K Number
K122278

Validate with FDA (Live)

Date Cleared
2012-11-07

(100 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Baylis Medical Company Radiofrequency Perforation Generator is indicated for use in general surgical procedures to cut and coagulate soft tissue.

Device Description

The Baylis Medical Company (BMC) Radiofrequency Perforation Generator is used with a compatible, separately-cleared radiofrequency (RF) device. The RF device is connected to the generator through the appropriate BMC Connector Cable, and delivers power in a monopolar mode between its distal tip electrode and a commercially available external dispersive electrode. An optional footswitch (Model: RFA-FS) may be used with the generator.

AI/ML Overview

The provided text is a 510(k) summary for the Baylis Medical Company Radiofrequency Perforation Generator. It describes the device, its intended use, and its substantial equivalence to a predicate device. However, it does not contain the detailed study information (like acceptance criteria, specific performance metrics, sample sizes for test/training sets, expert qualifications, or MRMC studies) that you've requested.

The document indicates that "substantial equivalence of predicate and current generators is supported by the device similarities and the results of verification and validation testing." It also lists the standards to which the device conforms (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, and ISO 14971).

To provide the information you've requested, one would need access to the full "verification and validation testing" report, which is not included in this 510(k) summary. These summaries are typically high-level documents highlighting the outcome of such testing, not the raw data or detailed methodologies.

Therefore, I cannot populate the table or answer the specific questions about the study that proves the device meets acceptance criteria based solely on the provided text. The document confirms that testing was done and conforms to standards, but does not describe the specifics of those tests, the acceptance criteria, or the results in the granular detail requested.

{0}------------------------------------------------

K122278

BMC Radiofrequency Perforation Generator

Baylis Medical Company Inc

6.0 510(k) Summary

V 7 ns

Submitter Information

A. Company Name: Baylis Medical Company Inc.

B. Company Address: 2645 Matheson Blvd. East Mississauga, Ontario L4W 5S4 Canada

C. Company Phone: (905) 602-4875; ext 252

D. Company Facsimile: (905) 602-5671

E. Contoct Person: Meghal Khakhar

F. Summary Prepared on: October 09, 2012

Device Identification

A. Device Trade Name: Baylis Medical Company Radiofrequency Perforation Generator, Model RFP-100A and optional footswitch (Model: RFA-FS)

B. Device Common Name: Electrosurgical Generator and optional footswitch

C. Classification Name: Electrosurgical Cutting and Coagulating Device 21 CFR 878.4400

D. Device Class: Class II

E. Device Code: GEI, General and Plastic Surgery

Identification of Predicate Device

  • Device: BMC Radio Frequency Perforation Generator, Model RFP-100 510(k) Number: K013904 Manufacturer: Baylis Medical Company Inc.

Device Description

The Baylis Medical Company (BMC) Radiofrequency Perforation Generator is used with a compatible, separately-cleared radiofrequency (RF) device. The RF device is connected to the generator through the appropriate BMC Connector Cable, and delivers power in a monopolar

{1}------------------------------------------------

mode between its distal tip electrode and a commercially available external dispersive electrode. An optional footswitch (Model: RFA-FS) may be used with the generator.

Intended Use

ﺎﺭ

The BMC Radiofrequency Perforation Generator is indicated for use in general surgical procedures to cut and coagulate soft tissue.

Substantial Equivalence

.

The indication for use and fundamental scientific technology of the current and predicate BMC Radiofrequency Perforation Generators is the same. The substantial equivalence of predicate · and current generators is supported by the device similarities and the results of verification and validation testing.

The BMC Radiofrequency Perforation Generator conforms to the following standards:

StandardTitle
IEC 60601-1Medical Electrical Equipment--Part 1: General Requirements for Safety
IEC 60601-1-2Medical Electrical Equipment--Part 1: General Requirements for Safety -Section 2: Collateral Standard--Electromagnetic Compatibility -Requirements and Tests
IEC 60601-2-2Medical Electrical Equipment--Part 2-2: Particular Requirements for theSafety of High Frequency Surgical Equipment
ISO 14971Medical Devices – Application of Risk Management to Medical Devices

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular fashion around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Letter Dated: November 7, 2012

Baylis Medical Company, Incorporated % Meghal Khakhar Director, Regulatory and Scientific Affairs 2645 Matheson Boulevard East Mississauga, Ontario L4W 5S4 Canada

Re: K122278

. •

Trade/Device Name: Baylis Medical Company Radiofrequency Perforation Generator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: October 09, 2012 Received: October 10, 2012

Dear Meghal Khakhar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Meghal Khakhar

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mark N. Melkerson

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: Baylis Medical Company Radiofrequency Perforation Generator

Indications for Use:

The Baylis Medical Company Radiofrequency Perforation Generator is indicated for use in general surgical procedures to cut and coagulate soft tissue.

Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Surgical, Orthop. and Restorative DevicesPage 1 of 1
510(k) NumberK122278

Vol 005-Page 2

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.