K Number
K123115
Date Cleared
2013-05-20

(229 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgical Face Mask of different colors (Blue, White, and Green) is a device intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operation room personnel from transfer of bacteria, body fluid and particulate material. This Surgical Face Mask is non-sterilized and disposable.

Device Description

Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop is flat pleated 3-ply (at least) masks with an inner and outer layer (spunbonded polypropylene) that sandwich a melt blown polypropylene filter material, also with elastic loops. The nosepiece for all Acme Filter Mask Inc. Surgical Face Masks with Ear-Loop are malleable aluminum wire. All the materials used in the construction of the Acme Filter Mask Inc. Surgical Face Masks with Ear-Loop are being used in currently marked devices.

AI/ML Overview

The Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop (K123115) demonstrates substantial equivalence to the predicate device, Modern Healthcare Corp. Surgical Face Mask (K063043), through a series of non-clinical performance tests.

1. Table of Acceptance Criteria and Reported Device Performance:

ItemAcceptance Criteria (Predicate Device K063043)Reported Device Performance (ACME Filter Mask Inc. K123115)
Similarity
Fluid ResistanceFluid ResistanceFluid Resistance
Flammability ClassClass I (No Flame Spread)Class I (No Flame Spread)
Regulatory ClassClass II (ASTM2100-04 Low Barrier)Class II
BFE (%) (Bacterial Filtration Efficiency)Higher than 99%Higher than 99.9%
Difference
Type (of mask fastening)Tie-on and Ear-loop (Green, White, Blue, Pink)Ear-Loop (Green, White, Blue)
Delta-P (Air Exchange Pressure)Average 2.6 $mmH_2O/cm^2$Average 3.33 $mmH_2O/cm^2$ for Air Exchange Pressure
Particulate Filtration Efficiency Performance (%)Average 96.8% at 0.1 micron (for general particulate filtration)Average 94.79% for Solid Aerosol Filtration Efficiency; More than 99.5% for Viral Filtration Efficiency

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample sizes used for each specific test (e.g., number of masks tested for BFE, Delta-P, etc.). The provenance of the data is from non-clinical bench testing performed by Acme Filter Mask Inc. The country of origin for the testing is implied to be Taiwan, R.O.C. where the submitter is located. The study is a retrospective comparison against established standards and a predicate device.

3. Number of Experts and Qualifications for Ground Truth:

This document describes a non-clinical device safety and effectiveness evaluation based on standardized testing. There is no mention of experts being used to establish a "ground truth" in the context of clinical interpretation, as this is a physical product performance assessment. The ground truth for these tests is defined by the technical specifications and performance criteria outlined in the referenced ASTM, MIL, and CFR standards.

4. Adjudication Method for the Test Set:

Not applicable. This is a non-clinical performance study against predefined standards and a predicate device's performance, not a study involving human interpretation or adjudication of results in a clinical setting.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not done. This device is a surgical face mask, and its evaluation focuses on physical performance characteristics rather than diagnostic or interpretative accuracy that would involve human readers.

6. Standalone Performance:

Yes, a standalone performance assessment was done. The reported performance metrics (BFE, Delta-P, Particulate Filtration Efficiency, Fluid Resistance, Flammability, Biocompatibility) are direct measurements of the device's physical properties and its ability to meet specific standards, independent of human interaction during use. These are "algorithm only" in the sense that they represent the intrinsic performance of the manufactured product.

7. Type of Ground Truth Used:

The ground truth used for these non-clinical tests is based on standardized test methods and acceptance criteria. These include:

  • ASTM F1862-05 for Fluid Resistance
  • ASTM F2101-01 for Bacterial Filtration Efficiency (BFE)
  • ASTM F1215 for Particulate Filtration Efficiency (latex Particle challenge)
  • MIL M 36954 C for Differential Pressure (Delta-P)
  • 16CFR 1610 for Flammability
  • ISO 10993-5 /-10 for Biocompatibility

The "truth" is whether the device's performance meets or exceeds the specified thresholds in these well-established industry standards and compares favorably to the predicate device.

8. Sample Size for the Training Set:

Not applicable. This is a physical product performance evaluation against standards, not an AI/machine learning device that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as no training set was used.

{0}------------------------------------------------

Acme Filter Mas

No.57, Hu Shan Road, Yingge District, New Taipei City, 23941, Taiwan, R.O.C.

