K Number
K121679
Manufacturer
Date Cleared
2012-10-11

(126 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACCU-CHEK Inform II Test strip is for use with the ACCU-CHEK Inform II meter to quantitatively measure glucose (sugar) in venous whole blood, arterial whole blood, neonatal heelstick, or fresh capillary whole blood samples drawn from the fingertip as an aid in monitoring the effectiveness of glucose control. The system is not for use in diagnosis or screening of diabetes mellitus, nor for testing neonate cord blood samples.

The ACCU-CHEK Inform II Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices.

The ACCU-CHEK Inform II Controls are intended for quality control performance checks on the ACCU-CHEK Inform II system with ACCU-CHEK Inform II test strips.

The ACCU-CHEK Inform II Linearity Test Kit is intended for use for periodic verification of linearity of the ACCU-CHEK Inform II system using ACCU-CHEK Inform II test strips.

Device Description

The ACCU-CHEK Inform II System includes the following components and accessories: Meter, Code Key Reader, Base Unit (with power supply) test strips, and controls.

When an ACCU-CHEK Inform II test-strip is inserted into the ACCU-CHEK Inform II meter, a small alternating current (AC) is applied until the application of blood causes a spike in the conductivity to be observed at the measurement and sample-sufficiency electrodes - both are used to assure an adequate sample has been applied.

The instrument then applies a series of AC voltages at four frequencies and reads the AC responses. These carry information about the sample type and environmental temperature; they also allow the system to perform various internal quality checks.

After the AC measures are completed, a small (DC) voltage is applied and current is observed which is proportionate to the glucose. The AC and DC information are then combined to provide a hematocrit and temperature compensated glucose result.

ACCU-CHEK Inform II System is the first point-of-care glucose meter featuring wireless connectivity functionality. Unlike the predicate device, ACCU-CHEK Inform II meter has the capability of direct data transfer right after each measurement, thus ensuring actual patient data to appear on the doctor's desk shortly after these data have been created at the bedside. It is understood that a transfer of data can only occur, when a stable communication with the Data Management System has been built up by the devices before transfer of data.

The glucose meter system has been designed in a way that its connectivity feature is virtually invisible to the operator. Only the appearance of an icon at the lower edge of the touch screen display indicates potential radio frequency activity to allow new information on control material or strip code files to be downloaded.

Even when there is no wireless WLAN available, the ACCU-CHEK Inform II meters work fine as long as the necessary information like operator list, strip code, and control lot is available. The device can store as many as 2000 measurement results in its memory which will be transferred as soon as there is a stable connection to the WLAN and no measurement operation pending.

AI/ML Overview

Acceptance Criteria and Device Performance for ACCU-CHEK Inform II Blood Glucose Monitoring System (K121679)

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for a blood glucose monitoring system are typically based on the ISO 15197:2003 standard for blood glucose monitoring systems for in vitro diagnostic testing. While the document doesn't explicitly state the numerical thresholds for each acceptance criterion, it indicates compliance with the standard. The reported device performance is presented as the percentage of glucose results that fit within accuracy requirement limits.

Performance CharacteristicAcceptance Criteria (Stated as ISO 15197:2003 Compliance)Reported Device Performance (ACCU-CHEK Inform II System)
Arterial Blood SamplesISO 15197:2003 accuracy requirement limits100% of glucose results fit within accuracy requirement limits
Capillary Blood SamplesISO 15197:2003 accuracy requirement limits100% of glucose results fit within accuracy requirement limits
Neonate Capillary Blood SamplesISO 15197:2003 accuracy requirement limits99.4% of glucose results fit within accuracy requirement limits
Venous Blood SamplesISO 15197:2003 accuracy requirement limits100% of glucose results fit within accuracy requirement limits
Consumer User Performance EvaluationISO 15197:2003 accuracy requirement limits98.9% of glucose results fit within accuracy requirement limits
PrecisionNot explicitly stated as a percentage, but predicate shows ≤ 3.2% variation (ISO 15197 typically has limits for repeatability and intermediate precision)Repeatability and reproducibility studies show a variation from strip to strip ≤ 3.4%
Measuring RangeImplied by predicate for substantial equivalence: 10-600 mg/dL10-600 mg/dL
Measuring TimeImplied by predicate for substantial equivalence: 5 seconds5 seconds
Sample VolumeImplied by predicate for substantial equivalence: 0.6 µL0.6 µL
Hematocrit RangeImplied by predicate for substantial equivalence: 20-65% (with sub-ranges)10-65%

2. Sample Sizes Used for the Test Set and Data Provenance

The document does not explicitly state the country of origin for the data or if it was retrospective or prospective. Typically, clinical studies for medical devices are prospective.

