K Number
K121931

Validate with FDA (Live)

Date Cleared
2012-10-19

(109 days)

Product Code
Regulation Number
874.1820
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This device is a manual surgical instrument for performance of dissection and nerve stimulation input to living tissue concurrently.

Device Description

The Stimulating Hemostat and other manual instruments are dissectors modified by insulation and application of a connecting pin for connection to nerve stimulation. The specific instruments cleared for this modification in this 510(k) are the following:

  • Scissors Hinge Hemostat, curved, fine
  • Box Hinge Hemostat, curved, fine
  • Scissors Hinge Hemostat, right angle
  • Mosquito Hemostat, curved, fine
  • Halsted Hemostat, curved
  • Crile Hemostat, curved
  • Kelly Hemostat, curved
  • Schnidt Hemostat, curved, fine
  • Gemini Hemostat, curved, fine
  • Petit Point Mixter, right angle
  • McCabe Dissector, curved, fine
  • Jacobson Mosquito, curved, fine
  • House Rosen Elevator, coarse
  • Sheehy "Weapon", Large (2.4 mm)
  • Probe, Ball tip
  • Suction Stimulator (Neurovision model)
AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Neurovision Dissection/Stimulating Accessory:

It's important to note that the provided documents (a 510(k) summary and the FDA clearance letter) describe a medical device that is a manual surgical instrument, not an AI/ML-based diagnostic or imaging device. Therefore, many of the typical acceptance criteria and study details relevant to AI devices, such as sensitivity, specificity, AUC, sample sizes for test/training sets, ground truth establishment by experts, and MRMC studies, are not applicable to this type of submission.

The "performance testing" described for this device focuses on reprocessing validation. The FDA clearance is based on the device being "substantially equivalent" to predicate devices, meaning it has similar technological characteristics and intended use, and does not raise new questions of safety or effectiveness.

Here's a breakdown based on your requested information, highlighting what is present and what is not applicable:


Neurovision Dissection/Stimulating Accessory Acceptance Criteria and Study Analysis

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (based on 510(k) regulatory pathway)Reported Device Performance (from submission)
Substantial Equivalence to Predicate Devices: - Same intended use. - Similar technological characteristics. - No new questions of safety or effectiveness raised.The device is "substantially equivalent" to the predicate devices (Stimulating Hemostat K895676; Nerve Locator Monitor K110140). Key Technological Similarities/Improvements: - Modified existing instruments with insulation and connection pin for nerve stimulation. - Uses same materials and method for insulating as predicate. - Insulation covers hinging and tips for more precise nerve localization and reduced shorting.
Reprocessing Validation: - Cleaning instructions validated. - Sterilization instructions validated.Reprocessing validation testing for cleaning and sterilization instructions was conducted according to: - ANSI/AAMI ST 79:2006 - AAMI TIR 12:2004 - AAMI TIR 30:2011 Conclusion: "These modified instruments present no additional issues with sterilization, safety or effectiveness."

2. Sample size used for the test set and the data provenance

  • Not Applicable. This is not an AI/ML or diagnostic device that uses a "test set" of data in the common sense. The "testing" referred to is engineering and reprocessing validation, not performance on a clinical dataset.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. See point 2. Ground truth in the context of expert review of images or diagnostic data is not relevant for this device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. See point 2.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is a manual surgical instrument, not an AI-assisted diagnostic or interpretative system. Therefore, MRMC studies and the concept of "human readers improving with AI assistance" are irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Not Applicable. This device is a manual surgical instrument that is always "human-in-the-loop," as its function is to be manually operated by a surgeon. It does not operate as an algorithm or standalone system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For the reprocessing validation, the "ground truth" would be regulatory standards and successful adherence to those standards, demonstrating the device can be safely cleaned and sterilized.
  • For the substantial equivalence claim, the "ground truth" is the established safety and effectiveness of the predicate devices and the demonstration that the new device does not alter these or introduce new risks, based on engineering analysis and comparison of materials/design.

