K Number
K121932
Date Cleared
2012-08-29

(58 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

Device Description

The device is an automatic non-invasive blood pressure system. The device is battery powered and can also be powered from an IEC 60601-1 compliant AC adaptor. The device inflates a cuff with an integral pump, then deflates the cuff via an electronically controllable valve. During deflation the cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic pressure. The device is intended to be used with specified Omron cuffs in three sizes to encompass arms ranging from 9 to 17 inches in circumference. The device also detects the appearance of irregular heartbeats during measurement.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Omron HEM-7200-Z (BP742) device, based on the provided 510(k) summary:

1. Acceptance Criteria and Reported Device Performance

Acceptance CriterionDevice Performance (Reported)
Accuracy (Pressure)+/- 3 mmHg or 2% of reading
Accuracy (Pulse Rate)+/- 5%
Measurement Range (Pressure)0-299 mmHg
Measurement Range (Pulse Rate)40 to 180 bpm
Compliance with ANSI/AAMI/ISO 81060-2Met the standard in clinical accuracy testing

Study Proving Device Meets Acceptance Criteria

The primary study mentioned to demonstrate clinical accuracy is:

  • Clinical Accuracy Testing in accordance with ANSI/AAMI/ISO 81060-2.

2. Sample Size and Data Provenance for the Test Set

  • Sample Size: 85 adults.
  • Data Provenance: Not explicitly stated whether retrospective or prospective, or country of origin, but given it's a clinical accuracy study, it's typically prospective.

3. Number of Experts and Qualifications for Ground Truth

This information is not explicitly provided in the given text. For blood pressure clinical accuracy studies, "ground truth" (reference measurements) is usually established by trained observers using a certified sphygmomanometer and stethoscope, following specific protocols (e.g., European Society of Hypertension, British Hypertension Society, or AAMI standards). The expertise would relate to their training in blood pressure measurement, not necessarily a specific clinical specialty like "radiologist with 10 years of experience."

4. Adjudication Method for the Test Set

This information is not explicitly provided in the given text. In blood pressure accuracy studies, adjudication often involves multiple observers taking independent measurements and their readings being averaged or compared. It's not typically framed as a "2+1" or "3+1" like in image interpretation studies.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done. This type of study is typically associated with AI in interpretation tasks (e.g., radiology) where human readers are assisted by AI. The device in question is a standalone blood pressure monitor, not an AI assistance tool for human interpretation.

6. Standalone (Algorithm Only) Performance Study

  • Yes, a standalone performance study was done in the form of "clinical accuracy testing of the device in accordance with ANSI/AAMI/ISO 81060-2." This standard assesses the algorithm's performance against reference measurements, without human-in-the-loop interpretation or assistance.

7. Type of Ground Truth Used

  • Expert Consensus / Reference Measurement: The "ground truth" for blood pressure accuracy testing according to ANSI/AAMI/ISO 81060-2 standards involves comparing the device's readings to those taken by trained observers using a validated reference method (e.g., auscultatory method with a mercury sphygmomanometer or an equivalent standard). This is analogous to an "expert consensus" on the true blood pressure.

8. Sample Size for the Training Set

  • The text does not provide information regarding a separate "training set" sample size. The device uses an "oscillometric" technology with a "software algorithm" that is stated to be "identical to that used in the predicate Omron HEM-741CRELN 510(k) K052153." This implies that the algorithm was likely developed and "trained" (or validated during development) on prior data, but details of that specific dataset are not presented for this submission. The 85 adults refers to the test set for clinical validation.

9. How Ground Truth for the Training Set Was Established

  • This information is not explicitly provided in the given text. As noted above, the algorithm is stated to be identical to a predicate device's algorithm, suggesting its development and "training" occurred prior to this submission, likely following similar clinical validation methodologies for its initial development.

{0}------------------------------------------------

K121932

Premarket Notification 510(k) Section 5 - 510(k) Summary

510(k) Summary
Page 1 of 5
6/27/2012

AUG 2 9 2012 Omron Healthcare, Inc. 1925 West Field Court Tel - 847-247-5626 Lake Forest, IL 60045 USA Fax - 847-247-5626 Official Contact: Mirna DiPano- Director, Quality & Regulatory Proprietary or Trade Name: Model HEM-7200-Z (BP742) Common/Usual Name: Noninvasive blood pressure measurement system. Classification Name/Code: DXN - Noninvasive blood pressure measurement system. 21CFR 870.1130 Class II Device: Model HEM-7200-Z (BP742) Predicate Device: Omron - HEM-741CRELN - K052153

Device Description:

The device is an automatic non-invasive blood pressure system. The device is battery powered and can also be powered from an IEC 60601-1 compliant AC adaptor. The device inflates a cuff with an integral pump, then deflates the cuff via an electronically controllable valve. During deflation the cuff pressure is monitored and pulse waveform data is extracted. The extracted pulse waveform data is then analyzed by software which determines pulse rate, as well as systolic and diastolic pressure.

The device is intended to be used with specified Omron cuffs in three sizes to encompass arms ranging from 9 to 17 inches in circumference.

The device also detects the appearance of irregular heartbeats during measurement.

Intended User Over the counter

Patient Population

This device is intended for use on adults

{1}------------------------------------------------

510(k) Summary Page 2 of 5 . 6/27/2012

Indications for Use:

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population.

