(94 days)
The DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500, APLIO 400 MODEL TUS-A400 and APLIO 300 MODEL TUS-A300 is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, and musculoskeletal (both conventional and superficial).
The DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500. APLIO 400 MODEL TUS-A400 and APLIO 300 MODEL TUS-A300 are mobile system. These systems are Track 3 devices that employ a wide array of probes that include flat linear array, convex linear array, and sector array with a frequency range of approximately 2 MHz to 12 MHz.
The provided text is a 510(k) summary for the Toshiba DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500, APLIO 400 MODEL TUS-A400, and APLIO 300 MODEL TUS-A300 Version 2.1. It primarily focuses on demonstrating substantial equivalence to a predicate device (Toshiba DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500 / APLIO 400 MODEL TUS-A400 / APLIO 300 MODEL TUS-A300 V2.0; 510(k) control number K110870) and outlining the intended uses for various transducers.
Based on the provided document, the information requested in your prompt about acceptance criteria and a study proving performance (specifically in the context of AI/ML or new clinical parameters) is not present. This document predates the widespread regulatory submissions for AI/ML devices in medical imaging, and its focus is on general ultrasound system functionality and traditional clinical applications.
Therefore, I cannot provide a table of acceptance criteria and reported device performance for AI features, nor can I provide details about sample sizes for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC comparative effectiveness studies, or standalone algorithm performance studies related to AI.
The document does include detailed tables for each transducer showing the "Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:", listing various clinical applications and modes of operation. For each application, it indicates whether it's a "P" (previously cleared by FDA) or "E" (added under this appendix) indication, or "N" (new indication) for a very few. This is essentially a declaration of the intended uses, but it does not specify performance acceptance criteria or provide study results to demonstrate performance for these applications.
The document mentions compliance with several standards, such as IEC 60601-1 and its parts, IEC 62304, and AIUM-NEMA UD2/UD3 standards. These are general safety and performance standards for medical electrical equipment and ultrasound output measurement/display. They are compliance standards, not specific acceptance criteria for AI or diagnostic performance in patient studies.
In summary, the provided content is a regulatory submission for device clearance based on substantial equivalence to a predicate device, focusing on intended uses and compliance with general safety and performance standards. It does not contain the kind of detailed information about acceptance criteria and performance studies you're asking for, particularly concerning AI or specific diagnostic efficacy metrics.
{0}------------------------------------------------
510(k) Summary of Safety and Effectiveness:
AUG 1 6 2012
| Submitter's Name: | Toshiba America Medical Systems, Inc. |
|---|---|
| Address: | PO Box 2068,2441 Michelle Drive Tustin, CA 92781-2068 |
| Contact: | Charlemagne Chua, Manager Regulatory Affairs |
| Telephone No.: | (714) 730-5000 |
| Device Proprietary Name: | DIAGNOSTIC ULTRASOUND SYSTEMAPLIO 500 MODEL TUS-A500 Version 2.1APLIO 400 MODEL TUS-A400 Version 2.1APLIO 300 MODEL TUS-A300 Version 2.1 |
| Common Name: | Diagnostic Ultrasound System |
| Classification: | |
| Regulatory Class: | II |
| Review Category: | Tier II |
| Ultrasonic Pulsed Doppler Imaging System - Product Code: 90-IYN |
[Fed.Reg.No.:892.1550] Ultrasonic Pulsed Echo Imaging System - Product Code: 90-IYO [Fed.Reg.No.:892.1560] Diagnostic Ultrasonic Transducer - Product Code: 90-ITX [Fed. Reg. No.: 892.1570]
Identification of Predicate Devices:
Toshiba America Medical Systems believes that this device is substantially equivalent to:
- l) Toshiba DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500 / APLIO 400 MODEL TUS-A400 / APLIO 300 MODEL TUS-A300 V2.0 ; 510(k) control number K110870
Device Description:
The DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500. APLIO 400 MODEL TUS-A400 and APLIO 300 MODEL TUS-A300 are mobile system. These systems are Track 3 devices that employ a wide array of probes that include flat linear array, convex linear array, and sector array with a frequency range of approximately 2 MHz to 12 MHz.
Indications for Use:
The DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500, APLIO 400 MODEL TUS-A400 and APLIO 300 MODEL TUS-A300 is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, and musculoskeletal (both conventional and superficial).
