K Number
K123548
Device Name
THE ARTAS SYSTEM
Date Cleared
2013-02-25

(98 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ARTAS™ System from Restoration Robotics is indicated for harvesting hair follicles from the scalp in men diagnosed with androgenic alopecia (male pattern hair loss) who have black or brown straight hair. The ARTAS System from Restoration Robotics is intended to assist physicians in identifying and extracting hair follicular units from the scalp during hair transplantation. The ARTAS System is also indicated for creating recipient sites for subsequent manual implantation of the harvested follicles.

Device Description

The ARTAS™ System implements the manual Follicular Unit Extraction (FUE) approach to harvesting follicular units and has the potential to solve the technical challenges inherent in the manual FUE technique. The ARTAS™ System is used, under the direction of a physician, to automate the manual FUE technique. The ARTASTM System is positioned over the patient's scalp by the physician and follicular units are identified. Follicular units are then harvested from the patient's scalp. The follicular units are stored until they are implanted into the patient's scalp by the physician or technician using current manual implantation techniques. Following harvesting, the ARTAS system is positioned over the patient's scalp by the physician and implantation sites are made in the designated area of the scalp. The ARTASTM System is an interactive, image-guided, computer-assisted system consisting of seven main subsystems: 1. Robotic Arm Subsystem 2. Imaging Subsystem 3. Needle Mechanism 4. Safety Subsystem 5. Computer 6. Accessory Kits (Disposable and Reusable) 7. Patient Chair These main subsystems, with the exception of the Accessory Kits and Patient Chair, are housed on or within a Control Cart.

AI/ML Overview

Here is an analysis of the ARTAS™ System's acceptance criteria and testing, based on the provided 510(k) summary:

1. Acceptance Criteria and Reported Device Performance

The provided 510(k) summary for the ARTAS™ System does not explicitly state quantitative acceptance criteria or detailed reported device performance metrics in a readily extractable table format. Instead, it describes general safety and effectiveness.

However, based on the summary, the implicit acceptance criteria are that the device should:

  • Perform consistently during its intended use.
  • Not raise new questions of safety or effectiveness compared to predicate devices.
  • Function to its specifications and perform as intended.

The reported performance is that the device meets these criteria.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The summary mentions a "multi-center, prospective, blinded, clinical study" that enrolled "healthy men, between the ages of 30 and 59, who had brown or black straight hair and a clinical diagnosis of androgenetic alopecia (male pattern hair loss)." However, the exact number of participants (sample size) in this clinical study is not specified in the provided text.
  • Data Provenance: The study was a "multi-center, prospective, blinded, clinical study." The country of origin is not explicitly stated, but the applicant (Restoration Robotics) is based in San Jose, California, USA, suggesting it was likely conducted in the USA or North America.

3. Number of Experts Used to Establish Ground Truth and Qualifications

  • The summary does not provide information on the number of experts used or their qualifications for establishing ground truth for the test set. It mentions the study compared the ARTAS System to the "manual hair follicle harvesting method," implying physicians or trained personnel performed both methods.

4. Adjudication Method for the Test Set

  • The summary does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing ground truth or evaluating outcomes in the clinical study. It states the study was "blinded," but the specifics of how blinding was achieved and how outcomes were assessed are not detailed beyond "post-procedural evaluation."

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC study was described. The clinical study was designed to compare the "safety and effectiveness of the ARTAS System to the manual hair follicle harvesting method." This implies a comparison of the device's performance against a standard manual procedure, but not specifically an "AI vs. Human with AI assistance" type study. The ARTAS System is the robotic assistance for the physician. Therefore, the study, in essence, compares the robotic-assisted method to the manual method.
  • Effect Size: Since an MRMC comparative effectiveness study of "human readers improve with AI vs. without AI assistance" was not described in this context, no effect size can be provided. The ARTAS system is the automated tool assisting physicians; it's not an AI model providing interpretations to human readers.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • No standalone (algorithm only) performance study was described. The ARTAS™ System is explicitly stated to assist physicians and operates "under the direction of a physician." Therefore, its intended use and evaluation intrinsically include a human-in-the-loop, where the physician operates and guides the automated system, validating its performance under this paradigm.

