K Number
K121656
Date Cleared
2012-09-24

(111 days)

Product Code
Regulation Number
876.5990
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Gemini is indicated for the fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones.

Device Description

The Gemini is a modular urological work station designed for extracorporeal shock wave lithotripsy ("ESWL") and for diagnostic and therapeutic procedures usual in Urology. The Gemini is composed of the following modules: (1) basic unit with integrated X-ray C-arm and therapy arm for shockwave treatment; (2) patient table; (3) control desk – user interface; and (4) ultrasound unit.

AI/ML Overview

The provided document is a 510(k) summary for the Gemini Lithotripter, detailing modifications to previously cleared predicate devices. It describes the device's intended use, technological characteristics, and conformance to various safety and performance standards. However, it does not include detailed acceptance criteria or a specific study proving the device meets those criteria in the way a clinical performance study would.

Here's an analysis based on the provided text, addressing your points:

1. A table of acceptance criteria and the reported device performance

The document does not specify quantitative acceptance criteria for clinical performance (e.g., stone fragmentation rates, stone-free rates, re-treatment rates). Instead, it relies on demonstrating substantial equivalence to predicate devices and compliance with a list of recognized standards and internal design control processes.

The "reported device performance" is primarily stated as: "In all instances, the Gemini Lithotripter functioned as intended and results observed were as expected." and "the system met all acceptance criteria" during internal design control.

Acceptance Criterion (Implicit)Reported Device Performance
Compliance with IEC 60601-1 (Electrical safety)Complied
Compliance with IEC 60601-1-2 (Electromagnetic compatibility)Complied
Compliance with IEC 60601-1-3 (Radiation protection)Complied
Compliance with IEC 60601-1-6 (Usability)Complied
Compliance with IEC 60601-2-7 (Safety of high-voltage X-ray)Complied
Compliance with IEC 60601-2-28 (X-ray source assemblies)Complied
Compliance with IEC 60601-2-36 (Extracorporeally induced Lithotripsy)Complied
Compliance with IEC 60601-2-32 (Safety of X-ray equipment)Complied
Compliance with ISO 13485 (Quality management system)Complied
Compliance with IEC 61846 (Ultrasonics - Pressure pulse lithotripters)Complied
Functionality of anti-collision systemVerified to perform as designed to mitigate the risk of injury.
Accuracy of stone localization methodologyTested during design verification testing; assessment demonstrated the system is capable of locating the shock wave focus area with sufficient accuracy.
Device manual reviewReviewed and approved as part of design control process; contains all necessary warnings, cautions, and instructions to mitigate potential injuries.
Overall performance vs. Intended Use and Predicates"In all instances, the Gemini Lithotripter functioned as intended and results observed were as expected." and "Performance data demonstrate that the Gemini Lithotripter is as safe and effective as the predicate devices." (This is a general statement rather than specific performance metrics.)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not mention a clinical "test set" in the context of human patients or image data for evaluating an AI/algorithm's performance. The testing described focuses on engineering verification and validation against standard requirements and internal acceptance criteria during the design control process. There is no indication of retrospective or prospective clinical study data provenance mentioned for the submitted data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. No clinical test set requiring expert ground truth is described. The "ground truth" for the engineering performance was established by the design specifications and regulatory standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No clinical test set requiring adjudication is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This document does not mention any multi-reader multi-case (MRMC) comparative effectiveness study or any AI component designed to assist human readers. The device is an extracorporeal shock wave Lithotripter, a physical medical device for stone fragmentation, not a diagnostic AI system.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. The device is a physical Lithotripter, not solely an algorithm. While it has software for control and image storage, the "performance" described pertains to the overall system's operation and safety, not a diagnostic algorithm operating in a standalone mode. The "assessment demonstrated that the system is capable of locating the shock wave focus area with sufficient accuracy," which is an algorithm-driven function, but its evaluation isn't detailed as a standalone algorithm performance study.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the engineering and safety tests, the "ground truth" refers to the pass/fail criteria defined by the recognized standards (IEC, ISO) and the device's design specifications. For the "accuracy of stone localization methodology," the ground truth would have been derived from physical measurements against known targets or possibly phantom studies. No patient outcomes data, pathology, or expert consensus on clinical findings is mentioned as ground truth for the provided performance data.

8. The sample size for the training set

Not applicable. This is not an AI/ML device that would typically have a "training set" in the context of deep learning models for image analysis or diagnosis. The device's underlying technology (electromagnetic shock wave emitter, X-ray, ultrasound) is well-established.

9. How the ground truth for the training set was established

Not applicable, as there is no mention of a training set for an AI/ML model.


