K Number
K123593
Date Cleared
2013-04-17

(147 days)

Product Code
Regulation Number
866.5100
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QUANTA Flash Sm is a chemiluminescent immunoassay for the semi-quantitative determination of lgG anti-Sm antibodies in human serum. The presence of antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus Erythematosus (SLE).

The QUANTA Flash RNP is a chemiluminescent immunoassay for the semi-quantitative determination of lgG anti-ribonucleoprotein (RNP) antibodies in human serum. The presence of anti-RNP antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus Erythematosus (SLE) and Mixed Connective Tissue Disease (MCTD).

QUANTA Flash Sm Calibrators are intended for use with the QUANTA Flash Sm chemiluminescent immunoassay for the determination of IgG anti-Sm antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash RNP Calibrators are intended for use with the QUANTA Flash RNP chemiluminescent immunoassay for the determination of IgG anti-RNP antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash Sm Controls are intended for use with the QUANTA Flash Sm chemiluminescent immunoassay for quality control in the determination of IgG anti-Sm antibodies in human serum.

QUANTA Flash RNP Controls are intended for use with the QUANTA Flash RNP chemiluminescent immunoassay for quality control in the determination of IgG anti-RNP antibodies in human serum.

Device Description

The QUANTA Flash Sm and RNP assays are designed to run on the BIO-FLASH instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash Sm and RNP assays utilize a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

Native Sm or RNP antigen that is purified from calf thymus is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspended using resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. A patient serum sample is prediluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(III)coproporphyrin in sodium hydroxide solution) and Trigger 2 (urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human lgG, which is in turn proportional to the amount of anti-RNP antibodies bound to the Sm or RNP on the beads.

For quantitation, the QUANTA Flash Sm and RNP assays utilize a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of reagent cartridge must be calibrated before first use with the QUANTA Flash Sm and RNP Calibrators. Based on the results obtained with the two Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.

The QUANTA Flash Sm kit contains the following materials:
One (1) QUANTA Flash Sm Reagent Cartridge
One (1) vial of Resuspension buffer
One (1) Transfer pipette

The QUANTA Flash Sm reagent cartridge contains the following reagents for 50 determinations:

  • Sm antigen coated paramagnetic beads, lyophilized.
  • Assay buffer colored pink, containing Tris-buffered saline, Tween 20, protein stabilizers and preservatives.
  • Tracer IgG Isoluminol labeled anti-human IgG antibodies in buffer, containing protein stabilizers and preservative.

The QUANTA Flash RNP kit contains the following materials:
One (1) QUANTA Flash RNP Reagent Cartridge

  • One (1) vial of Resuspension buffer
  • One (1) Transfer pipette

The QUANTA Flash RNP reagent cartridge contains the following reagents for 50 determinations:

  • RNP antigen coated paramagnetic beads, lyophilized.
  • Assay buffer colored pink, containing Tris-buffered saline, Tween 20, protein stabilizers and preservatives.
  • Tracer IgG Isoluminol labeled anti-human IgG antibodies in buffer, containing protein stabilizers and preservative.

The QUANTA Flash Sm Calibrators kit and the QUANTA Flash™ RNP Calibrators kit each contain 2 vials of Calibrators:

QUANTA Flash Sm Calibrators:

  • QUANTA Flash Sm Calibrator 1: Two (2) barcode labeled tubes containing 0.3 ml prediluted, ready to use reagent. Calibrators contain human antibodies to Sm in buffer, protein stabilizers, and preservatives.
  • QUANTA Flash Sm Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to Sm in buffer, protein stabilizers, and preservatives.

QUANTA Flash RNP Calibrators:

  • QUANTA Flash RNP Calibrator 1: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to RNP in buffer, protein stabilizers, and preservatives.
  • QUANTA Flash RNP Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to RNP in buffer, protein stabilizers, and preservatives.

The QUANTA Flash Sm Controls kit and the QUANTA Flash™ RNP Controls kit each contain 2 vials of Negative Control and two vials of Positive Control:

QUANTA Flash Sm Controls:

  • QUANTA Flash™ Sm Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to Sm in buffer, protein stabilizers, and preservatives.
  • QUANTA Flash™ Sm Positive Control: Two (2) barcode labeled tubes containing 0.5 ml, ready to use reagent. Controls contain human antibodies to Sm in buffer, protein stabilizers, and preservatives.

QUANTA Flash RNP Controls:

  • QUANTA Flash™ RNP Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to RNP in buffer, protein stabilizers, and preservatives.
  • QUANTA Flash™ RNP Positive Control: Two (2) barcode labeled tubes containing 0.5 ml, ready to use reagent. Controls contain human antibodies to RNP in buffer, protein stabilizers, and preservatives.
AI/ML Overview

Acceptance Criteria and Device Performance for QUANTA Flash® Sm and RNP Assays

This information is extracted from the provided 510(k) summary for the QUANTA Flash® Sm and QUANTA Flash® RNP assays.

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance CriteriaQUANTA Flash® Sm PerformanceQUANTA Flash® RNP Performance
PrecisionAll %CV values within 15%All %CV values within 15% (e.g., Total Precision: 8.7%-11.5%)All %CV values within 15% (e.g., Total Precision: 4.8%-10.8%)
ReproducibilityAll %CV values within 15%All %CV values within 15% (e.g., Total CV: 7.0%-8.3%)All %CV values within 15% (e.g., Total CV: 2.8%-2.9%)
Limit of Detection (LoD)Proportions of false positives (alpha) < 5% and false negatives (beta) < 5%803 RLU (< analytical measuring range)1128.8 RLU (< analytical measuring range)
Limit of Blank (LoB)Not explicitly stated, but determined in accordance with CLSI EP17-A540 RLU713 RLU
Analytical Measuring Range (AMR) LinearityLinear regression R² close to 1, slope close to 1, y-intercept close to 0Slope: 0.99 (0.97 to 1.00); Y-intercept: 0.81 (-3.2 to 4.8); R²: 0.99Slope: 1.00 (1.00 to 1.02); Y-intercept: -5.17 (-8.27 to -2.07); R²: 1.00
Auto-rerun Function (% Recovery)Within ± 20% acceptance limit87.6% - 97.7% (average 91.5%)82.5% - 102.2% (average 96.3%)
Interference (% Recovery/Difference)85% - 115% recovery, or ± 4 CU difference, whichever is greaterMet for bilirubin, hemoglobin, triglycerides, cholesterol, RF IgMMet for bilirubin, hemoglobin, triglycerides, cholesterol, RF IgM
High Concentration Hook EffectNo hook effect within tested range (up to highest concentration tested)No hook effect up to 2429 CUNo hook effect up to 3140 CU
Shelf Life (Controls & Calibrators)Regression analysis: lower 95% CI > 90% at 2 weeks, no individual data point ≤ 80% recovery; OR Individual data points: recovery values > 90% at day 14Fulfilled (1-year expiration dating assigned)Fulfilled (1-year expiration dating assigned)
Shelf Life (Microparticles)Regression analysis: lower 95% CI ≥ 85% at 2 weeks, no individual data point ≤ 75% recoveryFulfilled (1-year expiration dating assigned)Fulfilled (1-year expiration dating assigned)
In-use Stability (Calibrators)5 successful calibrations over 8.5 hours, RLU recovery 90%-110% compared to first use, patient samples within expected rangeMet (4 calibrations over 8 hours supported)Met (4 calibrations over 8 hours supported)
In-use Stability (Controls)All replicates within established range, linear regression line of %recovery between 85%-115% at run 15Met (%CV 5.9% for negative, 4.8% for positive)Met (%CV 5.9% for negative, 6.9% for positive)
In-use Stability (Reagent Cartridge)Lower 95% CI of regression line reaches 85% recovery, or 2% or more recovery data ≤ 75%33 days28 days
Cut-off EstablishmentEstablished at 99th percentile of reference subjects using non-parametric percentile method (CLSI C28-A3c)20 CU20 CU
Clinical Sensitivity (Sm for SLE)Not explicitly defined as an acceptance criterion within the context of a specific pre-market requirement but presented as performance data for diagnostic utility.14.4% (9.1-21.1%)N/A
Clinical Specificity (Sm for SLE)Not explicitly defined as an acceptance criterion within the context of a specific pre-market requirement but presented as performance data for diagnostic utility.97.9% (95.1-99.3%)N/A
Clinical Sensitivity (RNP for SLE/MCTD)Not explicitly defined as an acceptance criterion within the context of a specific pre-market requirement but presented as performance data for diagnostic utility.N/A37.1% (30.0%-44.6%)
Clinical Specificity (RNP for SLE/MCTD)Not explicitly defined as an acceptance criterion within the context of a specific pre-market requirement but presented as performance data for diagnostic utility.N/A95.1% (91.6%-97.5%)
Method Comparison (Sm vs. Predicate)Not explicitly defined as a numerical acceptance criterion, but data is presented to show substantial equivalence.Positive Agreement: 92.1% (78.6 - 98.3%); Negative Agreement: 92.6% (84.6 - 97.2%); Total Agreement: 92.4% (86.1 - 96.5%)N/A
Method Comparison (RNP vs. Predicate)Not explicitly defined as a numerical acceptance criterion, but data is presented to show substantial equivalence.N/APositive Agreement: 80.7% (68.1 - 90.0%); Negative Agreement: 94.5% (88.5 - 98.0%); Total Agreement: 89.8% (84.2 - 94.0%)

