K Number
K122705
Device Name
HUMICARE D900
Manufacturer
Date Cleared
2013-05-21

(259 days)

Product Code
Regulation Number
868.5450
Panel
AN
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Gründler HumiCare D900 system is intended to heat and humidify medical respiratory gases for patients. The device is for use during artificial ventilation or ventilatory support (e.g. invasive ventilation via tracheal tube or cannula), non-invasive ventilation, or respiratory therapy by means of a mask or other patient interface.

Heated breathing circuits are intended to provide warmed and/or humidified breathing gases before entering the patient airway reducing or eliminating water. They are accessories for the Gründler's HumiCare D900. The heated breathing circuits are used for flow rates greater than 3L/min.

The system is for use in hospital/institutional environment or in the home environment by medically trained healthcare users.

Device Description

The HumiCare D900 system is a respiratory gas humidifier according to 21 CER §868.5450. A respiratory gas humidifier is identified by the Food and Drug Administration (FDA) as a therapeutic device that is intended to warm and add humidity.

The HumiCare D900 including heated air tubings is an active heated humidifier which employs a passover humidification via an enhanced surface area for gas/water. The huge gas/water sufface area is resulting in an output of gas with a temperature almost identical to that of the water with a relative humidity of 100%.

The principle HumiCare D900 is to direct the gas mixture from the ventilator's outlet into the humidifier's water chamber via air tubing. There it is heated and humidified by means of heated water. A heated inspiratory tube is used to transport the conditioned gas from the water chamber's outlet to the patient.

Depending on the patient interface an inspiratory tube or inspiratory and expiratory tubes can be used for the humidification with HumiCare D900. The inspiratory limb can further include an antibacterial filter with filter heater to reduce condensation. In order to minimize condensation (rain out), both inspiratory and expiratory air tubing can be actively heated therefore temperature sensors are included in the inspiratory and expiratory air tubing,

The HumiCare D900 system consists of a heater base with external power supply, connector cables, a 30day disposable water chambers, heated tubes (inspiratory), connection tube, antibacterial filter and filter heater. The 30day disposable water chamber and air tubing circuits are intended for single patient use, with the exception of when an antibacterial filter is incorporated into the circuit allowing multi-patient use of the chamber. The HumiCare D900 heater base and filter heater are intended for multi-patient re-use.

AI/ML Overview

The HumiCare D900 is a respiratory gas humidifier. The provided text from the 510(k) summary (K122765) states that various design and verification activities were performed to ensure the device met predetermined acceptance criteria. These tests confirmed that the product met these criteria, specifically highlighting side-by-side testing for humidification performance and resistance.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document states: "All tests confirmed the product met the predetermined acceptance criteria. In particular side-by-side testing demonstrated that the HumiCare D900 essential performance specifications (humidification performance, resistance of the humidifier) are Substantially Equivalent to the predicate devices."

However, specific numerical acceptance criteria or detailed performance reports are not provided in this summary. The summary only generally states that the device met the criteria and was substantially equivalent to predicate devices for "humidification performance" and "resistance of the humidifier."

Acceptance CriteriaReported Device Performance
Specific numerical criteria for humidification performance not providedDemonstrated "Substantially Equivalent" humidification performance to predicate devices (K073706, K092256, K983112, K100104).
Specific numerical criteria for resistance of the humidifier not providedDemonstrated "Substantially Equivalent" resistance of the humidifier to predicate devices (K073706, K092256, K983112, K100104).
Biocompatibility (Genotoxicity, Cytotoxicity, Implantation, Sensitization, and Irritation) for materials in contact with warm wet air path.Met applicable requirements (ISO 10993-3, ISO 10993-5, ISO 10993-6, ISO10993-10).

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the test set. It also does not define the data provenance (e.g., country of origin, retrospective or prospective) for the performance testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not mention the use of experts or a ground truth established by them for the performance validation of the HumiCare D900. The testing appears to be of a technical/engineering nature, comparing the device's technical specifications and performance against predicate devices and relevant standards.

4. Adjudication Method for the Test Set

Since an expert-based ground truth is not mentioned, there is no adjudication method described in the provided text.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No. The document describes a medical device (a respiratory gas humidifier) and its engineering and performance validation. It does not involve human readers interpreting data or a multi-reader multi-case comparative effectiveness study in the context of AI assistance.

6. If a Standalone Performance Study Was Done

Yes, a standalone performance study was done in the sense that the device's performance was evaluated against predetermined acceptance criteria and compared to predicate devices. The text states: "All tests confirmed the product met the predetermined acceptance criteria. In particular side-by-side testing demonstrated that the HumiCare D900 essential performance specifications (humidification performance, resistance of the humidifier) are Substantially Equivalent to the predicate devices." This implies that the device (algorithm/system) was tested on its own to demonstrate its functionality and equivalence.

7. The Type of Ground Truth Used

The "ground truth" for the HumiCare D900's performance appears to be based on:

  • Engineering specifications and pre-determined acceptance criteria for device functionality.
  • The performance characteristics of legally marketed predicate devices, against which the HumiCare D900 was deemed "substantially equivalent."
  • Compliance with recognized standards (e.g., ISO 8185:2009 for humidifiers, ISO 10993 series for biocompatibility, IEC 60601-1 for safety).

8. The Sample Size for the Training Set

The concept of a "training set" is not applicable here, as this device is a physical medical device (a humidifier) and not an AI/machine learning model that undergoes training on data.

9. How the Ground Truth for the Training Set Was Established

As mentioned above, the concept of a "training set" is not applicable to this device. Therefore, no ground truth for a training set was established.

§ 868.5450 Respiratory gas humidifier.

(a)
Identification. A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.(b)
Classification. Class II (performance standards).