K Number
K122840

Validate with FDA (Live)

Date Cleared
2012-12-14

(88 days)

Product Code
Regulation Number
892.5700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Mick Radio-Nuclear Instruments, Inc. CT HDR Tandem/Ring Applicator with Rectal Retractor is indicated for High Dose Rate irradiation of the uterus and cervix. The modification that precipitated this Special 510(k) does not alter the indications for use.

Device Description

The Mick Radio-Nuclear Instruments, Inc. CT HDR Tandem/Ring Applicator with Rectal Retractor is intended for use in Brachytherapy. It is indicated for use where high dose rate (HDR) irradiation of the uterus and cervix is an accepted clinical practice. The applicator is designed to be compatible with the sealed sources found in High Dose Rate After-Loaders and do not modify or change the source, source packaging or remote source positioning mechanisms found on these after-loaders.

The applicator presents a stable and flexible delivery system for the administration of high dose rate irradiation of cancers of the uterus and cervix. The applicator offers several treatment options by including three different angles of Rings (30°, 45° & 60°) and nine different Tandems. The Tandems are offered in complementary angles to the three rings and in three different Tandem lengths for each angle (2cm, 4cm & 6cm from the ring axis). In addition, there are two sizes of Build-Up Caps (5mm & 7.5mm) that affix to the Rings that offer shielding to patient tissue and a Rectal Retractor is included. The entire applicator system is housed in a specially designed Sterilization Cassette, meant for both storage and sterilization. This applicator has been cleared by FDA since 2003.

The modification presented in this Special 510(k) allows for the removal of the paddle portion of the Rectal Retractor from the Rectal Retractor assembly. This facilitates easier cleaning by the user to prevent potential build-up of soil in the paddle. There are no other impacts to the applicator system in this modification. The intended use is unchanged.

The modification to a two-piece Rectal Retractor design is controlled through the Mick Radio-Nuclear design controls and is manufactured onsite at Mick Radio-Nuclear Instruments, Inc. The two-piece Rectal Retractor design does not introduce any new risks or potential negative impacts to the overall performance of the applicator.

AI/ML Overview

The provided document is a Special 510(k) Summary for a medical device (CT HDR Tandem/Ring Applicator with Rectal Retractor). This type of submission is for modifications to a legally marketed device that do not affect its intended use or fundamental scientific technology, and do not introduce new risks or alter performance.

Therefore, the document explicitly states that the in vitro and in vivo testing is "Not applicable to the modification for the device." This means there was no new study proving the device (or its modification) meets acceptance criteria in the traditional sense of clinical performance or diagnostic accuracy studies.

The acceptance criteria and performance are inherently assumed to be met via substantial equivalence to the predicate device because the modification is minor (a two-piece rectal retractor for easier cleaning) and does not impact performance.

Given this context, I cannot provide all the requested information, as a comprehensive performance study with specific acceptance criteria and detailed reporting on sensitivity, specificity, or similar metrics was not conducted or required for this type of submission.

Here's what can be extracted based on the document:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Criteria (Implicit for Special 510(k))Reported Device Performance (Implicitly Met)
Functional EquivalenceDevice performs its intended function (High dose rate brachytherapy of uterus/cervix) without degradation due to modification.The modification (two-piece rectal retractor) "does not introduce any new risks or potential negative impacts to the overall performance of the applicator." The "fundamental scientific technology is unchanged." The device continues to perform its indicated use.
SafetyModification does not introduce new safety risks."No new issues of safety or effectiveness are introduced by using this device." The same grade of titanium is used, and the modification facilitates easier cleaning, potentially reducing risk of soil build-up if anything. Biocompatibility is also deemed unchanged ("No new issues of biocompatibility are raised").
Material EquivalenceMaterials remain the same or are equivalent/safe.The modification incorporates a titanium mounting bracket and screws; the "same grade of titanium used in the previous Rectal Retractor assembly is used for the additional components." Other materials (Titanium for tandem/colpostat, Acetal for build-up cap/rectal retractor paddle) remain identical to the predicate.
Sterilizability/CleanabilityDevice remains capable of being cleaned and sterilized, ideally improved.The modification was specifically made to "facilitate easier cleaning by the user to prevent potential build-up of soil in the paddle." The device is housed in a "specially designed Sterilization Cassette, meant for both storage and sterilization," implicitly confirming sterilizability.
Intended UseIntended Use remains unchanged."The intended use is unchanged." The Indications for Use section explicitly states: "The modification that precipitated this Special 510(k) does not alter the indications for use."

