(217 days)
The GE CardiacVX for MRI is an analytical software tool, which provides reproducible tools for the review and reporting of medical images. CardiacVX can import medical images from a MR system and display them in a viewing area on the computer screen. The viewing area allows the access to multiple studies and series of multi-slice, multi-phase images. Multi-phase sequences of images can be displayed in a cine mode to facilitate visualization.
A report input interface is also available. Measurement tools on the report interface make it possible to quickly and reliably fill out a complete clinical report of an imaging exam. Available tools include: point, distance, area, and volume measurement tools such as ejection fraction, cardiac output, end-diastolic volume, end-systolic volume, and volume flow measurements.
Semi-automatic tools are available for left ventricular contour detection, valve plane detection, vessel contour detection for flow analysis, signal intensity analysis for myocardium and infarct sizing measurement, and T2 star analysis.
The results of the measurement tools are interpreted by the physician and can be communicated to referring physicians.
When interpreted by a trained physician these tools may be useful in supporting the determination of a diagnosis.
CardiacVX is a post-processing analytical software application, which provides tools for the review and reporting of cardiac MR datasets. Measurement tools on the report interface include: point, distance, area, and volume measurements (including ejection fraction, cardiac output, end-diastolic volume, end-systolic volume, and volume flow measurements). Semi-automatic tools are available for left ventricular contour detection, valve plane detection, vessel contour detection for flow analysis, signal intensity analysis for myocardium and infarct sizing measurement, T2 Star analysis, and patent foramen ovale (PFO) analysis.
The GE CardiacVX for MRI is an analytical software tool, which provides reproducible tools for the review and reporting of medical images. Cardiac VX can import medical images from a MR system and display them in a viewing area on the computer screen. The viewing area allows the access to multiple studies and series of multi-slice, multiphase images. Multi-phase sequences of images can be displayed in a cine mode to facilitate visualization.
A report input interface is also available. Measurement tools on the report interface make it possible to quickly and reliably fill out a complete clinical report of an imaging exam. Available tools include: point, distance, area, and volume measurement tools such as ejection fraction, cardiac output, end-diastolic volume, end-systolic volume, and volume flow measurements.
Semi-automatic tools are available for left ventricular contour detection, valve plane detection, vessel contour detection for flow analysis, signal intensity analysis for myocardium and infarct sizing measurement, and T2 Star analysis.
The provided text describes a 510(k) premarket notification for the GE Healthcare CardiacVX device, a post-processing analytical software application for cardiac MR datasets. However, it does not contain specific acceptance criteria, reported device performance metrics, or details of a comprehensive study proving the device meets acceptance criteria.
The document primarily focuses on establishing substantial equivalence to predicate devices and detailing the device's description, indications for use, and non-clinical tests (quality assurance measures).
Therefore, based only on the provided text, I cannot complete the requested tables and information.
Here's a breakdown of what can be extracted and what is missing:
What can be extracted:
- Non-Clinical Tests / Quality Assurance Measures: The document lists several quality assurance measures applied to the development, which are considered non-clinical tests. These include Risk Analysis, Requirements Reviews, Design Reviews, Testing on unit level (Module verification), Integration testing (System verification), Performance testing (Verification), Safety testing (Verification), and Simulated use testing (Validation). However, specific performance metrics or acceptance criteria for these tests are not provided.
- Data Provenance for Clinical Images (used for verification and validation testing):
- Images from GE Healthcare facilities under a non-significant risk investigation for internal volunteer scanning.
- Multi-vendor anonymized MR contrast-enhanced images obtained from a third-party clinical research study.
- Type of Ground Truth used for Clinical Images: Not explicitly stated, but for "internal volunteer scanning" and "third-party clinical research study," it's generally assumed that expert interpretation or a gold standard (e.g., pathology, clinical outcome for some studies, or expert consensus) would be used. However, the document does not specify how the ground truth was established for these particular images or their purpose beyond "verification and validation testing."