Summary of the Document's Approach:

The K121656 510(k) summary focuses on demonstrating substantial equivalence to predicate devices (Dornier Lithotripter (Doli) and Doli 140) based on:

  • Identical Intended Use/Indications for Use.
  • Similar Technological Characteristics/Principles of Operation: The core component (140f EMSE shock wave source) is identical to those in cleared devices. Other components (patient table, X-ray unit) are "similar to that of other cleared stationary Lithotripters" and "perform the same function and operate in the same manner."
  • Compliance with Recognized Standards: A comprehensive list of IEC and ISO standards for electrical safety, EMC, radiation protection, usability, and specific Lithotripter requirements were met.
  • Internal Design Control Testing: The company states that the system and its software were tested and validated, meeting "all acceptance criteria" during their design control process. Specific mention is made of the anti-collision system performing as designed and the stone localization methodology demonstrating sufficient accuracy.

This type of 510(k) submission, particularly a Special 510(k) as indicated, often relies heavily on demonstrating that modifications do not raise new questions of safety or effectiveness, rather than requiring extensive de novo clinical trials or AI performance studies with large, expert-adjudicated datasets. The "study" proving the device meets acceptance criteria is the sum of the documented engineering tests and validations against the listed standards and internal specifications, demonstrating its functional equivalence and safety compared to the predicates.

{0}------------------------------------------------

2121656 PE. I DF

510(k) SUMMARY

Gemini Lithotripter

SEP 24 2012

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Dornier MedTech America 1155 Roberts Blvd. Kennesaw, GA 30144 Phone: 770-514-6163 Fax: 770-514-6291

Contact Person: John Hoffer

Date Prepared: September 4, 2012

Device Trade Name

Gemini Lithotripter

Classification Name

Extracorporeal Shock Wave Lithotripters (21 C.F.R. § 876.5990)

Predicate Devices

Dornier Lithotripter (Doli) {P840008, S062)

Dornier Lithotripter 140 (Doli 140) (K011773)

Intended Use / Indications for Use

The Gemini is indicated for the fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones.

Purpose of the Special 510(k) Notice

The Gemini is a modification to Dornier's Lithotripter (P840008, S062) and the Doli 140 (K011773). Specifically, the Gemini is the same device as the cleared predicate Doli products with the exception of the modifications outlined below:

  • Minor changes to the patient table, including weight limit, travel range and speed; o
  • Minor changes to the X-ray system to include an optional larger Image Intensifier and 0 use of an alternate X-ray generator;
  • O Use of an alternate ultrasound unit; and
  • Use of an alternate image storage system. O

Technological Characteristics/Principles of Operation

The Gemini is a modular urological work station designed for extracorporeal shock wave lithotripsy ("ESWL") and for diagnostic and therapeutic procedures usual in Urology. The Gemini is composed of the following modules: (1) basic unit with integrated X-ray C-arm and

{1}------------------------------------------------

therapy arm for shockwave treatment; (2) patient table; (3) control desk – user interface; and (4) ultrasound unit.

As indicate above, the Gemini is an extracorporeal shock wave lithotripter used for fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones. The technology to perform this function involves use of an electromagnetic shock wave emitter ("EMSE"). In the case of the subject device and the predicate devices, the identical shock wave source is used. This source, the 140f EMSE, has a well-established record of performance and also is utilized in other Lithotripers that have been cleared by FDA (e.g., Dornier Compact Alpha (K002929) and Dornier Compact Delta (P840008, S066)).

The other elements of the Gemini, i.e., the patient table and the X-ray unit, are also are similar to that of other cleared stationary Lithotripters, the Doli 140 (K011773) and the Dornier Lithotriper (P840008, S062). They perform the same function and operate in the same manner during the procedures involved in the fragmenting of urological stones.

Performance Data

The company has complied with all of the requirements described in FDA's Guidance for the Content of Premarket Notifications (510k's) for Extracorporeal Shock Wave Lithotrioters Indicated for the Fragmentation of Kidney and Ureteral Calculi.

Device testing was performed to confirm compliance with the following standards:

o IEC 60601-1:2007Electrical safety of medical devices
o IEC60601-1-2:2007Electromagnetic compatibility
o IEC 60601-1-3:2008Radiation protection
o IEC 60601-1-6:2008Usability
o IEC 60601-2-7Safety of high-voltage generators of diagnostic X-raygenerator
o IEC 60601-2-28Particular requirements for the safety of X-raysource assemblies and X-ray tube assemblies formedical diagnosis
o IEC 60601-2-36:1997Extracorporeally induced Lithotripsy
o IEC 60601-2-32:1995Safety of X-ray equipment
o ISO 13485:2003+AC:2007Quality management system
o IEC 61846Ultrasonics - Pressure pulse lithotripterscharacteristic of fields

In all instances, the Gemini Lithotripter functioned as intended and results observed were as expected.