" 510(k) SUMMARY "

MAY 2 0 2013

September 23, 2012

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92

The assigned 510(k) number is : __ K123115

Submitter's Name: Acme Filter Mask Inc.

No.57, Hu Shan Road, Yingge District, New Taipei City, 23941,Taiwan, R.O.C.

Date summary prepared:

Device Name:

  • Mask, Surgical Classification name: .
  • Classification number: FXX, Class II ●
  • Regulation Number: 878.4040 .
  • Surgical Face Mask with Ear-Loop ● Proprietary name:

YN-501AB, YN-501AW, YN-501AG for Blue, White, Green Product model: One size for different colors (Blue, White, Green)

  • Common name of device: Surgical Face Mask, Disposable ●
  • Surgical Face Mask, Type: Tie-on, Ear-loop, K063043 . Predicate Device:
  • Official Correspondent: Dr. Jen. Ke-Min ●

E-mail: ceirs.jen@msa.hint.net (Tel) +886-3-5208829; (Fax) +886-3-5209783

Description of the device:

Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop is flat pleated 3-ply (at least) masks with an inner and outer layer (spunbonded polypropylene) that sandwich a melt blown polypropylene filter material, also with elastic loops. The nosepiece for all Acme Filter Mask Inc. Surgical Face Masks with Ear-Loop are malleable aluminum wire. All the materials used in the construction of the Acme Filter Mask Inc. Surgical Face Masks with Ear-Loop are being used in currently marked devices.

Labels/Labeling:

This device will be marked to medical device suppliers, Dentist and Doctor Officers, clinics, Emergency Response Professionals, Hospitals and other healthcare professional for the Intended Use purpose below:

{1}------------------------------------------------

Acme Filter Mask Inc.

No.57, Hu Shan Road, Yingge District, New Taipei City, 23941, Taiwan, R.O.C.

Intended Use:

Surgical Face Mask is Device that is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operation room personnel from transfer of microorganisms, body fluids and particulate material.

Comparison to Predicate Devices:

Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop is substantially equivalent for the safety and effectiveness to the Modern Healthcare Corp. Surgical Face Mask, type: Tie-on and Ear-loop:

Modern Healthcare Corp.Surgical Face Mask(K063043)ACME Filter Mask Inc.Surgical Face Mask(K123115)
Similarity:
Fluid ResistanceFluid ResistanceFluid Resistance
FlammabilityClassClass I(No Flame Spread)Class I(No Flame Spread)
Regulatory ClassClass II(ASTM2100-04 Low Barrier)Class II
BFE(%)Higher than 99%Higher than 99.9%
Difference:
TypeTie-on and Ear-loop(Green, White, Blue, Pink)Ear-Loop(Green, White, Blue)
Delta-PAverage 2.6Average 3.33 ( $mmH_2O/cm^2$ )for Air Exchange Pressure
ParticulateFiltrationEfficiencyPerformance (%)Average 96.8% at 0.1 micronAverage 94.79% for Solid AerosolFiltration EfficiencyMore than 99.5% for ViralFiltration Efficiency

Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

  • Fluid Resistance (ASTM F1862-05): Synthetic Blood Penetration Resistance Test 】.
  • Filtration Efficiency: Bacterial Filtration Efficiency (BFE) Test (ASTM F2101-01) II. and Particulate Filtration Efficiency (latex Particle challenge) (ASTM F1215)
  • Differential Pressure (Delta-P) Test (MIL M 36954 C) III.
  • Flammability Test (16CFR 1610) IV.
  • Biocompatibility per ISO 10993-5 /-10 V.

It was our conclusion that performance testing met all relevant requirements of the aforementioned test standard.

{2}------------------------------------------------

Acme Filter Mask Inc.

No.57, Hu Shan Road, Yingge District, New Taipei City, 23941, Taiwan, R.O.C.

Discussion of Clinical Tests Performed:

Not Applicable

Conclusions:

Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop has the same intended use and technological characteristics as the predicated devices Modern Healthcare Corp., Surgical Face Mask, type: Tie-on and Ear-loop (K063043). Especially, the predicate device's types are Tie-on and Ear-loop; and the subject device just for Ear-loop. Besides, the bench testing contained in this submission supplied demonstrates that the technological characteristics do not raise any new question of safety or effectiveness.

