K Number
K123365

Validate with FDA (Live)

Date Cleared
2013-03-01

(121 days)

Product Code
Regulation Number
884.1720
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

LTF-190-10-3D: This instrument is intended to be used with Olympus video system center, light source, documentation equipment, 3D processor, monitor, hand instruments, electrosurgical unit and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the thoracic and abdominal cavities including female reproductive organs. This instrument must not be used for observation or treatment of the heart and must not contact the heart or any area near the heart. In addition, this instrument must not contact with any device or therapeutic accessory that contacts the heart or any area near the heart.
CV-190: This video system center is intended to be used with Olympus camera heads, endoscopes, light source, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
MAJ-Y0154: This 3D processor is intended to be used with 3D videoscope and video system center for 3D observation.

Device Description

The subject system is intended for endoscopy and endoscopic surgery with three-dimensional view of endoscopic images and the system is mainly composed of following devices: LTF-190-10-3D (ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE), MAJ-Y0154 (3D PROCESSOR), CV-190 (EVIS EXERA III VIDEO SYSTEM CENTER). The ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-3D is a flexible video endoscope used for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs. The subject device, LTF-190-3D is utilized with the MAJ-Y0154, 3D Processor and the CV-190, EVIS EXERA III VIDEO SYSTEM CENTER. The connection of the devices enables the system for endoscopic observation in 3D view of the image. The 3D video observation of this system is implemented with following process. The captured signals from two CCDs which correspond to left and right eye incorporated in the distal end of the subject LTF-190-10-3D are transferred to the MAJ-Y0154 via two CV-190s convert the captured signals into video image signals, and transmits them to the MAJ-Y0154. The MAJ-Y0154 converts the video image signals into 3D video signal and transfers it to the LMD-2451MT, (K113203). On the LMD-2451MT, 3D video image is displayed as passive stereo type which has different polarizing angle in the left and right; therefore, the 3D glasses for LMD-2451MT are required to obtain 3D video image.

AI/ML Overview

The provided text describes a 510(k) summary for Olympus medical devices, specifically an endoscope, a 3D processor, and a video system center. The summary focuses on demonstrating substantial equivalence to predicate devices through design verification and non-clinical testing. It does not include information about AI/ML models, clinical studies, or acceptance criteria related to diagnostic performance or accuracy of an AI model.

Therefore, I cannot fulfill your request for:

  1. A table of acceptance criteria and reported device performance related to an AI model.
  2. Sample size and data provenance for an AI test set.
  3. Number and qualifications of experts for AI ground truth.
  4. Adjudication method for an AI test set.
  5. MRMC comparative effectiveness study results for AI.
  6. Standalone performance results for an AI algorithm.
  7. Type of ground truth for an AI model.
  8. Sample size for an AI training set.
  9. How ground truth for an AI training set was established.

This document describes technical aspects and safety testing for endoscopic hardware, not an AI-powered diagnostic device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "OLYMPUS" in bold, black letters. The letters are large and fill most of the frame. The background is plain white, which makes the text stand out. The font appears to be sans-serif.

510(k) SUMMARY

MAR 0 1 2013

OLYMPUS LTF-190-10-3D, ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE MAJ-Y0154, 3D PROCESSOR OLYMPUS CV-190, EVIS EXERA III VIDEO SYSTEM CENTER

November 12, 2012

1 General Information

  • Applicant: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Establishment Registration No: 8010047
  • l Official Correspondent: Sheri L. Musgnung Regulatory Affairs & Quality Assurance Olympus America Inc. 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 FAX: 484-896-7128 Email: sheri.musgnung@olympus.com
  • Manufacturer: .

(LTF-190-10-3D, MAJ-Y0154) OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant 34-3 Hirai, Hinode-cho, Nishitama-gun, Tokyo, 190-0182, Japan Establishment Registration No.:3003637092

(CV-190) SHIRAKAWA OLYMPUS CO., LTD. 3-1. Aza-Ookamiyama, Ooaza-Odakura, Nishigo-mura, Nishirakawa-gun, Fukushima, Japan 961-8061 Establishment Registration No: 3002808148

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "OLYMPUS" in a bold, sans-serif font. The letters are large and black, creating a strong visual impact. The word is horizontally oriented and appears to be a logo or brand name. The background is plain and white, which makes the text stand out.

