K Number
K121858
Manufacturer
Date Cleared
2012-11-05

(132 days)

Product Code
Regulation Number
884.4160
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Acessa System is indicated for use in percutaneous, laparoscopic coagulation and ablation of soft tissue, including treatment of symptomatic uterine fibroids under laparoscopic ultrasound guidance.

Device Description

The Acessa System includes the following system components: Generator (Model Number 1000): Provides RF energy to the Handpiece through the Handpiece Cable. Handpiece (Model Number 2000): Consisting of a disposable handle with a trocar-pointed shaft and 7 deployable needle electrodes. For use only disposable electrosurgical devices provided by Halt Medical, Inc. Handpiece Cable (Model Number 4200): Connects Handpiece to the Generator. This extension cable is provided with the Generator. Pads (Model Number 3000): A disposable set of 2 units, providing the return path for the RF energy applied by the Handpiece. Use only the Pads provided by Halt Medical, Inc. Pad Cable (Model Number 4300): Connects the Pads to the Generator. This extension cable is provided with the RF Generator. Power Cord (Model Number 4110): A medical grade power cord that provides AC power to the Generator. The Power cord is provided with the Generator. Foot Pedal (Model Number 4100): Pneumatic foot pedal with tubing used to turn RF energy on and off. The Foot Pedal with tubing is provided with the Generator. The Acessa System is designed to deliver up to 200 W of RF power at 460 kHz in three operational modes: Temperature Control, Manual Control and Coagulation Mode. A touch screen with a graphical user interface (GUI) enables selection of operational parameters such as the mode of operation, the ablation time, the target temperature, and the power delivery level. With the Handpiece placed in the tissue to be ablated and its electrodes deployed, RF power can be turned on. The system parameters are continuously monitored and displayed on the Generator. If the measured parameters are outside the acceptable limits, the RF energy delivery automatically stops and a message appears on the graphical user interface. RF energy during an ablation or coagulation can also be stopped at any time by the user by pressing the foot pedal. NOTE: The Acessa System must be used under laparoscopic ultrasound guidance. Laparoscopic ultrasound equipment is not included with the Acessa System.

AI/ML Overview

The provided document describes the Acessa System, a device for laparoscopic coagulation and ablation of soft tissue, specifically for symptomatic uterine fibroids. Here's a breakdown of the acceptance criteria and the studies conducted:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a tabulated format with specific thresholds. Instead, it describes various performance testing and clinical study outcomes that demonstrate the device's safety and effectiveness. Based on the provided information, the closest approximation of acceptance criteria and reported performance is:

Acceptance Criteria (Inferred)Reported Device Performance (as demonstrated by studies)
Safety - Electrical, Mechanical, Software, and BiocompatibilityThe Acessa System passed all electrical, mechanical, and software validation testing, as well as applicable safety requirements (EN/IEC 60601-1, EN/IEC 60601-1-2, EN/IEC 60601-2-2). EMC and safety testing were completed on the entire device and per component. Biocompatibility testing demonstrated materials are safe and in compliance with ISO 10993-1, ISO 10993-5, and ISO 10993-10. Device-related adverse events were less than 4% in the pivotal study, with only 2 serious adverse events (a pelvic abscess and a serosal colon injury) considered related to the device among 10 reported serious events. Only one (1.4%) serious adverse event (abdominal wall hematoma) was considered procedure-related in the Phase II studies.
Functional Performance - Ablation Capability & ControlAnimal and bench ablation testing successfully demonstrated the Acessa System performs as intended and per specifications. Ablation capability was confirmed, providing a reproducible, discretely demarcated zone of tissue necrosis. The peri-hysterectomy feasibility study showed that ex vivo bench testing ablation results are predictive of ablations in fibroid tissue in vivo, with no device-related adverse events. The system continuously monitors parameters, automatically stops RF energy if outside acceptable limits, and allows user-initiated stops.
Effectiveness - Reduction in Menstrual Blood Loss (for symptomatic uterine fibroids)Phase II Studies: At 12 months post-treatment, 3.5% (p<0.0001) of subjects reported heavy to very heavy bleeding, down from 77% at baseline. Pivotal Study: Mean reduction in menstrual blood flow at 12 months post-treatment was 103.6 ml. 83.9% of subjects experienced a reduction in blood loss. 40.2% (95% CI 31.6% - 48.7%) met the bleeding relief criterion (≥50% reduction in menstrual bleeding at 12 months), though this was below the prespecified study hypothesis of ≥45% lower bound.
Effectiveness - Symptom Severity Reduction (for symptomatic uterine fibroids)Phase II Studies: Mean percent improvement in Symptom Severity Scores (UFS-QOL) was 59.6% at 3 months, 65.0% at 6 months, and 79.0% at 12 months. Overall, 94.2% reported improved Symptom Severity Scores. Pivotal Study: Continuous decrease in Symptom Severity Scores (SSS) from a high 61 at baseline to a low 24 at 24 months post-treatment. At least 94% of subjects responded that the treatment had been somewhat, moderately, and very effective in eliminating their symptoms.
Effectiveness - Improvement in Health-Related Quality of Life (for symptomatic uterine fibroids)Phase II Studies: Mean percent change in HRQL Scores from baseline was 41.6% at 3 months, 39.7% at 6 months, and 42.9% at 12 months. Overall, 88.4% reported improved HRQL Scores. Pivotal Study: Steady improvement in mean Health-Related Quality of Life (HRQL) scores (from a low 37 at baseline to 80 after 24 months post-treatment). Subjects reported a decrease in disease burden and improved health state with mean EQ-5D scores starting at 71 at baseline and rising to 85 after 24 months.
Effectiveness - Reduction in Uterine and Fibroid VolumePhase II Studies: Mean uterine volume significantly reduced to 155.0 cm³ (p=0.012) at 6 months and 151.4 cm³ (p=0.009) at 12 months, from 204.4 cm³ at baseline. 79.3% of subjects showed reduced uterine volume at 12 months. Pivotal Study: At 12 months post-treatment, mean reduction in uterine volume was 25.1%, and mean reduction in fibroid volume was 44.3% (assessed by contrast-enhanced MRI).
Effectiveness - Low Reintervention RatePivotal Study: Within 24 months, a cumulative reintervention rate for fibroid-related bleeding was <6% (6 out of 107 subjects).
Outpatient Minimally Invasive Option with Rapid Recovery and High Patient SatisfactionTreatment with this device provides an outpatient minimally invasive option that allows rapid recovery. 96% of subjects in the pivotal study were treated on an outpatient basis. Subjects returned to work in a median of 5 days, and all subjects reported returning to normal in a median of 9 days. 94% of subjects were "very, moderately, or somewhat satisfied" with the treatment. 98% of subjects would "probably or definitely recommend" the procedure to friends with the same health problem at 12 months.
Feasibility in Varied Cases (e.g., fibroid types/locations, calcified fibroids) & Predictability of AblationThe peri-hysterectomy feasibility study encountered various fibroid locations and types. The Phase II and pivotal studies also treated various fibroid types and locations. Seven subjects (5.1%) with calcified fibroids were treated, and calcified fibroids did not appear to impact the physician's ability to insert the Handpiece tip. Bench testing was predictive of in vivo ablation.

2. Sample Sizes Used for the Test Set and Data Provenance

The document details three main clinical studies that serve as the "test set" for the device's performance:

  • Peri-hysterectomy feasibility study:
    • Sample Size: Not explicitly stated, but implies a small cohort of patients undergoing hysterectomy where the Acessa System was used.
    • Data Provenance: Not explicitly stated, but likely prospective and from a clinical setting where hysterectomies were being performed.
  • Prospective, non-randomized, longitudinal Phase II studies:
    • Sample Size: A total of 69 subjects (implied from the phrase "sixty-five subjects (94.2%)" and "sixty-one subjects (88.4%)" which point to a denominator around 69).
    • Data Provenance: Prospective, non-randomized, conducted at "two separate centers," likely in the US (not explicitly stated country, but common for IDE trials).
  • Device pivotal (Phase III) study:
    • Sample Size: 137 women with symptomatic fibroids.
    • Data Provenance: Prospective, conducted at 11 centers in the United States (9 centers) and Latin America (2 centers).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not describe a ground truth established by a specific number of experts for image interpretation or diagnosis in the traditional sense of AI studies. Instead, the "ground truth" for the clinical effectiveness studies was derived from:

  • Patient-reported outcomes: Validated questionnaires (UFS-QOL, EQ-5D), menstrual pad counts, patient surveys on satisfaction and symptom effectiveness, and self-reported return to normal activities.
  • Clinical measurements: Uterine volume, fibroid volume (assessed by contrast-enhanced MRI).
  • Pathology reports: From the peri-hysterectomy study, correlated with ablation zones.
  • Adverse event reporting: By investigators.
  • Investigators: 13 investigators participated in the pivotal study. Their specific qualifications (e.g., number of years of experience) are not detailed but they are presumed to be qualified clinicians (e.g., gynecologists) experienced in treating uterine fibroids.

4. Adjudication Method for the Test Set

The document does not mention an explicit adjudication method (like 2+1 or 3+1 consensus) for establishing a "ground truth" for the test set. The clinical trial data relies on patient reporting, objective clinical measurements, and investigator reporting of adverse events and outcomes based on standard clinical practice.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. The Acessa System is a therapeutic device (for ablation), not a diagnostic device that involves "human readers" interpreting images. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable to this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

No. The Acessa System is a physical medical device (generator, handpiece, etc.) used by a physician during a laparoscopic procedure. It is not an algorithm or AI system that operates in a standalone capacity without human interaction for its primary function. While it has internal software for control and monitoring, its performance is inherently tied to human use and guidance.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The ground truth for evaluating the Acessa System's performance was primarily a combination of:

  • Clinical Outcomes Data:
    • Patient-reported outcomes: From validated questionnaires (Uterine Fibroid Symptom Quality of Life (UFS-QOL), EuroQol 5 Dimensions (EQ-5D)), patient satisfaction surveys, symptom severity scales, and menstrual pad counts.
    • Objective clinical measurements: Reduction in uterine and fibroid volume (via contrast-enhanced MRI), reduction in menstrual blood loss, and reintervention rates.
    • Adverse event reporting: Clinical events reported by investigators and medical staff.
  • Pathology: In the peri-hysterectomy feasibility study, pathology reports of ablation zones were correlated with procedure documentation.
  • Device specifications: For electrical, mechanical, and software performance.

8. The Sample Size for the Training Set

The document does not describe a "training set" in the context of an AI/machine learning model. The studies described are clinical trials and engineering validations for a physical medical device. The "training" in this context would be the development and iterative testing of the device during its engineering phase, which is not quantified by a patient sample size but by laboratory and bench testing efforts.

9. How the Ground Truth for the Training Set Was Established

Since there is no "training set" in the AI/ML sense, this question is not directly applicable. For the device development itself, "ground truth" for design and engineering was established through:

  • Internal specifications and standards: Adherence to regulatory standards (EN/IEC 60601-1, EN/IEC 60601-1-2, EN/IEC 60601-2-2, ISO 10993).
  • Bench testing: To qualify and quantify mechanical and functional properties, ablation output, and safety features.
  • Animal studies: To demonstrate performance as intended and per specifications prior to human trials.