Note on "Acceptance Criteria" for Clinical Sensitivity/Specificity: For diagnostic device submissions like this, the criteria are often for demonstrating reasonable performance that supports the intended use and shows substantial equivalence to a predicate, rather than predefined absolute thresholds that must be met. The device sponsor presents the performance data to the FDA for review against the predicate and the defined intended use.

2. Sample Sizes Used for the Test Set and Data Provenance

For Clinical Sensitivity and Specificity:

  • QUANTA Flash Sm:

    • Test Set Size: 379 samples
      • 63 SLE patients (Neuss Center for Rheumatology, Neuss, Germany)
      • 83 SLE patients (Dr. Carlos von Mühlen, Germany)
      • 74 systemic sclerosis patients
      • 70 rheumatoid arthritis patients
      • 5 Sjögren's syndrome patients
      • 53 patients with other systemic rheumatic diseases
      • 2 patients with autoimmune liver disease
      • 19 patients with viral hepatitis
      • 10 patients with other infectious diseases (5 HIV, 5 syphilis)
    • Data Provenance: Mixed (Germany, likely other origins for disease controls). Retrospective, as these were existing samples used for validation.
  • QUANTA Flash RNP:

    • Test Set Size: 424 samples
      • 62 SLE patients (Neuss Center for Rheumatology, Neuss, Germany)
      • 32 MCTD patients (Neuss Center for Rheumatology, Neuss, Germany)
      • 84 SLE patients (Dr. Carlos von Mühlen, Germany)
      • 48 patients with other systemic rheumatic diseases
      • 2 patients with autoimmune liver disease
      • 6 patients with Sjögren's syndrome
      • 76 patients with systemic sclerosis
      • 70 patients with rheumatoid arthritis
      • 14 patients with polymyositis/dermatomyositis
      • 20 patients with viral hepatitis
      • 10 patients with other infectious diseases (5 HIV, 5 syphilis)
    • Data Provenance: Mixed (Germany, likely other origins for disease controls). Retrospective.

For Reference Range Establishment:

  • QUANTA Flash Sm:

    • Test Set Size: 232 subjects (129 apparently healthy blood donors, 41 non-autoimmune thyroid disease, 42 autoimmune thyroid disease, 20 infectious diseases)
    • Data Provenance: Not explicitly stated, but appears to be general reference populations. Retrospective.
  • QUANTA Flash RNP:

    • Test Set Size: 255 subjects (127 apparently healthy blood donors, 41 non-autoimmune thyroid disease, 42 autoimmune thyroid disease, 20 infectious diseases, 25 rheumatoid arthritis)
    • Data Provenance: Not explicitly stated, but appears to be general reference populations. Retrospective.

For Method Comparison:

  • QUANTA Flash Sm:

    • Test Set Size: 119 samples (from clinical validation studies, proficiency surveys (CAP and NEQAS), and apparently healthy blood donors)
    • Data Provenance: Mixed (includes proficiency testing samples, likely retrospective).
  • QUANTA Flash RNP:

    • Test Set Size: 167 samples (from clinical validation studies and apparently healthy blood donors)
    • Data Provenance: Mixed (likely retrospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly state the number or qualifications of "experts" used to establish the ground truth for the clinical test sets of patients with SLE, MCTD, and other conditions. For these types of diagnostic immunoassays, the "ground truth" for patient classification (e.g., "SLE patient," "MCTD patient") is typically based on a clinical diagnosis by treating physicians, often rheumatologists, according to established diagnostic criteria (e.g., ACR criteria for SLE). The specific details of how these clinical diagnoses were confirmed for each sample are not provided, nor is the number or specific qualifications of the clinicians who made the diagnoses. These are generally assumed to be standard clinical practice in the institutions mentioned (e.g., Neuss Center for Rheumatology).

4. Adjudication Method for the Test Set

No explicit adjudication method (e.g., 2+1, 3+1) is described for the clinical diagnoses that constitute the ground truth. The samples were collected from diagnosed patient populations.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. This device is an in-vitro diagnostic (IVD) assay, not an imaging or algorithmic-interpretation device that would involve human readers interpreting results with or without AI assistance. The output is a semi-quantitative antibody level.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, this is implicitly a standalone study. The device is an automated chemiluminescent immunoassay system (BIO-FLASH instrument with QUANTA Flash assays). The results (Chemiluminescent Units, CU) are generated by the algorithm/system and reported directly. While a human laboratory technician operates the instrument and interprets the final CU values against a cutoff, the performance metrics (sensitivity, specificity, precision, etc.) presented are characteristic of the assay system's inherent ability to detect and quantify the antibodies from the sample, without human-in-the-loop "improvement" on the detection itself.

7. Type of Ground Truth Used

The ground truth for the clinical studies (sensitivity and specificity) was clinical diagnosis (e.g., Systemic Lupus Erythematosus, Mixed Connective Tissue Disease) provided by the centers from which the patient samples were sourced. For analytical studies (precision, linearity, LoD, etc.), the ground truth was based on established analytical methods and reference materials.

8. Sample Size for the Training Set

The document does not mention a "training set" in the context of an AI/machine learning algorithm. For these types of immunoassay devices, the equivalent concept might be the samples used for assay development, initial optimization, and master curve establishment.

  • Master Curve: A predefined lot-specific Master Curve is uploaded onto the instrument. This curve is central to the assay's function. The data used to establish this master curve is not explicitly detailed in terms of sample size or how the ground truth was established, but it would typically involve a large set of characterized samples with known concentrations to define the relationship between RLU and CU.
  • Calibrator and Control Value Assignment: "at least two instruments, on at least two lots of reagent cartridge, in replicates of 10 to determine final value assignment." This is akin to a small-scale internal dataset used to finalize product specifications. There are currently no recognized international standards for the measurement of anti-RNP antibodies, so values are traceable to in-house standards.

9. How the Ground Truth for the Training Set was Established

As noted above, there is no explicit "training set" in the AI/ML sense. The functionality of the device relies on:

  • Reagent Development & Formulation: The components (antigens, antibodies, buffers, etc.) are developed and purified based on chemical and biological principles to ensure specific binding.
  • Master Curve Generation: While not explicitly detailed, the Master Curve that links RLU to CU for each lot of reagents is established using a set of reference samples with known concentrations of anti-Sm or anti-RNP antibodies. The "ground truth" for these reference samples would be determined through rigorous characterization using established analytical methods, potentially including comparative analysis with existing gold standard methods or highly characterized internal reference materials. The document states "Calibrator and Control values are directly traceable to in-house Standards that are used to create the Master Curve." This implies that internal, well-characterized standards form the basis of the "ground truth" for the quantitative aspect of the assay.

