AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Optecure and Optecure +CCC are intended for use in combination with autogenous bone as a bone graft extender (extremities, spine and pelvis) and as bone void fillers (extremities and pelvis) for bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure. Optecure and Optecure + CCC can be used with autogenous bone marrow. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.

Optecure and Optecure +CCC may be used with rigid fixation systems.

Device Description

Optecure and Optecure + CCC are aseptically processed, single use, ready-to-mix devices intended for use as bone graft extenders (extremities, spine and pelvis) and as bone void fillers (extremities and pelvis) for bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.

Optecure products are provided in the form of a kit with pre-measured polymer powder, demineralized bone matrix (DBM), corticocancellous bone chips (CCC), pre-measured mixing solution and all the tools necessary to mix the components. After the powder is hydrated, the resultant putty can then be handled and placed in the appropriate bone voids or gaps.

Optecure products gradually resorb and are replaced with new bone during the healing process.

The purpose of this 510(k) submission is to expand the current Indications for Use so that Optecure and Optecure + CCC may be used in combination with autogenous bone marrow as a bone graft extender.

AI/ML Overview

Here's an analysis of the provided text regarding the Exactech® Optecure® and Optecure® + CCC Traditional 510(k), focusing on acceptance criteria and supporting studies.

It's important to note that this document is a 510(k) summary, which by design provides a high-level overview. Therefore, specific details about acceptance criteria, numerical performance metrics, and detailed study methodologies are generally not present as they would be in a full study report or detailed submission.


Description of Acceptance Criteria and Proving Device Meets Criteria

The document describes the Exactech® Optecure® and Optecure® + CCC as bone void fillers and bone graft extenders. The primary purpose of this 510(k) submission is to expand the Indications for Use for Optecure and Optecure + CCC to include combination with autogenous bone marrow as a bone graft extender.

The "acceptance criteria" here are not explicitly stated as quantitative thresholds for performance (e.g., "device must achieve X% success rate"). Instead, acceptance is determined by demonstrating substantial equivalence to legally marketed predicate devices. This means that the device must be shown to have the same or similar intended use, design, material composition, and functions as existing devices. The studies conducted are used to support this claim of substantial equivalence.

The "study that proves the device meets the acceptance criteria" refers to the preclinical testing mentioned, which aims to demonstrate that the new indications for use or the device itself perform comparably to predicates.


1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of a 510(k) summary for a bone void filler/graft extender, specific quantitative acceptance criteria and performance metrics (like sensitivity, specificity, etc., common in AI/diagnostic devices) are not provided. The acceptance criterion is "substantial equivalence" to predicate devices.

Acceptance Criterion (Implicit)Reported Device Performance (Summary)
Substantial Equivalence"Optecure® and Optecure® + CCC are substantially equivalent to the legally marketed predicate devices identified above in that they share either same or similar intended use, design, material composition and functions."
Osteoinductive PotentialSamples from each lot of donor demineralized bone matrix (DBM) are formulated with the carrier and tested for osteoinductivity in an in vivo athymic mouse assay. (Note: "Osteoinduction assay results using the athymic mouse assay should not be interpreted to predict clinical performance in human subjects.")
Handling PropertiesUnderwent "several preclinical testing including... handling properties." The results are "substantially equivalent to those of the predicate devices."
Bone Healing EffectivenessUnderwent "several preclinical testing including... bone healing effectiveness of the formulations in both tibial defect and spinal fusion animal models." The results are "substantially equivalent to those of the predicate devices." (Note: This is specifically for the formulations, i.e., the Optecure products, not necessarily human clinical outcomes.)
Safety and BiocompatibilityImplicitly demonstrated through substantial equivalence to legally marketed predicates. (No specific studies mentioned, but standard for such devices.)

2. Sample Size Used for the Test Set and Data Provenance

The document describes preclinical testing using animal models, specifically an in vivo athymic mouse assay for osteoinduction and "tibial defect and spinal fusion animal models" for bone healing effectiveness.

  • Sample Size for Test Set: Not specified. The document states "Samples from each lot of donor demineralized bone matrix (DBM)" for the osteoinduction assay, and "animal models" without specifying the number of animals or defects.
  • Data Provenance: The studies are preclinical (in vivo animal models). The country of origin for the studies is not explicitly stated. They are retrospective in the sense that they were conducted prior to this 510(k) submission to support it, but they are not clinical human data (prospective or retrospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This section is not applicable to this type of device and submission. The "ground truth" for bone void fillers and graft extenders in preclinical animal studies is typically established through direct observation, histology, imaging, and biomechanical testing by trained researchers/veterinarians, not by human expert readers adjudicating diagnostic outputs. The assessment criteria are objective biological and physical measurements, not subjective expert interpretations in the same way an AI diagnostic algorithm would be evaluated.


4. Adjudication Method for the Test Set

This section is not applicable. The preclinical studies described (animal models, in vivo assays) involve objective measurements and observations of biological responses (e.g., bone formation, healing rates) rather than human interpretation that requires adjudication (like reading medical images).


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant to diagnostic devices, particularly those involving human interpretation of images, sometimes with AI assistance. The device in question (bone void filler) is a therapeutic implantable material, not a diagnostic tool requiring human reader interpretation in clinical practice in the way an MRMC study would assess.


6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

No, a standalone performance study (algorithm only) was not done, and is not relevant. The device is a physical bone void filler, not an algorithm or AI system.