7. Type of Ground Truth Used

  • The ground truth in the clinical study was based on clinical evaluation and post-procedural assessment over a nine-month period. This would likely involve visual inspection, assessment of hair growth, and identification of any adverse events, conducted by medical professionals within a clinical setting.

8. Sample Size for the Training Set

  • The 510(k) summary does not provide information on the sample size used for the training set for the ARTAS System's software or algorithms. The device uses an "interactive, image-guided, computer-assisted system," implying underlying algorithms for image analysis and robotic control. However, the details of their development data are not discussed.

9. How the Ground Truth for the Training Set Was Established

  • The 510(k) summary does not describe how the ground truth for any training set was established. Since the ARTAS System involves "identifying and extracting hair follicular units" and "avoiding existing hairs," it's highly probable that ground truth for training would involve manual annotation of follicular units and scalp features by experts on images or directly on patients. However, this is not explicitly stated in the provided text.

{0}------------------------------------------------

Restoration Robotics, Inc.

THE ARTAS™ SYSTEM FROM RESTORATION ROBOTICS

510(k) SUMMARY

510(k) Notification K123548

GENERAL INFORMATION

Applicant:

Restoration Robotics 128 Baytech Drive San Jose, Ca 95134 U.S.A. Phone: 408-883-6760 FAX: 408-883-6889

Contact Person:

Jim Talbot Restoration Robotics 128 Baytech Drive San Jose, Ca 95134 U.S.A. Phone: 408-883-6760 FAX: 408-883-6889

Date Prepared: November 1, 2012

DEVICE INFORMATION

Trade Name: The ARTAS System from Restoration Robotics

Generic/Common Name: Stereotaxic Instrument

Classification: 21CFR§882.4560, Stereotaxic Instrument

Product Code: ONA

$\rho_{i} & \psi$

{1}------------------------------------------------

RESTORATION ROBOTICS, INC.

THE ARTAS™ SYSTEM FROM RESTORATION ROBOTICS

510(k) SUMMARY

P224

PREDICATE DEVICE(S)

The proposed ARTASTM System is substantially equivalent to the following predicate devices:

1231-48

  • . ARTASTM System, K103428
  • Recipient Site Blades (Manual Surgical Instruments, Regulation No. 878.4800, . Class I, exempt)

INTENDED USE

The ARTASTM System from Restoration Robotics is indicated for harvesting hair follicles from the scalp in men diagnosed with androgenic alopecia (male pattern hair loss).who have black or brown straight hair. The ARTAS System from Restoration Robotics is intended to assist physicians in identifying and extracting hair follicular units from the scalp during hair transplantation. The ARTAS System is also indicated for creating recipient sites for subsequent manual implantation of the harvested follicles.

PRODUCT DESCRIPTION

The ARTAS™ System implements the manual Follicular Unit Extraction (FUE) approach to harvesting follicular units and has the potential to solve the technical challenges inherent in the manual FUE technique. The ARTAS™ System is used, under the direction of a physician, to automate the manual FUE technique. The ARTASTM System is positioned over the patient's scalp by the physician and follicular units are identified.

Follicular units are then harvested from the patient's scalp. The follicular units are stored until they are implanted into the patient's scalp by the physician or technician using current manual implantation techniques.

Following harvesting, the ARTAS system is positioned over the patient's scalp by the physician and implantation sites are made in the designated area of the scalp.

The ARTASTM System is an interactive, image-guided, computer-assisted system consisting of seven main subsystems:

    1. Robotic Arm Subsystem
    1. Imaging Subsystem
    1. Needle Mechanism
    1. Safety Subsystem
    1. Computer
    1. Accessory Kits (Disposable and Reusable)
    1. Patient Chair

These main subsystems, with the exception of the Accessory Kits and Patient Chair, are housed on or within a Control Cart.

{2}------------------------------------------------

RESTORATION ROBOTICS, INC.