{0}------------------------------------------------

K123593

APR 1 7 2013

510(k) Summary

This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter:INOVA Diagnostics, Inc9900 Old Grove Road,San Diego, CA, 92131
Purpose of submission:New device(s)
Devices in the submission:QUANTA Flash® SmQUANTA Flash® Sm CalibratorsQUANTA Flash® Sm ControlsQUANTA Flash® RNPQUANTA Flash® RNP CalibratorsQUANTA Flash® RNP Controls
Scientific contact:Gabriella Lakos, Director of Research, RheumatologyINOVA Diagnostics, Inc9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900/393Fax: 858-863-0025email: glakos@inovadx.com
Quality Systems contact:Tana Keivens, Director, Quality SystemsINOVA Diagnostics, Inc9900 Old Grove Road, San Diego, CA, 92131Phone: 858-586-9900Fax: 858-863-0025/351email: tkeivens@inovadx.com
Preparation date:11/19/2012
Device (1) name (assay kit):Proprietary name: QUANTA Flash® SmCommon name: Anti-Sm Chemiluminescent ImmunoassayClassification name: anti-Sm antibody, antigen and control
Regulation DescriptionAntinuclear antibody immunological test system
Regulation Medical SpecialtyImmunology
Review PanelImmunology
Product CodeLKP
Regulation Number866.5100
Device Class2
Common name:Sm Calibrators
Classification name:Calibrator, secondary
Regulation DescriptionCalibrator
Regulation Medical SpecialtyClinical Chemistry
Product CodeJIT
Regulation Number862.1150
Device Class2
Device (1) name (Controls):Proprietary name:QUANTA Flash® Sm Controls
Common name:Sm Controls
Classification name:Single (specified) analyte controls (assayed and unassayed)
Regulation DescriptionQuality control material (assayed and unassayed)
Regulation Medical SpecialtyClinical Chemistry
Product CodeJJX
Regulation Number862.1660
Device Class1
Predicate device (for Device 1):QUANTA Lite™ Sm ELISA, 510(k) number: K922831
Device (2) name (assay kit):Proprietary name:QUANTA Flash® RNP
Common name:Anti-RNP Chemiluminescent Immunoassay
Classification name:anti-RNP antibody, antigen and control
Regulation DescriptionAntinuclear antibody immunological test system
Regulation Medical SpecialtyImmunology
Review PanelImmunology
Product CodeLKO
Regulation Number866.5100
Device Class2
Device (2) name (Calibrators):Proprietary name:QUANTA Flash® RNP Calibrators
Common name:RNP Calibrators
Classification name:Calibrator, secondary
Regulation DescriptionCalibrator
Product CodeJIT
Regulation Number862.1150
Device Class2
Device (2) name (Controls):Proprietary name:Common name:Classification name:QUANTA Flash® RNP ControlsRNP Controlssingle (specified) analyte controls (assayed andunassayed)
Regulation DescriptionQuality control material (assayed and unassayed)
Regulation Medical SpecialtyClinical Chemistry
Product CodeJJX
Regulation Number862.1660
Device Class1

Device (1) name (Calibrators): Proprietary name:

QUANTA Flash® Sm Calibrators

{1}------------------------------------------------

and the comments of the comments of the comments of the comments of

.

:

{2}------------------------------------------------

Device (1) and (2) description:

The QUANTA Flash Sm and RNP assays are designed to run on the BIO-FLASH instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash Sm and RNP assays utilize a reagent cartridge format, which is compatible with the BIO-FLASH instrument.

Native Sm or RNP antigen that is purified from calf thymus is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspended using resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH instrument. Samples are also loaded onto the instrument in sample racks. A patient serum sample is prediluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at 37°C. The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at 37°C. The beads are magnetized and washed repeatedly. The isoluminol conjugate is oxidized when Trigger 1 (Fe(III)coproporphyrin in sodium hydroxide solution) and Trigger 2 (urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human lgG, which is in turn proportional to the amount of anti-RNP antibodies bound to the Sm or RNP on the beads.

For quantitation, the QUANTA Flash Sm and RNP assays utilize a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot number of

{3}------------------------------------------------

reagent cartridge must be calibrated before first use with the QUANTA Flash Sm and RNP Calibrators. Based on the results obtained with the two Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.

The QUANTA Flash Sm kit contains the following materials:

One (1) QUANTA Flash Sm Reagent Cartridge

One (1) vial of Resuspension buffer

One (1) Transfer pipette

The QUANTA Flash Sm reagent cartridge contains the following reagents for 50 determinations:

  • Sm antigen coated paramagnetic beads, lyophilized. a.
  • Assay buffer colored pink, containing Tris-buffered saline, Tween 20, protein b. stabilizers and preservatives.
  • Tracer IgG Isoluminol labeled anti-human IgG antibodies in buffer, containing protein C. stabilizers and preservative.

The QUANTA Flash RNP kit contains the following materials:

One (1) QUANTA Flash RNP Reagent Cartridge

  • One (1) vial of Resuspension buffer
  • One (1) Transfer pipette

The QUANTA Flash RNP reagent cartridge contains the following reagents for 50 determinations:

  • RNP antigen coated paramagnetic beads, lyophilized. a.
  • Assay buffer colored pink, containing Tris-buffered saline, Tween 20, protein b. stabilizers and preservatives.
  • Tracer IgG Isoluminol labeled anti-human IgG antibodies in buffer, containing protein C. stabilizers and preservative.

The QUANTA Flash Sm Calibrators kit and the QUANTA Flash™ RNP Calibrators kit each contain 2 vials of Calibrators:

QUANTA Flash Sm Calibrators:

  • QUANTA Flash Sm Calibrator 1: Two (2) barcode labeled tubes containing 0.3 ml prediluted, ready to use reagent. Calibrators contain human antibodies to Sm in buffer, protein stabilizers, and preservatives.
  • QUANTA Flash Sm Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to Sm in buffer, protein stabilizers, and preservatives.

QUANTA Flash RNP Calibrators:

  • QUANTA Flash RNP Calibrator 1: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to RNP in buffer,

{4}------------------------------------------------

protein stabilizers, and preservatives.

  • QUANTA Flash RNP Calibrator 2: Two (2) barcode labeled tubes containing 0.3 mL prediluted, ready to use reagent. Calibrators contain human antibodies to RNP in buffer, protein stabilizers, and preservatives.
    The QUANTA Flash Sm Controls kit and the QUANTA Flash™ RNP Controls kit each contain 2 vials of Negative Control and two vials of Positive Control:

QUANTA Flash Sm Controls:

  • QUANTA Flash™ Sm Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to Sm in buffer, protein stabilizers, and preservatives.
  • QUANTA Flash™ Sm Positive Control: Two (2) barcode labeled tubes containing 0.5 ml, ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ ready to use reagent. Controls contain human antibodies to Sm in buffer, protein stabilizers, and preservatives.

QUANTA Flash RNP Controls:

  • QUANTA Flash™ RNP Negative Control: Two (2) barcode labeled tubes containing 0.5 mL, ready to use reagent. Controls contain human antibodies to RNP in buffer, protein stabilizers, and preservatives.
  • QUANTA Flash™ RNP Positive Control: Two (2) barcode labeled tubes containing 0.5 ml, ready to use reagent. Controls contain human antibodies to RNP in buffer, protein stabilizers, and preservatives.

Intended use(s):

The QUANTA Flash Sm is a chemiluminescent immunoassay for the semi-quantitative determination of lgG anti-Sm antibodies in human serum. The presence of antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus Erythematosus (SLE).