8. The sample size for the training set

  • Not Applicable. This is not an AI/ML device that uses a "training set."

9. How the ground truth for the training set was established

  • Not Applicable. See point 8.

{0}------------------------------------------------

5. 510K Summary

K121931

P. 1 082

OCT 1 9 2012

In accordance with Title 21 of the Code of Federal Regulations, Part 807 and in particular 21 CFR §807.92, the following summary of information is provided:

Applicant Information

Christine Vergely Regulatory Affairs Manager Neurovision Medical Products, Inc. 2225 Sperry Ave., Suite 1000 Ventura, CA 93003 PH 805-866-6999 Fax: 413-410-4579 christie@neurovisionmedical.com

Date of Summary: 10/10/12

Device Identification:

Trade Name: Neurovision Stimulating Dissector Common Name: Neurovision Dissection/Stimulating Accessory Classification Name: Surgical nerve stimulator/locator, 21 CFR 874.1820 Class II, Product Code: ETN

Predicate Devices

Stimulating Hemostat (K895676); Nerve Locator Monitor (K110140)

Device Description

The Stimulating Hemostat and other manual instruments are dissectors modified by insulation and application of a connecting pin for connection to nerve stimulation. The specific instruments cleared for this modification in this 510(k) are the following:

  • Scissors Hinge Hemostat, curved, fine �
  • . Box Hinge Hemostat, curved, fine
  • Scissors Hinge Hemostat, right angle .
  • Mosquito Hemostat, curved, fine �
  • Halsted Hemostat, curved ●
  • Crile Hemostat, curved �
  • Kelly Hemostat, curved
  • Schnidt Hemostat, curved, fine ●
  • Gemini Hemostat, curved, fine ●
  • Petit Point Mixter, right angle �
  • McCabe Dissector, curved, fine .
  • Jacobson Mosquito, curved, fine ●
  • House Rosen Elevator, coarse ●
  • Sheehy "Weapon", Large (2.4 mm)
  • Probe, Ball tip
  • Suction Stimulator (Neurovision model)

{1}------------------------------------------------

K12.1931

p. 2 of 2

Intended Use

This device is a manual surgical instrument for performance of dissection and nerve stimulation input to living tissue concurrently.

Technological Characteristics of Device in Relation to Predicate Devices

The Stimulating Hemostat approved under K895676 was insulated only on its fingerings and arms but not over the box hinge or tip. Subsequent work has shown that hinged (both box and scissor type) devices can be insulated down to the tips, providing more precise localization of nerve and reduced problems with shorting of the instrument to the wound edges. For over 23 years, the additional instruments in this submission have been modified using the same materials to insulate, and the same method for insulating as the predicate. These modified instruments present no additional issues with sterilization, safety or effectiveness.

Performance Testing

The reprocessing validation testing used to create and validate the cleaning and sterilization instructions were conducted according to the standard ANSI/AAMI ST 79: 2006 and guidance documents AAMI TIR 12:2004 and AAMI TIR 30:2011.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

Neurovision Medical Products, Inc. c/o Ms. Christine Vergely Regulatory Manager 2225 Sperry Avenue, Suite 1000 Ventura, CA 93003

Re: K121931

Trade/Device Name: Neurovision Dissection/Stimulating Accessory Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: ETN Dated: September 18, 2012 Received: September 19, 2012

Dear Ms. Vergely:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

OCT 1 9 2012

{3}------------------------------------------------

Page 2 - Mr./Ms./Mrs./Dr.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Malvina B. Eydelman, M.D.

Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

4. Indications for Use:

Device Name: Neurovision Dissection/Stimulating Accessory

This device is a manual surgical instrument for performance of dissection and nerve stimulation input to living tissue concurrently.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K121931

Prescription Use
(Per 21 CFR 801.109)

§ 874.1820 Surgical nerve stimulator/locator.

(a)
Identification. A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.(b)
Classification. Class II.