THE ARTAS™ SYSTEM FROM RESTORATION ROBOTIC

510(k) SUMMARY

TECHNOLOGICAL CHARACTERISTICS

The technological characteristics of the ARTAS System from Restoration Robotics are similar to the predicate devices. No changes to the hardware were made to the subject modified ARTAS System, or even to the needle mechanism. Site making is performed simply by replacing one disposable (the harvest needle) with another (the site-making blade).

A new "operation mode" was created within the software for site-making. In this mode you can manually or automatically pick sites within the same tensioner bounds as used by harvesting. The operator can control site density, site depth, and angle at which the blade enters when it makes the site. The software automatically avoids existing hairs exactly as it does during harvest. Other than these few changes, the same software and user interface used by harvesting is used by site-making. Performance Data were provided to support the determination of substantial equivalence.

SUBSTANTIAL EQUIVALENCE

The indications for use for the predicate devices (ARTAS System and the Recipient Site Blades) are substantially equivalent to the proposed indications for use for the proposed ARTAS System. Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the ARTAS System is substantially equivalent to the predicate device.

TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION

All necessary bench and clinical testing was conducted on the ARTAS System to support a determination of substantial equivalence to the predicate device.

Nonclinical Testing Summary:

The nonclinical, bench testing included:

  • Software verification and validation data .
  • Hardware design verification data .
  • Implantation site-making performance testing .

The collective results of the nonclinical bench testing demonstrate that the materials chosen, the manufacturing processes, and design of the ARTAS System meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the ARTAS System does not raise new questions of safety or effectiveness when compared to the predicate devices.

Clinical Testing Summary:

A multi-center, prospective, blinded, clinical study was performed to compare the safety and effectiveness of the ARTAS System to the manual hair follicle harvesting method following a nine month period of post-procedural evaluation. The clinical study enrolled healthy men, between the ages of 30 and 59, who had brown or black straight hair and a

{3}------------------------------------------------

RESTORATION ROBOTICS, INC.

THE ARTAS™ SYSTEM FROM RESTORATION ROBOTICS

510(k) SUMMARY

clinical diagnosis of androgenetic alopecia (male pattern hair loss). The clinical report generated from this study supported safety and effectiveness of the predicate ARTAS System cleared under 510(k) No. K103428.

An additional safety analysis was performed on the data from this study to demonstrate that the intended use of creating follicular implantation sites do not raise any new issues of safety.

CONCLUSION

The ARTAS System from Restoration Robotics shares its design and mechanism of action as the predicate devices. The results of the nonclinical bench and clinical testing demonstrate that the ARTAS System functions to its specifications and performs as intended. The ARTAS System is substantially equivalent to the predicate devices in terms of technology characteristics, intended use, and performance. No new issues of safety or effectiveness are raised.

SUMMARY

The ARTAS System is substantially equivalent to the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 25, 2013

Restoration Robotics, Incorporated % Mr. Jim Talbot Senior Director, Quality and Regulatory Affairs 128 Baytech Drive San Jose, California 95134

Re: K123548

Trade/Device Name: The ARTAS" System from Restoration Robotics Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: ONA Dated: January 30, 2013 Received: February 01, 2013

Dear Mr. Talbot:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{5}------------------------------------------------

Page 2 - Mr. Jim Talbot

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

FOR

Peter D. Ryimm -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

RESTORATION ROBOTICS

THE ARTAS™SYSTEM FROM RESTORATION ROBOTICS 510(k) PREMARKET NOTIFICATION

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K123548

Device Name: The ARTASTM System from Restoration Robotics

Indications For Use:

The ARTAS™ System from Restoration Robotics is indicated for harvesting hair follicles from the scalp in men diagnosed with androgenic alopecia (male pattern hair loss) who have black or brown straight hair. The ARTAS System from Restoration Robotics is intended to assist physicians in identifying and extracting hair follicular units from the scalp during hair transplantation. The ARTAS System is also indicated for creating recipient sites for subsequent manual implantation of the harvested follicles.

X Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

(21 CFR 801 Subpart C)

Over-The-Counter Use _

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R Ogden: 2013.02.25 10:41:33 -05'00'

for MXM (Division Sign-Off)

Division of Surgical Devices

510(k) Number

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).