The QUANTA Flash RNP is a chemiluminescent immunoassay for the semi-quantitative determination of lgG anti-ribonucleoprotein (RNP) antibodies in human serum. The presence of anti-RNP antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus Erythematosus (SLE) and Mixed Connective Tissue Disease (MCTD).

QUANTA Flash Sm Calibrators are intended for use with the QUANTA Flash Sm chemiluminescent immunoassay for the determination of IgG anti-Sm antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash RNP Calibrators are intended for use with the QUANTA Flash RNP chemiluminescent immunoassay for the determination of IgG anti-RNP antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

QUANTA Flash Sm Controls are intended for use with the QUANTA Flash Sm chemiluminescent immunoassay for quality control in the determination of IgG anti-Sm antibodies in human serum.

QUANTA Flash RNP Controls are intended for use with the QUANTA Flash RNP chemiluminescent immunoassay for quality control in the determination of IgG anti-RNP antibodies in human serum.

{5}------------------------------------------------

Substantial equivalence:

The QUANTA Flash Sm, the QUANTA Flash Sm Calibrators and the QUANTA Flash Sm Controls have the same intended use and assay principle as the predicate device.

The QUANTA Flash RNP, the QUANTA Flash RNP Calibrators and the QUANTA Flash RNP Controls have the same intended use and assay principle as the predicate device.

Comparison to predicate device:

Similarities
ItemQUANTA Flash SmPredicate Device
Intended useSemi-quantitative determination ofanti-Sm antibodies in human serumSemi-quantitative detection of anti-Smantibodies in human serum
Assay methodologySolid phase (heterogenous)immunoassaySolid phase (heterogeneous)immunoassay
TraceabilityInternational Reference Preparation isnot availableResults are traceable to in-houseStandardsInternational Reference Preparation isnot available
AntigenNative Sm antigen, purified from calfthymusNative Sm antigen, purified from calfthymus
Sample typeSerumSerum
Shelf lifeOne yearOne year

QUANTA Flash Sm reagent kit

Differences
ItemQUANTA Flash SmPredicate Device
Detection/Operating principleChemiluminescent immunoassayEnzyme-linked immunosorbent assay
Solid phaseParamagnetic microparticles (beads)96-well plate
ConjugateIsoluminol conjugated anti-human IgGHRP conjugated anti-human IgG
CalibrationLot specific Master Curve + twoCalibrators (Sold separately)Sm ELISA Low Positive(Included in the kit)

QUANTA Flash RNP reagent kit

Similarities
ItemQUANTA Flash RNPPredicate Device
Intended useSemi-quantitative determination ofanti-RNP antibodies in human serumSemi-quantitative detection of anti-RNPantibodies in human serum
Assay methodologySolid phase (heterogenous)immunoassaySolid phase (heterogenous)immunoassay
TraceabilityInternational Reference Preparationis not availableResults are traceable to in-houseInternational Reference Preparation isnot available

{6}------------------------------------------------

Standards
AntigenNative RNP antigen, purified fromcalf thymusNative RNP antigen, purified from calfthymus
Sample typeSerumSerum
Shelf lifeOne yearOne year
Differences
ItemQUANTA Flash RNPPredicate Device
Detection/Operating principleChemiluminescent immunoassayEnzyme-linked immunosorbent assay
Solid phaseParamagnetic microparticles (beads)96-well plate
ConjugateIsoluminol conjugated anti-human IgGHRP conjugated anti-human IgG
CalibrationLot specific Master Curve + twocalibrators (Sold separately)RNP ELISA Low Positive(Included in the kit)
ItemQUANTA Flash Sm CalibratorsPredicate Device
Intended useFor use with the QUANTA Smchemiluminescentimmunoassay(CIA). Each calibrator establishes apoint of reference for the workingcurve that is used to determineChemiluminescent Unit (CU) valuesin the measurement of anti-Smantibodies in human serum.No separate intended use;calibrators are part of the kit.
AnalyteAnti-Sm antibodiesAnti-Sm antibodies
MethodQUANTA Flash Smchemiluminescent immunoassayQUANTA Lite Sm ELISA
UnitCU (Chemiluminescent units)(arbitrary)Units (arbitrary)
MatrixHuman serum, buffer, stabilizers,preservativeHuman serum, buffer, stabilizers,preservative
Physico-chemicalcharacteristicsLiquid, ready to useLiquid, ready to use
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year
In-use stabilityFour calibrations, maximum total 8 hours uncapped onboard the instrument.Calibrators can be used until the endof the shelf life when storedproperly

QUANTA Flash Sm Calibrators

QUANTA Flash RNP Calibrators

ItemQUANTA Flash RNP CalibratorsPredicate Device
------------------------------------------------------

{7}------------------------------------------------

Intended useFor use with the QUANTA Flash RNPchemiluminescentimmunoassay(CIA). Each calibrator establishes apoint of reference for the workingcurve that is used to determineChemiluminescent Unit (CU) valuesin the measurement of anti-RNPantibodies in serum.No separate intended use;calibrators are part of the kit.
AnalyteAnti-RNP antibodiesAnti-RNP antibodies
MethodQUANTA Flash RNPchemiluminescent immunoassayQUANTA Lite RNP ELISA
MatrixHuman serum, buffer, stabilizers,preservativeHuman serum, buffer, stabilizers,preservative
UnitCU (Chemiluminescent units)(arbitrary)Units (arbitrary)
Physico-chemicalcharacteristicsLiquid, ready to useLiquid, ready to use
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year
In-use stabilityFour calibrations, maximum total 8hours uncapped onboard theinstrument.Calibrators can be used until the endof the shelf life when storedproperly

QUANTA Flash Sm Controls

ItemQUANTA Flash Sm ControlsPredicate Device
Intended useQuality control purposes of theQUANTA Flash Smchemiluminescent immunoassay(CIA) kit.No separate intended use; controlsare part of the kit.
AnalyteAnti-Sm antibodiesAnti-Sm antibodies
MethodQUANTA Flash Smchemiluminescent immunoassayQUANTA Lite Sm ELISA
UnitCU (Chemiluminescent units)(arbitrary)Units (arbitray)
MatrixHuman serum, buffer, stabilizers,preservativeHuman serum, buffer, stabilizers,preservative
Physico-chemicalcharacteristicsLiquid, ready to useLiquid, ready to use
Levels2 (negative and positive)2 (ELISA negative, high positive)
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year
In-use stability15 uses, with a maximum time of 10minutes onboard the instrument peruse, or 2 ½ hours, total.Controls can be used until the endof the shelf life when storedproperly

:

.

{8}------------------------------------------------

ItemQUANTA Flash RNP ControlsPredicate Device
Intended useQuality control purposes of theQUANTA Flash RNPchemiluminescent immunoassay(CIA) kit.No separate intended use; controlsare part of the kit.
AnalyteAnti-RNP antibodiesAnti-RNP antibodies
MethodQUANTA Flash RNPchemiluminescent immunoassayQUANTA Lite RNP ELISA
MatrixHuman serum, buffer, stabilizers,preservativeHuman serum, buffer, stabilizers,preservative
UnitCU (Chemiluminescent units)(arbitrary)Units (arbitray)
Physico-chemicalcharacteristicsLiquid, ready to useLiquid, ready to use
Levels2 (negative and positive)2 (ELISA negative, high positive)
Storage2-8 °C2-8 °C
Shelf lifeOne yearOne year
In-use stability15 uses, with a maximum time of 10minutes onboard the instrument peruse, or 2 ½ hours, total.Controls can be used until the endof the shelf life when storedproperly

OUANTA Flash RNP Controls

Value assignment and traceability of Calibrators and Controls

The QUANTA Flash Sm and RNP Calibrators and Controls are manufactured by diluting human serum that contains high titer of anti-RNP antibodies with a buffer containing stabilizers and preservative. The human serum is obtained from commercial sources and it is tested for markers of infectious substances.

