K Number
K123075
Device Name
BUTTOCK MUSCLE STIMULATOR MODEL WL-2413B; BUTTOCK MUSCLE STIMULATOR MODE WL-2413E
Date Cleared
2013-07-11

(283 days)

Product Code
Regulation Number
890.5850
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles.
Device Description
The Buttock Muscle Stimulator, model WL-2413E, is a two channel battery operated muscle stimulation system specifically designed to exercise the buttocks' muscle. The device comprises mainly the electronic stimulator module which generates the required stimulation signals. The Buttock Muscle Stimulator, model WL-2413E, comprises a short with silicon pad electrodes which connect the signals from the stimulator to the skin. The electrodes are placed on the inner surface of the accessories. The product is supplied with a 510(K) cleared conductive gel, a user's manual, 3AAA batteries, 2 lead wires and a storage bag. The required power of the device is derived from 3 batteries located in a compartment protected by a removable battery cover. The electrodes are in the inner surface of the shorts which is worn as shown on the picture of user manual. There is no current passed from side to side. The user cannot access the wiring or connectors within the shorts.
More Information

Not Found

No
The description focuses on basic electrical muscle stimulation and does not mention any AI or ML components or functionalities.

No
Explanation: The device is intended for the "improvement of muscle tone of the buttocks muscles" in "healthy persons," which indicates an aesthetic or fitness purpose rather than a therapeutic medical purpose.

No
The device is described as a muscle stimulator intended for the "improvement of muscle tone of the buttocks muscles," which is a treatment or enhancement purpose, not a diagnostic one.

No

The device description explicitly states it comprises an "electronic stimulator module," "silicon pad electrodes," "lead wires," and is "battery operated," indicating it includes significant hardware components beyond just software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to apply transcutaneous electrical muscle stimulation for the improvement of muscle tone of the buttocks muscles. This is a physical therapy/rehabilitation type of application, not a diagnostic test performed on samples from the human body.
  • Device Description: The device description details a muscle stimulator that applies electrical signals through electrodes placed on the skin. It does not involve the analysis of biological samples (blood, urine, tissue, etc.).
  • Lack of IVD Characteristics: The description does not mention any components or processes typically associated with IVDs, such as reagents, assays, sample collection, or analysis of biological markers.

IVD devices are specifically designed to examine specimens derived from the human body to provide information for diagnostic, monitoring, or screening purposes. This device does not fit that definition.

N/A

Intended Use / Indications for Use

The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles.

Product codes (comma separated list FDA assigned to the subject device)

NGX

Device Description

The Buttock Muscle Stimulator, model WL-2413E, is a two channel battery operated muscle stimulation system specifically designed to exercise the buttocks' muscle. The device comprises mainly the electronic stimulator module which generates the required stimulation signals.

The Buttock Muscle Stimulator, model WL-2413E, comprises a short with silicon pad electrodes which connect the signals from the stimulator to the skin. The electrodes are placed on the inner surface of the accessories. The product is supplied with a 510(K) cleared conductive gel, a user's manual, 3AAA batteries, 2 lead wires and a storage bag.

The required power of the device is derived from 3 batteries located in a compartment protected by a removable battery cover. The electrodes are in the inner surface of the shorts which is worn as shown on the picture of user manual. There is no current passed from side to side. The user cannot access the wiring or connectors within the shorts.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

buttocks muscles

Indicated Patient Age Range

Not Found

Intended User / Care Setting

healthy persons

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Compliance to applicable voluntary standards includes IEC 60601-2-10, as well as IEC 60601-1, and IEC 60601-1-2 requirement.

In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.

Discussion: The Compliance to applicable voluntary standards as above mentioned indicates that the new device in this submission used the same standards as that of predicate device in addition to the software validation standard. For our new device, the software validation was prepared according to FDA guidance, however the predicate device used IEC 60601-1-4 to prepare their software validation report. For this difference we believe that we used more adequate reference tool to prepare software validation report than the predicate device. Therefore, we consider that the compliance of standards included in our submission is adequate for the determination of substantial equivalence.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K092142

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).

0

510(K) Summary

The assigned 510(k) number is: K123075

Submitter's Identifications: 1.