{0}------------------------------------------------

510(k) Summary

As Required By 21 CFR 807.92

5 2012 NOV

As Required By 21 CFR 807.92

Date of Summary PreparationNovember 2, 2012
Submitter and Owner's Name and AddressHalt Medical, Inc.131 Sand Creek Road, Suite BBrentwood, CA 94513Main: (925) 634-7943Fax: (925) 634-7841
Contact PersonClarisa A. TateSenior Director of RA & QA, Halt Medical, Inc.Office: (925) 271-0626e-mail: ctate@haltmedical.com
Trade NameThe Acessa System
Common NameThe Acessa System
Classification NameUnipolar endoscopic coagulator-cutter and accessories
ClassificationClass II
Product CodeHFG
Classification PanelObstetrics and Gynecology
Classification Regulation21 CFR §884.4160
Legally Marketed Device to whichsubstantial equivalence is claimedHalt 2000GITM Electrosurgical Radiofrequency AblationSystem (K094009)
Intended UseThe Acessa System is indicated for use in percutaneous, laparoscopic coagulationand ablation of soft tissue, including treatment of symptomatic uterine fibroidsunder laparoscopic ultrasound guidance.
DeviceDescriptionThe Acessa System includes the following system components:Generator (Model Number 1000): Provides RF energy to the Handpiecethrough the Handpiece Cable. Handpiece (Model Number 2000): Consisting of a disposable handlewith a trocar-pointed shaft and 7 deployable needle electrodes. For useonly disposable electrosurgical devices provided by Halt Medical, Inc. Handpiece Cable (Model Number 4200): Connects Handpiece to theGenerator. This extension cable is provided with the Generator. Pads (Model Number 3000): A disposable set of 2 units, providing thereturn path for the RF energy applied by the Handpiece. Use only thePads provided by Halt Medical, Inc. Pad Cable (Model Number 4300): Connects the Pads to the Generator.This extension cable is provided with the RF Generator. Power Cord (Model Number 4110): A medical grade power cord thatprovides AC power to the Generator. The Power cord is provided withthe Generator. Foot Pedal (Model Number 4100): Pneumatic foot pedal with tubingused to turn RF energy on and off. The Foot Pedal with tubing isprovided with the Generator.
The Acessa System is designed to deliver up to 200 W of RF power at 460 kHz inthree operational modes: Temperature Control, Manual Control and CoagulationMode. A touch screen with a graphical user interface (GUI) enables selection ofoperational parameters such as the mode of operation, the ablation time, thetarget temperature, and the power delivery level. With the Handpiece placed inthe tissue to be ablated and its electrodes deployed, RF power can be turned on.The system parameters are continuously monitored and displayed on theGenerator. If the measured parameters are outside the acceptable limits, the RFenergy delivery automatically stops and a message appears on the graphical userinterface. RF energy during an ablation or coagulation can also be stopped atany time by the user by pressing the foot pedal.NOTE: The Acessa System must be used under laparoscopic ultrasoundguidance. Laparoscopic ultrasound equipment is not included with the AcessaSystem.
Intended UseandTechnologicalCharacteristicsCompared toPredicateDevicesThe Acessa System and the Halt 2000GI System do not have the same Indicationsfor Use statement, as the Acessa System includes the specific indication of"treatment of symptomatic uterine fibroids."The difference between the Indications for Use statements of the Acessa Systemand Halt 2000GI System does not alter the intended therapeutic effect of thedevice. Like the Halt 2000GI System, the Acessa System uses coagulation orablation of soft tissue to achieve its therapeutic effect, (i.e., treatment ofsymptomatic uterine fibroids).The technological characteristics of the Acessa System are equivalent tocommercially available Halt 2000GITM System. It is the same product.
PerformanceTestingThe Acessa System was subjected to a battery of electrical, mechanical, andsoftware validation testing, as well as applicable safety requirements (EN/IEC60601-1, EN/IEC 60601-1-2, EN/IEC 60601-2-2). The system passed all testing.EMC and safety testing were completed on the device as a whole andper component as required within EN/IEC 60601-1, EN/IEC 60601-1-2,EN/IEC 60601-2-2. A series of tests were performed to qualify and quantify the mechanicaland functional properties of the device including shelf-life or service lifetesting. Ablation output and safety features were evaluated for theGenerator and its accessories. Retention forces and ultrasound visibilitywas evaluated for the Handpiece. Fluid adherence, separation, and pulltests were completed to evaluate the Pads. The software and hardware for the Generator has been developed,verified and validated to evaluate the graphical user interface (GUI),alerts, communication between components, real-time feedback to theuser via the device's sensors, power control, and software/hardwareinterface. A full system verification was completed to ensure safety is met in allrelated areas required by internal specifications, guidance documents,and standards.