Thus the new device is substantially equivalent to the predicate devices.

{3}------------------------------------------------

Acme Filter Mask Inc

No.57, Hu Shan Road, Yingge District, New Taipei City, 23941, Taiwan, R.O.C.

COMPARISON INFORMATION 7.

We place the 510K information for the predicate device thereafter this section.

Comparison to Predicate Devices:

Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop is substantially equivalent for the safety and effectiveness to the Modern Healthcare Corp. Surgical Face Mask, type: Tie-on and Ear-loop:

ItemModern Healthcare Corp.Surgical Face Mask(K063043)ACME Filter Mask Inc.Surgical Face Mask(K123115)
Similarity:
Fluid ResistanceFluid ResistanceFluid Resistance
FlammabilityClassClass I(No Flame Spread)Class I(No Flame Spread)
Regulatory ClassClass II(ASTM2100-04 Low Barrier)Class II
BFE(%)Higher than 99%Higher than 99.9%
Difference:
TypeTie-on and Ear-loop(Green, White, Blue, Pink)Ear-Loop(Green, White, Blue)
Delta-PAverage 2.6Average 3.33 (mmH2O/cm²)for Air Exchange Pressure
ParticulateFiltrationEfficiencyPerformance (%)Average 96.8% at 0.1 micronAverage 94.79% for Solid AerosolFiltration EfficiencyMore than 99.5% for ViralFiltration Efficiency

Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

  • I. Fluid Resistance (ASTM F1862-05): Synthetic Blood Penetration Resistance Test
  • Filtration Efficiency: Bacterial Filtration Efficiency (BFE) Test (ASTM F2101-01) X. and Particulate Filtration Efficiency (latex Particle challenge) (ASTM F1215)
  • differential Pressure (Delta-P) Test (MIL M 36954 C) XI.
  • Flammability Test (16CFR 1610) XII.
  • XIII. Biocompatibility per ISO 10993

It was our conclusion that performance testing met all relevant requirements of the aforementioned test standard.

{4}------------------------------------------------

Acme Filter Mask Inc.

No.57, Hu Shan Road, Yingge District, New Taipei City, 23941,Taiwan, R.O.C.

Discussion of Clinical Tests Performed:

Not Applicable

Conclusions:

Acme Filter Mask Inc. Surgical Face Mask with Ear-Loop has the same intended use and technological characteristics as the predicated devices Modern Healthcare Corp., Surgical Face Mask, type: Tie-on and Ear-loop (K063043). Especially, the predicate device's types are Tie-on and Ear-loop; and the subject device just for Ear-loop. The performance tests for the Delta-P and Particulate Filtration Efficiency Performance also have the similar effectiveness. Besides, the bench testing contained in this submission supplied demonstrates that the technological characteristics do not raise any new question of safety or effectiveness.

Thus the new device is substantially equivalent to the predicate devices.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized emblem with three abstract human figures intertwined, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged around the emblem in a circular fashion.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

May 20, 2013

Dr. Ke-Min Jen ACME Filter Mask No. 57 Hu Shan Road Yingge District, New Taipei City, Taiwan R.O.C. 23941

Re: K123115

Trade/Device Name: Surgical Face Mask with Ear-Loop Model: YN-501AB, YN-501AW, YN-501AG for Blue, White, Green Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: April 12, 2013 Received: April 17, 2013

Dear Dr. Jen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{6}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Image /page/6/Picture/6 description: The image shows a signature block. The text "Sincerely yours," appears at the top. Below that is a signature and the name "Tejashri Purohit-Sheth, M.D." followed by the title "Clinical Deputy Director" and the acronym "DAGRID". Finally, the word "FOR" is at the bottom right.

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510 (K) Number ( If Known ): _________________________________________________________________________________________________________________________________________________ K123115

Device Name: Surgical Face Mask with Ear-Loop Model: YN-501AB, YN-501AW, YN-501AG for Blue, White, Green

Indications for Use:

The Surgical Face Mask of different colors (Blue, White, and Green) is a device intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operation room personnel from transfer of bacteria, body fluid and particulate material.

This Surgical Face Mask is non-sterilized and disposable.

Prescription Use

AND/OR

Over-The-Counter Use_

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ri S. Purohitsheth -S 17 13:52:16

Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental

510(k) Number:

P 5

Page_1 of 1

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.