(65 days)
Joint replacement is indicated for patients suffering from disability due to:
- non inflammatory degenerative joint disease including osteoarthritis and avascular necrosis . of the natural femoral head
- rheumatoid arthritis ●
- correction of functional deformity ●
- femoral fracture .
This device may also be indicated in the salvage of previously failed surgical attempts.
The constrained acetabular component is indicated for primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuronuscular disease, or intra-operative instability and for who all other options to constrained acetabulati components have been considered.
The device is intended for cementless applications.
This device has an asymmetrical body with an anterior build up that requires right and left components. When a symmetrical implant is used a large amount of cancellus bone often remains in the anterior metaphaseal portion of the femur. The anterior built en paovodes a more complete fill of the proximal femur. The proximal body is trapezoidal in cross-sectional geometry and tapers proximal to distal and lateral to medial. The distal stem portion is slightly conical in shape. The Revelation stem is collarless and has a Morse type taper to receive modular heads. The purpose of this application is to add an alternate porous coating.
The provided 510(k) summary is for a medical device called the "Revelation Porous Coated Hip Stem, Size 8," which is a hip implant. This document focuses on demonstrating substantial equivalence to predicate devices through design features, materials, indications, sterilization, packaging, and intended use, rather than presenting a study that establishes acceptance criteria for "device performance" in terms of algorithm metrics or human-AI effectiveness.
Therefore, many of the requested categories for AI/algorithm-centric studies are not applicable to this 510(k) submission.
Here's the breakdown of the information that can be extracted or inferred from the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
This type of information is typically for software or AI-driven devices where performance metrics like sensitivity, specificity, or reader accuracy are assessed against pre-defined thresholds. For this hip stem, "acceptance criteria" are not framed in terms of performance metrics like those for an AI device. Instead, the acceptance is based on demonstrating substantial equivalence to legally marketed predicate devices through comparable features and non-clinical testing.
| Acceptance Criteria Category | Reported Device Performance (or Demonstration of Equivalence) |
|---|---|
| Porous Coating Characterization | Non-clinical testing performed. (Details of specific characteristic values and acceptance thresholds are not provided in this summary.) |
| Fatigue Testing | Non-clinical testing performed. (Details of specific fatigue limits and acceptance thresholds are not provided in this summary.) |
| Ceramic Burst Assessment | Non-clinical testing performed. (Details of specific burst pressures/strengths and acceptance thresholds are not provided in this summary.) |
| Design Features | Comparable to predicate devices. |
| Materials | Comparable to predicate devices. |
| Indications for Use | Comparable to predicate devices. |
| Sterilization | Comparable to predicate devices. |
| Packaging | Comparable to predicate devices. |
| Intended Use | Comparable to predicate devices. |
2. Sample size used for the test set and the data provenance
Not applicable. This device is a physical implant, not an AI or software device that uses a "test set" of data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. Ground truth as typically understood for AI/software (e.g., radiologist consensus on images) is not relevant for this physical device. The "ground truth" equivalent would be the established engineering and materials science standards for hip implants.
4. Adjudication method for the test set
Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a hip implant, not an AI-assisted diagnostic or therapeutic device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For this device, the "ground truth" for demonstrating safety and efficacy is based on:
- Established engineering and material standards for orthopedic implants, assessed through non-clinical laboratory testing (Fatigue testing, Porous coating characterization, Ceramic burst assessment).
- Predicate device history: The regulatory pathway relies heavily on the safety and effectiveness of already marketed "predicate devices" with similar designs, materials, and intended use.
8. The sample size for the training set
Not applicable. This is not an AI/ML device that requires a "training set."
9. How the ground truth for the training set was established
Not applicable.
Summary of Device and Regulatory Context:
This 510(k) submission is for a Revelation Porous Coated Hip Stem, Size 8, an orthopedic implant. The primary method for demonstrating its acceptance and marketability is through substantial equivalence to previously cleared predicate devices. This involves showing that the new device has comparable design features, materials, indications for use, sterilization, packaging, and intended use, along with supportive non-clinical testing (fatigue, porous coating characterization, ceramic burst assessment) to ensure it meets relevant physical and mechanical standards. No clinical testing was provided for this 510(k) submission, indicating that the FDA found the non-clinical data and comparison to predicates sufficient for clearance.
