K Number
K122086

Validate with FDA (Live)

Date Cleared
2012-08-23

(38 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MiSeal Reposable Thermal Ligating Shears are intended for the simultaneous cutting and cauterization of soft tissue during surgery, and cutting of natural or synthetic, nonmetallic sutures during surgery.

Device Description

The MiSeal Reposable Thermal Ligating Shears system consists of the following: MiScal Reusable Handpiece, MiScal Thermal Ligating Shears Kit, Universal Power Supply 200-006R. The MiSeal Reposable Thermal Ligating Shears are designed to provide thermal ligation and division in various surgical procedures. The MiSeal Reposable Thermal Ligating Shears consist of a reusable handpiece with heating elements at the distal tip which are activated by a finger switch located on the handpiece of the device. The MiSeal Reposable Thermal Ligating Shears are designed to allow the surgeon control of the heating element power of the device in order to accommodate the individual patient anatomy. An instrument cord connects the handpiece to the dedicated Microline Surgical Universal Power Supply K070871. The MiSeal Reposable Handpiece is supplied non-sterile in a foam cavity placed in a fiberboard carton and is for multiple patient uses following cleaning and steam sterilization procedures performed per the provided Instructions for Use. The handpiece can be used multiple times if cleaning and sterilization procedures are followed. The MiSeal Disposable Kit (the functional instrument Tip and power cable portion of the applied part of the system) is supplied sterile in a die-cut chipboard Packaging Insert and Tyvek/Mylar pouch and is labeled for single use only. The tip is intended to be used by a trained physician for a single patient use in open general surgery, open vascular surgery, and laparoscopic surgical procedures. The MiSeal single use power cable is connected to the instrument handpiece and terminates at the electrical connection of the Universal Power Supply. The system power supply is supplied non-sterile for reusable use outside the sterile field. The MiSeal device incorporates Hi and Low heating modes that are used to coagulate and cut soft tissue. The heating elements in the disposable tip are activated by a physician controlled finger switch located on the handpiece of the device. The MiSeal device is intended to provide general purpose dissection, spreading, and grasping of soft tissue during minimally invasive or open surgical procedures.

AI/ML Overview

Here's an analysis of the provided text regarding the MiSeal Reposable Thermal Ligating Shears, focusing on acceptance criteria and the study proving it:

Based on the provided document (K122086 510(k) Summary), the information is primarily about establishing substantial equivalence to a predicate device, as required for a 510(k) submission to the FDA. This type of submission generally does not include detailed acceptance criteria or extensive clinical study results in the same way a PMA (Pre-Market Approval) or de novo submission might.

The document states: "Preclinical and performance tests were performed to assure the MiSeal Reposable Thermal Ligating Shears functioned as intended and met all product specifications. Sufficient data was generated and analyzed to prove that the MiSeal Reposable Thermal Ligating Shears was substantially equivalent to the predicate device."

This indicates that internal performance specifications were met, and the key acceptance criterion for the FDA submission was demonstrating substantial equivalence to the predicate device (Starion Instruments Thermal Ligating Shears, K062257). However, the specific details of these internal "product specifications" and the data generated are not provided in this summary.

Therefore, I cannot populate all the requested fields with specific details. Here's what can be inferred or stated as "Not Provided" based on the document:


Acceptance Criteria and Device Performance Study

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Substantial Equivalence to Predicate Device (Starion Instruments Thermal Ligating Shears, K062257) in function, intended use, and indications for use."Sufficient data was generated and analyzed to prove that the MiSeal Reposable Thermal Ligating Shears was substantially equivalent to the predicate device." (Specific quantitative performance metrics are not detailed).
Functioned as intended."Preclinical and performance tests were performed to assure the MiSeal Reposable Thermal Ligating Shears functioned as intended." (Specific functional parameters or targets are not detailed).
Met all product specifications."Preclinical and performance tests were performed to assure the MiSeal Reposable Thermal Ligating Shears... met all product specifications." (Specific product specifications are not detailed).
Intended Use: Simultaneous cutting and cauterization of soft tissue, and cutting of natural or synthetic, nonmetallic sutures.The device is "intended for the simultaneous cutting and cauterization of soft tissue during surgery, and cutting of natural or synthetic, nonmetallic sutures during surgery." The performance testing supports this intended use through substantial equivalence.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not provided. The document mentions "Preclinical and performance tests" but does not specify sample sizes for these tests (e.g., number of tissue samples, number of devices tested).
  • Data Provenance: Not provided specific details. The tests were likely conducted internally by Microline Surgical or by a contracted lab. The document does not mention country of origin of data or whether it was retrospective or prospective. Given the nature of a 510(k), these would primarily be laboratory/bench tests or, at most, animal studies, rather than human clinical data in the traditional sense.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable/Not provided. The summary describes "Preclinical and performance tests" which are typically laboratory-based engineering evaluations (e.g., tensile strength, burst pressure, temperature profiles). Ground truth in this context would be objective physical measurements rather than expert human interpretation.

