(101 days)
The NeuroPace® Burr Hole Cover is intended for use following cranial surgery to cover a 14 mm burr hole. Secondarily, the NeuroPace Burr Hole Cover also can be used to support a 1.3 mm indwelling lead.
The NeuroPace® Burr Hole Cover (model 8110) (also referred to as "the device") includes a base (also referred to as a "retainer") that is screwed to the cranium (skull) using bone screws. The cap is an assembly comprised of a cap and a gasket. The cap is pressed into the base covering the opening in the base and securing a single 1.3 mm lead. The Burr Hole Cover requires three bone screws (1.5 to 1.8 mm).
The screws and driver are not included in the device's packaging. The contents of the unopened, undamaged package are sterile and non-pyrogenic.
The Burr Hole Cover is provided sterile (for single-use only) and consists of one model / size. The device is meant to be a permanent implant.
The Burr Hole Cover incorporates materials commonly found in medical devices that are known to be biocompatible. The cap and base are made from a synthetic polymer and the gasket is made from silicone.
The Burr Hole Cover is MR/CT scanning compatible and is sterilized using ethylene oxide gas (EtO).
The Burr Hole Cover does not include any software, incorporate any medicinal substances or contain any color additives.
The NeuroPace® Burr Hole Cover (model 8110) is a medical device designed to cover a 14 mm burr hole following cranial surgery and to support a 1.3 mm indwelling lead. The device's acceptance criteria and performance were assessed through non-clinical testing rather than clinical studies with human subjects.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document describes performance testing conducted to demonstrate compliance with product requirements. The acceptance criteria are implicit in the successful completion of these tests, indicating the device met the specified engineering and safety standards. Since the document does not present quantitative metrics for "performance" as one might expect from a diagnostic AI, the performance is stated as "compliance" or "demonstrated."
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Dimensional / Geometry | Compliant (demonstrated through performance testing) |
| Device Usability | Compliant (demonstrated through performance testing) |
| Lead Compatibility | Compliant (demonstrated through performance testing) |
| Lead Movement | Compliant (demonstrated through performance testing) |
| Fixation / Retention | Compliant (demonstrated through performance testing) |
| Cranial Rigidity | Compliant (demonstrated through performance testing) |
| Biocompatibility | Demonstrated in accordance with ISO 10993-1 |
| Sterility | Compliant (demonstrated through performance testing) |
| Pyrogenicity | Compliant (demonstrated through performance testing) |
| MR/CT Scanning Compatibility | Compliant (demonstrated through performance testing) |
| Packaging / Shelf Life | Compliant (demonstrated through performance testing) |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "No clinical testing was deemed necessary to support substantial equivalence." This means there was no human test set in the traditional sense of a clinical study. The "test set" consisted of physical devices undergoing various laboratory and engineering tests. The data provenance is from non-clinical laboratory testing conducted by NeuroPace Inc.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Since no clinical testing was performed, there were no human experts establishing ground truth for a patient-based test set. The "ground truth" here is based on engineering specifications, regulatory standards, and established material science properties. The experts involved would have been the engineers, material scientists, and regulatory affairs personnel responsible for designing, testing, and submitting the device, operating under the established guidelines and standards (e.g., ISO 10993-1).
4. Adjudication Method for the Test Set
Not applicable. As no clinical test set was used, there was no need for an adjudication method involving experts making clinical decisions.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
Not applicable. This device is a passive implantable medical device, not an AI or diagnostic software tool. Therefore, no MRMC study or assessment of human reader improvement with AI assistance was conducted or is relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Not applicable. The device is a physical implant, not an algorithm or AI.
7. The Type of Ground Truth Used
The ground truth used for this device's evaluation was based on:
- Engineering Specifications and Design Requirements: The device was designed to meet specific dimensions, material properties, and functional criteria (e.g., lead compatibility, fixation strength).
- Regulatory Standards: Compliance with international and national standards for medical devices, such as ISO 10993-1 for biocompatibility, ISO 11607-1 for packaging, and various ASTM and AAMI standards for material testing and sterilization.
- Predicate Device Characteristics: The device's performance was also assessed against the known characteristics and approved uses of the predicate device (Image-Guided Neurologics Burr Hole Cover, K001096) to demonstrate substantial equivalence.
8. The Sample Size for the Training Set
Not applicable. This is not an AI or machine learning device, so there is no concept of a "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this type of device.
In summary, the NeuroPace® Burr Hole Cover is a physical implantable device, and its acceptance criteria and "performance" were established and demonstrated through rigorous non-clinical laboratory testing against established engineering specifications and regulatory standards, rather than through clinical trials, expert consensus, or AI performance metrics. Its substantial equivalence to a legally marketed predicate device was a key aspect of its clearance.
