K Number
K122873
Date Cleared
2013-01-15

(118 days)

Product Code
Regulation Number
870.5800
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Caremed Supply Inc. VASOPOUSE® SQS and IPCS Decp Vein Thrombosis (DVT) Compression Devices are intended to increase venous blood flow in at risk patients in order to help prevent deep vein thrombosis.

Device Description

VASOPOUSE® that counteracts blood stasis and coagulation changes - two of the three major factors that promote deep vein thrombosis (DVT) - a potentially life threatening condition which can lead to pulmonary embolism. VASOPOUSE® is a non-invasive mechanical prophylactic system that massages the legs in a wavelike, milking motion that promotes blood flow and deters thrombosis, helping to empty pooled or static blood from the valve cusps of the femoral vein. Fibrinolytic activity is increased, stimulating the release of a plasminogen activator. This therapy typically complements other prophylactic measures, such as ant embolic stockings and anticoagulants.

VASOPOUSE® is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb. Therefore, VASOPOUSE® is identified as a compressible limb sleve.

AI/ML Overview

This document describes the 510(k) notification for the VASOPOUSE® DVT Compression Device Model: IPCS/SQS by Caremed Supply Inc. The submission aims to demonstrate substantial equivalence to a predicate device, the VESOFLOW (K110977), also manufactured by Caremed Supply Inc.

Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

Acceptance Criteria and Reported Device Performance

The document states that "All the test results demonstrate VASOPOUSE® meets the requirements of its pre-defined acceptance criteria and intended uses." However, the specific quantitative or qualitative acceptance criteria for each test performed are not explicitly provided in the document. The reported device performance is generally stated as meeting these unspecified requirements.

The non-clinical testing performed includes:

  • Function Test
  • Vibration Test
  • Shock Test
  • Free Fall Drop
  • Life Test

Additionally, the device was tested in accordance with several safety and biocompatibility standards:

  • EN/IEC60601-1-2:2007 (Electromagnetic compatibility for medical electrical equipment)
  • ISO10993-1:2009 (Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process)
  • ISO10993-5:2009 (Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity)
  • ISO10993-10:2002/AMD.1:2006(E) (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization)
  • IEC60068-2-06 (Environmental testing - Part 2-6: Tests - Test Fc: Vibration, sinusoidal)
  • IEC60068-2-27 (Environmental testing - Part 2-27: Tests - Test Ea and guidance: Shock)
  • ISO14971:2007 (Medical devices - Application of risk management to medical devices)

The document asserts that these non-clinical tests confirmed the device's safety and effectiveness and its substantial equivalence to the predicate device.

Table of Acceptance Criteria and Reported Device Performance:

Test NameAcceptance Criteria (Not explicitly stated in document)Reported Device Performance (Summary)
Function TestNot specifiedMeets requirements, functions as intended
Vibration TestNot specifiedMeets requirements
Shock TestNot specifiedMeets requirements
Free Fall DropNot specifiedMeets requirements
Life TestNot specifiedMeets requirements, demonstrates satisfactory operational lifespan
EN/IEC 60601-1-2:2007 TestingNot specifiedMeets standard requirements (EMC)
ISO 10993 series TestingNot specifiedMeets standard requirements (Biocompatibility)
IEC 60068-2-06 TestingNot specifiedMeets standard requirements (Vibration)
IEC 60068-2-27 TestingNot specifiedMeets standard requirements (Shock)
ISO 14971:2007 TestingNot specifiedMeets standard requirements (Risk Management)

Additional Requested Information:

Since this 510(k) pertains to a DVT compression device (mechanical, non-diagnostic), many of the detailed aspects typically associated with AI/software-as-a-medical-device (SaMD) validations (like sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth types) are not relevant or provided in this type of submission. This document focuses on demonstrating physical and functional equivalence to a predicate device, primarily through non-clinical bench testing.

Therefore, for the remaining points, the answer is "Not Applicable" based on the provided document, as the submission does not involve an AI/software component that would require such studies.

  1. Sample size used for the test set and the data provenance: Not Applicable (Bench testing of physical device, not data-driven software).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not Applicable (Bench testing, not expert-adjudicated data).
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not Applicable.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not Applicable (No AI component).
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not Applicable (No algorithm/AI).
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not Applicable (Performance evaluated against engineering specifications and industry standards for physical device functioning, not clinical ground truth in the context of diagnoses or outcomes).
  7. The sample size for the training set: Not Applicable (No training set as it's not a machine learning device).
  8. How the ground truth for the training set was established: Not Applicable.

{0}------------------------------------------------

122873

Caremed Supply Inc. 510(k) Notification

VASOPOUSE® DVT Compression Device Model: IPCS/SQS

JAN 1 5 2013

510(k) Summary

  • 5.1 Type of Submission: Traditional/Special/Abbreviated
  • Preparation Date: Sep 11, 2012 5.2
  • 5.3 Revised Date:
5.4 Submitter:Caremed Supply Inc.
Address:7F., No. 2, Lane 235, Bao Chiao Rd., Xin Tien Dist., NewTaipei City 23145, Taiwan, R.O.C.
Phone:+886-2-29179808
Fax:+886-2-29186505
Contact:TSUNG-HSUAN LIU
Establishment Registration Number:8022590

Identification of the Device: 5.5

Proprietary/Trade name:VASOPOUSE® Deep Vein Thrombosis CompressionDevice
Model Name: IPCS/SQS
Model No.: M6-9/M6-5
Common Name:Compression Therapy Device
Classification Name:Sleeve, Limb, Compression
Device Classification:Class II
Regulation Number:870.5800
Panel:Cardiovascular
Product Code:JOW

5.6 Identification of the Predicate Device:

Predicate Device Name:VESOFLOW
Manufacturer:CAREMED SUPPLY INC.
Product Code:JOW
510(k) Number:K110977

{1}------------------------------------------------

5.7 Intended Use and Indications for Use of the subject device

The Caremed Supply Inc. VASOPOUSE® SQS and IPCS Decp Vein Thrombosis (DVT) Compression Devices are intended to increase venous blood flow in at risk patients in order to help prevent deep vein thrombosis.

