(157 days)
The CO2 monitoring lines are intended to connect from a CO2 sampling port to the expired gas monitor.
The monitoring line tubing is single use, small diameter tubing intended to be connected to a port on a face mask or breathing circuit to allow for gas sampling from a patient's breath by gas sampling equipment. When used with a face mask the line connects between the female luer lock connector on the mask and a capnograph or similar gas monitoring device. The gas monitoring device will have a pump which pulls air from inside mask, along the monitoring line and into the gas monitoring equipment. The monitoring line is available with or without a hydrophobic filter, which prevents the transfer of water down the monitoring line and into the gas sampling equipment.
Here's an analysis of the provided text regarding the acceptance criteria and the study proving the device meets them:
Disclaimer: This document describes a medical device called "C02 Monitoring Line and C02 Monitoring Line with Filter." The "study" described is a submission for 510(k) premarket notification, which demonstrates substantial equivalence to a predicate device, rather than a clinical study establishing efficacy or a standalone algorithm performance study. Therefore, some of the requested categories (like MRMC studies, human reader improvement, and training set details) are not applicable or cannot be extracted from this type of regulatory document.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated as numerical thresholds in this document, but rather implied by the comparison to a predicate device. The performance is reported by comparing the new device's characteristics to those of the predicate device (K946044). The "acceptance criteria" here implicitly mean that the new device's performance for these characteristics should be comparable or substantially equivalent to the predicate device to justify market clearance.
| Characteristic | Acceptance Criteria (Implied: Comparable to Predicate) | Reported Device Performance (CO2 Monitoring Line) | Reported Device Performance (CO2 Monitoring Line + Filter) | Predicate Device Performance (Gas Sampling Line, K946044) | Predicate Device Performance (Gas Sampling Line + Filter, K946044) |
|---|---|---|---|---|---|
| Intended Use | Same as predicate | Connect from C02 sampling port to expired gas monitor | Connect from C02 sampling port to expired gas monitor | Connect from port in face mask breathing circuit to expired gas monitor | Connect from port in face mask breathing circuit to expired gas monitor |
| Target Population | Same as predicate | Any patient from which gas monitoring is required | Any patient from which gas monitoring is required | Any patient from which gas monitoring is required | Any patient from which gas monitoring is required during the patient's breathing cycle |
| Where Used | Same as predicate | Hospitals | Hospitals | Hospitals | Hospitals |
| Single Use/Reusable | Same as predicate | Single Use | Single Use | Single Use | Single Use |
| Resistance to Flow | Comparable to predicate | 8.1mbar at 100ml/min flow; 20.7mbar at 300ml/min flow (for non-filter) | 26.3mbar at 100ml/min flow; 82.0mbar at 300ml/min flow (for filter) | 12.5mbar at 100ml/min flow; 35.9mbar at 300ml/min flow (for non-filter) | 14.6mbar at 100ml/min flow; 47.3mbar at 300ml/min flow (for filter) |
| Leakage | <1.0ml/min | <1.0ml/min | <1.0ml/min | <1.0ml/min | <1.0ml/min |
| Connectors | Same as predicate | 2 x Luer Lock connectors | 2 x Luer Lock connectors (plus luer lock Male to Female for filter) | 2 x Luer Lock connectors | 2 x Luer Lock connectors (plus luer lock Male to Female for filter) |
| Outer Diameter | Comparable to predicate | 3.05mm | 3.05mm | 2.95mm | 2.95mm |
| Inner Diameter | Comparable to predicate | 1.47mm | 1.47mm | 1.34mm | 1.34mm |
| Static Water Load | Does not allow water passage | Not applicable (for non-filter) | Does not allow passage of water past filter media | Not applicable (for non-filter) | Does not allow passage of water past filter media |
| Energy Used | Same as predicate | Gas pulled by pump in gas sampling device | Gas pulled by pump in gas sampling device | Gas pulled by pump in gas sampling device | Gas pulled by pump in gas sampling device |
