(449 days)
The Air Barrier System is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty and posterior vertebral fusion and laminoplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical surface with no gap between the bottom of the nozzle and the surface, (2) the incision plane is parallel with the direction of air flow, and (3) the incision dimensions are within: 6" (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area.
The ABS device controls the airborne surgical environment at a specific and limited location adjacent to and over surgical incisions by emitting a non-turbulent flow of HEPA filtered air that displaces the presence of airborne colony-forming units and particulate matter within surgical site dimensions of 6" (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable outside of these dimensions and effectiveness depreciates beyond this specified area. The Air Barrier System (ABS) has two components: a Filter/Blower and an Air Delivery System. The ABS Filter/Blower filters ambient air found in a typical surgical operating room through a High Efficiency Particle Arresting (HEPA) filter. The HEPA filtered air exits the Filter/Blower from an exit port on the top of the unit. The Filter/Blower is nonsterile and reusable. The Air Delivery System, consisting of an Air Supply Hose and Nozzle, is sterile, single-use. The Nozzle portion is applied to the surgery site, on top of the incision drape and adjacent to a surgical incision. The end of the Air Supply Hose is plugged into the Filter/Blower's air exit port. HEPA filtered air then flows through the Air Supply Hose through the Nozzle and directly to the incision site.
Acceptance Criteria and Study Details for the Air Barrier System (ABS)
This document describes the acceptance criteria and the studies that demonstrate the Air Barrier System (ABS) meets these criteria, based on the provided 510(k) summary (K123006).
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria for the ABS are primarily defined by its ability to significantly reduce the presence of airborne particulate matter and microorganisms at the surgical site. The reported device performance is based on two clinical studies.
| Acceptance Criteria Category | Specific Acceptance Criteria | Reported Device Performance (Summary of Study Results) |
|---|---|---|
| Reduction of Airborne Microorganisms (CFU/m³) | Statistically significant reduction in airborne microorganism density (colony-forming units per cubic meter) at the surgical site compared to control/sham conditions. | Study 1 (Hip Arthroplasty): Mean microorganism density at ABS area of effect: 1.60 CFU/m³ (Experiment group) vs. 10.73 CFU/m³ (Outside ABS area). Significantly lower (P<0.001) than sham and control groups. Study 2 (Posterior Vertebral Fusion/Laminoplasty): Mean microorganism density at ABS area of effect: 1.55 CFU/m³ (Experiment group) vs. 5.05 CFU/m³ (Control group). Statistically significant reduction (p≤0.0001). |
| Reduction of Airborne Particulate Matter (particles/m³) | Statistically significant reduction in airborne particulate density (particles ≥5µm per cubic foot) at the surgical site compared to control/sham conditions. | Study 1 (Hip Arthroplasty): Mean particulate density (≥5µm) at ABS area of effect: 524 particles/ft³ (Experiment group) vs. 3853 particles/ft³ (Sham) and 4092 particles/ft³ (Control). Significantly lower (P<0.001). Study 2 (Posterior Vertebral Fusion/Laminoplasty): Mean particulate density (≥5µm) at ABS area of effect: 1325 particles/ft³ (Experiment group) vs. 4837 particles/ft³ (Control group). Statistically significant reduction (p≤0.0001). |
2. Sample Size Used for the Test Set and Data Provenance
The provided document describes two clinical studies, which serve as the test sets for demonstrating the device's efficacy.
Study 1 (Hip Arthroplasty):
- Sample Size: 29 patients, randomized into three groups (ABS active, ABS present but not active (sham), no ABS).
- Data Provenance: Not explicitly stated, but implies clinical setting in the US. This study was part of the previous submission [K092801]. It is retrospective information for the current submission, even if it was prospective when originally conducted for K092801.
Study 2 (Posterior Vertebral Fusion or Laminoplasty):
- Sample Size: 38 surgery cases (from 41 patients actually enrolled). Randomized into two groups (ABS device, no ABS device).
- Data Provenance: Conducted at the Michael E. DeBakey VA Medical Center (MEDVAMC), implying the United States. The study was prospective and randomized.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
For both clinical studies, the "ground truth" was established through direct measurement of airborne particulate and microorganism counts. This does not involve expert review or consensus in the typical sense of image analysis or diagnostic studies. Instead, it relies on quantitative measurements from environmental sampling.
Therefore, the concept of "number of experts used to establish the ground truth" and "qualifications of those experts" does not directly apply to this type of device performance study. The ground truth is objective, quantifiable measurements of environmental air quality parameters (CFU and particles) collected by trained personnel using standardized methods.
