(84 days)
Not Found
No
The document describes a surgical system for cutting, drilling, and driving screws, with no mention of AI or ML capabilities. The performance studies rely on engineering rationales and predicate device testing, not data-driven AI/ML model validation.
No
The device is used for surgical procedures involving cutting, removal, drilling, and sawing, or for drilling, tapping, and driving screws. These functions are primarily for surgical intervention rather than for therapeutic treatment.
No
Explanation: The device is described as an IPC System intended for incision, cutting, removal, drilling, and sawing of tissue and bone, and a POWEREASE System for drilling, tapping, and driving screws. These are all surgical actions, not diagnostic ones.
No
The device description explicitly details physical instruments (taps, drill bits, screwdrivers, etc.) and mentions compatibility with implant systems, indicating it is a hardware-based surgical system, not software only.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly describes the device as being used for surgical procedures involving the cutting, removal, drilling, and sawing of tissue and bone. This is a direct surgical intervention on the patient.
- Device Description: The device description details surgical instruments like taps, drill bits, screwdrivers, etc., used in surgical procedures.
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens (like blood, urine, tissue samples) in vitro (outside the body) to provide information about a patient's health status, diagnosis, or treatment.
IVD devices are used to perform tests on samples taken from the body, not to directly perform surgical actions on the body itself.
N/A
Intended Use / Indications for Use
IPC® System is indicated for the incision/cutting, removal, drilling and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.
The IPC® POWEREASE™ System is indicated for drilling, tapping and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in placement or cutting of screws, posts and rods.
Product codes (comma separated list FDA assigned to the subject device)
HBE, HWE, GWF
Device Description
The POWEREASE™ System Working Ends consists of instruments such as taps, drill bits, screwdrivers, post cutter, set screw break-off tool, reduction nut driver and sleeves. The working ends have a manual alternative. The working ends, listed above, are compatible with the CD HORIZON® SOLERA® and the TSRH® 3Dx Spinal System implants. Of the working ends, only the taps, screwdrivers, drill bits, and sleeves are also compatible with Medtronic's NIM-ECLIPSE® Spinal System.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Spinal
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Engineering rationales, based on engineering theoretical analysis, were completed using testing that had been performed on the predicate device. The predicate testing was in accordance with IEC 60601 for Medical Electrical Equipment. Information and data to support these engineering rationales, as well as surgeon validations and internal verification activities, were provided to demonstrate that the subject devices are substantially equivalent to the predicate devices.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
POWEREASE™ System K111520 (S.E. 10/26/2011)
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 882.4310 Powered simple cranial drills, burrs, trephines, and their accessories.
(a)
Identification. Powered simple cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments used on a patient's skull. The instruments are used with a power source but do not have a clutch mechanism to disengage the tip after penetrating the skull.(b)
Classification. Class II (performance standards).
0
510(k) SUMMARY MEDTRONIC Sofamor Danek POWEREASE™ System October 2012
JAN 1 1 2013
I. Medtronic Xomed, Inc. (PowerEase) Company: Medtronic Surgical Technologies 6743 Southpoint Drive North Jacksonville, Florida, 32216, USA (904) 296-9600
Medtronic Sofamor Danek, USA Inc. (Working Ends) 1800 Pyramid Place Memphis, Tennessee 38132 (901) 396-3133
Julie Bassett Principal Regulatory Affairs Specialist Telephone: (901) 399-3248 Fax: (901) 346-9738
IPC® System IPC® POWEREASE™ System POWEREASE™ System POWEREASE™ System Working Ends
Powered simple cranial drills, burrs, trephines, and their accessories; Surgical instrument motors and accessories/attachments; Evoked Response Electrical Stimulator
Class II
HBE, HWE, GWF
Product Code:
Class:
V. Description:
The POWEREASE™ System Working Ends consists of instruments such as taps, drill bits, screwdrivers, post cutter, set screw break-off tool, reduction nut driver and sleeves. The working ends have a manual alternative. The working ends, listed above, are compatible with the CD HORIZON® SOLERA® and the TSRH® 3Dx
II. Contact:
III. Proprietary Trade Name:
IV. Common &Classification Names:
1
Spinal System implants. Of the working ends, only the taps, screwdrivers, drill bits, and sleeves are also compatible with Medtronic's NIM-ECLIPSE® Spinal System.
VI. Indications for Use:
The IPC® System is indicated for the incision/cutting, removal, drilling and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Cranial, Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.
The IPC® POWEREASE™ System is indicated for drilling, tapping and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in the placement or cutting of screws, posts and rods.
Summary of the Technological Characteristics: VII.
The purpose of this application for Premarket Notification is to expand the previously cleared [POWEREASE™ System K111520 (S.E. 10/26/2011)] indications to include placement of not only screws, but rods as well. The indications are changing to include the intended use of the nut reduction driver, as well as, update the indications since screws, posts, and rods may all be "cut" or "placed" during the procedure.
In addition, two modified Taps and a Reduction Nut Driver are being added to the POWEREASETM System.
The instrument modifications detailed in this submission have no impact on the technological characteristic of the existing instruments. The working end Taps are intended for tapping during spinal surgery, including both open and minimally invasive procedures. The working ends are used to facilitate the placement of the
2
rods. Like the predicate POWEREASE™ System instruments, the subject instruments are manufactured from stainless steel.
IX. Discussion of Non-Clinical Testing:
Engineering rationales, based on engineering theoretical analysis, were completed using testing that had been performed on the predicate device. The predicate testing was in accordance with IEC 60601 for Medical Electrical Equipment. Information and data to support these engineering rationales, as well as surgeon validations and internal verification activities, were provided to demonstrate that the subject devices are substantially equivalent to the predicate devices.
Conclusion: X. .
Based on the supporting documentation provided in this premarket notification, Medtronic believes the subject devices demonstrate substantial equivalence to the predicate POWEREASE™ System K111520 (S.E. 10/26/2011).
3
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling a caduceus, with three wavy lines representing the staff and a single line representing the snake.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Medtronic Sofamor Danek USA, Incorporated % Ms. Julie Bassett Principal Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132
January 11, 2013
Re: K123270
Trade/Device Name: POWEREASE™ System Regulation Number: 21 CFR 882.4310 Regulation Name: Powered simple cranial drills, burrs, trephines, and their accessories Regulatory Class: Class II Product Code: HBE, HWE, GWF Dated: October 18, 2012 Received: October 19, 2012
Dear Ms. Bassett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
4
Page 2 – Ms. Julie Bassett
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
510(k) Number (if known): _K123270
Device Name: POWEREASE™ System
Indications for Use:
IPC® System is indicated for the incision/cutting, removal, drilling and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.
The IPC® POWEREASE™ System is indicated for drilling, tapping and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in placement or cutting of screws, posts and rods.
Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Peter D. Rumm -S
(Division Sign-Off)
Division of Surgical Devices
K123270 510(k) Number_