K Number
K123116
Date Cleared
2013-02-11

(131 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Chlorinated Powder Free Nitrile Examination Gloves (Black Colour) is a single use device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between the healthcare personnel and the patient.

Device Description

The Chlorinated Powder Free Nitrile Examination Gloves (Black Colour) meets all the Nitrile requirements of ASTM Specification D6319-10 Standard Specification for Examination Gloves for Medical Application.

AI/ML Overview

The provided document pertains to a 510(k) premarket notification for "Chlorinated Powder Free Nitrile Examination Gloves (Black Colour)". This is a medical device, and the submission focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than a novel AI/software-based device.

Therefore, many of the requested criteria (e.g., sample size for test set, data provenance, expert qualifications, adjudication method, MRMC study, standalone performance, training set details, ground truth for training set) are not applicable to this type of medical glove submission. These criteria are typically relevant for AI/ML-based diagnostic or prognostic devices, clinical decision support systems, or other software as a medical device (SaMD).

However, I can extract the acceptance criteria and device performance as reported in the summary:

1. A table of acceptance criteria and the reported device performance

CharacteristicsStandards (Acceptance Criteria)Device Performance
DimensionsASTM D6319-10Meets standard requirements
Physical PropertiesASTM D6319-10Meets standard requirements
ThicknessASTM D6319-10Meets standard requirements
Biocompatibility (Irritation)ISO 10993-10:2002/Amd 1:2006(E) Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivityPass (Not a primary skin irritant)
Biocompatibility (Sensitization)ISO 10993-10:2002/Amd 1:2006(E) Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivityPass (Not a contact sensitizer)
Watertight (1000ml)21 CFR 800.20Pass
Residual Powder TestBelow 2mg/glove (Implied acceptance based on predicate)Passes - Below 2mg/glove
Freedom from Holes21 CFR 800.20: Gloves Free of Holes at quality level of AQL 1.5 (AQL 2.5 required per standard ASTM D6319-10)Meets Requirements per 21CFR800.20

Additionally, the "Substantial Equivalence Comparison" table provides more detailed performance characteristics, with the acceptance criteria being conformance to ASTM D6319-10 or specific thresholds:

Characteristic and parametersAcceptance Criteria (from predicate or ASTM D6319-10 references)Device Performance (Worldmed Manufacturing Sdn. Bhd.)
Width (Size Large)Meets ASTM D6319-10 (e.g., XS - 70 ± 10, S - 80 ± 10, M - 95 ± 10, L - 110 ± 10, XL - 120 ± 10)Meets ASTM D6319-10 (XS - 70 ± 10, S - 80 ± 10, M - 95 ± 10, L - 110 ± 10, XL-120 ± 10)
Overall lengthLength ≥ 240mm (Meets ASTM D6319-10)Length ≥ 240mm
Palm thicknessMin 0.05mm (Meets ASTM D6319-10)Min 0.05mm
Finger thicknessMin 0.05mm (Meets ASTM D6319-10)Min 0.05mm
Tensile Strength before aging min.14.0 MPa (Meets ASTM D6319-10)14.0 MPa
Tensile Strength after aging min14.0 MPa (Meets ASTM D6319-10)14.0 MPa
Ultimate elongation before aging min500% (Meets ASTM D6319-10)500%
Ultimate elongation after aging400% (Meets ASTM D6319-10)400%
Meets BiocompatibilityYesYes
Skin irritationPassesPasses
Dermal sensitizationPassesPasses
Residual powder testBelow 2mg/glove (Implied based on predicate)Passes - Below 2mg/glove
Freedom from HolesAQL 1.5 (as per 21CFR800.20, which is stricter than ASTM D6319-10's AQL 2.5)AQL 1.5

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the sample sizes for each test performed (e.g., for dimensions, physical properties, thickness, biocompatibility, watertightness, or tensile strength tests). The testing is generally understood to be performed on manufactured batches of gloves. The manufacturing facility is in Malaysia. The studies are non-clinical, likely prospective testing of manufactured samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This is for a physical medical device (examination gloves), not an AI/software device requiring expert interpretation of results for ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/software-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the device's performance is established by standardized material and performance tests as defined by ASTM D6319-10 and 21 CFR 800.20. For biocompatibility, it refers to the results of standard irritation and sensitization tests.