  • Arterial Blood Samples: N = 214
  • Capillary Blood Samples: N = 200
  • Neonate Capillary Blood Samples: N = 191
  • Venous Blood Samples: N = 439
  • Consumer User Performance Evaluation: N = 174
  • Precision (Repeatability and Reproducibility studies): Sample sizes for these studies are not specified, only the resulting variation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not provide information on the number or qualifications of experts used to establish the ground truth for the test set. For blood glucose monitoring systems, the ground truth is typically established using a laboratory reference method (e.g., a central laboratory analyzer traceable to a recognized standard), not directly by expert consensus in the same way an imaging study would.

4. Adjudication Method for the Test Set

The document does not mention an adjudication method. For blood glucose monitoring systems, ground truth is usually established by a laboratory reference method, eliminating the need for reader adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

This is not applicable. The ACCU-CHEK Inform II Blood Glucose Monitoring System is a standalone diagnostic device for measuring glucose levels, not an AI-assisted diagnostic tool that would involve human readers interpreting AI output.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, a standalone performance study was conducted. The data presented in the table under "Reported Device Performance (ACCU-CHEK Inform II System)" represents the algorithm's performance in measuring glucose from various blood sample types without human intervention (beyond standard operation of the device). The system provides a direct numerical glucose reading.

7. The Type of Ground Truth Used

The ground truth for the glucose measurements is implicitly established by comparing the device's readings against a reference laboratory method. This is indicated by the regression equations (y = mx + b) and correlation coefficients (r) provided, where 'x' typically represents the reference method value and 'y' represents the device's reading. While not explicitly stated, this is the standard practice for evaluating blood glucose meters. This would fall under the category of comparison to an established reference method or laboratory standard.

8. The Sample Size for the Training Set

The document does not provide information on the sample size for any training set. Blood glucose meters do not typically involve "training sets" in the same way machine learning algorithms do. Their calibration and performance are based on analytical methods and extensive testing during development, rather than machine learning on a large dataset.

9. How the Ground Truth for the Training Set Was Established

As noted above, the concept of a "training set" and establishing "ground truth for the training set" is not applicable to this type of device in the context of machine learning. Device calibration and analytical performance are established through rigorous laboratory experiments and comparisons to reference methods, which are part of the device's design and manufacturing process.

{0}------------------------------------------------

510(k) Summary

IntroductionAccording to the requirements of 21 CFR 807.92, the following informationprovides sufficient detail to understand the basis for a determination ofsubstantial equivalence.
1) Submittername.address.contactRoche Diagnostics Corporation9115 Hague Rd.Indianapolis, IN 46250Contact Person: Mike FlisDate Prepared: 11 October 2012
2) DevicenameACCU-CHEK Inform II Blood Glucose Monitoring System consisting of:Meter: ACCU-CHEK Inform II Meter●Test Strip: ACCU-CHEK Inform II Test StripControls: ACCU-CHEK Inform II Control Solutions●Linearity Kit: ACCU-CHEK Inform II Linearity Test Kitwhole blood glucose test systemCommon name:Classification name: Glucose dehydrogenase, glucose (21 C.F.R. § 862.1345)LFR, Glucose DehydrogenaseProduct Codes:JJX, Single (specified) analyte controls (assayed andunassayed)
3) PredicatedeviceThe ACCU-CHEK Inform System presented in K003846 is the predicate forthe ACCU-CHEK Inform II system.The controls solutions presented in the ACCU-CHEK Aviva 510(k)submission (K043474) are the predicate for the ACCU-CHEK Inform IIcontrol solutions.The Nova StatStrip Glucose Linearity Kit (receiving clearance in K060345) isthe predicate for the ACCU-CHEK Inform II Linearity Kit.

Continued on next page

K121679

OCT 1 1 2012

Roche

r "...

.

{1}------------------------------------------------

510(k) Summary, Continued

The ACCU-CHEK Inform II System includes the following components and 4) Device Description accessories: Meter, Code Key Reader, Base Unit (with power supply) test strips, and controls.

When an ACCU-CHEK Inform II test-strip is inserted into the ACCU-CHEK Inform II meter, a small alternating current (AC) is applied until the application of blood causes a spike in the conductivity to be observed at the measurement and sample-sufficiency electrodes - both are used to assure an adequate sample has been applied.