4. Adjudication method for the test set:

  • Not applicable/Not provided. This concept of "adjudication" (e.g., 2+1, 3+1) is typically relevant for studies involving human interpretation or clinical endpoints that require consensus among experts, such as in diagnostic imaging or pathology. For device performance testing described, direct measurements against specifications would be the primary evaluation.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This device is an electrosurgical system, not an AI software/diagnostic device. MRMC studies are not relevant for this type of product.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is a surgical instrument, not an algorithm or AI system. Its performance is inherent in its physical and electrical properties when operated by a surgeon.

7. The type of ground truth used:

  • Laboratory/Bench Test Results (Objective Measures): Likely based on defined physical and electrical property measurements (e.g., tissue sealing strength, burst pressure, cutting time, temperature, power output, device integrity under stress) as per specified testing protocols.
  • Substantial Equivalence: The ultimate "ground truth" for the FDA submission is a comparison against the performance and characteristics of the predicate device.

8. The sample size for the training set:

  • Not applicable. This is a hardware medical device, not a machine learning or AI model that requires a "training set." Device design and development typically involve iterative testing and refinement, but not in the sense of a machine learning training set.

9. How the ground truth for the training set was established:

  • Not applicable. (See #8).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Microline Surgical. The logo has the word "MICROLINE" in large, bold, black letters. Below that is the word "SURGICAL" in smaller, black letters. There is a symbol between the two words that looks like four leaves arranged in a circle.

800 Cummings Center, Suite 166T Beverly, MA 01915, USA Tel: 978.922.9810 Fax: 978.922.9209 Web: www.microlinesurgical.com

KI22046 510(k) Summary

Page 1 of 3

Contact Person:Dean Ciporkin, Sr. Director of Regulatory Affairs and Quality Assurance
Submitted By:Microline Surgical800 Cummings Center,Beverly, MA 01915Tel: 978-922-9810Fax: 978-922-9209
AUG 2 3 2012
Common Name:Electrosurgical System
Device Product Code:GEI
Classification Name:Electrosurgical Cutting and Coagulation Device and accessories21 CFR § 787.4400
Device Panel:General Surgery/Restorative Devices :
Date Prepared:August 17, 2012
Proprietary Name:MiSeal™ Reposable Thermal Ligating Shears
Device Classification:Class II
Predicate Device:This product is similar in design, composition, and function to the:Starion Instruments Thermal Ligating Shears (K062257) cleared October 10, 2006
Establishment Registration Number:1223422

Device description and technological Characteristics:

The MiScal Reposable Thermal Ligating Shears system consists of the following:

MiScal Reusable Handpiece
MiScal Thermal Ligating Shears Kit
Universal Power Supply 200-006R

The MiSeal Reposable Thermal Ligating Shears are designed to provide thermal ligation and division in various surgical procedures. The MiSeal Reposable Thermal Ligating Shears consist of a reusable handpiece with a

surgical procedures. The MiSeal Reposable Thermal Ligating Shears surgious proceeding the reas heating elements at the distal tip which are activated by a finger switch located on

Page 1 of 3 Confidence, simply delivered.

Image /page/0/Picture/8 description: The image contains two logos. The logo on the left is the SGS logo, which features the letters "SGS" in a stylized font. The logo on the right is the UKAS logo, which features a crown and a checkmark. Below the checkmark are the letters "UKAS" and the number "006".

{1}------------------------------------------------

K 122086

Page 2 of ③

the handpiece of the device. The MiSeal Reposable Thermal Ligating Shears are designed to allow the surgeon control of the heating clement power of the device in order to accommodate the individual patient anatomy. An instrument cord connects the handpiece to the dedicated Microline Surgical Universal Power Supply K070871.