{1}------------------------------------------------
Declaration of Conformity
p
This device is designed and manufactured in conjunction with the Quality System Regulation, IEC 60601-1 (applicable portions), IEC 60601-1-1 (applicable portion), IEC 60601-1-2 (applicable portion), IEC 60601-1-4 (applicable portion), IEC60601-2-37 (applicable portions), IEC 62304 (applicable portion) and the AIUM-NEMA UD2 Output Measurement Standard as applied to Track 3 Ultrasound systems and the AIUM-NEMA UD3 Output Display Standard.
Standards form:
Please see the attached standard form of the IEC 60601-1-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-37 and IEC 62304.
イ
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a stylized emblem. The emblem consists of a white eagle with its wings spread, superimposed over a blue background. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the emblem.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Toshiba Medical Systems Corporation, Japan % Mr. Charlemagne Chua Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780-2068
AUG 16 2012
Re: K121422
Trade/Device Name: Aplio™ 500/400/300 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: July 13, 2012 Received: July 16, 2012
Dear Mr. Chua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Aplio™ 500/400/300 Diagnostic Ultrasound System, as described in your premarket notification:
Transducer Model Number
| PST-25BT | PVT-661VT | PVT-770RT | PLT-1204BT |
|---|---|---|---|
| PST-30BT | PVT-674BT | PLT-604AT | PLT-1204BX |
| PST-50BT | PVT-675MV | PLT-704AT | PLT-1204MV |
| PST-65AT | PVT-681MV | PLT-704SBT | PET-508MA |
| PVT-375BT | PVT-712BT | PLT-705BTF | PET-510MB |
| PVT-375MV | PVT-745BTF | PVT-705BTH | PET-512MC |
| PVT-382BT | PVT-745BTH | PLT-805AT | PC-20M |
| PVT-382MV | PVT-745BTV | PLT-1202S | PC-50M |
{3}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Joshua Nipper at 1 (301) 796-6524.
Sincerely Yours,
Michael D O'Shea for
Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name:
Aplio™ 500 /400/300 Diagnostic Ultrasound System
Indications for Use:
The DIAGNOSTIC ULTRASOUND SYSTEM APLIO 500 MODEL TUS-A500, APLIO 400 MODEL TUS-A400 and APLIO 300 MODEL TUS-A300 is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, and musculoskeletal (both conventional and superficial).
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number K124222
{5}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: __________________________________________________________________________________________________________________________________________________________________ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | P | 2 | P | P | P | P | 5,7 | |||||
| Fetal | P | P | P | P | P | 2 | P | P | P | P | 5,7 | |
| Abdominal | P | P | P | P | P | 2,3 | P | P | P | P | 5,7,12 | |
| Intra-operative (Abdominal) | P | P | P | P | P | 2 | P | P | P | 4,5 | ||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | P | 2,3 | P | P | P | P | 5,7,12 | |
| Small Organ (Note 1) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,11 | ||
| Neonatal Cephalic | P | P | P | P | P | 3 | P | P | P | |||
| Adult Cephalic | P | P | P | P | P | 3 | P | P | P | |||
| Trans-rectal | P | P | P | P | P | 2 | P | P | P | P | 4,5,7,11 | |
| Trans-vaginal | P | P | P | P | P | 2 | P | P | P | P | 4,5,7 | |
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,11 | ||
| Musculo-skeletal (Superficial) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,11 | ||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | P | P | P | P | P | 3 | P | P | P | P | 4 | |
| Cardiac Pediatric | P | P | P | P | P | 3 | P | P | P | P | 4 | |
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | P | P | P | P | P | 3 | P | 4 | ||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,11 | ||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI
- Note 5 ApliPure
Note 13 2D WMT
Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro ation and Safety
510K K12.1422
Prescription Use Only (Per 21 CRF801.109)
{6}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: ______________________________________________________________________________________________________________________________________________________________________ Transducer: PST-25BT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | P | P | P | P | P | 3 | P | P | P | 11 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | P | 3 | P | P | P | |||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | P | P | P | P | P | 3 | P | P | P | |||
| Adult Cephalic | P | P | P | P | P | 3 | P | P | P | |||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | . | |||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | P | P | P | P | P | 3 | P | P | P | P | 4,13 | |
| Cardiac Pediatric | P | P | P | P | P | 3 | P | P | P | P | 4,13 | |
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| (Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
516
Prescription Use Only (Per 21 CRF801.109)
(Division Sign-Off)
{7}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PST-30BT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | P | P | P | P | P | 3 | P | P | P | 11 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | P | 3 | P | P | P | |||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | P | P | P | P | P | 3 | P | P | P | |||
| Adult Cephalic | P | P | P | P | P | 3 | P | P | P | |||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | P | P | P | P | P | 3 | P | P | P | P | 4,13 | |
| Cardiac Pediatric | P | P | P | P | P | 3 | P | P | P | P | 4,13 | |
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle.