The instrument then applies a series of AC voltages at four frequencies and reads the AC responses. These carry information about the sample type and environmental temperature; they also allow the system to perform various internal quality checks.

After the AC measures are completed, a small (DC) voltage is applied and current is observed which is proportionate to the glucose. The AC and DC information are then combined to provide a hematocrit and temperature compensated glucose result.

ACCU-CHEK Inform II System is the first point-of-care glucose meter featuring wireless connectivity functionality. Unlike the predicate device, ACCU-CHEK Inform II meter has the capability of direct data transfer right after each measurement, thus ensuring actual patient data to appear on the doctor's desk shortly after these data have been created at the bedside. It is understood that a transfer of data can only occur, when a stable communication with the Data Management System has been built up by the devices before transfer of data.

The glucose meter system has been designed in a way that its connectivity feature is virtually invisible to the operator. Only the appearance of an icon at the lower edge of the touch screen display indicates potential radio frequency activity to allow new information on control material or strip code files to be downloaded.

Even when there is no wireless WLAN available, the ACCU-CHEK Inform II meters work fine as long as the necessary information like operator list, strip code, and control lot is available. The device can store as many as 2000 measurement results in its memory which will be transferred as soon as there is a stable connection to the WLAN and no measurement operation pending.

Continued on next page

{2}------------------------------------------------

510(k) Summary, Continued

  1. Intended use

ACCU-CHEK Inform II System

The ACCU-CHEK Inform II Test strip are for use with the ACCU-CHEK Inform II meter to quantitatively measure glucose (sugar) in venous whole blood, arterial whole blood, neonatal heelstick, or fresh capillary whole blood samples drawn from the fingertips as an aid in monitoring the effectiveness of glucose control. The system is not for use in diagnosis or screening of diabetes mellitus, nor for testing neonate cord blood samples.

The ACCU-CHEK Inform II Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices.

The multiple-patient use ACCU-CHEK Inform II Blood Glucose Monitoring System will consist of:

  • . Meter: ACCU-CHEK Inform II Meter
  • Test Strip: ACCU-CHEK Inform II Test Strip .
  • . Controls: ACCU-CHEK Inform II Control Solutions

The ACCU-CHEK Inform II Controls are intended for quality control performance checks on the ACCU-CHEK Inform II system with ACCU-CHEK Inform II test strips.

The ACCU-CHEK Inform II Linearity Test Kit is intended for use for periodic verification of linearity of the ACCU-CHEK Inform II system using ACCU-CHEK Inform II test strips.

  1. Substantial equivalence The ACCU-CHEK Inform II Test System is substantially equivalent to the ACCU-CHEK Inform Test System.

Continued on next page

ﻨﺎ

{3}------------------------------------------------

510(k) Summary, Continued

Performance testing on the ACCU-CHEK Inform II System demonstrated 7) Data demonstrati that the device meets the performance requirements for its intended use. The data demonstrates that the test strip is substantially equivalent to the predicate ng substantial device. equivalence

ACCU-CHEKInform IISystemACCU-CHEKInform II SystemISO 15197:2003ACCU-CHEK Inform(predicate)
Arterial BloodSamplesN = 214y = 1.038x - 3.5r = 0.990Range = 58 to 322 mg/dLHCT range = 19 to 54%100% of glucose results fitwithin accuracyrequirement limitsN = 143y = 1.045x - 1.3r = 0.984Range = 75 to 388 mg/dL
Capillary BloodSamplesN = 200y = 1.018x - 3.4r = 0.980Range = 22 to 572 mg/dL100% of glucose results fitwithin accuracyrequirement limitsN = 187y = 0.991x + 8.4r = 0.980Range = 41 to 503 mg/dL
NeonateCapillary BloodSamplesN = 191y = 1.011x + 1.5r = 0.976Range = 28 to 152 mg/dLHCT range = 23 to 58%99.4% of glucose resultsfit within accuracyrequirement limitsN = 105y = 0.999x - 4.4r = 0.981Range = 33 to 153 mg/dLHCT Range = 30 to 64%
Neonate CordBlood SamplesNot claimedNot claimedSame as above
Venous BloodSamplesN = 439y = 1.01x -2.2r = 0.995Range = 38-546 mg/dLHCT range = 20-63%100% of glucose results fitwithin accuracyrequirement limitsN = 218y = 0.975x + 3.0r = 0.987Range = 41 to 503 mg/dL
PrecisionRepeatability andreproducibility studiesshow a variation fromstrip to strip ≤ 3.4%Not applicableRepeatability and reproducibilitystudies show a variation from stripto strip ≤ 3.2%
Consumer UserPerformanceEvaluationN = 174y = 0.933x + 6.6r = 0.990Range = 56-579 mg/dL98.9% of glucose resultsfit within accuracyrequirement limitsN = 187y = 0.985x + 7.5r = 0.977Range = 41 to 503 mg/dL
Measuring Range10-600 mg/dL---10- 600 mg/dL
Measuring Time5 seconds---5 seconds
Sample Volume0.6 µL---0.6 µL
HematocritRange10-65%---20-65% for glucose <200 mg/dL20-55% for glucose >200 mg/dL