8. The sample size for the training set
Not applicable. This device does not involve a "training set" in the context of AI/ML.

9. How the ground truth for the training set was established
Not applicable.

{0}------------------------------------------------

K123116

Chlorinated Powder Free Nitrile Examination Gloves (Black Colour)

Attachment 15

510(k) SUMMARY

FEB 1 1 2013

Submitter : 1.0

Name:Worldmed Manufacturing Sdn. Bhd.
Address:Lot 18873, Jalan Perusahaan 3,Kamunting Industrial Estate,34600 Taiping, Perak,Malaysia.
Phone No.:605-892 5555
Fax No .:605-829 5590
Contact Person:Ooi Loon Seng (Madam)

September 5, 2012 Date of Preparation :

2.0 Name of the Device

Chlorinated Powder Free Nitrile Examination Gloves (Black Colour)

Common Name: Patient Examination Gloves
-------------------------------------------
Classification Name :Patient Examination Gloves (CLASS I - 21CFR 880.6250)
------------------------------------------------------------------------------

510(K) Number

Identification of The Legally Marketed Devices That equivalency is claimed: 3.0

Primary Predicate:

RS BLACK Black Nitrile Medical Examination Gloves (Powder Free) Company : Riverstone Resources Sdn. Bhd. 510(K) : K112924

Additional Predicate:

Non-Sterile, Powder-Free, Black Nitrile Examination Glove Company : YTY Industry (Manjung) Sdn. Bhd. 510(K) : K061553

Description of the Device: 4.0

The Chlorinated Powder Free Nitrile Examination Gloves (Black Colour) meets all the Nitrile requirements of ASTM Specification D6319-10 Standard Specification for Examination Gloves for Medical Application.

5.0 Intended Use of the Device

The Chlorinated Powder Free Nitrile Examination Gloves (Black Colour) is a single use disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between the healthcare personnel and the patient.

{1}------------------------------------------------

Summary of the Technological Characteristics of the Device: 6.0

The Chlorinated Powder Free Nitrile Examination Gloves (Black Colour) are summarized with the following technological characteristics compared to ASTM Specification D6319-10 Standard Specification for Nitrile Examination Gloves for Medical Application or equivalent standards.

CHARACTERISTICSSTANDARDSDEVICE PERFORMANCE
DimensionsASTM D6319-10Meets standard requirements
Physical PropertiesASTM D6319-10Meets standard requirements
ThicknessASTM D6319-10Meets standard requirements
BiocompatibilityISO 10993-10:2002/Amd1:2006(E) Biologicalevaluation of medicaldevices - Part 10: Tests forirritation and delayed-typehypersensitivity-Amendment 1:2006-07-15ISO 10993-10:2002/Amd1:2006(E) Biologicalevaluation of medicaldevices - Part 10: Tests forirritation and delayed-typehypersensitivity-Amendment 1:2006-07-15Pass (Not a primary skinirritant)Pass (Not a contact sensitizer)
Watertight (1000ml)21 CFR 800.20Pass

7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data

The performance test data of non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above.

Substantial Equivalent Based on Assessment of Clinical Performance Data 8.0

Clinical data is not needed for gloves or most devices cleared by the 510(k) process.

{2}------------------------------------------------

Attachment 15

9.0 Substantial Equivalence Comparison

.

