K Number
K122600
Date Cleared
2012-10-11

(45 days)

Product Code
Regulation Number
870.5310
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Defibrillation is a recognized means of terminating certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. .

The defibrillator is to be used in AED mode only on patients who are in cardiopulmonary arrest. The patient must be unresponsive, not breathing normally, and showing no signs of circulation.

The defibrillator may be used with standard defibrillation pads only on adults and children who are 8 years old or more or who weigh more than 25 kg (55lbs). The defibrillator may be used on children who are less than 8 years old or weigh less than 25kg (55lbs) with Infant/Child Reduced Energy Defibrillation Electrodes.

ECG Monitoring is for use on conscious and unconscious patients of all ages for the purpose of ECG rhythm recognition and heart rate monitoring

Device Description

The LIFPAK 1000 defibrillator is a semi-automatic defibrillator with optional manual mode and ECG display. It is intended for use by personnel who are authorized by a physician/medical director and are trained in CPR and the use of the LIFEPAK 1000 defibrillator. There are 3 modes of operation for the LIFEPAK 1000 defibrillator: AED mode (automated external defibrillation), manual mode (operator ECG interpretation, operator control of charge and shock functions), and ECG mode (ECG display allows for rhythm and heart rate monitoring). The currently marketed LIFEPAK 1000 defibrillator is powered by a non-rechargeable lithium-manganese battery.

The proposed LIFEPAK 1000 defibrillator has the same features and functions as the predicate LIFEPAK 1000 defibrillator with an additional battery option - a rechargeable lithium-ion battery.

AI/ML Overview

The provided document is a 510(k) summary for the LIFEPAK 1000 defibrillator, seeking substantial equivalence to a previously cleared device. The key information is that the only change is the addition of a rechargeable lithium-ion battery option. The document explicitly states that the "defibrillation waveform, ECG analysis system, and monitoring and defibrillation features of the proposed LIFEPAK 1000 defibrillator are unchanged from the previously cleared LIFEPAK 1000 defibrillator."

This directly implies that no new clinical studies were conducted to prove device performance or meet acceptance criteria related to its core defibrillation or ECG analysis functions. The substantial equivalence is based on the fact that these critical functional components remain identical to the predicate device, and new testing was likely focused on the safety and performance of the new battery option (though specifics of this testing are not detailed in this summary).

Therefore, this document does not contain the information required to answer most of your questions about acceptance criteria and associated studies for the device's functional performance.

Here's a breakdown of what can and cannot be answered based on the provided text:

1. A table of acceptance criteria and the reported device performance:

  • Cannot be provided. The document states that the defibrillation waveform and ECG analysis system are unchanged from the predicate device. It does not provide acceptance criteria or performance metrics for these functions for the current submission, nor does it refer to specific studies done for this submission.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Cannot be provided. No clinical performance test set details are presented for the updated device in this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Cannot be provided. No clinical performance test set details are presented.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Cannot be provided. No clinical performance test set details are presented.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Cannot be provided. This device is an external defibrillator, not an AI-assisted diagnostic tool for human readers. No MRMC study is mentioned or relevant here.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Cannot be provided. While defibrillators have algorithms for AED mode, no specifics of standalone algorithm performance studies are described in this summary for the updated device. The document emphasizes the unchanged nature of the ECG analysis system from the predicate.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Cannot be provided. No specific studies requiring ground truth establishment are detailed for this submission.

8. The sample size for the training set:

  • Cannot be provided. No training set for an algorithm is discussed in this submission.

9. How the ground truth for the training set was established:

  • Cannot be provided. No training set is discussed.

Summary of available information:

  • Device: LIFEPAK 1000 defibrillator
  • Change: Addition of a rechargeable lithium-ion battery option.
  • Justification: Substantial equivalence based on the core functional components (defibrillation waveform, ECG analysis system, monitoring and defibrillation features) being unchanged from the predicate device (K042404).
  • Conclusion: The 510(k) notification argues that due to the only change being a battery option and the core functions remaining identical to a previously cleared device, the updated device is substantially equivalent to the predicate, implying that specific clinical performance studies for defibrillation efficacy or ECG analysis were not required for this particular submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo for Physio Control. The logo is in a black box with white text. The word "PHYSIO" is on the top line, and the word "CONTROL" is on the bottom line.

122600 Physio-Control Inc. LIFEPAK 1000 defibrillator 510(k) Premarket Notification

SECTION E -- 510(K) SUMMARY

OCT 11 2012

Submitter:

Physio-Control, Inc. 11811 Willows Road Northeast P.O. Box 97006 Redmond, Washington 98073-9706

Contact Person:

Michelle Ackermann Principal Regulatory Affairs Specialist Phone: 425-867-4744, Fax: (425) 867-4154

Date Summary Prepared:

August 24th, 2012

Device Trade Naíne:

LIFEPAK®1000 defibrillator

Device Common Name:

External defibrillator

Device Classification:

Classification NameClass
Low Energy DC-Defibrillator (Including Paddles), (21CFR870.5300, Product Code LDD)II
Automatic External Defibrillators (Non-Wearable) (21CFR870.5310, Product Code MKJ)III

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a black square with the words "PHYSIO CONTROL" stacked on top of each other in white, sans-serif font. The word "PHYSIO" is on the top line, and the word "CONTROL" is on the bottom line. The letters are all capitalized and evenly spaced within the square.