5.8 Device Description

VASOPOUSE® that counteracts blood stasis and coagulation changes - two of the three major factors that promote deep vein thrombosis (DVT) - a potentially life threatening condition which can lead to pulmonary embolism. VASOPOUSE® is a non-invasive mechanical prophylactic system that massages the legs in a wavelike, milking motion that promotes blood flow and deters thrombosis, helping to empty pooled or static blood from the valve cusps of the femoral vein. Fibrinolytic activity is increased, stimulating the release of a plasminogen activator. This therapy typically complements other prophylactic measures, such as ant embolic stockings and anticoagulants.

VASOPOUSE® is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb. Therefore, VASOPOUSE® is identified as a compressible limb sleve.

5.9 Statement of Substantial Equivalence

The VASOPOUSE® SQS and IPCS Deep Vein Thrombosis (DVT) Compression Devices are substantially equivalent in all aspects, e.g., technological characteristics, modes of operation, performance characteristics, intended use, etc., to the commercially available VESOFLOW SQS and IPCS DVT Compression Devices. The changes between the two systems include the exclusion of battery back up and the dimension of the pump.

5.10Non-clinical Testing

A series of safety tests were performed to assess the safety and effectiveness of VASOPOUSE . The safety tests were conducted in accordance with EN/IEC60601-1-2:2007, ISO10993-1:2009, ISO10993-5:2009, and ISO10993-10:2002/AMD.1:2006(E). The performance testing conducted on subject device and predicate device are listed below:

  • · Function Test
  • · Vibration Test

{2}------------------------------------------------

Caremed Supply Inc. 510(k) Notification

VASOPOUSE® DVT Compression Device Model: IPCS/SQS

  • Shock Test
  • Free Fall Drop
  • Life Test

All the test results demonstrate VASOPOUSE® meets the requirements of its pre-defined acceptance criteria and intended uses.

5.11Substantial Equivalence Determination

The VASOPOUSE® submitted in this 510(k) file is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the cleared VESOFLOW (K110977). Differences between the devices cited in this section do not raise any new issues of safety or effectiveness.

ItemVESOPOUSE®VESOFLOW (K110977)
Similarities
Regulation Number870.5800870.5800
ClassificationClass IIClass II
Product CodeJOWJOW
Prescription UseYesYes
Indications for UseThe Caremed Supply Inc.VASOPOUSE® SQS and IPCSDeep Vein Thrombosis (DVT)Compression Devices are intendedto increase venous blood flow in atrisk patients in order to help preventdeep vein thrombosis.The Caremed Supply Inc. VesoflowSQS and IPCS Deep VeinThrombosis (DVT) CompressionDevices are intended to increasevenous blood flow in at risk patientsin order to help prevent deep veinthrombosis.
ModelSQS/IPCSSQS/IPCS
Pressure RangeSame40/45 and 130 mmHg
SterilizationN/AN/A
ReferencedSameIEC60601-1:1988+A1:1991+A2:1995EN/IEC60601-1-2:2007EN/IEC60601-1-4
Standards

{3}------------------------------------------------

Caremed Supply Inc.510(k) NotificationVASOPOUSE® DVT Compression DeviceModel: IPCS/SQS
IEC60068-2-06
IEC60068-2-27
ISO14971:2007
ISO10993-1:2009
ISO10993-5:2009
ISO10993-10:2002/AMD.1:2006(E)
Differences
Battery PackNoYes

5.12Conclusion

After analyzing bench tests, safety testing data, it can be concluded that VASOPOUSE® IPCS and SQS DVT Compression Device are substantially equivalent to the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

JAN 1 5 2013

Caremed Supply Inc. c/o Mr. Michael Lee Consultant 7F., No. 2, LANE 235, BAO CHIAO RD. NEW TAIPEI CITY, XIN TIEN DIST. TAIWAN, R.O.C, 23145

Re: K122873

Trade/Device Names: VASOPOUSE® Deep Vein Thrombosis Compression Device Regulatory Number: 21 CFR 870.5800 Regulation Name: Sleeve, Limb, Compression Regulatory Class: Class II Product Code: JOW Dated: September 11, 2012 Received: September 19, 2012

Dear Mr. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Michael Lee

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Matthew G. Hillebrenner

for

· Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices

Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Caremed Supply Inc. 510(k) Notification

VASOPOUSE® DVT Compression Device Model: IPCS/SQS

Indications for Use

877 122 510(k) Number (if known):

Device Name: VASOPOUSE® IPCS/SQS Deep Vein Thrombosis Compression Device

Indications for Use:

The Caremed Supply Inc. VASOPOUSE® SQS and IPCS Deep Vein Thrombosis (DVT) Compression Devices are intended to increase venous blood flow in at risk patients in order to help prevent deep vein thrombosis.

Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ovascular Devices

M.A. Hillel

Page 1 of

510(k) Number K122873

15

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).