CharacteristicandparametersWorldmedManufacturingSdn. Bhd.RiverstoneResources Sdn.Bhd.K112924YTY Industry(Manjung) Sdn.Bhd.K061553SubstantialEquivalence(SE)
Product CodeLZALZALZA
Intended useThe ChlorinatedPowder Free NitrileExaminationGloves (BlackColour) is a singleuse deviceintended formedical purposesthat is worn on thehand of healthcareand similar topersonnel preventcontaminationbetween thehealthcarepersonnel and thepatient.RS BLACK BlackNitrile MedicalExaminationGloves PowderFree (Non-Sterile)is a disposabledevice intended formedical purposesthat is worn on theexaminer's hand orfinger to preventcontaminationbetween patientand examiner. Thedevice is for over-the-counter use.The Non-Sterile,Powder-Free, BlackNitrile ExaminationGlove is a disposabledevice intended formedical purpose thatis worn on theexaminer's hand orfinger to preventcontaminationbetween patient andexaminer.SE
Width(Size Large)Meets ASTM D6319-10 (mm)XS - 70 ± 10S - 80 ± 10M - 95 ± 10L - 110 ± 10XL-120 ± 10Meets ASTM D6319-10Palm Width (mm)Size XS: 73-78S : 83-88M : 93-98L : 103-107SE
Overall lengthLength ≥ 240mmMeets ASTM D6319-10240mm minimum forall sizesSE
Palm thicknessMin 0.05mmMeets ASTM D6319-10Min 0.08SE
FingerthicknessMin 0.05mmMeets ASTM D6319-10Min 0.08
TensileStrength beforeaging min.14.0 MPaMeets ASTM D6319-1015-21 MPa
TensileStrengthafteraging min14.0 MPaMeets ASTM D6319-1014-22 MPa
Ultimateelongationbeforeagingmin500%Meets ASTM D6319-10550 % - 630 %
Ultimateelongation afteraging400%Meets ASTM D6319-10520 % - 610 %
MeetsBiocompatibilityYesYesYesSE
Skin irritationPassesPassesPasses
DermalsensitizationPassesPassesPasses
Residualpowder testPasses - Below2mg/gloveMeets ASTM D6319-10Below 2 mg/glove
Freedom fromHolesMeetsRequirements per21CFR800.20:Gloves Free ofHoles at qualitylevel of AQL 1.5(AQL 2.5 requiredper standard)Meets ASTM D6319-10Multiple Normal GIIAQL=2.5Yes, SE
MaterialsNitrile LatexNitrile Latexcompound-Yes,SubstantialEquivalence

.

{3}------------------------------------------------

Chlorinated Powder Free Nitrile Examination Gloves (Black Colour)

Attachment 15

K123116

10.0 Conclusion

Chlorinated Powder Free Nitrile Examination Gloves (Black Colour) will perform according to the gloves performance standards referenced in section 6.0 above and meets ASTM standards and FDA requirements for water leak test on pinhole AQL. Consequently, the device is substantially equivalent to currently marketed devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, sans-serif font. The text is centered in the image. The words are stacked on top of each other.

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Inside the circle is a stylized symbol that resembles a person embracing another person, or a symbol representing care and support. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 11, 2013

Ms. Ooi Loon Seng Regulatory Affairs Manager Worldmed Manufacturing Sdn. Bhd. Lot 18873, Jalan Perusahaan 3, Kamunting Industrial Estate Kamunting Perak, Malaysia 34600

Re: K123116

Trade/Device Name: Chlorinated Powder Free Nitrile Examination Gloves (Black Colour) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: 1 Product Code: LZA Dated: September 24, 2012 Received: November 13, 2012

Dear Ms. Seng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

hoi

heth
Harshit.D.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Chlorinated Powder Free Nitrile Examination Gloves (Black Colour)

:

:

Attachment 2

INDICATIONS FOR USE

Applicant

WORLDMED MANUFACTURING SDN. BHD. Lot 18873, Jalan Perusahaan 3, Kamunting Industrial Estate, 34600 Taiping, Perak, Malaysia.

510(k) Number (if known)

K123116

Device Name

CHLORINATED POWDER FREE NITRILE EXAMINATION GLOVES (BLACK COLOUR)

Indications For Use

Chlorinated Powder Free Nitrile Examination Gloves (Black Colour) is a single use device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between the healthcare personnel and the patient.

Concurrence of CDRH Office of Device Evaluation (ODE )

Prescription UsePer 21 CFR 801.109OR Over-The-CounterX
Elizabeth F. Claverie
2013.02.11 13:32:02-05'00'
(Division Sign-Off)Division of Anesthesiology, General HospitalInfection Control, Dental Devices
510(k) Number:K123116Page 1 of 1

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.