{0}------------------------------------------------
JAN 1 8 2013
ﮩ ﮩ
Premarket Notification [510(k)] Summary
| Category | Comments |
|---|---|
| Date Summary Prepared: | October 8, 2012 |
| Applicant: | NeuroPace Inc.455 N. Bernardo AvenueMountain View, CA 94043USATel: 650-237-2700.FAX: 815-352-0788www.neuropace.com |
| Applicant's Contact Information: | Isabella AbatiVP, Regulatory Affairs |
| Device Trade/ Proprietary Name: | NeuroPace® Burr Hole Covermodel 8110 |
| Device Common Name: | Cover, Burr Hole |
| Device Classification: | II |
| Device Classification Name: | Burr Hole Cover (21 CFR 882.5250, Product Code GXR) |
Substantial Equivalence Device Information
| Predicate Device(s): | Burr Hole Cover |
|---|---|
| Predicate Device Manufacturers: | Image-Guided Neurologics, Inc. (Medtronic) |
| 510(k) Number: | K001096 |
| Predicate Device Common Name: | Cover, Burr Hole |
| Predicate Device Classification Name & Citation: | Burr Hole Cover (21 CFR 882.5250, Product Code GXR) |
510(k) Summary - Predicate Device Information
Description of the Device
The NeuroPace® Burr Hole Cover (model 8110) (also referred to as "the device") includes a base (also referred to as a "retainer") that is screwed to the cranium (skull) using bone screws. The cap is an assembly comprised of a cap and a gasket. The cap is pressed into the base covering the opening in the base and securing a single 1.3 mm lead. The Burr Hole Cover requires three bone screws (1.5 to 1.8 mm).
The screws and driver are not included in the device's packaging. The contents of the unopened, undamaged package are sterile and non-pyrogenic.
{1}------------------------------------------------
Device Characteristics
The Burr Hole Cover is provided sterile (for single-use only) and consists of one model / size. The device is meant to be a permanent implant.
The Burr Hole Cover incorporates materials commonly found in medical devices that are known to be biocompatible. The cap and base are made from a synthetic polymer and the gasket is made from silicone.
The Burr Hole Cover is MR/CT scanning compatible and is sterilized using ethylene oxide gas (EtO).
The Burr Hole Cover does not include any software, incorporate any medicinal substances or contain any color additives.
Intended Use
The NeuroPace® Burr Hole Cover is intended for use following cranial surgery to cover a 14 mm burr hole. Secondarily, the NeuroPace Burr Hole Cover also can be used to support a 1.3 mm indwelling lead.
The predicate device is also intended for use as an implantable 14 mm burr hole cover following cranial surgery and can also be used to support indwelling catheters and electrodes (leads). (The intended use of the subject Burr Hole Cover is encompassed within the intended use of the predicate device and is narrower in scope than the intended use of the predicate device. Therefore, no safety or efficacy issues are presented by the difference between the intended use of the subject Burr Hole Cover and the intended use of the predicate device.
Comparison to Predicate Device
The Burr Hole Cover is substantially equivalent in its intended use, technology, target population, materials, and anatomical site to the Image-Guided Neurologics Burr Hole Cover (K001096). Image-Guided Neurologics (a.k.a. IGN) was acquired by Medtronic, Inc. in 2005. A side-by-side comparison of the NeuroPace Burr Hole Cover to the predicate device follows.
{2}------------------------------------------------
| Description | Subject Device:NeuroPace, Inc.Burr Hole Cover | Predicate Device:Image-Guided Neurologics, Inc.Burr Hole Cover | |
|---|---|---|---|
| Intended Use | Intended to cover a 14 mm burrhole following cranial surgery.Secondarily, the NeuroPaceBurr Hole Cover can be used tosupport 1.3 mm indwellingleads. | Intended to be used as animplantable 14 mm burr holecover following cranial surgery.The device can also be used tosupport indwelling catheters andelectrodes. The device is MR/CTscanning compatible, and isprovided sterile, and is for onetime use only. | |
| Device Classification Nameand Product Code | Burr Hole Cover (21 CFR882.5250, Product Code GXR) | Same | |
| Environment of Use andPrincipal Operator | Hospital/Healthcare FacilityNeurosurgeon | Same | |
| Target Population | To be used following cranialsurgery in patients to cover ahole drilled into the cranium andto support a lead (electrode) | To be used following cranialsurgery in patients to cover a holedrilled into the cranium and tosupport electrodes or catheters | |
| Anatomical Site | Permanent implant in cranium(skull) | Same | |
| Method of Sterilization | Eto (Ethylene Oxide) | Same | |
| Labeled as non-pyrogenic? | Yes | Same | |
| Single Use? | Yes | Same | |
| Biocompatible? | Yes | Same | |
| Performance Testing: | |||
| Method of Fixation toCranium | Titanium Screws | Same | |
| Lead (Electrode)Support? | Yes | Same | |
| MR/CT Compatible? | Yes | Same |
510(k) Summary - Device Comparison
The NeuroPace® Burr Hole Cover materials differ from the predicate device. The NeuroPace device is composed of a synthetic polymer and silicone material. The synthetic polymer material is used to make a commercially available cranial permanent implant [Kelyniam Custom Skull Implant (K103582)]. The silicone material has a long standing history of use in implantable medical devices. Furthermore, biocompatibility of the Burr Hole Cover has been demonstrated in accordance with ISO 10993-1.
