K Number
K123516
Device Name
CRYO-TOUCH IV
Manufacturer
Date Cleared
2013-01-10

(57 days)

Product Code
Regulation Number
882.4250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Myoscience Cryo---Touch IV is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo--- Touch IV is not indicated for treatment of central nervous system tissue.

Device Description

The Myoscience Cryo-Touch IV is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through application of extreme cold to the selected site. The device is based on introduction of a Smart Tip internally cooled by the cryogenic fluid (nitrous oxide, N-O) to a selected area. The Smart Tip is cooled by the Joule-Thomson Effect and/or Latent Heat of Vaporization. The Cryo-Touch IV may be used in conjunction with a standard off-the-shelf nerve stimulator device in applications where precise nerve location is desired. The device is comprised of four main components: A reusable Handpiece, A Charging Dock, A single-patient use Smart Tip, and A Cartridge (Nitrous Oxide).

AI/ML Overview

The Myoscience Cryo-Touch IV is a cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through the application of extreme cold. The provided document describes nonclinical testing performed to demonstrate that the product met design requirements for system performance.

1. Table of Acceptance Criteria and Reported Device Performance:

Test PerformedAcceptance Criteria (Implied)Reported Device Performance
Temperature reproducibilitySubstantially equivalent to predicate devicesPASS, Substantially equivalent to predicate
Mechanical Integrity for SystemSubstantially equivalent to predicate devicesPASS, Substantially equivalent to predicate
Nitrous ExposureSubstantially equivalent to predicate devicesPASS, Substantially equivalent to predicate
Cryozone SizeSubstantially equivalent to predicate devicesPASS, Substantially equivalent to predicate
Needle IntegritySubstantially equivalent to predicate devicesPASS, Substantially equivalent to predicate
Device DurabilitySubstantially equivalent to predicate devicesPASS, Substantially equivalent to predicate
Sterilization and Shelf Life TestingIn compliance with pertinent standards (ISO 11137-1) and substantially equivalent to predicate devicesPASS, Substantially equivalent to predicate
Electrical Safety TestingIn compliance with pertinent standards (IEC 60601-1, IEC 60601-1-2) and substantially equivalent to predicate devicesPASS, Substantially equivalent to predicate
Software TestingSubstantially equivalent to predicate devicesPASS, Substantially equivalent to predicate
Safety TestingSubstantially equivalent to predicate devicesPASS, Substantially equivalent to predicate
Biocompatibility TestingIn compliance with pertinent standards (ISO 10993-1) and substantially equivalent to predicate devicesPASS, Substantially equivalent to predicate

2. Sample Size Used for the Test Set and Data Provenance:

The document mentions "Verification testing was performed on the Cryo-Touch IV device," but does not specify the sample size used for the test set. The data provenance is not explicitly stated. These were nonclinical tests, likely performed in
a laboratory setting, rather than on human subjects or clinical data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

This information is not provided in the document. Given that these are nonclinical device performance tests, the "ground truth" would likely be established by engineering specifications, validated measurement techniques, and comparison to predicate device performance, not clinical expert consensus.

4. Adjudication Method for the Test Set:

This information is not provided in the document. As these are nonclinical tests, an adjudication method in the human-in-the-loop sense is not applicable. The assessment would be based on meeting predefined engineering and performance specifications.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

No, an MRMC comparative effectiveness study was not done. The tests described are nonclinical performance validations of the device itself, not studies comparing human reader performance with and without AI assistance. The Cryo-Touch IV is a physical surgical device, not an AI or imaging diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

The concept of a "standalone" algorithm performance is not applicable to the Cryo-Touch IV. It is a physical medical device. The "standalone" evaluation would refer to its own functional standalone performance as measured by the nonclinical tests listed, which were indeed performed.

7. The Type of Ground Truth Used:

The ground truth for these nonclinical tests appears to be based on:

  • Engineering specifications and design requirements: The document states the tests "demonstrate that the product met the design requirements for system performance."
  • Pertinent consensus standards: Compliance with standards like IEC 60601-1, IEC 60601-1-2, ISO 10993-1, and ISO 11137-1 serves as a form of ground truth or benchmark.
  • Performance of legally marketed predicate devices: The "substantially equivalent to predicate" outcome for many tests indicates that the performance of existing, approved devices served as a critical benchmark or ground truth for acceptable performance.

8. The Sample Size for the Training Set:

This information is not applicable and not provided. The Cryo-Touch IV is a physical cryogenic surgical device. There is no mention or indication of any AI/ML components that would require a "training set" in the context of machine learning. The nonclinical tests described are for device validation, not algorithm training.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable and not provided for the same reasons as point 8.