Well Life Healthcare Limited 1F., No. 16, Lane 454, Jungjeng Road, Yunghe City, Taipei County 234, Taiwan, ROC

Contact: Jenny Hsieh Telephone: + 886 2 2928 2112

Date of Summary Preparation: May 25, 2013

2. Name of the Device:

Trade/Device Name: Buttock Muscle Stimulator, model: WL-2413E. Regulation Number: 21 CFR 890.5850 Requlation Name: Power muscle stimulator. Regulatory Class: II Product Code: NGX

3. Information of the 510(k) Cleared Device (Predicate Device):

SPORT-ELEC models Body Control System "BS"( K092142)

Device Description: 4.

The Buttock Muscle Stimulator, model WL-2413E, is a two channel battery operated muscle stimulation system specifically designed to exercise the buttocks' muscle. The device comprises mainly the electronic stimulator module which generates the required stimulation signals.

The Buttock Muscle Stimulator, model WL-2413E, comprises a short with silicon pad electrodes which connect the signals from the stimulator to the skin. The electrodes are placed on the inner surface of the accessories. The product is supplied with a 510(K) cleared conductive gel, a user's manual, 3AAA batteries, 2 lead wires and a storage bag.

The required power of the device is derived from 3 batteries located in a compartment protected by a removable battery cover. The electrodes are in the inner surface of the shorts which is worn as shown on the picture of user manual. There is no current passed from side to side. The user cannot access the wiring or connectors within the shorts.

The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles

1

5. Intended Use:

The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles

Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are റ. as follows:

Compliance to applicable voluntary standards includes IEC 60601-2-10, as well as IEC 60601-1, and IEC 60601-1-2 requirement.

In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.

Discussion: The Compliance to applicable voluntary standards as above mentioned indicates that the new device in this submission used the same standards as that of predicate device in addition to the software validation standard. For our new device, the software validation was prepared according to FDA guidance, however the predicate device used IEC 60601-1-4 to prepare their software validation report. For this difference we believe that we used more adequate reference tool to prepare software validation report than the predicate device. Therefore, we consider that the compliance of standards included in our submission is adequate for the determination of substantial equivalence.

7. Comparison of Significant device features

| Comparison

FeaturesModels to be Compared
ModelSPORT-ELEC Body Control System
"BS"WL-2413E
510(K) No.K092142K123075
Prescription or OTCOTCSame
Indication for useintended for use by healthy persons to
apply transcutaneous electrical muscle
stimulation(EMS) through skin contact
electrode for the purpose of improvement
of muscle tone of the buttocks musclesSame
FDA product codeNGXSame
Electrode UsedSelf Adhesive Electrode( 5 X 5 cm)K082065, Silicon Pad Electrode
( 7.0 cm diameter)
ManufacturerSPORT-ELECTWell-Life