{1}------------------------------------------------

{2}------------------------------------------------

A series of biocompatibility testing also demonstrated that the device materials are safe, suitable, and appropriate for their intended use and in compliance with ISO 10993-1, ISO 10993-5, and ISO 10993-10.

Animal and bench ablation testing also successfully demonstrates that the Acessa System performs as intended and per specifications. The ablation capability was confirmed and the radiofrequency ablation provides a reproducible, discretely demarcated zone of tissue necrosis which was surrounded by normal tissue perfused with blood.

The Acessa System has undergone an IDE clinical trial under G080163 with identifier NCT00874029 under clinicaltrials.gov.

Clinical trials results demonstrated that radiofrequency volumetric ablation with the Acessa System effectively reduced the amount of menstrual blood loss, decreased the severity of fibroid symptoms, and improved health-related quality of life for up to 12 and 24 months post-treatment. The results also demonstrated the Acessa System's safety profile is acceptable with a high benefits to low risk ratio. Treatment with this device provides an outpatient minimally invasive option that allows rapid recovery and that avoids the risks, costs, and discomfort of major surgery.

  • Peri-hysterectomy feasibility study was conducted to evaluate the device . for acute procedural and device safety. Various fibroid locations and types were encountered during this study. Pathology reports of the ablation zones were correlated per patient, per fibroid, with procedure documentation. From these data, it is shown that the ex vivo bench testing ablation results are predictive of the ablations in fibroid tissue invivo. There were no device-related adverse events in this study.
  • . Prospective, non-randomized, longitudinal Phase II studies using the Acessa System were conducted at two separate centers to establish early safety and effectiveness data. Various fibroid types and locations were treated in this study. The patients were followed to 12 months using validated questionnaire for fibroid symptoms (UFS-QOL) as well as menstrual pad counts. The mean uterine volume at baseline was 204.4 cm3. At 6 months and 12 months post-treatment, the mean uterine volume was significantly reduced to 155.0 cm3 (p= 0.012) and 151.4 cm3 (p = 0.009), respectively. The percent of subjects demonstrating reduced uterine volume in contrast to baseline was 82.2%, 80.0%, and 79.3% at the 3-, 6-, and 12-month visits, respectively. At baseline, 77% of subjects reported heavy to very heavy bleeding prior to the treatment. After 12 months post-treatment, 3.5% (p<0.0001) reported heavy to very heavy bleeding based on the patient reported survey. The UFS-QOL questionnaire was administered at baseline and at all follow-up visits. The mean percent improvement in Symptom Severity Scores from baseline to post-treatment was 59.6% at 3 months, 65.0% at 6 months, and 79.0%, at 12 months. The mean percent change in HRQL Scores

{3}------------------------------------------------

K121858

Page 4/5

from baseline to 3 months post-procedure was 41.6%, from baseline to 6 months post-procedure was 39.7%, and from baseline to 12 months post-procedure was 42.9%. Overall, sixty-five subjects (94.2%) reported improved Symptom Severity Scores and sixty-one subjects (88.4%) reported improved HRQL Scores. In terms of safety, there was only one (1.4%) serious adverse event (i.e., abdominal wall hematoma), which was determined to be related to the procedure. It was treated by laparotomy and vessel ligation.