2 Device Identification

3D 를 Device Trade Name: OLYMPUS LTF-190-10-3D ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE 3D PROCESSOR MAJ-Y0154 3D PROCESSOR OLYMPUS CV-190 EVIS EXERA III VIDEO SYSTEM CENTER ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE ■ Common Name: 3D PROCESSOR EVIS EXERA III VIDEO SYSTEM CENTER 21 CFR 884.1720 트 Regulation Number: 21 CFR 876.1500 ■ Gynecologic laparoscope and accessories Regulation Name: Endoscope and Accessories 제 ll Regulatory Class: ■ Classification Panel: General and plastic surgery, Obstetrics/Gynecology Gastroenterology and urology ■ FGB, HET, GCJ, and NWB Product Code:

Predicate Device Information 3

Subject Device(Part of this submission)Predicate DevicePD's510(k)No.
OLYMPUS LTF-190-10-3DENDOEYE FLEX 3D DEFLECTABLEVIDEOSCOPE(Hereinafter referred to as LTF-190-10-3D)LTF-Y00093D Laparo-Thoraco VideoscopeK102379
LTF-S190-10ENDOEYE FLEX DEFLECTABLEVIDEOSCOPEK111425
MAJ-Y01543D PROCESSOR(Hereinafter referred to as MAJ-Y0154)MAJ-Y00413D Video MixerK102379
OLYMPUS CV-190EVIS EXERA III VIDEO SYSTEM CENTER(Hereinafter referred to as CV-190)CV-190EVIS EXERA III VIDEO SYSTEMCENTERK112680
OTV-S190VISERA ELITE VIDEO SYSTEM CENTERK111425

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, black letters. The font is sans-serif and the letters are closely spaced together. A thin line is present underneath the word, emphasizing the brand name.

4 Device Description

The subject system is intended for endoscopy and endoscopic surgery with three-dimensional view of endoscopic images and the system is mainly composed of following devices.

  • LTF-190-10-3D (ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE)
  • MAJ-Y0154 (3D PROCESSOR)
  • CV-190 (EVIS EXERA III VIDEO SYSTEM CENTER)

The ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-3D is a flexible video endoscope used for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs.

The subject device, LTF-190-3D is utilized with the MAJ-Y0154, 3D Processor and the CV-190, EVIS EXERA III VIDEO SYSTEM CENTER. The connection of the devices enables the system for endoscopic observation in 3D view of the image.

The 3D video observation of this system is implemented with following process.

The captured signals from two CCDs which correspond to left and right eye incorporated in the distal end of the subject LTF-190-10-3D are transferred to the MAJ-Y0154 via two CV-190s convert the captured signals into video image signals, and transmits them to the MAJ-Y0154. The MAJ-Y0154 converts the video image signals into 3D video signal and transfers it to the LMD-2451MT, (K113203). On the LMD-2451MT, 3D video image is displayed as passive stereo type which has different polarizing angle in the left and right; therefore, the 3D glasses for LMD-2451MT are required to obtain 3D video image.

5 Indications for Use

LTF-190-10-3D

This instrument is intended to be used with Olympus video system center, light source, documentation equipment, 3D processor, monitor, hand instruments, electrosurgical unit and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the thoracic and abdominal cavities including female reproductive organs.

This instrument must not be used for observation or treatment of the heart and must not contact the heart or any area near the heart. In addition, this instrument must not contact with any device or therapeutic accessory that contacts the heart or any area near the heart.

CV-190

This video system center is intended to be used with Olympus camera heads, endoscopes, light source, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation.

MAJ-Y0154

This 3D processor is intended to be used with 3D videoscope and video system center for 3D observation.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, black letters. The font is sans-serif and the letters are closely spaced together. The word is centered and takes up most of the frame.

Comparison of Technological Characteristics હ

The OLYMPUS LTF-190-10-3D, MAJ-Y0154, OLYMPUS CV-190 are basically identical to the predicate device in intended use, and similar in specifications.

The subject LTF-190-10-3D is nearly identical to the predicate LTF-Y0009. It has identical spefications to the predicate device except for, (1) addition information in a memory, (2) NBI observation in 3D view and (3) addition of sterilization method.

