outcome. Therefore, the information provided primarily pertains to performance testing designed to ensure the device operates according to its specifications and is safe, rather than a study proving clinical effectiveness against defined acceptance criteria for a medical outcome.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document lists performance tests but does not explicitly state quantitative acceptance criteria or their corresponding reported device performance values. The general statement is that "Bench and laboratory testing was performed and assures that the product meets its specifications."

Acceptance Criteria CategoryReported Device Performance
Electrical Safety- Dielectric Strength Test: Performed. (Implicitly met standards, but no data provided.)- Leakage Current Test: Performed. (Implicitly met standards, but no data provided.)
Pressure Control- Pressure Calibration for Pressure Sensor: Performed. (Implicitly met specifications, but no data provided.)- Pressure Gradient Performance: Performed. (Implicitly met specifications, but no data provided.)
Timing Accuracy- Inflation & Deflation Time Performance: Performed. (Implicitly met specifications, but no data provided.)- Treatment Time Performance: Performed. (Implicitly met specifications, but no data provided.)
Biocompatibility- ISO 10993-5 (in vitro cytotoxicity): Conforms to standard.- ISO 10993-10 (irritation and delayed-type hypersensitivity): Conforms to standard.
Risk Management- ISO 14971 (risk management): Conforms to standard.
General Safety & Performance- IEC 60601-1 / AAMI ES 60601-1 (general requirements for basic safety and essential performance): Conforms to standards.
Electromagnetic Compatibility- IEC 60601-1-2 (collateral standard for EMC): Conforms to standard.

2. Sample Size Used for the Test Set and the Data Provenance:

The document describes bench and laboratory testing, not a clinical study on human subjects. Therefore, the concept of a "test set" in the context of clinical data provenance (country of origin, retrospective/prospective) and sample size for human subjects does not apply. The tests mentioned are likely conducted on a limited number of device units in a controlled lab environment.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This question is not applicable as the testing described is technical product performance testing, not a study requiring expert clinical ground truth assessment on a test set.

4. Adjudication Method for the Test Set:

Not applicable, as this was not a clinical study requiring human adjudications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is a medical device, not an AI diagnostic or assistive tool, and no MRMC study was conducted or referenced.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. This is a physical medical device, not an algorithm. The device operates independently according to its programmed functions.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

For the performance testing, the "ground truth" would be established engineering and safety standards (e.g., IEC 60601-1, ISO 10993) and the device's own internal design specifications. The tests verify that the device's performance aligns with these predefined technical requirements.

8. The sample size for the training set:

Not applicable. This is a physical medical device. The concept of a "training set" is relevant for machine learning algorithms, which are not described here.

9. How the ground truth for the training set was established:

Not applicable, as there is no training set mentioned for this device.

{0}------------------------------------------------

FEB 1 4 2013

510(K) SUMMARY

Submitter:

Devon Medical Products, Inc.

Contact Person:

Ruth Wu, COO 1100 First Avenue, Suite 202 King of Prussia, PA 19406 Phone: 610.757.4103 610.930.4035 Fax:

Common Classification & Proprietary Names:

Sequential Compression Device Common Names: Proprietary Name: CircuFlow 5208

Date Prepared:

July 23'd, 2012

Classification

The classification name, 21 CFR Part and Paragraph number, product code and classification of the CircuFlow 5208.

Classification Name21SectionCFR ProductCodeClass
Compressible LimbSleeve870.5800JOWII

Predicate Devices:

The CircuFlow 5208 Sequential Compression Device is substantially equivalent to the following.

Predicate DeviceManufacturer510(k)#
CircuFlow 5200Devon Medical, Inc.K101523
Lymphapress Model 201MAXMego Afek LtdK100677
Proact 1Proact CompanyK003909
CircuFlow 5100Devon Medical, Inc.K100446

Device Description

The CircuFlow 5208 is a digitally controlled sequential pneumatic compression device designed to apply compression to a limb. The CircuFlow 5208 enables different treatment pressures and treatment times that should be used according to physician prescription. When activated, air flows into garments chambers, the pump provides gradient pressurization to the chambers (sequential inflation of distal to proximal, with distal chambers inflated to a greater pressure than the proximal ones).

