Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K133766
    Device Name
    HUMICARE D900
    Manufacturer
    Date Cleared
    2014-11-24

    (348 days)

    Product Code
    Regulation Number
    868.5450
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Gründler's HumiCare D900 system is intended to heat and humidify medical respiratory gases for patients. The device is for use during artificial ventilatory support (e.g. invasve ventilation via tracheal tube or cannula), non-invasive ventilation, or respiratory therapy by means of a mask or other patient interface.

    Heated breathing circuits are intended to provide warmed and/or humidified breathing gases before entering the patient airway reducing or eliminating water. They are accessories for the Gründler's HumiCare D900. The heated breath or flow rates greater than 3L/min.

    The system is for use in hospital/institutional environment or in the home environment by medically trained healthcare users.

    Device Description

    The HumiCare D900 system is a respiratory gas humidifier according to 21 CFR §868.5450. A respiratory gas humidifier is identified by the Food and Drug Administration (FDA) as a therapeutic device that is intended to warm and add humidity.

    The HumiCare D900 including heated air tubings is an active heated humidifier which employs a pass-over humidification via an enhanced surface area for gas/water. The huge gas/water surface area is resulting in an output of gas with a temperature almost identical to that of the water with a relative humidity of 100%.

    The principle operation of the HumiCare D900 is to direct the gas mixture from the ventilator's outlet into the humidifier's water chamber via air tubing. There it is heated and humidified by means of heated water. A heated inspiratory tube is used to transport the conditioned gas from the water chamber's outlet to the patient.

    Depending on the patient interface an inspiratory tube or inspiratory and expiratory tubes can be used for the humidification with HumiCare D900. The inspiratory limb can further include an antibacterial filter with filter heater to reduce condensation. In order to minimize condensation (rain out), both inspiratory and expiratory air tubing can be actively heated. Therefore a temperature sensor is included in the inspiratory air tubing to control the temperature regulation of the air tubing.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the HumiCare D900 respiratory gas humidifier, which is a modification of an already cleared device (K122705). The submission primarily focuses on demonstrating substantial equivalence to the predicate device through non-clinical bench testing.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria and reported device performance in the format of specific numerical targets vs. achieved results. Instead, it indicates that the modified device was tested against the "predetermined pass/fail requirements of the HumiCare D900 System Specification".

    However, some performance characteristics are mentioned and compared between the predicate and modified device, which can be inferred as meeting acceptance criteria due to the "Equivalent" or "Substantially Equivalent" comments.

    Acceptance Criterion (Inferred from comparison)Reported Device Performance (Modified HumiCare D900)
    Humidification - Invasive mode:
    Flow up to 60 L/minFlow up to 60 L/min
    > 33 mg/L> 33 mg/L
    Humidification - Non-invasive mode:
    Flow up to 100 L/minFlow up to 100 L/min
    > 10 mg/L> 10 mg/L
    Accuracy of displayed gas temperatureMet predetermined pass/fail requirements
    Surface temperature of the tubesMet predetermined pass/fail requirements
    System performance according to ISO 8185Met predetermined pass/fail requirements
    Absence of condensate formationMet predetermined pass/fail requirements
    Biocompatibility (ISO 10993-1, -3, -5, -6, -10)Passed all requirements
    Absence of volatiles, particulates, leachablesPassed all requirements
    Cleaning and reuse validationCompleted, establishing reusability of T2 plug-in sensor

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily discusses non-clinical bench testing.

    • Sample size: The document does not specify the exact sample size for the bench tests (e.g., number of devices tested, number of test runs). It generally refers to "Side-by-Side bench testing" and "Design and Verification activities."
    • Data provenance: The testing was "Side-by-Side bench testing" and "Design and Verification activities" performed by the manufacturer, ResMed Corp., which is based in San Diego, California, USA (though the original submitter, Dr. Christoph Gründler, and the device name "Gründler's HumiCare D900" suggest a European origin for the base technology). The nature of the testing is non-clinical/laboratory, not human subject data. Therefore, the terms "retrospective" or "prospective" are not applicable in this context.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable as the study did not involve human interpretation or subjective judgment to establish ground truth. The acceptance criteria were based on "predetermined pass/fail requirements of the HumiCare D900 System Specification" and international standards like ISO 8185 and ISO 10993. These are objective measurements and engineering specifications, not expert consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable as the testing was non-clinical bench testing and did not involve human observers or interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not done. This device is a respiratory gas humidifier, not an AI-powered diagnostic or assistive technology that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The concept of "standalone algorithm only" is not applicable to this device. It is a physical medical device (humidifier) that performs its function directly, not through an algorithm interpretation. The device does contain software for control and monitoring, but its performance is assessed via its physical output (temperature, humidity), not through an inferential algorithm's output.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" or reference for evaluating performance was based on:

    • Predetermined pass/fail requirements from the HumiCare D900 System Specification.
    • International standards: ISO 8185 for system performance and ISO 10993 for biocompatibility.
    • Objective physical measurements: Such as temperature accuracy, humidification output (mg/L), and absence of condensation.

    8. The sample size for the training set

    This information is not applicable. This is a physical medical device, not a machine learning model, so there is no "training set."

    9. How the ground truth for the training set was established

    This information is not applicable as there is no training set for a machine learning model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1