| Compatibility | Same as predicate | Designed for use with gas monitoring device (e.g., capnograph) | Designed for use with gas monitoring device (e.g., capnograph) | Designed for use with gas monitoring device (e.g., capnograph) | Designed for use with gas monitoring device (e.g., capnograph) |
| Materials | Comparable to predicate | PVC (main line), PC (luer connectors) | PVC (main line), ABS (luer connectors & filter housing), Nylon (filter media) | PVC (main line), ABS (luer connectors) | PVC (main line), ABS (luer connectors) |
| Biocompatibility | Compliant with ISO 10993 | Compliant with ISO 10993 | Compliant with ISO 10993 | Compliant with ISO 10993 | Compliant with ISO 10993 |
| Sterility | Same as predicate | Non-sterile | Non-sterile | Non-sterile | Non-sterile |
| Luer Lock Conformance | Conform to ISO 594-2:1998 (clauses 4.2-4.7) | Conform to clauses 4.2-4.7 of ISO 594-2:1998 | Conform to clauses 4.2-4.7 of ISO 594-2:1998 | Not explicitly stated for predicate in this document, but implied by regulatory standard for such devices. | Not explicitly stated for predicate in this document, but implied by regulatory standard for such devices. |
Study Proving Device Meets Acceptance Criteria:
The study proving the device meets the "acceptance criteria" (which in this context is "substantial equivalence") is the nonclinical testing described in the document, specifically the "Summary of Testing" section.
2. Sample Sizes Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document does not specify exact sample sizes (e.g., number of lines tested) for each physical characteristic (resistance to flow, leakage, etc.). These are typically described in detailed test reports, not usually summarized in the 510(k) filing itself beyond stating that the tests were performed.
- Data Provenance: The data appears to be prospective testing conducted by the manufacturer, Intersurgical Incorporated, on their new devices. The country of origin for the data is not specified, but the manufacturer is based in Liverpool, NY, USA. However, given the predicate is from "Catheter Research Inc.," it implies the predicate data is also likely from the US, or at least from a study accepted by the FDA.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts
This information is not applicable and not provided in this document. The "ground truth" for nonclinical performance characteristics like resistance to flow or leakage is established by validated testing methodologies and metrology, not by expert consensus. This is a technical device characteristic evaluation, not an diagnostic or interpretative one.
4. Adjudication Method for the Test Set
This information is not applicable and not provided. Adjudication methods (like 2+1, 3+1) are relevant for studies where human interpretation of data (e.g., medical imaging, clinical symptoms) is involved, and discrepancies among observers need to be resolved to establish ground truth. For physical device performance tests, the outcome is typically measured objectively and does not require adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Reader Improvement with AI vs. Without AI Assistance
This information is not applicable. This document describes a C02 monitoring line, which is a physical accessory for gas monitoring equipment, not an AI-powered diagnostic or interpretive device. Therefore, MRMC studies, AI assistance, and human reader performance are not relevant to this submission.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This information is not applicable. As stated above, this is a physical medical device accessory, not an algorithm or software-as-a-medical-device (SaMD).
7. Type of Ground Truth Used
The "ground truth" for the performance characteristics like resistance to flow, leakage, dimensions, and biocompatibility is established through objective physical measurements and standardized laboratory testing (e.g., in accordance with ISO standards) using established metrology and chemical/biological analysis. It is not based on expert consensus, pathology, or outcomes data in the traditional sense.
8. Sample Size for the Training Set
This information is not applicable and not provided. As this is not an AI/machine learning device, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
This information is not applicable and not provided. There is no training set for this type of device.