{0}------------------------------------------------

Cryo-Touch IV

Premarket Notification

Section 5: 510(k) Summary

Device Information:

CategoryComments
Sponsor / Submitter:Myoscience, Inc1600 Seaport BlvdNorth Lobby, Suite 450Redwood City, CA 94063(650) 474-2600(650) 474-2700
Correspondent ContactInformation:Tracey HenryVice President RAQA, Operations1600 Seaport BlvdNorth Lobby, Suite 450Redwood City, CA 94063(650) 474-2600(650) 474-2900
Device Common Name:Cryogenic surgical device
Device Classification & Code:Class II, GXH
Device Classification Name:Cryosurgical unit and accessories (21 CFR 882.4250)
Device Proprietary Name:Myoscience Cryo-Touch IV

a. Predicate Device Information:

The Cryo-Touch IV is substantially equivalent to the following currently legally marketed devices:

510(k) NumberProductSponsor
K102021Cryo-Touch IIMyoscience, Inc
K120415Cryo-Touch IIIMyoscience, Inc

b. Date Summary Prepared

November 6, 2012

c. Description of Device

The Myoscience Cryo-Touch IV is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through application of extreme cold to the selected site. The device is based on introduction of a Smart Tip internally cooled by the cryogenic fluid (nitrous oxide, N-O) to a selected area. The Smart Tip is cooled by the Joule-Thomson Effect and/or Latent Heat of Vaporization. The Cryo-Touch IV may be used in conjunction with a standard off-the-shelf nerve stimulator device in applications where precise nerve location is desired.

Device Design

The device is comprised of four main components:

  • A reusable Handpiece 1.
    1. A Charging Dock
    1. A single-patient use Smart Tip
    1. A Cartridge (Nitrous Oxide)

The Cryo-Touch IV Handpiece is battery powered and provides feedback to the user during device preparation and use. The Handpiece connects to both the Cartridge and to the Smart Tip. The user

{1}------------------------------------------------

myoscience

Cryo-Touch IV

Premarket Notification

activates a treatment cycle through a control on the Handpiece, which starts and stops the treatment. The Handpiece also contains LEDs for providing feedback to the user when the device is ready to use. The Charging Dock stores the Handpiece between uses and provides power for charging the battery.

An assortment of Smart Tips is available for the Cryo-Touch IV. All Smart Tips needles are made of stainless steel and have a closed-end that fully contains the cryogen so that it does not enter the target tissue. The Smart Tip is the only patient contacting component of the Cryo-Touch IV. The user removes the Smart Tip from the sterile packaging and attaches it to the Handpiece.

The Cryo-Touch IV uses a commercially available nitrous oxide cylinder. The Cartridge is filled with pure N20.

Device Functionality/Scientific Concepts

The device functionality is based on the user introducing the Smart Tip to the selected treatment area: unwanted tissue or the target nervous tissue. The user then initiates the flow of cryogen by pressing the on/off button. Liquid cryogen flows from the Handpiece into the closed-end Smart Tip is cooled by the Joule-Thomson Effect and/or Latent Heat of Vaporization; as the liquid cryogen expands into a gas, the temperature drops around the external surface of the Smart Tip causing the surrounding tissue to freeze. The treatment is completed after a pre-programmed amount of time at which time the user can safely remove the Smart Tip.

d. Intended Use

The Myoscience Cryo-Touch IV is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo-Touch IV is not indicated for treatment of central nervous system tissue.

e. Comparison to Predicate Devices

The Myoscience Cryo-Touch IV is substantially equivalent in intended use, technology, design and materials to the above listed legally marketed predicate devices.

f. Summary of Supporting Data

Nonclinical testing:

Verification testing was performed on the Cryo-Touch IV device to demonstrate that the product met the design requirements for system performance. These specific tests are listed below.

Test PerformedResult
Temperature reproducibilityPASS, Substantially equivalent to predicate
Mechanical Integrity for SystemPASS, Substantially equivalent to predicate
Nitrous ExposurePASS, Substantially equivalent to predicate
Cryozone SizePASS, Substantially equivalent to predicate
Needle IntegrityPASS, Substantially equivalent to predicate
Device DurabilityPASS, Substantially equivalent to predicate
Sterilization and Shelf Life TestingPASS, Substantially equivalent to predicate
Electrical Safety TestingPASS, Substantially equivalent to predicate
Software TestingPASS, Substantially equivalent to predicate

Section 5: 510k Summary

Page 16

{2}------------------------------------------------

myoscience

Cryo-Touch IV

Premarket Notification

Test PerformedResult
Safety TestingPASS, Substantially equivalent to predicate
Biocompatibility TestingPASS, Substantially equivalent to predicate

This performance testing demonstrated that the device is in compliance with pertinent standards (IEC 60601-1, IEC 60601-1-2, ISO 10993-1, and ISO 11137-1), the product labeling, and is substantially equivalent to the predicates.

..------------------------------------------------------------------------------------------------------------------------------------------------------------------------------FILES AND POST OF CO., CO., AND AND------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

g. Conclusion

Myoscience concludes that the Cryo-Touch IV described in this substantially equivalent to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with a more abstract and modern design.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

January 10, 2013

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Myoscience, Inc. % Tracey Henry Vice President RAQA, Operations 1600 Seaport Boulevard North Lobby, Suite 450 Redwood City, CA94063

. Re: K123516

Trade/Device Name: MyoScience Cryo-Touch IV Regulation Number: 21 CFR 882.4250 Regulation Name: Cryogenic Surgical Device Regulatory Class: II Product Code: GXH Dated: November 06, 2012 Received: November 14, 2012

Dear Tracey Henry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

Page 2 - Tracey Henry

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Victor Krauthamer -S

Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): _ K123516

Device Name: _ MyoScience Cryo-Touch IV

Indications For Use:

The Myoscience Cryo---Touch IV is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. The Cryo--- Touch IV is not indicated for treatment of central nervous system tissue.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Vith Saath

ision of Neurological and Physical Medicine Devices (DNPMD)

510(k) Number K123516

§ 882.4250 Cryogenic surgical device.

(a)
Identification. A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.(b)
Classification. Class II (performance standards).