SECTION E - 510(K) SUMMARY

Predicate Devices:

The features and functions of the proposed LIFEPAK 1000 defibrillator are substantially equivalent to the previously cleared LIFEPAK 1000 defibrillator. The 510(k) number for the predicate LIFEPAK 1000 defibrillator is K042404 (cleared on 12/22/05).

Description:

The LIFPAK 1000 defibrillator is a semi-automatic defibrillator with optional manual mode and ECG display. It is intended for use by personnel who are authorized by a physician/medical director and are trained in CPR and the use of the LIFEPAK 1000 defibrillator. There are 3 modes of operation for the LIFEPAK 1000 defibrillator: AED mode (automated external defibrillation), manual mode (operator ECG interpretation, operator control of charge and shock functions), and ECG mode (ECG display allows for rhythm and heart rate monitoring). The currently marketed LIFEPAK 1000 defibrillator is powered by a non-rechargeable lithium-manganese battery.

The proposed LIFEPAK 1000 defibrillator has the same features and functions as the predicate LIFEPAK 1000 defibrillator with an additional battery option - a rechargeable lithium-ion battery.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo for Physio Control. The logo is in a black box with white text. The text is stacked vertically, with "PHYSIO" on top and "CONTROL" on the bottom.

SECTION E - 510(K) SUMMARY

Intended Use:

The LIFEPAK 1000 defibrillator is intended for use by personnel who are authorized by a physician/medical director and are trained in CPR and the use of the LIFEPAK 1000 defibrillator. The manual and ECG modes of the LIFEPAK 1000 defibrillator are intended for use by health care providers trained in ECG rhythm recognition.

Indications for Use:

Defibrillation

Defibrillation is a recognized means of terminating certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia.

The defibrillator is to be used in AED mode only on patients who are in cardiopulmonary arrest. The patient must be unresponsive, not breathing normally, and showing no signs of circulation.

The defibrillator may be used with standard defibrillation pads only on adults and children who are 8 years old or more or who weigh more than 25 kg (55lbs). The defibrillator may be used on children who are less than 8 years old or weigh less than 25kg (55lbs) with Infant/Child Reduced Energy Defibrillation Electrodes.

ECG Monitoring

ECG monitoring is for use on conscious and unconscious patients of all ages for the purpose of ECG rhythm recognition and heart rate monitoring

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo for Physio-Control. The logo is in a black square. The text "PHYSIO" is on the top line, and the text "CONTROL" is on the bottom line.

SECTION E - 510(K) SUMMARY

Technological characteristics of the proposed and predicate device:

The only difference between the predicate LIFEPAK 1000 defibrillator and the proposed LIFEPAK 1000 defibrillator is a new battery option. The defibrillation waveform, ECG analysis system, and monitoring and defibrillation features of the proposed LIFEPAK 1000 defibrillator are unchanged from the previously cleared LIFEPAK 1000 defibrillator.

Conclusion of Testing

The information in this 510(k) notification demonstrates that the LIFEPAK 1000 defibrillator is substantially equivalent to the predicate LIFEPAK 1000 device with respect to safety, effectiveness, and performance.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings spread, rendered in blue lines. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAR 2 0 2013

Ms. Michelle Ackermann Principal Regulatory Affairs Specialist Physio-Control, Inc. 11811 Willows Road Northeast P.O. Box 97006 Redmond, WA 98073-9706

Re: K122600

Trade/Device Name: LIFEPAK 1000 defibrillator Regulation Number: 21 CFR 870.5310 Regulation Name: Automatic External Defibrillator (Non-Wearable) Regulatory Class: Class III Product Code: MKJ Dated: August 24, 2012 Received: August 27, 2012

Dear Ms. Ackermann:

This letter corrects our substantially equivalent letter of October 11, 2012.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{5}------------------------------------------------

Page 2 - Ms. Michelle Ackermann

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

M.g. Hilleben

Ja Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Physio Control. The logo is in black and white, with the words "PHYSIO" stacked on top of the word "CONTROL". The text is in a simple, sans-serif font. The logo is simple and recognizable.

SECTION D - STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known): not yet assigned K ) ≥2600

Device Name: LIFEPAK® 1000 defibrillator

Indications For Use:

Defibrillation

Defibrillation is a recognized means of terminating certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia. .

The defibrillator is to be used in AED mode only on patients who are in cardiopulmonary arrest. The patient must be unresponsive, not breathing normally, and showing no signs of circulation.

The defibrillator may be used with standard defibrillation pads only on adults and children who are 8 years old or more or who weigh more than 25 kg (55lbs). The defibrillator may be used on children who are less than 8 years old or weigh less than 25kg (55lbs) with Infant/Child Reduced Energy Defibrillation Electrodes.

ECG Monitoring

ECG monitoring is for use on conscious and unconscious patients of all ages for the purpose of ECG rhythm recognition and heart rate monitoring

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1(Division Sign-Off)Division of Cardiovascular Devices
510(k) NumberK122600D-1

§ 870.5310 Automated external defibrillator system.

(a)
Identification. An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.(b)
Classification. Class III (premarket approval)(c)
Date PMA or notice of completion of PDP is required. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED that was in commercial distribution before May 28, 1976. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976. Any other AED and AED accessory described in paragraph (a), shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.