Page 1 of 3

{1}------------------------------------------------

SECTION 5

After each chamber is inflated, the pressure is held constant until all chambers are inflated, in order to prevent reverse gradient flow. Once all chambers are inflated, they are then all released simultaneously, and the cycle repeats. Pressure within each chamber can be programmed and individual chambers can be skipped or the pressure decreased in the case of a wound.

Intended Use:

The CircuFlow 5208 Sequential Compression Device is a compression device based on sequential pneumatic compression technique which is intended for the treatment of the following conditions:

  • . Lymphedema
  • Venous stasis ulcers -
  • Venous insufficiency -
  • Peripheral edema -

The device is intended for home and hospital use.

Technological Characteristics:

The manufacturer believes that the technological characteristics of the CircuFlow 5208 are substantially similar to those of the predicate devices.

The CircuFlow 5208 has very similar components to its predicate devices and has very similar principles of operation. The device consists of an electrically generated source of compressed air, tubing to convey the pressurized air to the sleeve, like the predicates, pressure is applied cyclically for a specified period of time, according to the physician's prescription.

Performance Testing

Bench and laboratory testing was performed and assures that the product meets its specifications. The manufacturer believes that the technological characteristics of the CircuFlow 5208 are substantially similar to those of the predicate devices. The performance testing includes the following tests:

List of Performance Tests
Test 1Dielectric Strength Test
Test 2Leakage Current Test
Test 3Pressure Calibration for Pressure Sensor
Test 4Pressure Gradient Performance
Test 5Inflation & Deflation Time Performance
Test 6Treatment Time Performance

Page 2 of 3

{2}------------------------------------------------

SECTION 5

Standards

The CircuFlow 5208 conforms to the following standards:

IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard:

AAMI ES 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

ISO 10993-5 Biological evaluation of medical devices - Part 1: Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity

ISO 10993-10 Biological evaluation of medical devices --Part 10:Tests for irritation and delayed-type hypersensitivity

ISO 14971 Medical devices - Application of risk management to medical devices

Statement of Substantial Eguivalence

The CircuFlow 5208 is substantially equivalent in technology, function, operating parameters, and intended use to predicate devices that are currently commercially available and in distribution, and does not raise any new questions of safety or effectiveness.

Conclusions

In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, Devon Medical Products, believes that the CircuFlow 5208, is substantially equivalent to the predicate devices as described herein.

Page 3 of 3

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure with three flowing lines representing the department's mission to promote health and well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

FEB 1 4 2013

Devon Medical Inc. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313

Re: K123647

Trade/Device Name: CircuFlow 5208 Sequential Compression Device Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II (two) Product Code: JOW Dated: January 31, 2013 Received: February 1, 2013

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 -- Mr. Mark Job

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Matthew G. Hillebrenner

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

SECTION 4

INDICATIONS FOR USE STATEMENT

K1236

K12364 510(k) Number (if known):

Device Name: CircuFlow 5208 Sequential Compression Device

Indications for Use:

The CircuFlow 5208 Sequential Compression Device is a compression device based on sequential pneumatic compression technique which is intended for the treatment of the following conditions:

  • Lymphedema 1
  • Venous stasis ulcers -
  • Venous insufficiency -
    • Peripheral edema

The device is intended for home and hospital use.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ilə qalında da bir sənədlər Şaman Şamar Şaman Şaman Şaman Şamanın Şaman Şamanın Şamanın Şamanın Şamanın Şamanın Şamanın Şamanın Şamanın Şamanın Şamanın Şa

on of of Cardiovassualized to noisivil Mislon Sign-Un-Un-Un-Un-Un-

UA-Kelle
4.1

(Division Sign On)
Division of Cardiovascular Devices

510/12 Number14ﻟ
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).