The MiSeal Reposable Handpiece is supplied non-sterile in a foam cavity placed in a fiberboard carton and is for multiple patient uses following cleaning and steam sterilization procedures performed per the provided Instructions for Use. The handpiece can be used multiple times if cleaning and sterilization procedures are followed. The MiSeal Disposable Kit (the functional instrument Tip and power cable portion of the applied part of the system) is supplied sterile in a die-cut chipboard Packaging Insert and 'I'yvek/Mylar pouch and is labeled for single use only. The tip is intended to be used by a trained physician for a single patient use in open general surgery, open vascular surgery, and laparoscopic surgical procedures.

The MiSeal single use power cable is connected to the instrument handpiece and terminates at the electrical connection of the Universal Power Supply. The system power supply is supplied non-sterile for reusable use outside the sterile field.

The MiSeal device incorporates Hi and Low heating modes that are used to coagulate and cut soft tissue. The heating elements in the disposable tip are activated by a physician controlled finger switch located on the handpiece of the device. "The MiSeal device is intended to provide general purpose dissection, spreading, and grasping of soft tissue during minimally invasive or open surgical procedures.

To coagulate and cut tissue, the physician grasps the desired tissue between the jaws of the MiScal Reposable Thermal Ligating Shears and gently squeezes the thumb trigger and handpiece to close the jaws. Depressing the finger switch and squeczing the thumb trigger activates the heating clements in the distal tip. An audible low frequency tone accompanies the activation of the heating clement in the variable mode to notify the physician the power is being applied. An audible high frequency tone will accompany the activation of the heating element in High power mode. Depressing the finger switch on the top of the handle provides cither Hi or Low heating mode. Generally, a lower UPS output setting improves the scaling capabilities and lengthens the time required to divide tissue. Conversely, a higher output setting reduces the time to divide tissue and may result in decreased vessel seal integrity

Indications for Use:

The MiSeal Reposable Thermal Ligating Shears are intended for the simultaneous cutting and cauterization of soft tissue during surgery, and cutting of natural or synthetic, nonmetallic sutures during surgery.

Performance Testing:

Page 2 of 3

Confidence, simply delivered.

Image /page/1/Picture/13 description: The image contains two logos. The logo on the left is the SGS logo, which features a stylized spiral design with the letters "SGS" underneath. The logo on the right includes a crown above the letters "UNAS" and "QVS".

{2}------------------------------------------------

K122081

Page 3 of 3

Preclinical and performance tests were performed to assure the MiSeal Reposable Thermal Ligating Shears functioned as intended and met all product specifications. Sufficient data was generated and analyzed to prove that the MiSeal Reposable Thermal Ligating Shears was substantially equivalent to the predicate device.

Summary:

The information provided dcmonstrates that the MiSeal Reposable Thermal Ligating Shears is substantially equivalent to the Starion Instruments Thermal Ligating Shears in function, intended use and indications for use.

Image /page/2/Picture/7 description: The image shows a circular design with four leaf-like shapes arranged symmetrically around a central point. Each leaf shape is curved and pointed at the ends, resembling petals or stylized leaves. The overall design has a simple, geometric quality, and the shapes are evenly spaced within the circle.

Image /page/2/Picture/8 description: The image contains two certification logos. On the left is the SGS certification mark, featuring a stylized eye-like shape with the letters "SGS" underneath. To the right is the UKAS (United Kingdom Accreditation Service) logo, which includes a crown-like symbol above the UKAS acronym and additional text below it. The logos suggest that the product or organization associated with them has met certain standards and has been accredited by these certification bodies.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three lines representing its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Microline Surgical % Mr. Dean Ciporkin Sr. Director of Regulatory Affairs and Quality Assurance 800 Cummings Center, Suite 166T Beverly, Massachusetts 01915

ANG 2 3 2012

Re: K122086

Trade/Device Name: MiSeal Reposable Thermal Ligating Shears Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 13, 2012 Received: July 16, 2012

Dear Mr. Ciporkin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration: Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Dean Ciporkin

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerso Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number (if known): K122086

Device Name: MiSeal Reposable Thermal Ligating Shears

Indications for Use:

The MiSeal Reposable Thermal Ligating Shears are intended for the simultaneous cutting and cauterization of soft tissue during surgery, and cutting of natural or synthetic, nonmetallic sutures during surgery.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTIÑUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R.P. Ogden mrm
Division Sign Off

sión Sign-( Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

510(k) Number K122086

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.