(89 days)
The TRC-NW300 intended for use in capturing images of the retina and the anterior segment of the eye and presenting the data to the eye care professional, without the use of a mydriatic.
The Topcon TRC-NW300 is a fundus camera designed to observe, photograph and record the fundus oculi of a patient's eye with or without the use of a mydriatic. The TRC-NW300 does not come into contact with the patient's eye and provides the fundus oculi image information as an electronic image for later analysis. The TRC-NW300 houses a color LCD monitor used for observation and display of a photographed image and a digital photography unit used for recording images. A photographed image may be recorded on a commercially available memory card inserted in the memory card slot of the TRC-NW300, to a commercially available image filing system, or on a commercially available storage device (such as a flash memory, a hard disk or a card reader/writer) connected to the TRC-NW300. A photographed image may also be printed on a commercially available digital printer connected to the TRC-NW300.
The provided 510(k) summary for the Topcon TRC-NW300 Non-Mydriatic Retinal Camera primarily focuses on demonstrating substantial equivalence to predicate devices rather than presenting a detailed study with specific acceptance criteria and performance metrics for the device itself.
Therefore, many of the requested details, particularly regarding specific numerical performance metrics, sample sizes for test and training sets, expert qualifications, and adjudication methods for a diagnostic effectiveness study, are not explicitly present in the provided document. The document mainly highlights compliance with safety and electrical standards and modifications from previous models.
However, based on the information provided, here's what can be gathered:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not provide a table with specific, quantifiable acceptance criteria (e.g., sensitivity, specificity, image quality scores) and corresponding reported device performance metrics in the way a diagnostic study would.
Instead, the performance data section states:
- "The TRC-NW300 has been tested and found in compliance with the following recognized consensus standards: IEC 60601-1, IEC 60601-1-2, ISO 15004-1, ISO 15004-2."
- "Additionally, design verification tests were performed as a result of the risk analysis assessment. These test results demonstrated that the TRC-NW300 met predetermined acceptance criteria."
The "acceptance criteria" here refer to meeting the requirements of these standards and the predetermined criteria of internal design verification tests, which are not detailed in this summary.
2. Sample Size Used for the Test Set and Data Provenance
Not applicable/Not provided. The document does not describe a clinical performance study with a test set of images or patients for diagnostic accuracy. The "test results" refer to engineering design verification.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable/Not provided. There is no mention of a test set with ground truth established by experts.
4. Adjudication Method for the Test Set
Not applicable/Not provided.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
Not applicable/Not provided. The document does not describe an MRMC study comparing human readers with and without AI assistance. The device is a retinal camera, not an AI diagnostic tool.
6. Standalone Performance Study (Algorithm Only)
Not applicable/Not provided. The device is a retinal camera for image capture, not an AI algorithm for standalone diagnosis. The "performance data" refers to the camera's technical performance and safety compliance.
7. Type of Ground Truth Used
Not applicable/Not provided in the context of a diagnostic performance study. The ground truth for the device's functional integrity was established through compliance with recognized safety and performance standards (e.g., electrical safety, light hazard protection) and internal design verification against "predetermined acceptance criteria."
8. Sample Size for the Training Set
Not applicable/Not provided. The device is a camera, and there is no mention of a training set as would be used for a machine learning algorithm.
9. How the Ground Truth for the Training Set Was Established
Not applicable/Not provided.
Summary of the Study that Proves the Device Meets Acceptance Criteria:
The "study" or rather, the evidence provided to demonstrate that the TRC-NW300 meets its acceptance criteria, is based on:
- Compliance with Recognized Consensus Standards: The device was tested and found compliant with:
- IEC 60601-1: General Requirements for Safety of Medical Electrical Equipment.
- IEC 60601-1-2: Electromagnetic Compatibility.
- ISO 15004-1: General requirements applicable to all ophthalmic instruments.
- ISO 15004-2: Light hazard protection for ophthalmic instruments.
- Design Verification Tests: These tests were performed as a result of a risk analysis assessment. The document states that "These test results demonstrated that the TRC-NW300 met predetermined acceptance criteria." These are internal engineering and functional tests to ensure the device performs as designed and safely, especially considering the modifications from its predicate.
- Substantial Equivalence Argument: The core of the 510(k) submission is to argue that the TRC-NW300 is substantially equivalent to its predicate devices (TRC-NW200 and TRC-NW8F). This argument is based on the device having the "same intended use and similar indications, principles of operation, and technological characteristics." The modifications (e.g., higher megapixel camera, smaller pupil photography, auto-focus, automatic flash, USB memory card capability) were evaluated, and Topcon determined "they do not affect the safety or effectiveness of the device." Many of the new features were already cleared in the TRC-NW8F.
In essence, the "acceptance criteria" here are primarily functional, safety, and performance standards as an ophthalmologic imaging device, and the "proof" is through conformity to these standards and a comparison to previously cleared, substantially equivalent devices. The document does not describe a clinical study focused on diagnostic accuracy or reader performance.
{0}------------------------------------------------
FEB 0 6 2013
510(k) SUMMARY
TOPCON TRC-NW300 Non-Mydriatic Retinal Camera
Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Topcon Medical Systems, Inc. 111 Bauer Drive Oakland, NJ 07436 Phone: (201) 599-5553 Facsimile: (201) 599-5240 Contact Person: Michael Gusel
Additional Correspondent:
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 Phone: (978) 207-1245 Facsimile: (978) 824-2541
February 1, 2013 Date Prepared:
Name of Device and Name/Address of Sponsor
Topcon TRC-NW300 Non-Mydriatic Retinal Camera Topcon Medical Systems, Inc. 111 Bauer Drive Oakland, NJ 07436
Common or Usual Name Retinal Camera
Classification Name
Camera, Ophthalmic, AC-Powered 21 C.F.R. 886.1120 Product Code: HKI
{1}------------------------------------------------
Predicate Devices Topcon Medical Systems TRC-NW200 (K041367) Topcon Medical Systems TRC-NW8F (K100207)
Intended Use
The TRC-NW300 intended for use in capturing images of the retina and the anterior segment of the eye and presenting the data to the eye care professional, without the use of a mydriatic.