  • . The device pivotal (Phase III) study included the treatment of 137 women with symptomatic fibroids including menorrhagia. A total of 11 centers in the United States (9 centers) and Latin America (2 centers) and 13 investigators participated in the study. The primary objectives of the study were to confirm the safety and efficacy of the device for the treatment of symptomatic uterine fibroids. The secondary objectives of the study were to evaluate the change in uterine and fibroid volume, symptom severity, health related quality of life, general health status, and subject satisfaction at 12 months post treatment compared to baseline. UFS-QOL and EQ-5D questionnaire are followed-up to 3 years post-treatment.
    • o Various fibroid types and locations were treated in this study totaling 674. The mean number of fibroids treated in the study is 5 per patient (median of 4) with the maximum at 29.
    • There are less than 4% device-related adverse events in the 0 study. Of the 10 (7%) reported serious adverse events, only the pelvic abscess and the serosal colon injury were considered to be related to the device.
    • The mean reduction in menstrual blood flow at 12 months post O treatment was 103.6 ml, with 83.9% of subjects experiencing a reduction in blood loss while 15.3% experiencing an increase in bleeding at 12 months post-treatment. When including the subjects lost to follow up as treatment failures, 40.2% (95% Cl 31.6% - 48.7%) met the protocol criterion for bleeding relief (defined as ≥50% reduction in menstrual bleeding at 12 months post-treatment) though this did not meet the prespecified study hypothesis that the lower bound of the 95% Cl would be ≥45%.
    • O Within 24 months, a total of 6 subjects (6/107) have undergone surgical reintervention for fibroid-related bleeding for a cumulative reintervention rate of <6%. Complete 24-month outcomes data on reintervention will be available after March 2013.
    • RFA treatment resulted in a reduction from baseline in total o uterine and fibroid volume, as assessed by pretreatment and post treatment contrast-enhanced MRI, at 3 and 12 months post treatment. At 12 months post treatment, the mean reduction in uterine volume was 25.1% and the mean reduction in fibroid volume was 44.3%.
    • Subjects showed a steady improvement in mean Health-Related 0

{4}------------------------------------------------

Quality of Life (HRQL) scores (from a low 37 at baseline to 80 after 24 months post-treatment) and a continuous decrease in Symptom Severity Scores (SSS) scores (from a high 61 at baseline to a low 24 at 24 months post-treatment). Subjects reported a decrease in disease burden and improved O health state with mean EQ-5D scores starting at 71 at baseline and rising to 85 after 24 months. The results of the OTE surveys showed that 94% of the subjects 0 responded that they were very satisfied, moderately satisfied, or somewhat satisfied with the treatment. At 12-months post treatment, 98% of the subjects reported that they would probably or definitely recommend the procedure to their friends with the same health problem. When asked about effectiveness of the treatment, at least 94% of the subjects responded that the treatment had been somewhat, moderately, and very effective in eliminating their symptoms. A total of 88 subjects reported that they were working. Subjects 0 returned to work in a median of 5 days and all subjects reported that they returned to normal in a median of 9 days. Ninety-six percent of the subjects in the Halt study were treated on an outpatient basis. 0 Seven subjects (5.1%) had one or more calcified fibroids that were treated during the RF ablation procedure. Calcified fibroids did not appear to impact the physician's ability to insert the tip of the Handpiece but the data are minimal regarding calcified fibroids at this time. Conclusion The Acessa System is substantially equivalent to its proposed predicate device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the emblem.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Letter Date: November 5, 2012

Halt Medical, Inc. % Ms. Clarisa Tate Senior Director Regulatory Affairs and Quality Assurance 131 Sand Creek Road, Suite B BRENTWOOD CA 94513

K121858 Re:

Trade/Device Name: The Acessa System Regulation Number: 21 CFR§ 884.4160 Regulation Name: Unipolar endoscopic coagulator-cutter and accessories Regulatory Class: II Product Code: HFG Dated: October 8, 2012 Received: October 24, 2012

Dear Ms. Tate:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 – Ms. Clarisa Tate

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known): K121858

Device Name: Acessa System

Indications For Use: The Acessa System is indicated for use in percutaneous, laparoscopic coagulation and ablation of soft tissue, including treatment of symptomatic uterine fibroids under laparoscopic ultrasound guidance.

The Acessa System consists of the following components:

  • Generator (Model Number 1000) .
  • Handpiece (Model Number 2000)
  • Handpiece Cable (Model Number 4200)
  • Pads (Model Number 3000)
  • Pad Cable (Model Number 4300)
  • Power Cord Model Number (4110)
  • Foot Pedal (Model Number 4100)

ਮ Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) NumberK121858

Page 1 of

§ 884.4160 Unipolar endoscopic coagulator-cutter and accessories.

(a)
Identification. A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.(b)
Classification. Class II (performance standards).