{0}------------------------------------------------
Section 5 510(k) Summary of Safety and Effectiveness
DEC 2 0 2012
Name: INTERSURGICAL INCORPORATED Address: 417 Electronics Parkway Liverpool, NY 13088 Date: December 19, 2012 Contact Person: Michael Zalewski - Assistant VP - RA/CS Phone Number: 315-451-2900 Fax Number: 315-451-3696
Trade Name: C02 Monitoring Line and C02 Monitoring Line with Filter
Common Name: C02 Monitoring Line and C02 Monitoring Line with Filter
Classification: 21 CFR 868.1400, Classification Name: Analyzer, Gas, Carbon Dioxide, Gaseous Phase - Accessory - Classification Product Code: CCK
Predicate Device:
Catheter Research Inc., Disposable Gas Sampling Lines, K946044. The CRI monitoring lines are designed for connection between a gas monitoring device and a patient interface device in which gas monitoring is required. CRI gas sampling lines are available with or without a hydrophobic filter for the prevention of water transfer along the sampling line and into the gas sampling equipment.
Description:
C02 Monitoring Line and Monitoring Line with Filter
The monitoring line tubing is single use, small diameter tubing intended to be connected to a port on a face mask or breathing circuit to allow for gas sampling from a patient's breath by gas sampling equipment. When used with a face mask the line connects between the female luer lock connector on the mask and a capnograph or similar gas monitoring device. The gas monitoring device will have a pump which pulls air from inside mask, along the monitoring line and into the gas monitoring equipment. The monitoring line is available with or without a hydrophobic filter, which prevents the transfer of water down the monitoring line and into the gas sampling equipment.
Indications for Use:
The CO2 monitoring lines are intended to connect from a CO2 sampling port to the expired gas monitor
{1}------------------------------------------------
Section 5 510(k) Summary of Safety and Effectiveness (cont.)
Technological Characteristics Summary:
The intended use of the Intersurgical C02 Monitoring Line and Monitoring Line with Filter is comparable to the referenced predicate devices. The comparison of the data shows similar values for the key performance characteristics. Proposed devices show similar values for resistance to flow, leakage, connectors, outer diameter, inner diameter and static water load.
| CO2 Monitoring Line | Gas Sampling Line | |
|---|---|---|
| K946044 | ||
| Intended Use: | ||
| Target population | Any patient from which gasmonitoring is required | Any patient from which gasmonitoring is required |
| Indications for use | The C02 monitoring lines areintended to connect from a C02sampling port to the expired gasmonitor. | The gas sampling lines areintended to connect from a portin the face mask breathingcircuit to the expired gasmonitor. |
| Where used | Hospitals | Hospitals |
| Product Labelling | CO2 Monitoring Line | Gas Sampling Line |
| Single Use or Reusable? | Single Use | Single Use |
| Design and Performance: | ||
| Resistance to Flow | 8.1mbar at 100ml/min flow20.7ml/min at 300ml/min flow | 12.5mbar at 100ml/min flow35.9ml/min at 300ml/min flow |
| Leakage | <1.0ml/min | <1.0ml/min |
| Connectors | 2 x Luer Lock connectors | 2 x Luer Lock connectors |
| Outer Diameter | 3.05mm | 2.95mm |
| Inner Diameter | 1.47mm | 1.34mm |
| Energy Used/Delivered: | Gas is pulled from one end ofthe tube to the other by a pumpin the gas sampling device | Gas is pulled from one end ofthe tube to the other by a pumpin the gas sampling device |
| Compatibility: | Designed for use with gasmonitoring device (for examplea capnograph) with luerconnections to gas samplingtubing | Designed for use with gasmonitoring device (for examplea capnograph) with luerconnections to gas samplingtubing |
| Materials: | PVC (main line) and PC (luerconnectors) | PVC (main line) and ABS (luerconnectors) |
| Biocompatibility: | Compliant with ISO 10993 | Compliant with ISO 10993 |
| Sterility: | Non-sterile | Non-sterile |
| CO2 Monitoring Line + filter | Gas Sampling Line + filterK946044 | |
| Intended Use:Target population | Any patient, from which gas monitoring is required during the patient's breathing cycle | Any patient from which gas monitoring is required during the patient's breathing cycle |