4. Adjudication Method for the Test Set
The studies described are not based on subjective interpretation or classification that would require an adjudication method like 2+1 or 3+1. The results are quantitative measurements of airborne particles and microorganisms. Therefore, no adjudication method was used or needed.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is an environmental control system, not a diagnostic aid that would involve human readers interpreting output. The studies directly measure the physical effect of the device on the surgical environment.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
Yes, in essence. The clinical studies performed are standalone evaluations of the device's ability to reduce airborne contamination. The device (ABS) operates without continuous human intervention during the procedure to achieve its effect. Its performance is measured directly by environmental sampling, independent of human operators' perception or interpretation.
7. The Type of Ground Truth Used
The ground truth used in both clinical studies was objective, quantitative measurements of environmental parameters:
- Colony-forming units (CFU) per cubic meter: Measured microorganisms (bacteria and fungal spores).
- Particles per cubic foot: Measured particulate matter of a specific size (≥5µm).
These measurements were collected at specific locations within and outside the ABS area of effect during actual surgical procedures.
8. Sample Size for the Training Set
The document does not describe a "training set" in the context of an algorithm or AI model development. The Air Barrier System (ABS) is a physical device that functions based on mechanical and filtration principles, not a software algorithm that requires training data. The studies conducted are performance validations, not model training.
9. How the Ground Truth for the Training Set Was Established
Since there is no "training set" for an algorithm, this question is not applicable. The device's efficacy is demonstrated through performance studies rather than machine learning training.
{0}------------------------------------------------
510(k) Summary
Date Prepared: December 9, 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92 for the Air Barrier System (ABS).
1. Company making the submission:
| Owner/Submitter: | Nimbic Systems, Inc.4910 Wright Road, Suite 170Stafford, Texas 77477 |
|---|---|
| Contact Person: | Sean SelfPresident281-565-5715281-565-5712 faxself@nimbicsystems.com |
2. Device Name and Classification:
| Common/Usual Name | : | Air-handing apparatus for a surgicaloperation room |
|---|---|---|
| Proprietary Name | : | Air Barrier System (ABS) |
| Device Class | : | Class II |
| Regulation Number | : | 21 CFR 878.5070 |
| Product Code | : | ORC |
| 510(k) Number | : | K123006 |
3. Predicate Device
This submission for the Air Barrier System (ABS) Model 5001 is for new claims only. The Air Barrier System (ABS) Model 5001 specifications are the same as the model 1001 released to market [K092801]. The Air Barrier System (ABS) Model 1001 is the primarily predicate device.
4. Intended Use Statement
The Air Barrier System is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty and posterior vertebral fusion and laminoplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical surface with no gap between the bottom of the
{1}------------------------------------------------
nozzle and the surface. (2) the incision plane is parallel with the direction of air flow, and (3) the incision dimensions are within: 6" (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area.
5. Description of Device
The ABS device controls the airborne surgical environment at a specific and limited location adjacent to and over surgical incisions by emitting a non-turbulent flow of HEPA filtered air that displaces the presence of airborne colony-forming units and particulate matter within surgical site dimensions of 6" (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable outside of these dimensions and effectiveness depreciates beyond this specified area.
Clinical studies at two sites by different investigators have been conducted to demonstrate the performance characteristics of the Air Barrier System (ABS) during surgical procedures:
In the first study, twenty-nine (29) patients were randomized into one of three groups: with the ABS active (experiment), with the ABS present but not active (sham), and with no ABS device present (control). Airborne particulate and microorganism samples were collected simultaneously at a discrete single location at the surgical incision within the ABS area of effect and from a discrete single location within the sterile field, but not within the ABS area of effect. The airborne particulate and microorganism counts observed in the experiment group were significantly lower (P<0.001) than that observed in the sham and control groups. In the experiment group, the mean microorganism density at the sampling location within the ABS area of effect was 1.60 colony-forming units per cubic meter compared to 10.73 at the location outside the ABS area of effect. For particulate of size ≥5um, the mean observed particulate density in the experiment group was 524 particles per cubic foot compared with 3853 and 4092 particles per cubic foot in the sham and control groups, respectively.