2

8. Comparison of Unit Characteristics & Output Specification

Mode or Program NamePredicate DeviceNew Device
Waveform (e.g., pulsed monophasic, biphasic)BiphasicBiphasic
Shape (e.g., rectangular, spike, rectified
sinusoidal)RectangularRectangular
Maximum Output Voltage (volts)-
(+/- 20%)31.2V @500Ω
66.0V @2KΩ
66.0V @10KΩ40.8V @500Ω
70.0V @2KΩ
106.0V @10KΩ
Maximum Output Current (mA)-
(+/- 20%)62.4A @500Ω
33.0A @2KΩ
6.6A @10KΩ81.6mA @500Ω
35.0mA @2KΩ
10.6mA @10KΩ
Duration of primary phase (usec)200 Max300 Max
Pulse Duration (usec)540 Max.720 Max.
Frequency (Hz) [or Rate (pps)]70 Max70 Max
For
multiphasic
waveforms
only:Symmetrical phases?
Phase Duration (include units),
(Stage range, if applicable),
(both phases, if asymmetrical)Yes
Not applicableYes
Not applicable
510(K) NumberK092142K123075
Device Name and ModelBSWL-2413E
ManufacturerSPORT-ELECTWell-Life
Power Source(s)1.5Vx3
(AAA Size)1.5Vx3
(AAA Size)
- Method of Line current Isolation
  • Patient Leakage Current | Type BF
    --- | Type BF
    --- | | | | |
    | - Normal condition (uA) | Under 0.1 | Under 0.1 | | | | |
    | - single Fault condition (uA) | Under 0.5 | Under 0.5 | | | | |
    | Average DC current through electrodes when
    device is on but no pulses are being applied ( uA) | Not applicable | Not applicable | | | | |
    | Number of Output Modes | 4 | 4 | | | | |
    | Number of Output
    Channels: | Synchronous or
    Alternating?
    Method of Channel
    Isolation | Synchronous
    Output Coil | Synchronous
    Output Coil | | | |
    | Regulated Current or Regulated Voltage? | Voltage | Voltage | | | | |
    | Software/Firmware/Microprocessor control? | Yes | Yes | | | | |
    | Automatic Overload Trip? | No | No | | | | |
    | Automatic No-Load Trip? | Yes | Yes | | | | |
    | Automatic Shut Off? | Yes | Yes | | | | |
    | User Override control? | No | No | | | | |
    | Indicator Display: | On/Off Status?
    Low Battery?
    Voltage/Current
    Level? | Yes
    Yes
    Yes | Yes
    Yes
    Yes | | | |
    | Timer Range (Minutes) | 20-40 | 20-40 | | | | |
    | Compliance with Voluntary Standards? | IEC 60601-2-10 | IEC 60601-2-10 | | | | |
    | Compliance with 21 CFR 898? | Yes | Yes | | | | |
    | Weight (g) including battery | 171.6 | 80 | | | | |
    | Dimensions (mm.) [W x H x D] | 87×43×69 | 64×90×20 | | | | |
    | Housing Materials and construction | ABS | ABS | | | | |
    | Pulse per burst | Same for each program | Same for each program | | | | |
    | Burst per second | Same for each program | Same for each program | | | | |
    | Bust duration | Same for each program | Same for each program | | | | |
    | Duty Cycle | Same for each program | Same for each program | | | | |
    | Method of achieving zero net charge for net | Method of achieving zero net charge/pulse | Biphasic symmetric | | Biphasic symmetric | Biphasic symmetric wave for each pulse | Biphasic symmetric wave for each pulse |

·

3

Comparison featureModelRemarks
SPORT-ELECT BSWL-2413E
Net charge00Note 1
Max. phase charge14 uc24 uc
Max. current Density0.0784 mA/cm²0.0997 mA/cm²
Max. Average
current38.43mA38.644mA
500 Ω
2K Ω14.892mA16.907 mA
(RMSA)
10 Ω4.996mA5.120 mA
Max. Power Density0.00274 Watts/ cm²0.00399 Watts/ cm²
Burst ModeYesYes

9. Significant output characteristics comparison table:

Note 1: Net Charge is Zero for the symmetrical stimulation biphasic wave

10. Summary for the technology comparison.

Basically the WL-2413E has the similar technological characteristics with the predicate device in the product design, material, energy source type, main program mode and the main output waveform... etc. There exist some difference in the detailed output parameters( mainly in the pulse duration and electrode sizes). Through the detailed calculation comparison of stimulation output energy for each operation mode(in particular the output current density), we found the output level in each operation mode for our WL-2413E and predicate device are very close and within the acceptable range as specified in FDA guidance. So we believe the difference in detailed output parameters does not affect the determination of substantial equivalence.

11. Conclusions

The Buttock Muscle Stimulator, model: WL-2413E have the same intended use and the similar technological characteristics as the cleared device- SPORT-ELEC models Body Control System "BS"( K092142). Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device.

In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.

4

Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles a person embracing another person, or a bird in flight, composed of three curved lines.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 11, 2013

Well-Life Healthcare Ltd C/O Ms. Jennv Hsieh 1F. No. 16, Lane 454, Jungjeng Road Yunghe City, Taipei County, Taiwan R.O.C.

Rc: K123075

Trade/Device Name: Well-Life Buttock Muscle Stimulator. model W1.-2413E Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX Dated: June 2. 2013 Received: June 5, 2013

Dear Ms. Hsieh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (sec above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

5

Page 2 - Ms. Jenny Hsieh

CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Joyce M. Whang -S

Victor Krauthamer, Ph.D. for Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

6

Indications for Use

510(k) Number (if known): K123075

Device Name: Well-Life Buttock Muscle Stimulator, model: WL 2413E

Indications For Use:

The Buttock Muscle Stimulator, model WL-2413E, is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation(EMS) through skin contact electrode for the purpose of improvement of muscle tone of the buttocks muscles. .

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Joyce M. Whang -S

(Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)

510(k) Number _ K123075_

Page 1 of 1