- Ground Truth for Training Set (implicitly): The document mentions "clinical images that were used for the completion of verification and validation testing." It's possible these same images or a subset were used for training/development, but the document doesn't explicitly state a "training set" or how its ground truth was established.
What is missing and cannot be provided from the text:
- A table of acceptance criteria and the reported device performance: This is the most significant missing piece. The document states "Performance testing (Verification)" was done, but gives no details about what performance was measured, what the acceptance criteria were, or what the results were.
- Sample sizes used for the test set: While it mentions "clinical images" were used, it doesn't quantify the sample size of these images for testing.
- Number of experts used to establish the ground truth for the test set and their qualifications: Not mentioned.
- Adjudication method for the test set: Not mentioned.
- Multi reader multi case (MRMC) comparative effectiveness study information: Not mentioned. The document only mentions the device being interpreted by a trained physician, but not a study comparing performance with and without AI assistance.
- Standalone (algorithm-only) performance: While the device is a "software application," the performance details are not provided in a standalone manner. The indications for use state "The results of the measurement tools are interpreted by the physician," implying human-in-the-loop.
- Sample size for the training set: Not mentioned.
- How the ground truth for the training set was established: Not mentioned.
Conclusion based on provided text:
While the document confirms that verification and validation testing using clinical images were performed and that quality assurance measures were applied, it does not provide the specific details required to answer the prompt regarding acceptance criteria, reported performance, sample sizes for training/test sets, ground truth establishment, expert qualifications, or comparative effectiveness studies. These details would typically be found in a more comprehensive study report or a different section of the 510(k) submission, not in this 510(k) summary.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. The logo is black and white, with the letters and border in black and the background in white. The logo is simple and recognizable.
AN 1 8 2013
Page 1 of
GE Healthcare 510(k) Premarket Notification Submission .......
:: : : :
:
510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: December 11, 2012
GE Medical Systems, LLC (doing business as GE Submitter: Healthcare) 3200 N. Grandview Blvd . Waukesha, WI 53188 FDA Registration Number: 2183553
Primary Contact Person:
Michelle Huettner Regulatory Affairs Leader, MR GE Medical Systems, LLC (doing business as GE Healthcare) Phone (262) 521-6102 Fax (262) 546-0902
Secondary Contact Person:
Glen Sabin Regulatory Affairs Director, MR GE Medical Systems, LLC (doing business as GE Healthcare) Phone (262) 521-6848 Fax (262) 364-2785
Device: Trade Name: Cardiac VX ... ...
Common/Usual Name:
System, image processing, radiological- Picture archiving and communications system.
lassification Names: 21CFR 892.2050
Product Code:
{1}------------------------------------------------
K121762
Page 2 of
GE Healthcare 510(k) Premarket Notification Submission
GE Delta Software Option for MRI (marketed as Predicate Device(s): ReportCARD) (K031927) manufactured by GE
Device Description: ::
Indications for Use:
......
READY View (K110573) manufactured by GE
CardiacVX is a post-processing analytical software application, which provides tools for the review and reporting of cardiac MR datasets. Measurement tools on the report interface include: point, distance, area, and volume measurements (including ejection fraction, cardiac output, end-diastolic volume, end-systolic volume, and volume flow measurements). Semi-automatic tools are available for left ventricular contour detection, valve plane detection, vessel contour detection for flow analysis, signal intensity analysis for myocardium and infarct sizing measurement, T2 Star analysis, and patent foramen ovale (PFO) analysis.
The GE CardiacVX for MRI is an analytical software tool, which provides reproducible tools for the review and reporting of medical images. Cardiac VX can import medical images from a MR system and display them in a viewing area on the computer screen. The viewing area allows the access to multiple studies and series of multi-slice, multiphase images. Multi-phase sequences of images can be displayed in a cine mode to facilitate visualization.