15123460
Device Description
The Topcon TRC-NW300 is a fundus camera designed to observe, photograph and record the fundus oculi of a patient's eye with or without the use of a mydriatic. The TRC-NW300 does not come into contact with the patient's eye and provides the fundus oculi image information as an electronic image for later analysis.
The TRC-NW300 houses a color LCD monitor used for observation and display of a photographed image and a digital photography unit used for recording images. A photographed image may be recorded on a commercially available memory card inserted in the memory card slot of the TRC-NW300, to a commercially available image filing system, or on a commercially available storage device (such as a flash memory, a hard disk or a card reader/writer) connected to the TRC-NW300. A photographed image may also be printed on a commercially available digital printer connected to the TRC-NW300.
Performance Data
The TRC-NW300 has been tested and found in compliance with the following recognized consensus standards:
IEC 60601-1:1988 +Amd 1: 1991 + Amd 2: 1995, Medical Electrical Equipment - Part 1: General Requirements for Safety;
IEC 60601-1-2: 2007 Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance: Collateral Standard: Electromagnetic Capability - Requirements and Tests;
{2}------------------------------------------------
15123460
ISO 15004-1:2006 Ophthalmic Instruments-Fundamental requirements and test methods-Part 1: General requirements applicable to all ophthalmic instruments;
ISO 15004-2:2007, Ophthalmic Instruments-Fundamental requirements and test methods-Part 2: Light hazard protection.
Additionally, design verification tests were performed as a result of the risk analysis assessment. These test results demonstrated that the TRC-NW300 met predetermined acceptance criteria.
Technological Characteristics and Substantial Equivalence
The TRC-NW300 has very similar technological characteristics as the TRC-NW200, to which it is a modification. The overall principle of operation of the TRC-NW300 is unchanged compared to the previously cleared TRC-NW200. Both devices illuminate the fundus oculi using infrared light emitted from the illumination optical system and present an image of the fundus oculi formed on the observation CCD camera housed in the main unit to the color LCD monitor. Both devices have a built in camera for photography and a built in camera for observation. Both devices have a color LCD monitor for observation. The observation light source and the photography light source are the same in both devices. Additionally, both devices have an infrared filter for retinal observation that can be removed from the optical path. Thus, the TRC-NW300 is the same device as the TRC-NW200 except for the following modifications:
- TRC-NW300 incorporates the capability to use a USB Memory Card or . an external digital printer
- · TRC-NW300 had a different diaphragm design which allows photography of patients with pupils as small as 3.3 mm
- TRC-NW300 incorporates an 8 mega pixel CCD Camera vs. a 3.15 . mega pixel CCD Camera in the TRC-NW200
- . TRC-NW300 incorporates an automatic focusing function
- TRC-NW300 changes the position of internal fixation targets .
- TRC-NW300 employs an automatic flash (capture) function (when the . camera is in the proper position)
- · TRC-NW300 has the capability to use alpha as well as numeric patient identification characters
{3}------------------------------------------------
- TRC-NW300 includes an automatic small pupil detection system
- TRC-NW300 includes an illumination system for the control panel
- TRC-NW300 utilizes a motorized chinrest mechanism
- TRC-NW300's power supply's normal power has increased from 80 VA . to 100 VA
- TRC-NW300 includes minor changes to the physical dimension of the . device
Each of these modifications was evaluated by Topcon to determine if they could affect the safety or effectiveness of the device and it was determined that they do not. Additionally, the vast majority of these new features have already been cleared by the Agency in the Topcon TRC-NW8F Retinal Camera (K100207). Design verification tests were performed as a result of a risk analysis assessment and demonstrated that the TRC-NW300 met predetermined acceptance criteria.
The TRC-NW300 has the same intended use and similar indications, principles of operation, and technological characteristics as the TRC-NW200 and the TRC-NW8F. The minor differences in the TRC-NW300's technological characteristics do not raise any new questions of safety or effectiveness. Performance data demonstrates that the TRC-NW300 is as safe and effective as the TRC-NW200 and the TRC-NW8F. Thus, the TRC-NW300 is substantially equivalent to its predicate devices.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
February 6, 2013
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Topcon Medical Systems, Inc. % Ms. Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864
Re: K123460
Trade/Device Name: Topcon TRC-NW300 Non-Mydriatic Retinal Camera Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: November 8, 2012 Received: November 9, 2012
Dear Ms. O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. . The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Ms. Maureen O'Connell
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Kesia YAlexander -S
for Malvina B. Eydelman, M.D. Director Division of Opthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use Statement
5123460 510(k) Number (if known):
Device Name: TRC-NW300
Indications for Use:
The TRC-NW300 intended for use in capturing images of the retina and the anterior segment of the eye and presenting the data to the eye care professional, without the use of a mydriatic.
Prescription Use × (Per 21 C.F.R. 801.109) AND/OR
Over-The-Counter Use (Per 21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
CLC
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices 510(k) Number
§ 886.1120 Ophthalmic camera.
(a)
Identification. An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.(b)
Classification. Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.