The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

Environment of Use:

Home

Contraindications

There are no known contraindications.

Summary of substantial equivalence

The HEM-7200-Z (BP742) was compared to the predicate HEM-741CRELN (K052153 ) as in the device comparison table below.

The software algorithm used for determining blood pressure and pulse rate in the HEM-7200-Z (BP742) is identical to that used in the predicate Omron HEM-741CRELN 510(k) K052153.

{2}------------------------------------------------

K121932

Premarket Notification 510(k)

Section 5—510(k) Summary

Omron HEM-7200-Z (BP742

510(k) Summary

Device Compariso

Omron Hem-7200-Z (BP742) 510(k) not knownOmron HEM-741CRELN 510(k) K052153Comment
Indications for UseThe device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population.The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population.The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.Identical
Patient PopulationAdultAdultIdentical
Environment of UseHomeHomeIdentical
PrescriptiveOTCOTCIdentical
Patient ConnectionYes via cuffYes via cuffIdentical
TechnologyOscillometricOscillometricIdentical
Measurement rangePressure: 0-299mmHgPulse rate: 40 to 180 bpmPressure: 0-299 mmHgPulse rate: 40 to 180 bpmIdentical
Accuracy or pressure indicator+/- 3 mmHg or 2% of reading+/-3 mmHg or 2% of readingidentical
Accuracy Pulse Rate+/-5%+/-5%Identical
Inflation MethodDC rolling pumpDC rolling pumpIdentical
Deflation MethodDynamic linear deflationDynamic linear deflationIdentical
Display TypeLCDLCDIdentical
Irregular pulse detectionYesYesIdentical
AuscultatoryNoNoIdentical
Power SourceAA battery or AC adapterAA battery or AC adapterIdentical
Operating ConditionsTemperature: 10° to 40° CHumidity: 15 to 90% RHTemperature: 10° to 40° CHumidity: 30 to 85% RHSimilar
Storage ConditionsTemperature: -20° to 60° CHumidity: 10 to 95% RHTemperature: -20° to 60° CHumidity: 10 to 95% RHIdentical
Dimensions141 mm (L) x 123 mm (W) × 85 mm (H)141 mm (L) x 121 mm (W) × 86 mm (H)Similar
Weight12 oz12 3/4 ozSimilar

PDF Page 17 of 526

{3}------------------------------------------------

510(k) Summary Page 4 of 5 6/27/2012

Differences Between Other Legally Marketed Predicate Devices

The OmronHEM-7200-Z (BP742) is viewed as substantially equivalent to the predicate device because: The HEM-7200-Z (BP742) uses the exact same technology and has substantially equivalent indications for use. The differences that exist between the devices are insignificant in the terms of safety or effectiveness.

Indications -

The indications for use are identical

Prescriptive - The HEM-7200-Z (BP742) and H HEM-741CRELN are both OTC.

Design and Technology - The HEM-7200-Z (BP742) has equivalent design and features as the predicate and has the identical technology to the predicate.

Performance and Specifications - The HEM-7200-Z (BP742) has equivalent specifications of performance as the predicate.

Compliance with standards - The HEM-7200-Z (BP742) and predicate device declare compliance with IEC 60601-1 and IEC 60601-1-2.

Materials -

The patient contact materials of the device (the cuffs) have been cleared in other 510(k) submissions as described in Section 15 - Biocompatibility. Materials are detailed in Section 15 as well.

Environment of Use -

The HEM-7200-Z (BP742) and predicate are both intended for home use

Patient Population -

The HEM-7200-Z (BP742) and predicate are both for adult populations

{4}------------------------------------------------

510(k) Summary Page 5 of 5 6/27/2012

Performance Testing:

We have performed bench tests and found that the HEM-7200-Z (BP742) met all requirements specifications and standards requirements and were found to be equivalent in comparison to the predicate. Testing includes the following:

  • . Verification Testing
  • . Testing for compliance to IEC 60601-1
  • . Testing for compliance to IEC 60601-1-2
  • . Testing for compliance to AAMI SP10
  • Comparative Testing to the Predicate .

Testing to insure clinical accuracy of the device in accordance with ANSI/AAMI/ISO 81060-2. This testing was performed on 85 adults with results showing compliance to the standard.

Conclusion

Omron maintains that the HEM-7200-Z (BP742) is substantially equivalent to the predicate HEM-741CRELN (K052153 ) in indications for use, patient population, and environment for use, technology characteristics, specifications / performance and compliance with international standards

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" surrounding the top half of the circle. The text is written in a sans-serif font and is arranged to follow the curve of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AUG 29 2012

Omron Healthcare, Inc. c/o Paul Dryden ProMedic, Inc. 24301 Woodsage Drive Bonita Springs, FL 34134

Re: K121932

Trade/Device Name: HEM-7200-Z (BP742) Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: II (two) Product Code: DXN Dated: June 27, 2012 Received: July 2, 2012

Dear Mr. Paul Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{6}------------------------------------------------

Page 2 - Mr. Paul Dryden

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

K121932

Indications for Use Statement

Page 1 of 1

510(k) Number: (To be assigned)

Device Name: HEM-7200-Z (BP742)

Indications for Use:

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population.

The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

Prescription Use

(Part 21 CFR 801 Subpart D)

or

Over-the-counter use XX (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) Number

Page 4.2 PDF Page 14 of 520

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).