2. Sample size used for the test set and the data provenance:

  • None. As explicitly stated in the document: "In vitro Testing Not applicable to the modification for the device." and "In vivo Testing Not applicable for the modification for the device." This type of submission relies on design changes and substantial equivalence to a predicate device, not new performance studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. No test set or ground truth was established for a new performance study.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Not Applicable. No test set was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This device is a mechanical applicator, not an AI or imaging diagnostic tool that would involve human readers or AI assistance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This device is a mechanical applicator; there is no algorithm for standalone performance testing in this context.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not Applicable. No new performance study requiring ground truth was conducted. The ground truth for the predicate device's safety and effectiveness was established through its prior clearance (K030110), likely including clinical acceptance, material testing, and design validation appropriate for that time.

8. The sample size for the training set:

  • Not Applicable. No training set was used, as this is not an AI/machine learning device.

9. How the ground truth for the training set was established:

  • Not Applicable. No training set was used.

{0}------------------------------------------------

K122840

DEC 1 4 2012

Special 510(k) Summary for compliance with CFR 807.92

September 12, 2012

1. General Provisions

Common Name:Remote Controlled Radionuclide Applicator System (21 CFR 892.5700 - Product Code JAQ)
Proprietary Name:CT HDR Tandem/Ring Applicator with Rectal Retractor
Owner NameMick Radio-Nuclear Instruments, Inc.
Address521 Homestead Avenue
Mount Vernon, New York 10550
Telephone(914) 667-3999
Fax(914) 665-8834
Contact NameChuck Smith
Contact emailchuck@micknuclear.com

2. Name of Predicate Device:

DeviceManufacturerK Number
CT HDR Tandem/RingApplicator with Rectal RetractorMick Radio-Nuclear Instruments, Inc.K030110

Reason for this Special 510(k) Premarket Notification 3.

The purpose of this Special 510(k) submission is to obtain clearance to market the CT HDR Tandem/Ring Applicator with Rectal Retractor (Catalog # 0407), with a modified two-piece Rectal Retractor.

ﺎﻧﻬﺎ ﻓﻬ

4. Classification

This device is classified as a class II device according to 21 CFR 892.5700 .

{1}------------------------------------------------

5. Intended Use and Device Description

The Mick Radio-Nuclear Instruments, Inc. CT HDR Tandem/Ring Applicator with Rectal Retractor is intended for use in Brachytherapy. It is indicated for use where high dose rate (HDR) irradiation of the uterus and cervix is an accepted clinical practice. The applicator is designed to be compatible with the sealed sources found in High Dose Rate After-Loaders and do not modify or change the source, source packaging or remote source positioning mechanisms found on these after-loaders.

The applicator presents a stable and flexible delivery system for the administration of high dose rate irradiation of cancers of the uterus and cervix. The applicator offers several treatment options by including three different angles of Rings (30°, 45° & 60°) and nine different Tandems. The Tandems are offered in complementary angles to the three rings and in three different Tandem lengths for each angle (2cm, 4cm & 6cm from the ring axis). In addition, there are two sizes of Build-Up Caps (5mm & 7.5mm) that affix to the Rings that offer shielding to patient tissue and a Rectal Retractor is included. The entire applicator system is housed in a specially designed Sterilization Cassette, meant for both storage and sterilization. This applicator has been cleared by FDA since 2003.

The modification presented in this Special 510(k) allows for the removal of the paddle portion of the Rectal Retractor from the Rectal Retractor assembly. This facilitates easier cleaning by the user to prevent potential build-up of soil in the paddle. There are no other impacts to the applicator system in this modification. The intended use is unchanged.