The target CU is achieved through trial dilutions on small scale. Once a dilution is selected, the Calibrators and Control are bulked, tested, and adjusted. Upon completion of the manufacturing process, the Calibrators and Controls are tested on at least two instruments, on at least two lots of reagent cartridge, in replicates of 10 to determine final value assignment.

There are currently no recognized international standards for the measurement of anti-RNP antibodies.

Calibrator and Control values are directly traceable to in-house Standards that are used to create the Master Curve for the QUANTA Flash Sm and QUANTA Flash RNP assays.

Performance characteristics

Precision

The precision of the QUANTA Flash Sm assay was evaluated on 5 samples containing various concentrations of Sm antibodies in accordance with CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Procedures - Approved Guideline: samples were run in duplicates, twice a day, for at least 20 days. Data were analyzed with the Analyse-it for Excel method evaluation software,

{9}------------------------------------------------

Within-RunPrecision(repeatability)Between-RunBetween-DayPrecisionTotal Precision
SampleIDNMean(CU)SD%CVSD%CVSD%CVSD%CV
18013.10.96.50.21.50.75.61.18.7
28022.02.19.71.04.60.94.22.511.5
38493.06.06.40.50.54.85.27.78.3

16.8

ਰੇ ਰੇ

7.1

2.9

6.4

19.7

2.7

5.8

27.3

28.5

11.5

8.4

and within run, between run, between day and total precisions are summarized in the Table below. All %CV values were within the acceptance limit, 15%.

The precision of the QUANTA Flash RNP assay was evaluated on 7 samples containing various concentrations of RNP antibodies in accordance with CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Procedures - Approved Guideline: samples were run in duplicates, twice a day, for at least 20 days. Data were analyzed with the Analyse-it for Excel method evaluation software, and within run, between run, between day and total precision are summarized in the Table below. All %CV values were within the acceptance limit, 15%.

Within-RunPrecision(repeatability)Between-RunPrecisionBetween-DayPrecisionTotal Precision
SampleIDNMean (CU)SD%CVSD%CVSD%CVSD%CV
1926.70.23.50.00.00.45.30.46.4
28424.80.93.70.00.00.83.01.24.8
38831.61.13.40.72.22.78.53.09.4
492120.74.94.14.53.74.53.78.06.6
588218.69.34.27.83.620.39.323.610.8
688319.512.03.714.04.411.63.621.86.8
796409.619.64.818.24.523.35.735.58.7

Reproducibility

4

5

88

84

237.7

338.6

20.6

18.1

8.7

5.3

Three samples for Sm and RNP each were tested on two different regent lots, using two different lots of Calibrators, by two operators. Samples were run in quadruplicates, two times a day, for 10 days, to generate 80 data points. Data were analyzed with the Analyse-it for Excel method evaluation software, and within run, between reagent lots, between calibrator lots, between operators and total precision were calculated and the results are summarized in the Tables below. All %CV values were within the acceptance limit, 15%.

{10}------------------------------------------------

QUANTA Flash Sm

Within RunBetweenReagent LotsBetweenCalibratorLotsBetweenOperatorsTotal
SampleMeán(CU)NumberofreplicatesSD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)
Sample#115.2800.42.81.49.41.16.91.16.91.17.0
Sample#223.3800.62.42.510.71.87.61.77.31.87.7
Sample#3157.3805.53.515.810.115.19.613.18.413.08.3

QUANTA Flash RNP

Within RunBetweenReagent LotsBetweenCalibratorLotsBetweenOperatorsTotal
SampleMean(CU)NumberofreplicatesSD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)SD(CU)CV(%)
Sample#115.3800.42.90.53.10.53.30.31.80.42.8
Sample#223.1800.52.30.72.80.73.10.41.70.62.6
Sample#3173.1805.43.16.23.65.12.93.11.85.12.9

Limit of Blank, Limit of Detection

QUANTA Flash Sm:

The Limit of Detection (LoD) of the QUANTA Flash Sm assay is 803 RLU, which is below the analytical measuring range of the assay. It was determined consistent with CLSI EP17-A guideline with proportions of false positives (alpha) less than 5% and false negatives (beta) less than 5%; based on 140 determinations, with 60 measurements on blank samples and 80 measurements of low level samples. The LoB is 540 RLU. Because the curve ends at 3.3 RLU (which is equal to approximately 2430 RLU), it was not possible to calculate the LoB (540 RLU) and LoD ( 803 RLU) in CUs, only in RLUs (instrument signal).

QUANTA Flash RNP:

The Limit of Detection (LoD) of the QUANTA Flash RNP assay is 1128.8 RLU, which is below the analytical measuring range of the assay. It was determined consistent with CLSI EP17-A guideline with proportions

{11}------------------------------------------------

of false positives (alpha) less than 5% and false negatives (beta) less than 5%; based on 140 determinations, with 60 measurements on blank samples and 80 measurements of low level samples. The LoB is 713 RLU. Because the curve ends at 3.5 RLU (which is equal to approximately 3450 RLU), it was not possible to calculate the LoB (713 RLU) and LoD (1128.8 RLU) in CUs, only in RLUs (instrument signal).

Analytical Measuring Range

QUANTA Flash Sm:

The analytical measuring range (AMR) of the assay (determined by the lowest and highest points of the Master Curve) is 3.3 CU to 693.5 CU, which corresponds to the linear range of the assay. If a patient result is less than 3.3 CU, the BIO-FLASH system will report it as "<3.3 CU". Since this is less than 20 CU, it is considered a negative result. If a patient result is greater than 693.5 CU, the BIO-FLASH system will report it as ">693.5 CU". This is considered a positive result. The BIO-FLASH software has an Auto-Rerun option available (see description and validation in the next paragraph). If this option is selected, the instrument will automatically rerun any sample that has a result of >693.5 CU after additional 10 fold dilution, thereby bringing the measured value within the AMR.

The linearity of the AMR was evaluated by a study according to CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. Five serum samples with various Sm antibody concentrations were diluted with a low negative serum to obtain values that covered a range approximately 20% wider than the AMR. All specimens showed dilution linearity individually, and the combined data yielded the following results with linear regression:

SampleSlope (95% Cl)Y-intercept (95% Cl)R2
All Samples (n=5)0.99 (0.97 to 1.00)0.81 (-3.2 to 4.8)0.99

QUANTA Flash RNP:

The analytical measuring range (AMR) of the assay (determined by the lowest and highest points of the Master Curve) is 3.5 CU to 643.8 CU, which corresponds to the linear range of the assay. If a patient result is less than 3.5 CU, the BIO-FLASH system will report it as "<3.5 CU". Since this is less than 20 CU, it is considered a negative result. If a patient result is greater than 643.8 CU, the BIO-FLASH system will report it as ">643.8 CU". This is considered a positive result. The BIO-FLASH software has an Auto-Rerun option available (see description and validation in the next paragraph). If this option is selected, the instrument will automatically rerun any sample that has a result of >643.8 CU after additional 10 fold dilution, thereby bringing the measured value within the AMR.

The linearity of the AMR was evaluated by a study according to CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. Five serum samples with various anti-RNP concentrations were diluted with a low negative serum to obtain values that covered a range approximately 20% wider than the AMR. All six specimens showed dilution linearity individually, and the combined data yielded the following results with linear regression:

SampleSlope (95% CI)Y-intercept (95% CI)R2
All Samples (n=5)1.00 (1.00 to 1.02)-5.17 (-8.27 to -2.07)1.00

12

{12}------------------------------------------------

Auto-rerun function

The BIO-FLASH software has an Auto-rerun option available. If this option is selected, the instrument will automatically rerun any sample that has a result of >693.5 CU for Sm and > 643.8 CU for RNP after additional 10 fold dilution, thereby bringing the measured value within the AMR. The final result will be calculated by the software. As the highest value that can be measured is 693.5 CU for Sm and 643.8 CU for RNP, the highest value that can be reported is 6935 CU and 6438 CU, respectively.