In addition, the system and its software was tested and validated during the company's design control process and the system met all acceptance criteria. The anti-collision system of the subject device was verified to perform as designed to mitigate the risk of injury. The methodology for accurately locating the stone was tested during the design verification testing and the assessment demonstrated that the system is capable of locating the shock wave focus area with sufficient accuracy. Lastly, the device manual was reviewed and approved as part of the design control process. It contains all necessary warnings, cautions and instructions to mitigate potential injuries.

{2}------------------------------------------------

K121656 PG.3 VF- 3

Substantial Equivalence

The Gemini Lithotripter is as safe and effective as the identified predicate devices listed above. The Gemini Lithotripter has the same intended uses / indications for use, technological characteristics, and principles of operation as its predicate devices. Performance data demonstrate that the Gemini Lithotripter is as safe and effective as the predicate devices. Thus, the Gemini Lithotripter is substantially equivalent.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's name in a circular arrangement around the edge. In the center of the seal is a stylized image of an eagle or other bird-like figure, with its wings spread and head turned to the side. The design is simple and monochromatic.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SEP 24 2012

Mr. John Hoffer VP Quality, Regulatory, Clinical Dornier MedTech America 1155 Roberts Blvd. KENNESAW GA 30144

Re: K121656

Trade/Device Name: Gemini Regulation Number: 21 CFR§ 876.5990 Regulation Name: Extracorporeal shock wave lithotripter Regulatory Class: II Product Code: LNS Dated: September 4, 2012 Received: September 5, 2012

Dear Mr. Hoffer:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 3.10(t) premier on substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosure) to regard management date of the Medical Device Amendments, or to commerce prior to May 26, 1776, the enaounters with the provisions of the Federal Food, Drug, devices that have been reclassinou in access approval of a premarket approval application (PMA). and Cosmelle Act (Act) that do not require apple of the general controls provisions of the Act. The You may, dielefore, market the device, basics to the ments for annual registration, listing of general controls provisions of the rict labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability additeration. Thease note: ODICT acce how care labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (SCC above) into othing major regulations affecting your device can be
it may be subject to additional controls. Existing major regulations affe It may be subject to additional conseiler. This 21, Parts 800 to 898. In addition, FDA may found in the Code of I cacal resguing your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issuation of a backer with other requirements of the Act
that FDA has made a determination that your device complies with of the Act that FDA has made a decemination that your accessed by other Federal agencies. You must of ally Federal statutes and roganatelle , including, but not limited to: registration and listing comply with an the Act 3 requirements, mondaing, Frinately, Frances
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2-

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set device-related adverse events) (21 CFR Part 820); and if applicable, the electronic forul in the quality systems (QD) rogalications (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please IT you desire specific advice for your do need on the corress.
go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/com/15809.html/sear go to million.rwww.ida.gov/rtloudio.org/w/index.php?tice of Compliance. Also, please the Concertion Borroos and Ramorogious of reference to premarket notification" (21CFR Patt note the regulation onlined, "informating of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the I ou thay of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely yours,

Benjamin K. Twitchell

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

K121656 510(k) Number (if known):

Device Name: Gemini

Indications for Use:

The Gemini is indicated for the fragmentation of urinary tract stones, i.e. renal calyceal stones, renal pelvic stones, and upper ureteral stones.

Prescription Use _ × (Per 21 C.F.R. 801.109) AND/OR

Over-The-Counter Use_ (Per 21 C.F.R. 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number [K121656](https://510k.innolitics.com/search/K121656)

§ 876.5990 Extracorporeal shock wave lithotripter.

(a)
Identification. An extracorporeal shock wave lithotripter is a device that focuses ultrasonic shock waves into the body to noninvasively fragment urinary calculi within the kidney or ureter. The primary components of the device are a shock wave generator, high voltage generator, control console, imaging/localization system, and patient table. Prior to treatment, the urinary stone is targeted using either an integral or stand-alone localization/imaging system. Shock waves are typically generated using electrostatic spark discharge (spark gap), electromagnetically repelled membranes, or piezoelectric crystal arrays, and focused onto the stone with either a specially designed reflector, dish, or acoustic lens. The shock waves are created under water within the shock wave generator, and are transferred to the patient's body using an appropriate acoustic interface. After the stone has been fragmented by the focused shock waves, the fragments pass out of the body with the patient's urine.(b)
Classification. Class II (special controls) (FDA guidance document: “Guidance for the Content of Premarket Notifications (510(k)'s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi.”)