Note 1 Small organ includes BM; B/PWD; BDF/PWD; BDF/PWD; BDF/A/DF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion
Note 13 2D WMT Prescription Use Only (Per 21 CRF801.109)
(Division Sign-Off)
of Radiological Devices
510
{8}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:__ PST-50BT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other(Note) |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | E | E | E | E | E | 3 | E | E | E | 11 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | E | E | E | E | E | 3 | E | E | E | |||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | E | E | E | E | E | 3 | E | E | E | |||
| Adult Cephalic | E | E | E | E | E | 3 | E | E | E | |||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | E | E | E | E | E | 3 | E | E | E | E | 4,13 | |
| Cardiac Pediatric | E | E | E | E | E | 3 | E | E | E | E | 4,13 | |
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K090158
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 2 Combined mode includes BAN, BR WD, BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 MicroPure
- Note 7 Precision Imaging
- Note 8 STIC
- Note 9 3D Color (Volume Color) Note 10 STIC Color
- Note 11 Elastography
- Note 12 Fusion
- Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
(Division Sign-Off)
Division of Radiological Devices
510k K121492
MAY 2012
{9}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PST-65AT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | TIII | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | |||||||||||||
| Fetal | |||||||||||||
| Abdominal | E | E | E | E | E | 3 | E | E | E | 11 | |||
| Intra-operative (Abdominal) | |||||||||||||
| Intra-operative (Neuro) | |||||||||||||
| Laparoscopic | |||||||||||||
| Pediatric | E | E | E | E | E | 3 | E | E | E | ||||
| Small Organ (Specify) (1) | |||||||||||||
| Neonatal Cephalic | E | E | E | E | E | 3 | E | E | E | ||||
| Adult Cephalic | E | E | E | E | E | 3 | E | E | E | ||||
| Trans-rectal | |||||||||||||
| Trans-vaginal | |||||||||||||
| Trans-urethral | |||||||||||||
| Trans-esoph. (non-Card.) | |||||||||||||
| Musculo-skeletal (Conventional) | |||||||||||||
| Musculo-skeletal (Superficial) | |||||||||||||
| Intravascular | |||||||||||||
| Other (Specify) | |||||||||||||
| Cardiac Adult | E | E | E | E | E | 3 | E | E | E | E | 4,13 | ||
| Cardiac Pediatric | E | E | E | E | E | 3 | E | E | E | E | 4,13 | ||
| Intravascular (Cardiac) | |||||||||||||
| Trans-esoph. (Cardiac) | |||||||||||||
| Intra-cardiac | |||||||||||||
| Other (Specify) | |||||||||||||
| Peripheral vessel | |||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645
Note I Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color
Note 11 Elastography · Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
BealEBS
(Division Sign-Off)
on of Radiological
ﻟﻘﺎﺩ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ
MAY 2012
{10}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: __Aplio 300, Aplio 400, Aplio 500 v2,1 Transducer: _ PVT-375BT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | |||||||||||
| Fetal | P | P | P | P | 2 | P | P | P | 5,7 | ||
| Abdominal | P | P | P | P | 2 | P | P | P | 5, 7,11,1 | ||
| Intra-operative (Abdominal) | |||||||||||
| Intra-operative (Neuro) | |||||||||||
| Laparoscopic | |||||||||||
| Pediatric | P | P | P | P | 2 | P | P | P | 5, 7,12 | ||
| Small Organ (Specify) (1) | |||||||||||
| Neonatal Cephalic | |||||||||||
| Adult Cephalic | |||||||||||
| Trans-rectal | |||||||||||
| Trans-vaginal | |||||||||||
| Trans-urethral | |||||||||||
| Trans-esoph. (non-Card.) | |||||||||||
| Musculo-skeletal (Conventional) | |||||||||||
| Musculo-skeletal (Superficial) | |||||||||||
| Intravascular | |||||||||||
| Other (Specify) | |||||||||||
| Cardiac Adult | |||||||||||
| Cardiac Pediatric | |||||||||||
| Intravascular (Cardiac) | |||||||||||
| Trans-esoph. (Cardiac) | |||||||||||
| Intra-cardiac | |||||||||||