To validate the Auto-rerun function, five (Sm) and six (RNP) high positive specimens with results above the analytical measuring range were selected for the Sm and the RNP assays, respectively. The samples were run with the Auto-rerun function enabled on the BIO-FLASH. Then the specimens were manually diluted the same way as it happens in the Auto-rerun function (10 fold dilution), and tested on the BIO-FLASH. The results were within the analytical measuring range after auto-rerun or manual dilution for all specimens. The % recovery values for results obtained with the auto-rerun results compared to the results obtained by manual dilution were between 87.6% and 97.7% for Sm (average 91.5%), and between 82.5% and 102.2% for RNP (average 96.3%), respectively (within the ± 20% acceptance limit).

Interference

QUANTA Flash Sm:

The interference study was performed according to CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. Three specimens were tested (negative: 16.3 CU); around the cutoff weak positive: 28.6 CU; high positive: 190.7 CU). Interfering substances were spiked into every specimen at three different concentrations in 10% of total specimen volume, and the resulting samples were assessed in triplicates with the Sm assay. Recovery of the unit values was calculated compared to control samples spiked with the same volume of diluents. Acceptance criteria for the interference studies were 85% - 115% recovery, or ± 4 CU difference, whichever is greater.

No interference was detected with bilirubin up to 10 mg/dL (recovery: 103% to 108% or within ± 4 CU), hemoglobin up to 200 mg/dL (recovery: 97% to 108% or within ± 4 CU), triglycerides up to 1000 mg/dL (recovery: 105% to 111% or within ± 4 CU), cholesterol up to 224.3 mg/dL (recovery: 105% to 111% or within ± 4 CU), and RF IgM up to 500 IU/mL (recovery: 85% to 95% or within ± 4 CU).

QUANTA Flash RNP:

The interference study was performed according to CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. Three specimens were tested (negative: 8.8 CU; weak positive: 55.5 CU; high positive: 280.3). Interfering substances were spiked into every specimen at three different concentrations in 10% of total specimen volume, and the resulting samples were assessed in triplicates with the RNP assay. Recovery of the unit values was calculated compared to control samples spiked with the same volume of diluents. Acceptance criteria for the interference studies were 85% -115% recovery, or ± 4 CU difference, whichever is greater.

No interference was detected with bilirubin up to 10 mg/dL (recovery: 96% to 112%), hemoglobin up to 200 mg/dL (recovery: 88% to 108%), triglycerides up to 1000 mg/dL (recovery: 86% to 109%), cholesterol up to 224.3 mg/dL (recovery: 86% to 109%), and RF IgM up to 500 IU/mL (recovery: 88% to 113%).

{13}------------------------------------------------

High concentration hook effect

To assess hook effect, the measurement signal (relative light units, RLU) was examined for the samples that were used to validate the Auto-rerun function (five Sm and six RNP high positive specimens), before and after automatic or manual dilution. All sera produced significantly higher RLU values (above the AMR) when used "as is" compared to the manually or automatically diluted ones, thereby confirming that high positive specimens above the analytical measuring range do not show hook effect up to 2429 CU in the Sm assay and up to 3140 CU in the RNP assay (the highest concentrations tested).

Cross-reactivity

QUANTA Flash Sm:

To test potential cross-reactivity with autoantibodies and infection-induced antibodies, 233 patient samples with various antibodies to autoimmune or infectious disease markers were tested. Three samples from patients with scleroderma, one sample with HIV antibodies, and one sample with MCTD/primary biliary cirrhosis (PBC) overlap were positive on the QUANTA Flash Sm. All other serum samples were negative:

Patient GroupNPositivePercent Positive
Autoimmune liver disease200%
Viral hepatitis1900%
Scleroderma7434.1%
Sjögren's Syndrome500%
Rheumatoid arthritis7000%
Systemic rheumatic disease, other5311.9%
Infectious disease (HIV + syphilis)10110%
Total23352.1%

QUANTA Flash RNP:

To test potential cross-reactivity with autoantibodies and infection-induced antibodies, 246 patient samples were tested from patients with infectious diseases and connective tissue diseases. Eight samples from patients with scleroderma, two samples from patients with RA, one sample from a patient with PM/DM, and one sample with HIV antibodies were positive on the QUANTA Flash RNP. As anti-RNP antibodies can sometimes be detected in patients with systemic sclerosis, and as all of these scleroderma samples were positive on the predicate RNP ELISA, too, it is highly possible that those samples are true positives, and not cross-reacting samples. All other serum samples were negative:

Patient GroupNPositivePercent Positive
Autoimmune liver disease200%
Viral hepatitis2000%
Scleroderma76810.5%
Sjögren's Syndrome600%
Rheumatoid arthritis7022.9%

{14}------------------------------------------------

Systemic rheumatic disease, other4800%
Poly- and Dermatomyositis1417.1
Infectious disease (HIV + syphilis)10110%
Total246124.9%

Stability

Shelf life

To establish the initial claim for shelf life, accelerated stability studies were performed.

Accelerated stability testing was performed on each of the following sealed components in the QUANTA Flash Sm and RNP to establish initial stability claim: the beads, the two calibrators, and the negative and positive controls. Each week a new sealed component was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5º ± 3ºC. The recovery of the measured values was calculated for each time point (compared to those obtained with 5º ± 3ºC stored reagent). All calculations were performed by comparing results of sealed components stored at 5º ± 3ºC (control) to those stored at 37º ± 3ºC (test) for 1, 2, 3, and 4 weeks, where one week is equal to six months at 5º ± 3ºC. Linear regression analysis was performed between recovery values and the number of days in those cases when at least 3 data points were available at each time point; otherwise, individual data points were analyzed.

Acceptance criteria for one year preliminary expiration dating were:

- Controls and Calibrators:

a) if regression analysis is used, the lower 95% Cl interval of the regression line is > 90% at 2 weeks, and no individual data point has ≤ 80% recovery,

b) If individual data points are analyzed, recovery values are > 90% at day 14.

  • Microparticles:

with regression analysis, the lower 95% Cl interval of the regression line is ≥ 85% at 2 weeks, and no individual data point has ≤ 75% recovery.

Sm beads, RNP beads, Sm Calibrators, RNP Calibrators, Sm Controls and RNP Controls each fulfilled the acceptance criteria above, so one year expiration dating was assigned to each component.

In-use (onboard) stability

Sm Calibrators

During assessing on-board stability, Calibrators were placed, uncapped, onboard the instrument, and calibration was performed altogether five times, then a panel of characterized patient specimens were run on each calibration curve.

Acceptance criteria were: Calibrators are considered stable if all five calibrations performed in the 8.5 hour period are successful, and Calibrator RLU recovery values are between 90% and 110% compared to the first use.

{15}------------------------------------------------

A total of 5 successful calibrations were performed over an 8.5 hour period. Calibrator RLU values remained within the 90-110% range. Moreover, all patient panel samples ran within their expected range. This supports the claim that calibrators can be used for up to 4 calibrations over an 8 hour period.

Sm Controls

During assessing on-board stability, Controls were assayed twice each day over 10 days, for a total of 20 runs. The controls were left uncapped, onboard the instrument for 15 minutes per run. When not in use, the controls were capped, and stored at 2-8°C.

Acceptance criteria: Controls are considered stable when all replicates run within their established range, moreover, and the linear regression line obtained by plotting %recovery values against the number of runs stays between 85% and 115% at day run 15.

Both controls ran within their respective acceptable range for all 20 runs, resulting in a %CV of 5.9% for the Negative Control and 4.8% for the Positive Control. These results support the claim that controls can be used for 15 times, up to 2.5 hours total.

