K Number
K121505
Device Name
DISPOSABLE HEMOSTASIS CHIP
Date Cleared
2012-08-31

(102 days)

Product Code
Regulation Number
876.4400
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
This device is used for endoscopic clip placement within the gastrointestinal tract for the purpose of endoscopic marking, hemostasis for mucosal/submucosal defects less than 3 cm in the upper GI tract, bleeding ulcers, arteries less than 2 mm, and polyps less than 1.5 cm in diameter in the GI tract. This device is not intended for the repair of GI tract lumenal perforations.
Device Description
The Disposable Hemostasis Clip is a sterile, single use device that consists of a metal clip that detaches from the introducer to maintain approximation of tissue to achieve hemostasis in the gastrointestinal tract. The device is compatible with endoscopes with a minimum accessory channel of 2.8 mm. The Disposable Hemostasis Clip is 230 cm long. The deployed clip portion of the Disposable Hemostasis Clip is stainless steel and nitinol while the introducer is nylon, stainless steel and nitinol.
More Information

Not Found

No
The summary describes a purely mechanical device for endoscopic clip placement and does not mention any software, algorithms, or data processing that would indicate the use of AI or ML.

Yes
The device is used for hemostasis, which is treating a medical condition (bleeding) in the gastrointestinal tract.

No
The device is described as being used for endoscopic clip placement for therapeutic purposes (hemostasis, marking) and describes no diagnostic capabilities.

No

The device description clearly states it is a physical device consisting of a metal clip and introducer made of stainless steel, nitinol, and nylon. It is a hardware device used for endoscopic procedures.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to provide information for diagnosis, monitoring, or screening.
  • Device Description and Intended Use: The description clearly states that this device is used for endoscopic clip placement within the gastrointestinal tract for purposes like marking, hemostasis, and treating bleeding. This is a surgical or interventional procedure performed directly on the patient's tissue, not a test performed on a sample outside the body.

The device is a medical device used for therapeutic and procedural purposes within the body, not for diagnostic testing of biological samples.

N/A

Intended Use / Indications for Use

This device is used for endoscopic clip placement within the gastrointestinal tract for the purpose of endoscopic marking, hemostasis for mucosal/submucosal defects less than 3 cm in the upper GI tract, bleeding ulcers, arteries less than 2 mm, and polyps less than 1.5 cm in diameter in the GI tract. This device is not intended for the repair of GI tract lumenal perforations.

Product codes

MND, PKL

Device Description

The Disposable Hemostasis Clip is a sterile, single use device that consists of a metal clip that detaches from the introducer to maintain approximation of tissue to achieve hemostasis in the gastrointestinal tract. The device is compatible with endoscopes with a minimum accessory channel of 2.8 mm. The Disposable Hemostasis Clip is 230 cm long. The deployed clip portion of the Disposable Hemostasis Clip is stainless steel and nitinol while the introducer is nylon, stainless steel and nitinol.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

gastrointestinal tract

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Performance testing consisting of non-clinical bench testing demonstrates that the Disposable Hemostasis Clip met the performance requirements to fulfill the intended use of the device. The device is substantially equivalent to currently cleared predicate devices.

Key Metrics

Not Found

Predicate Device(s)

K023903

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 876.4400 Hemorrhoidal ligator.

(a)
Identification. A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.(b)
Classification. Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

0

Image /page/0/Picture/0 description: The image shows the logo for Cook Medical. The logo is white text on a black background. The word "COOK" is on the top line, and the word "MEDICAL" is on the second line.

of 2 COOK ENDOSCOPY 4900 BETHANIA STATION ROAD WINSTON-SALEM, NC 27105 U.S.A. PHOHE: 336.744.0157 TOLL FREE: 800.245.4707 WWW.COOKMEDICAL.COM

. ﺎ ﻓﻲ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ

AUG 3 1 2012

121505

5. 510(k) Summary

Name:Wilson-Cook Medical, Inc. /Cook Endoscopy
Address:4900 Bethania Station Road
Winston-Salem, North Carolina 27105
Phone:(336)744-0157
Fax:(336)201-5994
Contact:Scottie Fariole, Global Regulatory Affairs Specialist
Date:May 18, 2012
Trade Name:Disposable Hemostasis Clip
Common Name:Hemostasis Clip
Classification Name:Hemorrhoidal ligator, Gastroenterology-Urology (21 CFR 876.4400,
Product Code MND)
Legally Marketed Devices:Wilson-Cook Endoscopic Clipping Device (K023903)
Description of the Device:The Disposable Hemostasis Clip is a sterile, single use device that
consists of a metal clip that detaches from the introducer to
maintain approximation of tissue to achieve hemostasis in the
gastrointestinal tract. The device is compatible with endoscopes
with a minimum accessory channel of 2.8 mm. The Disposable
Hemostasis Clip is 230 cm long. The deployed clip portion of the
Disposable Hemostasis Clip is stainless steel and nitinol while the
introducer is nylon, stainless steel and nitinol.
Intended Use:This device is used for endoscopic clip placement within the
gastrointestinal tract for the purpose of endoscopic marking,
hemostasis for mucosal/submucosal defects less than 3 cm in the
upper GI tract, bleeding ulcers, arteries less than 2 mm, and
polyps less than 1.5 cm in diameter in the GI tract. This device is
not intended for the repair of GI tract lumenal perforations.
Technological Characteristics:The Disposable Hemostasis Clip has similar technological
characteristics to the Wilson-Cook Endoscopic Clipping Device
(K023903) in terms of general design and function but differs in
terms of materials, dimensions, clip geometry and handle
operation. Additionally, the deployed clip portion of the device is
MR Conditional per ASMT F2503.

1

121505
pg 2 of 2

Performance Data:

Performance testing consisting of non-clinical bench testing demonstrates that the Disposable Hemostasis Clip met the performance requirements to fulfill the intended use of the device. The device is substantially equivalent to currently cleared predicate devices.

2

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2015

Wilson-Cook Medical, Inc. / Cook Endoscopy Scottie Fariole Global Regulatory Affairs Specialist 4900 Bethania Station Road Winston-Salem, NC 27105

Re: K121505

Trade/Device Name: Disposable Hemostasis Clip Regulation Number: 21 CFR§ 876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: II Product Code: PKL Dated (Date on orig SE ltr): August 21, 2012 Received (Date on orig SE ltr): August 22, 2012

Dear Scottie Fariole,

This letter corrects our substantially equivalent letter of August 31, 2012.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

3

Page 2 -

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Joyce M. Whang -S

for Benjamin R. Fisher, Ph.D.

Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

4. Indications for Use

510(k) Number (if known): K121505

Device Name: _ Disposable Hemostasis Clip

Indications for Use:

This device is used for endoscopic clip placement within the gastrointestinal tract for the purpose of endoscopic marking, hemostasis for mucosal/submucosal defects less than 3 cm in the upper GI tract, bleeding ulcers, arteries less than 2 mm, and polyps less than 1.5 cm in diameter in the GI tract. This device is not intended for the repair of GI tract lumenal perforations.

Prescription Use _ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Richard Leuu

inn Sidne tive, Gastro-Renal, and S10/k) Numr

Page __ of 1