7. The Type of Ground Truth Used

The ground truth used in the preclinical studies would be based on:

  • Histology: Microscopic examination of tissue samples for evidence of new bone formation.
  • Imaging: X-rays, CT scans, or other imaging modalities to assess bone bridging, density, and defect filling in animal models.
  • Biomechanical Testing: Measuring the mechanical strength of the healed bone.
  • Direct Observation: Gross assessment of healing in animal models.
  • Biochemical markers: Where applicable for specific aspects of bone regeneration.

For the osteoinduction assay, the ground truth would be the observation of ectopic bone formation in the athymic mouse.


8. The Sample Size for the Training Set

This section is not applicable. The device is a medical product (bone void filler), not an AI algorithm that requires a training set in the machine learning sense. The "training" for such devices would be the iterative development and testing of the material formulation itself by the manufacturer.


9. How the Ground Truth for the Training Set Was Established

This section is not applicable for the same reasons as point 8.

{0}------------------------------------------------

K121989

Exactech® Optecure® and Optecure® + CCC Traditional 510(k)

510(k) Summary

Company:Exactech®, Inc2320 NW 66th CourtGainesville, FL 32653NOV 2 7 2012
Tel: 352-377-1140Fax: 352-378-2617E-mail: gina.cassidy@exac.com
Date of Submission:September 25, 2012
Contact Person:Gina CassidyDirector, Regulatory Affairs
Proprietary Name:Exactech® Optecure® and Optecure® + CCC
Common Name:Bone Void Filler
Classification Name:21CFR §888.3045 Resorbable calcium salt bone
Regulatory Class:II
Product Code:MQV, MBP
Classification Panel:Orthopedic

Legally Marketed Devices for Substantial Equivalence Comparison:

Produc CodeManufacturer510(k) NumberProduct
MQV, MBPExactech Inc.K050806Optecure
MQV, MBPExactech, Inc.K061668Optecure+CCC
MQV, MBPMusculoskeletalTransplantFounationK110003K113167MTF New Bone VoidFiller
MQV, MBPMusculoskeletalTransplantFounationK053218K063676K080399K103784K103795DBX DemineralizedBone Matrix Putty

{1}------------------------------------------------

Exactech® Optecure® and Optecure® + CCC Traditional 510(k)

Device Description/Purpose of Premarket Notification

Optecure and Optecure + CCC are aseptically processed, single use, ready-to-mix devices intended for use as bone graft extenders (extremities, spine and pelvis) and as bone void fillers (extremities and pelvis) for bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.

Optecure products are provided in the form of a kit with pre-measured polymer powder, demineralized bone matrix (DBM), corticocancellous bone chips (CCC), pre-measured mixing solution and all the tools necessary to mix the components. After the powder is hydrated, the resultant putty can then be handled and placed in the appropriate bone voids or gaps.

Optecure products gradually resorb and are replaced with new bone during the healing process.

The purpose of this 510(k) submission is to expand the current Indications for Use so that Optecure and Optecure + CCC may be used in combination with autogenous bone marrow as a bone graft extender.

Indications for Use

Optecure and Optecure +CCC are intended for use in combination with autogenous bone as a bone graft extender (extremities, spine and pelvis) and as bone void fillers (extremities and pelvis) for bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure. Optecure and Optecure + CCC can be used with autogenous bone marrow. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.

Optecure and Optecure +CCC may be used with rigid fixation systems.

Osteoinductive Potential

Samples from each lot of donor demineralized bone matrix (DBM) are formulated with the carrier and tested for osteoinductivity in an in vivo athymic mouse assay.

Osteoinduction assay results using the athymic mouse assay should not be interpreted to predict clinical performance in human subjects.

Preclinical testing

Optecure and Optecure + CCC formulations underwent several preclinical testing including, but not limited to, testing on the handling properties, osteoinductive potential, and bone healing effectiveness of the formulations in both tibial defect and spinal fusion animal models. The results of these studies are substantially equivalent to those of the predicate devices identified in this 510(k) Summary.

{2}------------------------------------------------

K121989

Exactech® Optecure® and Optecure® + CCC Traditional 510(k)

Substantial Equivalence Conclusion

Optecure® and Optecure® + CCC are substantially equivalent to the legally marketed predicate devices identified above in that they share either same or similar intended use, design, material composition and functions.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

November 27, 2012

Exactech. Incorporated Ms. Gina Cassidy Director, Regulatory Affairs 2320 Northwest 66th Court Gainesville, Florida 32653

. Re: K121989

Trade/Device Name: Exactech® Optecure® and Optecure® + CCC Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV, MBP Dated: September 25, 2012 Received: September 26, 2012

Dear Ms. Cassidy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 " CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Gina Cassidy

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Ronald P. Jean

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Exactech® Optecure® and Optecure® + CCC Traditional 510(k)

Indications for Use Statement

510(k) Number:

121189

Device Name: Exactech® Optecure® and Optecure® + CCC

INDICATIONS FOR USE:

Optecure and Optecure +CCC are intended for use in combination with autogenous bone as a bone graft extender (extremities, spine and pelvis) and as bone void fillers (extremities and pelvis) for bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure. Optecure and Optecure + CCC can be used with autogenous bone marrow. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone.

Optecure and Optecure +CCC may be used with rigid fixation systems.

and/or

Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

Please do not write below this line - use another page if needed.

Concurrence of CDRH, Office of Device Evaluation (ODE)

510(k) Number

(Division Sign-Off)
Division of Orthopedic Devices

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.