Sm Reagent Cartridge

To determine the in-use stability of the QUANTA Flash Sm reagent pack, three lots of cartridges were tested by using five serum specimens (with different reactivity levels). The specimens were tested periodically up to 48 days, twice a day. Recoveries were calculated compared to the day zero average values, and linear regression analysis was performed. The claim was established using the following criteria (using the one that is fulfilled first):

a)The stability claim is established at the day where the lower 95% confidence interval of the regression line reaches 85% recovery, or

b) When 2% or more of the recovery data is ≤ 75%.

The onboard stability results of the three lots are the following: RP0005: 33 days RP0006: 40 days 121005: 39 days

Using these criteria, the in-use (onboard) stability of the Sm reagent cartridge was set at 33 days.

RNP Calibrators

During assessing on-board stability, Calibrators were placed, uncapped, onboard the instrument, and calibration was performed altogether five times, then a panel of characterized patient specimens were run on each calibration curve.

Acceptance criteria were: Calibrators are considered stable if all five calibrations performed in the 8.5 hour period are successful, and Calibrator RLU recovery values are between 90% and 110% compared to the first use.

{16}------------------------------------------------

A total of 5 successful calibrations were performed over an 8.5 hour period. Calibrator RLU values remained within the 90-110% range. Moreover, all patient panel samples ran within their expected range. This supports the claim that calibrators can be used for up to 4 calibrations over an 8 hour period.

RNP Controls

During assessing on-board stability, Controls were assayed twice each day over 10 days, for a total of 20 runs. The controls were left uncapped, onboard the instrument for 15 minutes per run. When not in use; the controls were capped, and stored at 2-8℃.

Acceptance criteria: Controls are considered stable when all replicates run within their established range, moreover, and the linear regression line obtained by plotting %recovery values against the number of runs stays between 85% and 115% at day run 15.

Both controls ran within their respective acceptable range for all 20 runs, resulting in a %CV of 5.9% for the Negative Control and 6.9% for the Positive Control. These results support the claim that controls can be used for 15 times, up to 2.5 hours total.

RNP Reagent Cartridge

To determine the in-use stability of the QUANTA Flash RNP reagent pack, three lots of cartridges were tested by using four to eight serum specimens (with different reactivity levels). The specimens were tested periodically up to 41 days, twice a day. Recoveries were calculated compared to the day zero average values, and linear regression analysis was performed. The claim was established using the following criteria (using the one that is fulfilled first):

alThe stability claim is established at the day where the lower 95% confidence interval of the regression line reaches 85% recovery, or

b) When 2% or more of the recovery data is ≤ 75%.

The onboard stability results of the three lots are the following: 111001: 28 days 121002: 28 days 121003: 31 days

Using these criteria, the in-use (onboard) stability of the RNP reagent cartridge was set at 28 days.

The Stability Claims are tabulated below for easy review. The 1 year shelf life begins from the date of manufacture.

Sm stability claims

ProductUnopened shelf lifeAfter being opened
QUANTA Flash SmReagent CartridgeUntil the expirationdate on the label(currently 1 year).33 days (onboard only – cannot beresealed).

{17}------------------------------------------------

Sm Calibrator 1 and 2Until the expirationdate on the label(currently 1 year).4 uses, up to 8 hours onboard total
Sm Negative Controland Positive ControlUntil the expirationdate on the label(currently 1 year).15 uses, up to 2.5 hours onboard total

RNP stability claims

ProductUnopened shelf lifeAfter being opened
QUANTA Flash RNPReagent CartridgeUntil the expirationdate on the label(currently 1 year).28 days (onboard only – cannot beresealed).
RNP Calibrator 1 and 2Until the expirationdate on the label(currently 1 year).4 uses, up to 8 hours onboard total
RNP Negative Controland Positive ControlUntil the expirationdate on the label(currently 1 year).15 uses, up to 2.5 hours onboard total

Cut-off, reference range

QUANTA Flash Sm:NegativePositive<20 CU≥20 CU
QUANTA Flash RNP :NegativePositive<20 CU≥20 CU

QUANTA Flash Sm:

The reference population for establishing the reference interval for the Sm assay consisted of 232 subjects:

Apparently healthy blood donors129
Non-autoimmune thyroid disease41
Autoimmune thyroid disease42
Samples for patients with infectious diseases20

All specimens were the same matrix (serum) as specified in the Intended Use. All specimens were unaltered. The cut off was established in accordance to CLSI C28-A3c: Defining, Establishing, and

{18}------------------------------------------------

Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition. The Analyseit for Excel software was used to make the calculations. The distribution of the results was non-normal (Saphiro-Wilk p<0.0001), so the non-parametric percentile method was used. The cut-off was established at the 99" percentile of the results obtained on the reference subjects, and was assigned a value of 20 CU.

QUANTA Flash RNP:

The reference population for establishing the reference interval for the RNP assay consisted of 255 subjects:

Apparently healthy blood donors127
Non-autoimmune thyroid disease41
Autoimmune thyroid disease42
Samples for patients with infectious diseases20
Rheumatoid arthritis25

All specimens were the same matrix (serum) as specified in the Intended Use. All specimens were unaltered. The cut off was established in accordance to CLSI C28-A3c: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition. The Analyseit for Excel software was used to make the calculations. The distribution of the results was non-normal (Saphiro-Wilk p<0.0001), so the non-parametric percentile method was used. The cut-off was established at 99" percentile of the results obtained on the reference subjects, and was assigned a value of 20 CU.

Clinical sensitivity, specificity

QUANTA Flash Sm:

A separate set of samples, none of which were used in establishing the reference range, was used to validate the clinical performance of the Sm CIA.

A total of 379 samples were included in the clinical sensitivity, specificity calculations for the QUANTA Flash Sm. This cohort includes:

  • 63 samples from SLE patients from Neuss Center for Rheumatology, Neuss, Germany .
  • · 83 samples from SLE patients from Dr. Carlos von Mühlen, Germany
  • 74 samples from systemic sclerosis patients
  • 70 samples from rheumatoid arthritis patients .
  • 5 samples from Sjogren's syndrome patients ●
  • 53 samples from patients with other systemic rheumatic diseases .
  • 2 samples from patients with autoimmune liver disease .
  • 19 samples from patients with viral hepatitis .
  • 10 samples from patients with other infectious diseases( 5 HIV, 5 syphilis) .

· All samples were run on the QUANTA Flash Sm ClA. The results were analyzed to calculate sensitivity and specificity for SLE. Results obtained on proficiency testing specimens were excluded from the clinical sensitivity/specificity calculations.

{19}------------------------------------------------

Clinical AnalysisN=379DiagnosisAnalysis(95% confidence)
SLENot SLETotal
QUANTA FlashSmPositive21526Sensitivity = 14.4% (9.1-21.1%)
Negative125228353Specificity = 97.9% (95.1-99.3%)
Total146233379

To assess diagnostic efficiency, ROC analysis was performed on the validation sample pool for SLE (excluding apparently healthy blood donors). The results are below:

TestArea95% CISEZp
QUANTAFlash Sm0.620.57 to 0.670.0254.81<0.0001

QUANTA Flash RNP:

A separate set of samples, none of which were used in establishing the reference range, was used to validate the clinical performance of the RNP CIA.

A total of 424 samples were included in the Validation Set for the QUANTA Flash RNP. This Validation Set includes:

  • 62 samples from SLE patients from Neuss Center for Rheumatology, Neuss, Germany .
  • 32 samples from MCTD patients from Neuss Center for Rheumatology, Neuss, Germany ●
  • 84 samples from SLE patients from Dr. Carlos von Mühlen, Germany ●
  • 48 samples from patients with other systemic rheumatic diseases (not SLE and not . MCTD)
  • 2 patients with autoimmune liver disease .
  • 6 patients with Sjögren's syndrome
  • 76 samples from patients with systemic sclerosis .
  • 70 samples from rheumatoid arthritis patients .
  • 14 samples from patients with polymyositis/dermatomyositis .
  • . 20 samples from patients with viral hepatitis
  • . 10 samples from patients with other infectious diseases( 5 HIV, 5 syphilis)

All samples were run on the QUANTA Flash RNP CIA. The results were analyzed to calculate sensitivity and specificity for SLE and MCTD separately, as well as combined.

