(43 days)
The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free blue nitrile patient examination glove, that meets all the requirements of ASTM standard D6319, except for sterility requirements.
The provided text describes the regulatory filing for a medical device: "UTB Blue Powder Free Nitrile Patient Examination Gloves." The information focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of AI performance, clinical effectiveness, or diagnostic accuracy. Therefore, many of the requested categories related to AI studies, reader performance, and ground truth establishment are not applicable.
Here's the breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Standard (Acceptance Criteria) | Device Performance |
|---|---|---|
| Dimension | ASTM D6319-10 | Meets |
| Physical Properties | ASTM D6319-10 | Meets |
| Freedom from Pinholes | ASTM D5151-06 & D6319-10 | Meets |
| Powder Residual | ASTM D6124-06 | Meets; Results generated values below 2mg of residual powder |
| Biocompatibility | Primary Skin Irritation in rabbits (ISO 10993-10:2010), Dermal Sensitization in the guinea pig (ISO 10993-10:2010) | Gloves are 'non-irritating'; Gloves do not display any potential for sensitization |
2. Sample size used for the test set and the data provenance
The document does not specify exact sample sizes for each test mentioned (e.g., number of gloves tested for pinholes, number of animals for biocompatibility). The data provenance is described by the standards used, which are international (ASTM, ISO). There is no explicit mention of the country of origin of the data or whether the tests were conducted retrospectively or prospectively, except that they generally refer to evaluations of the product itself for regulatory compliance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The "ground truth" here is compliance with established international standards for patient examination gloves, rather than interpretation of medical images or data by human experts. The tests are laboratory-based.
4. Adjudication method for the test set
Not applicable. This concept is relevant for studies involving human interpretation or subjective assessments. Here, the tests involve objective measurements against predefined standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device is a patient examination glove, not an AI-powered diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an algorithm or AI device. The tests performed are on the physical and biological properties of the glove.
7. The type of ground truth used
The ground truth used is the standards themselves (ASTM D6319-10, ASTM D5151-06, ASTM D6124-06, and ISO 10993-10:2010). The device performance is compared against the requirements stipulated in these recognized industry and international standards for each characteristic.
8. The sample size for the training set
Not applicable. This is not a machine learning or AI device. There is no concept of a "training set" in this context.
9. How the ground truth for the training set was established
Not applicable. As above, there is no training set for this type of medical device.
{0}------------------------------------------------
510(k) SUMMARY
JUL 3 2012
| DATE: | February 17, 2012 |
|---|---|
| OWNER: | Northstar Healthcare Holdings |
| 70 Sir John Rogerson's Quay | |
| Dublin 2, Ireland | |
| OFFICIAL CORRESPONDENT: | Alex Nagy |
| Manager, Quality Systems and Regulatory Affairs | |
| Cypress Medical Products LLC | |
| 1200 South Route 31 |
Telephone: 815-385-0100 Fax: 815-385-0114 DEVICE NAME: Trade Name: UTB Blue Powder Free Nitrile Patient Examination Gloves Common Name: Patient Examination Gloves Patient Examination Gloves Classification:
Product Code:
Class:
McHenry, IL 60050
PREDICATE DEVICE(S):
| Predicate510(k) | Device Name | Indication | ClearanceDate | Company |
|---|---|---|---|---|
| K091652 | Nitrile PowderFree ExaminationGloves (Blue) | The nitrile examination gloves is a disposabledevice intended for medical and dentalpurposes that is worn on the examiner's hand toprevent contamination between patient andexaminer. | 4 Sep 2009 | HartalegaSDN BHN,BestariJaya,Selengar45600Malaysia |
Class I
LZA
| DEVICEDESCRIPTION: | Powder free blue nitrile patient examination glove,that meets all the requirements of ASTM standardD6319, except for sterility requirements. |
|---|---|
| STATEMENT OFINTENDED USE: | The nitrile examination glove is a disposable deviceintended for medical and dental purposes to be worn onthe examiner's hand or finger to prevent contaminationbetween patient and examiner. |
{1}------------------------------------------------
TECHNOLOGICAL CHARACTERISTICS:
The UTB Blue Nitrile Powder Free Patient Exam Glove is substantially equivalent to the predicate device with regard to physical characteristics, design, product features, and intended use. Both gloves are made with nitrile using similar manufacturing processes.
ASSESSMENT OF NONCLINICAL DATA:
| Characteristic | Standard | Device Performance |
|---|---|---|
| Dimension | ASTM D6319-10 | Meets |
| Physical Properties | ASTM D6319-10 | Meets |
| Freedom from Pinholes | ASTM D5151-06 & D6319-10 | Meets |
| Powder Residual | ASTM D6124-06 | MeetsResults generatedvalues below 2mg ofresidual powder |
| Biocompatibility | Primary Skin Irritation in rabbits(ISO 10993-10:2010)Dermal Sensitization in the guineapig (ISO 10993-10:2010) | Gloves are'non-irritatingGloves do not displayany potential forsensitization |
CONCLUSIONS:
The UTB Blue Powder Free Nitrile Patient Exam Gloves meet the requirements of established standards ASTM D6319-10, ASTM D6124-06, ASTM D5151-06, and ISO 10993-10:2010.
Based on the comparison of intended use, design, materials and performance, the UTB Blue Powder Free Nitrile Patient Exam Gloves are substantially equivalent to the predicate device.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Northstar Healthcare Holdings C/O Mr. Ned Devine Responsible Third Party Official Underwriters Laboratories, Incorporated 333 Pfingsten Road Northbrook, Illinois 60062
3 2012 JUL
Re: K121506
Trade/Device Name: UTB Blue Powder Free Nitrile Patient Exam Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: June 18, 2012 Received: June 20, 2012
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucml 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
R. Shem
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
INDICATIONS FOR USE
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
UTB Blue Powder Free Nitrile Patient Exam Glove Device Name:
Indications for Use: The examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use __ X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Q (Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K121506
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
19
..............................................................................................................................................................................
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.