A report input interface is also available. Measurement tools on the report interface make it possible to quickly and reliably fill out a complete clinical report of an imaging exam. Available tools include: point, distance, area, and volume measurement tools such as ejection fraction, cardiac output, end-diastolic volume, end-systolic volume, and volume flow measurements.
Semi-automatic tools are available for left ventricular contour detection, valve plane detection, vessel contour detection for flow analysis, signal intensity analysis for myocardium and infarct sizing measurement, and T2 Star analysis.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circle. The letters and the circle are in black, creating a strong contrast. The logo is a well-known symbol associated with the General Electric company.
GE Healthcare 510(k) Premarket Notification Submission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The results of the measurement tools are interpreted by the physician and can be communicated to referring physicians.
When interpreted by a trained physician these tools may be useful in supporting the determination of a diagnosis.
Technology:
Determination of
Substantial Equivalence:
The proposed medical device, Cardiac VX, employs the same fundamental scientific technology as its predicate devices. The proposed device (CardiacVX) is substantially equivalent to the predicate devices because it is a post-processing software option for use on cardiac MR image datasets.
Summary of Non-Clinical Tests:
CardiacVX complies with voluntary standards, including the following process standards: ISO 13485, ISO 14971, and IEC 62304. Because CardiacVX is a software-only application which runs on the Advantage Workstation . (AW), there are no applicable mandatory performance standards.
The following quality assurance measures were applied to the development of the system:
- Risk Analysis .
- Requirements Reviews
- Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification),
- Performance testing (Verification)
- Safety testing (Verification)
Simulated use testing (Validation)
Summary of Clinical Tests:
The MR acquired clinical images that were used for the completion of verification and validation testing for CardiacVX was obtained from the GE Healthcare facilities under the provisions of a non-significant risk investigation for internal volunteer scanning. In addition,
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white.
:
Page 4 of 4
GE Healthcare
510(k) Premarket Notification Submission . . . .
:
multi-vendor anonymized MR contrast-enhanced images were obtained from a third-party clinical research study.
Conclusion: GE Healthcare considers the CardiacVX application to be as safe, as effective, and performance is substantially equivalent to its predicate devices.
. . . .
:::::
:
:: : : :
........
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle symbol, with three curved lines representing the eagle's body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol. The logo is simple and recognizable, representing the department's role in protecting the health of all Americans.
Food and Drug Administration 10903 New Hampshire A venue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
January 18, 2013
GE Medical Systems, LLC % Michelle Huettner 3200 N. Grandview Blvd. WAUKESHA WI 53188
Re: K121762
Trade/Device Name: CardiacVX Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: December 11, 2012 Received: December 12, 2012
Dear Ms. Huettner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Trou may, therefore, maxisions of the Act include requirements for annual registration, listing of The general vonations practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any 1 outhall the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Or K F at 607), lacemig (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sean M. Boyd -S
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
for
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known):
Cardiac VX Device Name:
Indications for Use:
The GE CardiacVX for MRI is an analytical software tool, which provides reproducible tools for the review and reporting of medical images. CardiacVX can import medical images from a MR system and display them in a viewing area on the computer screen. The viewing area allows the access to multiple studies and series of multi-slice, multi-phase images. Multi-phase sequences of images can be displayed in a cine mode to facilitate visualization.
A report input interface is also available. Measurement tools on the report interface make it possible to quickly and reliably fill out a complete clinical report of an imaging exam. Available tools include: point, distance, area, and volume measurement tools such as ejection fraction, cardiac output, end-diastolic volume, end-systolic volume, and volume flow measurements.
Semi-automatic tools are available for left ventricular contour detection, valve plane detection, vessel contour detection for flow analysis, signal intensity analysis for myocardium and infarct sizing measurement, and T2 star analysis.
The results of the measurement tools are interpreted by the physician and can be communicated to referring physicians.
When interpreted by a trained physician these tools may be useful in supporting the determination of a diagnosis.
Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Sean M. Boyd -S
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
K121762 510(k)
Page 1 of
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).