Clinical AnalysisN=424DiagnosisAnalysis(95% confidence)
SLE/MCTDControlTotal
QUANTAPositive661278Sensitivity = 37.1% (30.0%-44.6%)
Flash™Negative112234346Specificity = 95.1% (91.6%-97.5%)
RNPTotal178246424

Clinical sensitivity and specificity of the OUANTA Flash RNP in MCTD+SLE:

{20}------------------------------------------------

Clinical sensitivity and specificity of the QUANTA Flash RNP in SLE:
Clinical AnalysisN=392DiagnosisAnalysis(95% confidence)
SLEControlsTotal
QUANTAFlashRNPPositiveNegativeTotal421041461223424654338392Sensitivity = 28.8% (21.6-36.8%)Specificity = 95.1% (91.6-97.5%)

Clinical sensitivity and specificity of the QUANTA Flash RNP in MCTD:

Clinical AnalysisN=278DiagnosisAnalysis(95% confidence)
MCTDControlsTotal
QUANTAFlashRNPPositiveNegativeTotal248321223424636242278Sensitivity = 75.0% (56.6-88.5%)Specificity = 95.1% (91.6-97.5%)

To assess diagnostic efficiency, ROC analysis was performed on the validation sample pool for SLE. The results are below:

TestArea95% CISEZp
QUANTAFlash RNP0.610.55 to 0.670.0303.570.0002

To assess diagnostic efficiency, ROC analysis was performed on the validation sample pool for MCTD. The results are below:

TestArea95% CISEZp
QUANTAFlash RNP0.870.79 to 0.950.0418.95<0.0001

Expected values

Sm

The expected value in the normal population is "negative". Anti-Sm autoantibody levels were analyzed

{21}------------------------------------------------

on a panel of 101 apparently healthy blood donors (71 females and 30 males, ages 18 to 55 years, with an average and median age of 34 years) using the QUANTA Flash Sm. With the cut-off of 20 CU, 1 (1.0 %) of the samples was positive on the QUANTA Flash Sm. The mean concentration was 3.6 CU, and the values ranged from <3.3 to 20.5 CU.

RNP

The expected value in the normal population is "negative". Anti-RNP autoantibody levels were analyzed using the QUANTA Flash RNP on a panel of 101 apparently healthy blood donors (71 females/30 males, ages 18 to 55 years, with an average and median age of 34 years). With a cut-off of 20 CU, one (1%) of the samples was positive on the QUANTA Flash RNP. The mean concentration was 5.7 CU, and the values ranged from <3.5 to 196.8 CU.

Comparison with predicate device

QUANTA Flash Sm:

Samples for the method comparison analysis included samples from the clinical validation studies (SLE patients and disease controls), as well as samples from proficiency surveys (CAP and NEQAS) and samples from apparently healthy blood donors within the AMR of the assay. These samples were tested on both the QUANTA Flash Sm and on the predicate ELISA.

Method Comparison(N=119)Sm ELISAPercent Agreement(95% Confidence)
PositiveNegativeTotal
QUANTA FlashSm CIAPositive35641Pos. Agreement = 92.1% (78.6 - 98.3%)
Negative37578Neg. Agreement = 92.6% (84.6 - 97.2%)
Total3881119Total Agreement = 92.4% (86.1 - 96.5%)

QUANTA Flash RNP:

Samples for the method comparison analysis included samples from the clinical validation studies (SLE, MCTD patients and disease controls), as well as samples from apparently healthy blood donors with results within the AMR of the assay. These samples were tested on both the QUANTA Flash RNP and on the predicate ELISA.

Method Comparison(N=167)RNP ELISAPercent Agreement(95% Confidence)
PositiveNegativeTotal
QUANTA FlashRNP CIAPositive46652Pos. Agreement = 80.7% (68.1 - 90.0%)
Negative11104115Neg. Agreement = 94.5% (88.5 - 98.0%)
Total57110167Total Agreement = 89.8% (84.2 - 94.0%)

22

{22}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/22/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 17, 2013

INOVA DIAGNOSTICS, INC. c/o GABRIELLA LAKOS, Ph.D. DIRECTOR, RHEUMATOLOGY RESEARCH 9900 OLD GROVE ROAD SAN DIEGO CA 92131

Re: K123593

Trade/Device Name: QUANTA Flash® Sm

QUANTA Flash® Sm Calibrators QUANTA Flash® Sm Controls QUANTA Flash® RNP OUANTA Flash® RNP Calibrators QUANTA Flash® RNP Controls

Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: II Product Code: LKP, LKO, JIT, JJX Dated: April 15. 2013 Received: April 16, 2013

Dear Dr. Lakos:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your, device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements

{23}------------------------------------------------

Page 2-Gabriella Lakos, Ph.D.

as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.

Sincerely yours.

Maria M. Chan -S

Maria M. Chan, Ph.D.

Director

Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health

Center for Devices and Radiological Health

Enclosure

{24}------------------------------------------------

510(k) Number (if known): K123593

Device Name: QUANTA Flash® Sm

Indications for Use:

The QUANTA Flash Sm is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-Sm antibodies in human serum. The presence of anti-Sm antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus Erythematosus (SLE).

Prescription Use __ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Maria M. Chan -S

Division Sign-Off

Office of In Vitro Diagnostics and Radiological Health

510(k)_k123593

Page 1 of _ 1 _

{25}------------------------------------------------

510(k) Number (if known): K123593

QUANTA Flash® Sm Calibrators Device Name: .

Indications for Use:

QUANTA Flash Sm Calibrators are intended for use with the QUANTA Flash Sm chemiluminescent immunoassay for the determination of IgG anti-Sm antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Maria M. Chan -S

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k)_k123593

Page 1 of _ 1 _

{26}------------------------------------------------

510(k) Number (if known): K123593 :

QUANTA Flash® Sm Controls Device Name:

Indications for Use:

QUANTA Flash Sm Controls are intended for use with the QUANTA Flash Sm chemiluminescent immunoassay for quality control in the determination of IgG anti-Sm antibodies in human serum.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OlR)

Maria M. Chan -S

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k) K123593

Page 1 of _ 1 _

{27}------------------------------------------------

K123593 510(k) Number (if known):

Device Name: QUANTA Flash® RNP

Indications for Use:

The QUANTA Flash RNP is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-ribonucleoprotein (RNP) antibodies in human serum. The presence of anti-RNP antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus Erythematosus (SLE) and Mixed Connective Tissue Disease (MCTD).

Prescription Use X (Part 21 CFR 801 Subpart D)

.AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Maria M. Chan -S

Division Sign-Off

Office of In Vitro Diagnostics and Radiological Health

510(k)_k123593

Page 1 of _ 1_

{28}------------------------------------------------

510(k) Number (if known): K123593

Device Name: QUANTA Flash® RNP Calibrators

Indications for Use:

QUANTA Flash RNP Calibrators are intended for use with the QUANTA Flash RNP chemiluminescent immunoassay for the determination of IgG anti-RNP antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Maria M. Chan -S

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k) K123593

Page 1 of 1 _

{29}------------------------------------------------

510(k) Number (if known): K123593

QUANTA Flash® RNP Controls Device Name:

Indications for Use:

QUANTA Flash RNP Controls are intended for use with the QUANTA Flash RNP chemiluminescent immunoassay for quality control in the determination of IgG anti-RNP antibodies in human serum.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Maria M. Chan -S

. Division Sign-Off Office of In Vitro Diagnostics and Radiological Health

510(k)_k123593

Page 1 of _1

§ 866.5100 Antinuclear antibody immunological test system.

(a)
Identification. An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).(b)
Classification. Class II (performance standards).