{3}------------------------------------------------
Technological characteristics, such as materials and design (for example dimensions) are demonstrated by performance testing to not raise new issues of safety or effectiveness; thus demonstrating that the NeuroPace® Burr Hole Cover is substantially equivalent to the legally marketed predicate device.
Summary of Supporting Data
Performance testing was conducted on the Burr Hole Cover to demonstrate compliance with the product requirements specification either by design analysis or functional testing. The following product requirements are supported: dimensional / geometry, device usability, lead compatibility, lead movement. fixation / retention, cranial rigidity, biocompatibility, sterility, pyrogenicity, MR/CT scanning compatibility and packaging / shelf life. The following guidances and standards were used for testing.
- · ANSI/AAMI/ISO 10993-1:2009, Biological evaluation of medical devices Part 1; Evaluation and testing.
- · ANSI/AAMI/ISO 10993-7:2008, Biological evaluation of medical devices, Part 7: Ethylene oxide sterilization residuals
- · ANSI/AAMI/ISO 11607-1: 2006 Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging
- ANSI/AAMI/ISO 11135:2007, Sterilization of health care products Ethylene oxide - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices.
- ASTM F88 Standard Test Methods for Seal Strength of Flexible Barrier Materials
- · ASTM F-1980, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- · ASTM F2096-04, Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test)
- · ASTM · D 4169-09, Standard Practice for Performance Testing of Shipping Containers and Systems
- · ST67:2003/(R) 2008, Sterilization of health care products Requirements for products labeled 'sterile', 1st Ed.
- · United States Pharmacopoeia 28 & NF 23, Bacterial Endotoxins Test <85>
- United States Pharmacopoeia 28 & NF 23, Transfusion and Infusion Assemblies and Similar Medical Devices <161>
- · FDA Guidance, A Primer on Medical Device Interactions with Magnetic Resonance Imaging Systems, Draft Document: 1997.
- · AAMI TIR 28:2009 Product adoption and process equivalence for ethylene oxide sterilization
{4}------------------------------------------------
- Guidance for Industry Medical Device Tracking; Guidance for Industry and FDA Staff - Labeling
- · ASTM D-4332-01, Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing
- · FDA Guideline On Validation Of The Limulus Amebocyte Lysate Test As An End-Product Endotoxin Test For Human And Animal Parenteral Drugs, Biological Products, And Medical Devices, December 1987
No clinical testing was deemed necessary to support substantial equivalence.
This 510(k) notification for the Burr Hole Cover concludes that the device is considered to be substantially equivalent to the legally-marketed predicate device (as shown in the 510(k) Summary – Device Comparison Table). The successful completion of the performance testing further supports the subject NeuroPace® Burr Hole Cover's substantial equivalence to the predicate device. No issues of safety or effectiveness are raised by the NeuroPace ® Burr Hole Cover.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
January 18, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
NeuroPace. Inc Ms. Isabella Abati Vice President of Regulatory Affairs 455 N. Bernardo Avenue Mountain View, CA 94043
Re: K123163
Trade Name: NeuroPace® Burr Hole Cover (Model 8110) Regulation Number: 21 CFR 882.5250 Regulation Name: Burr Hole Covers Regulatory Class: Class II Product Code: GXR Dated: December 5, 2012 Received: December 6, 2012
Dear Ms. Abati:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Ms. Isabella Abati
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use Statement
510(k) Number (if known): Not Known
NeuroPace® Burr Hole Cover Device Name:
Indications for Use:
The NeuroPace® Burr Hole Cover is intended for use following cranial surgery to cover a 14 mm burr hole. Secondarily, the NeuroPace Burr Hole Cover also can be used to support a 1.3 mm indwelling lead.
Prescription Use × (21 CRF 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sui Nguyen
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
K123163 510(k) Number.
§ 882.5250 Burr hole cover.
(a)
Identification. A burr hole cover is a plastic or metal device used to cover or plug holes drilled into the skull during surgery and to reattach cranial bone removed during surgery.(b)
Classification. Class II (performance standards).