{0}------------------------------------------------
510(k) Summary
Date: October 2, 2012 -
�>
Manufacturer: Encore Medical, L.P. 9800 Metric Blvd Austin, TX 78758
NOV 2 2012
L
, TX 78758 .
Contact Person: Teffany Hutto Manger, Regulatory Affairs Phone: (512) 834-6255 Fax: (512) 834-6313 Email: teffany.hutto@djoglobal.com
| Product | Classification | Product Codes | Product Code | Regulation and Classification Name |
|---|---|---|---|---|
| Revelation Porous Coated Hip Stem, Size 8 | Class II | LPH, MBL, LZO, KWZ | LPH | Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis per 21 CFR888.3358 |
| MBL | Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis per 21 CFR888.3358 | |||
| LZO | Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.per 21 CFR 888.3353 | |||
| KWZ | Hip joint metal/polymer constrained cemented or uncemented prosthesis per 21 CFR 888.3310 |
Description: This device has an asymmetrical body with an anterior build up that requires right and left components. When a symmetrical implant is used a large amount of cancellus bone often remains in the anterior metaphaseal portion of the femur. The anterior built en paovodes a more complete fill of the proximal femur. The proximal body is trapezoidal in cross-sectional geometry and tapers proximal to distal and lateral to medial. The distal stem portion is slightly conical in shape. The Revelation stem is collarless and has a Morse type taper to receive modular heads. The purpose of this application is to add an alternate porous coating.
Indications for Use:
Joint replacement is indicated for patients suffering from disability due to:
- non inflammatory degenerative joint disease including osteoarthritis and avascular necrosis . of the natural femoral head
- rheumatoid arthritis ●
- correction of functional deformity ●
- femoral fracture .
This device may also be indicated in the salvage of previously failed surgical attempts.
The constrained acetabular component is indicated for primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuronuscular disease, or intra-operative instability and for who all other options to constrained acetabulati components have been considered.
The device is intended for cementless applications.
{1}------------------------------------------------
K122636 (pg. 2 of 2)
Predicate Device:
ಿ
DJO Surgical Revelation Hip Stem, Size 8 – K994070 DJO Surgical Revelation Hip Stem with P2 Coating – K081679
Comparable Features to Predicate Device(s): Features comparable to predicate devices include the same design features, materials, indications, sterilization, packaging and intended use.
Non-Clinical Testing: Fatigue testing, Porous coating characterization, Ceramic burst assessment.
Clinical Testing: None provided.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is written in all capital letters.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Letter Dated: November 2, 2012
Encore Medical, L.P. % Ms. Teffany Hutto Manager, Regulatory Affairs 9800 Metric Boulevard Austin, Texas 78758
Re: K122636
Trade/Device Name: Revelation Porous Coated Hip Stem, Size 8 Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, MBL, LZO, KWZ Dated: October 2, 2012
Received: October 3, 2012
Dear Ms. Hutto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 – Ms. Teffany Hutto
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K122636
Device Name: Revelation® Porous Coated Hip Stem, Size 8
Indications for Use:
Revelation® Porous Coated Hip Stem, Size 8 Indications for Use
Joint replacement is indicated for patients suffering from disability due to:
- non inflammatory degenerative joint disease including osteoarthritis and avascular necrosis . of the natural femoral head
- rheumatoid arthritis ●
- correction of functional deformity .
- femoral fracture .
This device may also be indicated in the salvage of previously failed surgical attempts
The constrained acetabular component is indicated for primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for who all other options to constrained acetabular components have been considered.
The device is intended for cementless applications.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Oft) (Division Sign-On)
Division of Surgical, Orthopedic.
Division of Surgical, Dovices Division or and Restorative Devices
510(k) Number .
N/A