{4}------------------------------------------------

The ACCU-CHEK Aviva control solutions (receiving clearance in K043474) 7) Data are the predicate control solutions for the ACCU-CHEK Inform II control demonstrati ng solutions. Below is a comparison of the ACCU-CHEK Inform II control substantial solutions to the predicate, and the regulation numbers and product codes have equivalence, been included. continued

ACCU-CHEK Inform IIControl Solutions (K121679)ACCU-CHEK Aviva ControlSolutions (K043474)
Intended UseStatementFor performance checks on theACCU-CHEK Inform II systemwith ACCU-CHEK Inform II teststrips.For performance checks on the ACCU-CHEK Aviva system with ACCU-CHEK Aviva test strips
ClassificationRegulation862. 1660862. 1660
Product CodeJJXJJX
AnalyteGlucoseGlucose
Number ofSolutions2 Levels2 Levels
Low LevelTarget45 mg/dL45 mg/dL
High LevelTarget307 mg/dL307 mg/dL

{5}------------------------------------------------

The Nova StatStrip Glucose Linearity Kit (receiving clearance in K060345) is 7) Data the predicate for the ACCU-CHEK Inform II Linearity Kit. Below is a demonstrati ng
substantial comparison of the ACCU-CHEK Inform II Linearity Kit to its predicate, and the regulation numbers and product codes have been included. equivalence, continued

ACCU-CHEK Inform II LinearityKit (K121679)Nova StatStrip Glucose LinearityKit (K060345)
Intended UseStatementFor periodic verification of linearityof the ACCU-CHEK Inform IIsystem using ACCU-CHEK InformII test strips.Nova StatStrip Glucose LinearityKit solutions are used to check thelinearity of the Nova StatStripGlucose Hospital System.
ClassificationRegulation862. 1660862. 1660
Product CodeJJXJJX
AnalyteGlucoseGlucose
Number ofSolutions6 Levels5 Levels
Low Level Target28 mg/dL31 mg/dL
High Level Target559 mg/dL514 mg/dL

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three curved lines representing the staff and a serpent intertwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The logo is in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

November 20, 2012

Roche Diagnostics c/o Mike Flis 9115 Hague Road Indianapolis, Indiana 46250

Re: K121679 Trade Name: Accu-chek Inform II blood glucose monitoring system Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW. LFR, JJX Dated: September 19, 2012 Received: September 21, 2012

Dear Mr. Flis:

This letter corrects our substantially equivalent letter of October 11, 2012.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice

{7}------------------------------------------------

requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carol C. Benson

for

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

{8}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K121679 Device Name: ACCU-CHEK Inform II Blood Glucose Monitoring System

The ACCU-CHEK Inform II Test strip is for use with the ACCU-CHEK Inform II meter to quantitatively measure glucose (sugar) in venous whole blood, arterial whole blood, neonatal heelstick, or fresh capillary whole blood samples drawn from the fingertip as an aid in monitoring the effectiveness of glucose control. The system is not for use in diagnosis or screening of diabetes mellitus, nor for testing neonate cord blood samples.

The ACCU-CHEK Inform II Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings. This system should only be used with single-use, auto-disabling lancing devices.

The ACCU-CHEK Inform II Controls are intended for quality control performance checks on the ACCU-CHEK Inform II system with ACCU-CHEK Inform II test strips.

The ACCU-CHEK Inform II Linearity Test Kit is intended for use for periodic verification of linearity of the ACCU-CHEK Inform II system using ACCU-CHEK Inform II test strips.

Prescription Use __ XX (Part 21 CFR 801 Subpart D)

Over-The-Counter Use _ XX AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Division Sign Off

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k): K121679

Page 1 of 1

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.