| Indications for use | The C02 monitoring lines are intended to connect from a C02 sampling port to the expired gas monitor. | The gas sampling lines are intended to connect from a port in the face mask breathing circuit to the expired gas monitor. |
| Where used | Hospitals | Hospitals |
| Product Labelling | CO2 Monitoring Line2.45m + hydrophobic filter | Gas Sampling Line + filter |
| Single Use or Reusable? | Single Use | Single Use |
| Design and Performance:Resistance to Flow | 26.3mbar at 100ml/min flow82:0mbar at 300ml/min flow | 14.6mbar at 100ml/min flow47.3mbar at 300ml/min flow |
| Leakage | <1.0ml/min | <1.0ml/min |
| Connectors | 2 x Luer Lock connectors (plus luer lock Male to Luer lock Female connection between monitoring line and filter) | 2 x Luer Lock connectors (plus luer lock Male to Luer lock Female connection between monitoring line and filter) |
| Outer Diameter | 3.05mm | 2.95mm |
| Inner Diameter | 1.47mm | 1.34mm |
| Static water load | Does not allow passage of water past filter media | Does not allow passage of water past filter media |
| Energy Used/Delivered: | Gas is pulled from one end of the tube to the other by a pump in the gas sampling device | Gas is pulled from one end of the tube to the other by a pump in the gas sampling device |
| Compatibility: | Designed for use with gas monitoring device (for example a capnograph) with luer connections for gas sampling tubing | Designed for use with gas monitoring device (for example a capnograph) with luer connections for gas sampling tubing |
| Materials: | PVC (main line), ABS (luer connectors and filter housing) and Nylon (filter media) | PVC (main line) and ABS (luer connectors) |
| Biocompatibility: | Compliant with ISO 10993 | Compliant with ISO 10993 |
| Sterility: | Non sterile | Non sterile |
3
{2}------------------------------------------------
Section 5 510(k) Summary of Safety and Effectiveness (cont.)
4
{3}------------------------------------------------
Section 5 510(k) Summary of Safety and Effectiveness (cont.)
Summary of Testing:
Nonclinical tests submitted to demonstrate substantial equivalence for the Monitoring Lines and Monitoring Lines with Filter include resistance to flow, leakage, connectors, outer diameter, inner diameter and static water load. All materials used in the C02 Monitoring Lines and Monitoring Lines with Filter have been evaluated according to tests outlined in ISO 10993-1 and meet the requirements of Bluebook Memo, General Program Memorandum G95-1 biocompatibility testing for cytotoxicity, sensitization, and irritation. The monitoring line and hydrophobic filter conform to clauses 4.2 to 4.7 of ISO 594-2:1998, which are considered the 'functional' aspects of the luer lock design, ranging from separation and assembly forces to durability.
Substantial Equivalence:
Intersurgical Incorporated has demonstrated that the proposed devices are safe and effective. They are considered to be substantially equivalent to the currently marketed predicate devices which have been previously reviewed for market clearance by the FDA.
K122075
Premarket Notification [510(k)] Number
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002
December 20, 2012
Mr. Michael Zalewski Vice President-RA/OA/CS Intersurgical, Incorporated 417 Electronics Parkway LIVERPOOL NY 13077
Re: K122075
Trade/Device Name: Line and C02 Monitoring Line with Filter Regulation Number: 21 CFR 868.1400 Regulation Name: Carbon Dioxide Gas Analyzer Regulatory Class: II Product Code: CCK . . Dated: December 4, 2012 Received: December 11, 2012
Dear Mr. Zalewski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease be devices that I Dr over device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthong Lo m
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Section 4 Indications for Use Statement
510(k) Number (if known): K122075
Device Name:
C02 Monitoring Line and C02 Monitoring Line with Filter
Indications For Use:
The CO2 monitoring lines are intended to connect from a CO2 sampling port to the expired gas monitor.
Prescription Use XXX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lester W. Schultheis Jr
2012.12.19 16:44:53-05.00,
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
12075 510(k) Number,
§ 868.1400 Carbon dioxide gas analyzer.
(a)
Identification. A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.(b)
Classification. Class II (performance standards).