In the second study, thirty-eight (38) patients undergoing instrumented posterior vertebral fusion or laminoplasty were randomized into one of two groups: with the ABS device (experiment) and with no ABS device (control). Airborne particulate and microorganism samples were collected simultaneously at a discrete single location at the surgical incision in both groups at ten-minute intervals for the first 100 minutes of the procedures. The airborne particulate and microorganism counts observed in the experiment group were significantly lower (P<0.001) than that observed in the control group. The mean experiment group microorganism density at the sampling location within the ABS area of effect was 1.55 colony-forming units per cubic meter compared to 5.05 in the control group. The mean experiment group density of particles sized ≥5um at the sampling location within the ABS area of effect was 1325 particles per cubic foot compared to 4837 measured in the control group.
The Air Barrier System (ABS) has two components: a Filter/Blower and an Air Delivery System.
{2}------------------------------------------------
The ABS Filter/Blower filters ambient air found in a typical surgical operating room through a High Efficiency Particle Arresting (HEPA) filter. The HEPA filtered air exits the Filter/Blower from an exit port on the top of the unit. The Filter/Blower is nonsterile and reusable.
The Air Delivery System, consisting of an Air Supply Hose and Nozzle, is sterile, single-use. The Nozzle portion is applied to the surgery site, on top of the incision drape and adjacent to a surgical incision. The end of the Air Supply Hose is plugged into the Filter/Blower's air exit port. HEPA filtered air then flows through the Air Supply Hose through the Nozzle and directly to the incision site.
6. Summary of the technological characteristics of the device compared to predicate device
The difference from the predicate Air Barrier System (ABS) Model 1001 [K092801], is new and extended claims of application to posterior vertebral fusion and laminoplasty procedures that are in covenant with Indications for Use statement.
The technology, design and method of construction of the Air Barrier System (ABS) Model 5001 are identical to the Air Barrier System (ABS) Model 1001, predicate device.
| Parameter | This submission [K123006] Nimbic Systems' Air Barrier System with new claims | Predicate [K092801] Nimbic Systems Air Barrier System |
|---|---|---|
| Product Code | ORC | ORC |
| Regulation No. | 878.5070 | 878.5070 |
| Review Panel | General Hospital | General Hospital |
| Indications for Use | The Air Barrier System is a portable device for use in a surgical operating room that produces a directed, non- turbulent flow of air to the surgical site. The air flow from the device is HEPA- filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty and posterior vertebral fusion and laminoplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical | The Air Barrier System (ABS) is indicated as a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site. The air flow form the device is HEPA-filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical surface with no gap between the bottom of the nozzle and |
| Parameter | This submission [K123006] Nimbic Systems' Air Barrier System with new claims | Predicate [K092801] Nimbic Systems Air Barrier System |
| surface with no gap between the bottom of the nozzle and the surface, (2) the incision plane is parallel with the direction of air flow, and (3) the incision dimensions are within: 6" (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area. | the surface, (2) the incision plane is parallel with the direction of air flow, and (3) the incision dimensions are within: 6" (15.2 cm) in width and 20" (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area. Patient age restrictions: none. | |
| Physical | ||
| Housing Dimensions | 11 X 11 X 25 inches | 11 X 11 X 25 inches |
| Weight in pounds | 48 | 48 |
| Material type | Stainless Steel | Stainless Steel |
| Filter/Blower Output (CFM) | 150 | 150 |
| Blower Recertification Interval | Annual | Annual |
| Electrical | ||
| Voltage (VAC) | 120 | 120 |
| Frequency (Hertz) | 60 | 60 |
| Max Current (Amps) | 5 | 5 |
| HEPA Filtration Specifications | ||
| Filtration Efficiency | 99.97% @ 0.3 micron | 99.97% @ 0.3 micron |
| HEPA Filter Media | Continuous pleated glass microfiber | Continuous pleated glass microfiber |
| Air Inlet Pre-Filter Media | Polyurethane Foam | Polyurethane Foam |
| Measured average particulate density @ surgical incision (0.5 μm particles per cubic foot) | 68,122 | 29,992 |
| Parameter | This submission [K123006] | Predicate [K092801] |
| Nimbic Systems'Air Barrier Systemwith new claims | Nimbic SystemsAir Barrier System | |
| CFU's @ SurgicalIncision (ColonyForming Units/m³) | 1.57 | 1.60 |
| Air VolumeDelivered toincision in CFM | 41 | 41 |
| Hose and Nozzle Specifications | ||
| Sterility | Sterile | Sterile |
| Use | Single-Use | Single-Use |
| Sterility AssuranceLevel (SAL) | 10-6 SAL | 10-6 SAL |
| Sterilization Method | ETO | ETO |
Comparison table of characteristics between current submission and predicate:
{3}------------------------------------------------
{4}------------------------------------------------
7. Testing
The Air Barrier System (ABS) Model 5001 physical and operational specifications are identical to Air Barrier System (ABS) Model 1001 [K092801]. No changes have been made to the design or method of construction between Models 5001 and 1001.