Compared to the predicate device, the MAJ-Y0154 incorporates the following features: (1) Communication with the video system center, (2) 3G-SDI image signal output, (3) Recognition of synchronization of input image signals, (4) Menu, (5) Automatic switching of 20/3D view of endoscopic image, (6) Switching of 2D/3D output for recording .

The subject CV-190 has almost the same in design and function to its predicate CV-190 except for following features: (1)Link between the video system centers. (2)Synchronization with the 3D processor MAJ-Y0154 (GenLock). (3)Communication with the 3D processor MAJ-Y0154.

7 Summary of non-clinical testing

Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verifications tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.

The software validation activities were performed in accordance with the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The device software is considered a "Moderate Level of Concern."

The following bench tests were conducted to demonstrate the substaintal equivalence and safety and effectiveness of the subject devices:

  • Mechanical durability testing .
  • Electrical safety testing .
  • . Electromagnetic compatibility testing
  • Thermal safety testing .
  • Bench testing to support promotional claims .

The following standards have been applied to the subject system :

  • · IEC 60601-1
  • · IEC 60601-1-1
  • · IEC 60601-2-18
  • · IEC 60601-1-2
  • · ISO 14971
  • · ISO 10993-7 ·
  • ANSI/AAMI/ISO 11135-1

8 Conclusion

When compared to the predicate device, the OLYMPUS LTF-190-10-3D, MAJ-Y0154 and OLYMPUS CV-190 do not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device and therefore is Substantially Equivalent to the identified predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 1, 2013

OLYMPUS MEDICAL SYSTEMS CORP. % Ms. Sheri L. Musgnung Regulatory Affairs & Quality Assurance Olympus Corporation of the Americas 3500 Corporate Parkway, P.O. Box 610 CENTER VALLEY PA 18034-0610

Re: K123365

Trade/Device Name: OLYMPUS LTF-190-10-3D ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE MAJ-Y0154, 3D PROCESSOR OLYMPUS CV-190, VIDEO SYSTEM CENTER

Regulation Number: 21 CFR& 884.1720 Regulation Name: Gynecologic laparoscope and accessories Regulatory Class: II Product Code: HET, GCJ, NWB, FGB Dated: January 29, 2013 Received: January 30, 2013

Dear Ms. Musgnung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

HerbertPLerner-S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K123365

Device Name:

OLYMPUS LTF-190-10-3D, ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE Indications For Use:

This instrument is intended to be used with Olympus video system center, light source, documentation equipment, 3D processor, monitor, hand instruments, electrosurgical unit and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the thoracic and abdominal cavities including female reproductive organs.

This instrument must not be used for observation or treatment of the heart and must not contact the heart or any area near the heart. In addition, this instrument must not come into contact with any device or therapeutic accessory that contacts the heart or any area near the heart.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 3 of _ 1

Herbert P. Lerner -S

(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K123365 510(k) Number _

{7}------------------------------------------------

Indications for Use

510(k) Number (if known): K123365

Device Name: MAJ-Y0154, 3D PROCESSOR

Indications For Use:

This 3D processor is intended to be used with 3D videoscope and video system center for 3D observation

Prescription Use______________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

2

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner -S

(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K123365 510(k) Number

{8}------------------------------------------------

Indications for Use

K123365 510(k) Number (if known): Device Name: OLYMPUS CV-190, VIDEO SYSTEM CENTER Indications For Use:

This video system center is intended to be used with Olympus camera heads, endoscopes, light source, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation.

Prescription Use_ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use_ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Herbert Prefect Prefection (ODE)

(Division Sign-Off) Division of Reproductive, Gastro-Renal, and 30 30 _3 Urological Devices K123365 510(k) Number

§ 884.1720 Gynecologic laparoscope and accessories.

(a)
Identification. A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for gynecologic laparoscope accessories that are not part of a specialized instrument or device delivery system, do not have adapters, connector channels, or do not have portals for electrosurgical, lasers, or other power sources. Such gynecologic laparosope accessory instruments include: the lens cleaning brush, biopsy brush, clip applier (without clips), applicator, cannula (without trocar or valves), ligature carrier/needle holder, clamp/hemostat/grasper, curette, instrument guide, ligature passing and knotting instrument, suture needle (without suture), retractor, mechanical (noninflatable), snare, stylet, forceps, dissector, mechanical (noninflatable), scissors, and suction/irrigation probe. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.