The modification to a two-piece Rectal Retractor design is controlled through the Mick Radio-Nuclear design controls and is manufactured onsite at Mick Radio-Nuclear Instruments, Inc. The two-piece Rectal Retractor design does not introduce any new risks or potential negative impacts to the overall performance of the applicator.

6. Drawings

Schematic diagrams of the CT HDR Tandem / Ring Applicator with Rectal Retractor are provided in Tab I.

7. Manufacturing Process

This device is manufactured according to Good Manufacturing Practices (GMPs) as defined in 21CFR part 820. The processes used to fabricate these devices are similar to those used for the predicate device described in this 510(k) notification.

Biocompatibility 8.

No new issues of biocompatibility are raised with regard to the modification of the Rectal Retractor of this device.

{2}------------------------------------------------

9. Summary of Substantial Equivalence

This device is similar in design and construction, utilizes the identical materials, and has the same intended use and performance characteristics to the predicate device. The modification to a two-piece Rectal Retractor incorporates a titanium mounting bracket and two additional titanium screws to affix the paddle to the Rectal Retractor assembly. The same grade of titanium used in the previous Rectal Retractor assembly is used for the additional components. The fundamental scientific technology is unchanged from the original predicate device. No new issues of safety or effectiveness are introduced by using this device.

Performance Test Results

  • a. In vitro Testing Not applicable to the modification for the device. By introducing this device, no new issues of safety or effectiveness are raised.
  • b. In vivo Testing Not applicable for the modification for the device.

10. Summary of Similarities and Differences

The design similarities/differences between the Rectal Retractor modification in this device and the original predicate device is/are:

The devices have the same basic design with the exception of the removable paddle component in the Rectal Retractor;

The devices have the same intended use: The devices use the same materials:

A summary of the design controls in the modification of the Rectal Retractor are contained in Tab K.

{3}------------------------------------------------

· 11. Comparison Table

The Table below compares the modification of this device to the predicate device. Included in these tables are a comparison of the materials and intended uses of these devices.

CT HDR Tandem/Ring Applicatorwith Rectal RetractorModification to the CT HDRTandem/Ring Applicator with RectalRetractor
K NumberK030110To Be Determined
Intended UseHigh dose rate brachytherapy treatmentof the uterus and cervix.High dose rate brachytherapy treatmentof the uterus and cervix.
ShieldingNoNo
Tandem MaterialTitaniumTitanium
ColpostatMaterialTitaniumTitanium
Build-Up CapMaterialAcetalAcetal
Rectal RetractorPaddle MaterialAcetalAcetal
Rectal RetractorPaddle MountMaterialNot ApplicableTitanium

1

.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized image of an eagle with its wings spread, overlaid with three wavy lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Mick Radio-Nuclear Instruments, Inc. % Mr. Chuck Smith Manager, Oulaity Assurance 521 Homestead Avenue MOUNT VERNON NY 10550

DEC 1 4 2012

Re: K122840

Trade/Device Name: CT HDR Tandem/Ring Applicator with Rectal Retractor Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radio-nuclide applicator system Regulatory Class: II Product Code: JAQ Dated: September 12, 2012 Received: September 17, 2012

Dear Mr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2-Mr. Smith

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mary S. Pastel

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): K122840

Device Name: CT HDR Tandem/Ring Applicator with Rectal Retractor

Indications for Use:

The Mick Radio-Nuclear Instruments, Inc. CT HDR Tandem/Ring Applicator with Rectal Retractor is indicated for High Dose Rate irradiation of the uterus and cervix. The modification that precipitated this Special 510(k) does not alter the indications for use.

Prescription Use (Part 21 CFR 801 Subpart D) · AND/OR

Over-The-Counter Use . (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Mary S pastel

(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health

510(k)_

Page 1 of 1

§ 892.5700 Remote controlled radionuclide applicator system.

(a)
Identification. A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.(b)
Classification. Class II.