The Electrical Safety and Electromagnetic Compatibility testing results are identical to Air Barrier System (ABS) Model 1001 {K092801}. For completeness the testing documentation is presented in this submission in the appropriate Sections.
Bench Testing - The bench performance testing submitted in this submission is identical to that submitted in the predicate submission IK0928011. Specifically, it includes two feasibility studies (Ambient and Challenge Tests), and the final bench testing reports (Validation and Performance Qualification). The Validation and Performance Qualification test results demonstrate that the ABS successfully met the criteria for achieving reduction of airborne contamination at the simulated incision site and over the specified control area.
Clinical Testing - Clinical testing was provided in the previous submission [K092801] demonstrating clinical performance of the ABS during hip surgery. Additional clinical data is provided in this submission demonstrating clinical performance of the ABS during posterior cervical and lumbar fusion procedures at the Michael E. DeBakey VA Medical Center (MEDVAMC). The goal of this submission is to incorporate these additional procedures into the ABS' indications for use.
The objective of the MEDVAMC study was to determine the degree to which the Air Barrier System (ABS), which deploys directed HEPA air flow across the surgical field, reduces airborne particulate and airborne colony forming units (e.g., bacteria and fungal spores) at surgical incisions during posterior instrumented lumbar and cervical vertebral fusion procedures. This was a prospective randomized trial with one control
{5}------------------------------------------------
group and one experiment group with an enrollment of up to 40 patients (41 patients actually enrolled). Statistically valid conclusions were able to be made regarding the ABS intended use after analyzing 38 surgery cases.
The clinical trial results demonstrated a statistically significant reduction (p≤0.0001. a=0.05) of both CFU and particulate sized ≥5um in the ABS group versus control group. CFU were reduced from a mean of 5.05 to 1.55 CFU/m³. Particulate sized ≥5pm were reduced from a mean of 4837 to 1325 particles/m3.
8. Conclusion
The results of all of the tests performed, including clinical and nonclinical tests, demonstrate that the ABS-5001 is as safe, and as effective as the predicate device.
9. Rx or OTC
The Air Barrier System (ABS) Model 5001 is a Rx Prescription device per 21 CFR Subpart D. The indication for use is for clinical settings only as stated in device labeling.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 20, 2013
Nimbic Systems, Incorporated C/O Mr. Harvey Knauss Contract Consultant DELPHI Consulting Group 11874 South Evelyn Circle HOUSTON TX 77477
Re: K123006
Trade/Device Name: Air Barrier System Regulation Number: 21 CFR 878.5070 Regulation Name: Air-handling Apparatus For A Surgical Operating Room Regulatory Class: II Product Code: ORC Dated: November 26, 213 Received: December 17, 2013
Dear Mr. Knauss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. Knauss
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
Sincerely vours.
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Image /page/7/Picture/7 description: The image shows a signature and some text. The text includes the name "Tejashri Purohit-Sheth, M.D." and the title "Clinical Deputy Director". The acronym "DAGRID" is also present. The word "FOR" is at the bottom right.
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
、1、1、1、1、1、2、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、1、
Enclosure
{8}------------------------------------------------
Indications for Use Statement
510(k) Number: K123006
Device Name: Air Barrier System (ABS)
The Air Barrier System is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter and microorganisms at the surgical site during hip arthroplasty and posterior vertebral fusion and laminoplasty. The ABS Nozzle is intended to be used only where: (1) it can be placed on an anatomical surface with no gap between the bottom of the nozzle and the surface, (2) the incision plane is parallel with the direction of air flow. and (3) the incision dimensions are within: 6° (15.2 cm) in width and 20° (50.8 cm) in length. Device effectiveness may not be reliably detectable at a distance of 20 inches from the Nozzle, and effectiveness depreciates beyond this specified area.
Prescription Use YES (Parl 21 CFR 801 Subpart D)
AND/OR
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Elizabeth F. Claverie -S 2013.12.18 18:45:25 -05'00'
indications for use Statement Al response 4-18-13.doc
Delphi Consulting Group
Houston, TX 77071
Section 4 510(k) Submission Page #
48
§ 878.5070 Air-handling apparatus for a surgical operating room.
(a)
Identification. Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.(b)
Classification. Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.