| Other (Specify) | |||||||||||
| Peripheral vessel | |||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note I Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
(Division Sign-Off)
510(k): APLIO 500/400/300, v2.1 Page 194 of 3006
{11}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PVT-375MV
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | P | P | P | P | P | 2 | P | P | P | N | 5,7,8,9,10 | |
| Abdominal | P | P | P | P | P | 2 | P | P | P | N | 5,7,8,9,10 | |
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | P | 2 | P | P | P | N | 5,7,8,9,10 | |
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645
Note 1 Small organ includes thyroid, breast and testicle. Note 1 Small organ alstades arfrenz, e. BRWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
A-7
{12}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PVT-382BT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | ||||||||||||||
| Fetal | P | P | P | P | P | 2 | P | P | P | 5,7 | ||||
| Abdominal | P | P | P | P | P | 2 | P | P | P | 5,7 | ||||
| Intra-operative (Abdominal) | ||||||||||||||
| Intra-operative (Neuro) | ||||||||||||||
| Laparoscopic | ||||||||||||||
| Pediatric | P | P | P | P | P | 2 | P | P | P | 5,7 | ||||
| Small Organ (Specify) (1) | ||||||||||||||
| Neonatal Cephalic | ||||||||||||||
| Adult Cephalic | ||||||||||||||
| Trans-rectal | ||||||||||||||
| Trans-vaginal | ||||||||||||||
| Trans-urethral | ||||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||||
| Intravascular | ||||||||||||||
| Other (Specify) | ||||||||||||||
| Cardiac Adult | ||||||||||||||
| Cardiac Pediatric | ||||||||||||||
| Intravascular (Cardiac) | ||||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||||
| Intra-cardiac | ||||||||||||||
| Other (Specify) | ||||||||||||||
| Peripheral vessel | ||||||||||||||
| Other (Specify) |
N= new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography
Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
Ruch D. 82
{13}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: __Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PVT-382MV
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | |||||||||||||
| Fetal | P | P | P | P | P | 2 | P | P | P | P | 5, 7,9 | ||
| Abdominal | P | P | P | P | P | 2 | P | P | P | P | 5, 7,9 | ||
| Intra-operative (Abdominal) | |||||||||||||
| Intra-operative (Neuro) | |||||||||||||
| Laparoscopic | |||||||||||||
| Pediatric | P | P | P | P | P | 2 | P | P | P | P | 5,7,9 | ||
| Small Organ (Specify) (1) | |||||||||||||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Trans-rectal | |||||||||||||
| Trans-vaginal | |||||||||||||
| Trans-urethral | |||||||||||||
| Trans-esoph. (non-Card.) | |||||||||||||
| Musculo-skeletal (Conventional) | |||||||||||||
| Musculo-skeletal (Superficial) | |||||||||||||
| Intravascular | |||||||||||||
| Other (Specify) | |||||||||||||
| Cardiac Adult | |||||||||||||
| Cardiac Pediatric | |||||||||||||
| Intravascular (Cardiac) | |||||||||||||
| Trans-esoph. (Cardiac) | |||||||||||||
| Intra-cardiac | |||||||||||||
| Other (Specify) | |||||||||||||
| Peripheral vessel | |||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix ·Previous 510(k) of the transducer: K103645
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color
Note 11 Elastography Note 12 Fusion
Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
(Division Sign-Off)
Division of Radiological Devices
and
A-9 -
{14}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: ______________________________________________________________________________________________________________________________________________________________________ Transducer:_ PVT-661VT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | P | P | P | P | P | 2 | P | P | P | 4,5,7,11 | ||
| Trans-vaginal | P | P | P | P | P | 2 | P | P | P | 4,5,7,11 | ||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle. .
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT .
Prescription Use Only (Per 21 CRF801.109)
vision Sion-Off
ological Devices
{15}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PVT-674BT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] | |
| Ophthalmic | ||||||||||||
| Fetal | P | P | P | P | 2 | P | P | P | 5,7 | |||
| Abdominal | P | P | P | P | 2 | P | P | P | 5,7 | |||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | 2 | P | P | P | 5,7 | |||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes BM; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF 801.109)
(Division Sign-Off)
{16}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: _Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PVT-675MV ___
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | P | P | P | P | 2 | P | P | P | P | 5,7,8,9,10 | ||
| Abdominal | P | P | P | P | 2 | P | P | P | P | 5,7,8,9,10 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | 2 | P | P | P | P | 5,7,8,9,10 | ||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K 110870
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
(Division Sign-Division of Radiological Devices
510k
MAY 2012
{17}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1)TUS-A500 Ultrasound System
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:___ PVT-681MV
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | P | P | P | P | 2 | P | P | P | P | 4,5,7,9,1 | ||
| Trans-vaginal | P | P | P | P | 2 | P | P | P | P | 4,5,7,9,1 | ||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
- Note 5 ApliPure
- Note 6 MicroPure
Note 7 Precision Imaging
-
Note 8 STIC Note 9 3D Color (Volume Color)
Note 10 STIC Color -
Note 11 Elastography
Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
Michael D. O'Keefe
(Division Sign-Off)
Division of Radiological Devices
2012
{18}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1)TUS-A500 Ultrasound System
System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PVT-712BT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | P | P | P | P | 2 | P | P | P | P | 5,7 | ||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | P | P | P | P | 2 | P | P | P | 5,7 | |||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | P | P | P | P | 2 | P | P | P | 5,7 | |||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication: P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CFR801.109)
(Division Sign-Off Division of Radiological Devices
STOR
MAY 20
{19}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 PVT-745BTF Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | E | E | E | E | E | 2 | E | E | E | 5,7 | ||
| Intra-operative (Abdominal) | E | E | E | E | E | 2 | E | E | E | 5,7 | ||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | E | E | E | E | E | 2 | E | E | E | 5,7 | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: None
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Notc 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only (Per
(Division Sign-Off)
Division of Radiological Devices
OIVD
{20}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PVT-745BTH
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | E | E | E | E | 2 | E | E | E | 5,7 | |||
| Intra-operative (Abdominal) | E | E | E | E | 2 | E | E | E | 5,7 | |||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | E | E | E | E | 2 | E | E | E | 5,7 | |||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: None
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
A-16
100 rescription Use On.
510
{21}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
Toshiba America Medical Systems, Inc.
System: Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:__ PVT-745BTV
Intended Use; Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | P | P | P | P | P | 2 | P | P | P | 5,7 | ||
| Intra-operative (Abdominal) | P | P | P | P | P | 2 | P | P | P | 5,7 | ||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
cal Devices
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M: B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only (Per 21 CRF80
Prescription Use Only (Per 21 CRF801.109)
(Division Sign-Off)
MAY 2012
{22}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PVT-770RT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | E | E | E | E | 2 | E | E | E | 4,5,7,1 | |||
| Trans-vaginal | E | E | E | E | 2 | E | E | E | 4,5,7,1 | |||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 2 Combined mode includes BMF, BF WD, BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
Mulin D'then
(Division Sign-Off)
of Radiological Devices
{23}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
Toshiba America Medical Systems, Inc.
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PLT-604AT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | E | E | PWD | E | 2 | E | E | E | E | 5,7 | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | E | E | E | E | 2 | E | E | E | E | 5,7 | ||
| Musculo-skeletal (Superficial) | E | E | E | E | 2 | E | E | E | E | 5,7 | ||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | E | E | E | E | 2 | E | E | E | E | 5,7 | ||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645
Note 1 Small organ includes thyroid, breast and testicle. Note 1 Shartorgan motiades arrivate, 6 R/F/FWD; BDF/PWD; BDF/A/DF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only (Per 21 CRF801.109)
Prescription Use Only (Per 21 CFR 801.109)
Division
MAY 2012
{24}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 PLT-704AT Transducer:_
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | ||||||||||||||
| Fetal | ||||||||||||||
| Abdominal | ||||||||||||||
| Intra-operative (Abdominal) | ||||||||||||||
| Intra-operative (Neuro) | ||||||||||||||
| Laparoscopic | ||||||||||||||
| Pediatric | ||||||||||||||
| Small Organ (Specify) (1) | E | E | E | E | 2 | E | E | E | 5,7 | |||||
| Neonatal Cephalic | ||||||||||||||
| Adult Cephalic | ||||||||||||||
| Trans-rectal | ||||||||||||||
| Trans-vaginal | ||||||||||||||
| Trans-urethral | ||||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||||
| Musculo-skeletal(Conventional) | E | E | E | E | 2 | E | E | E | 5,7 | |||||
| Musculo-skeletal (Superficial) | E | E | E | E | 2 | E | E | E | 5,7 | |||||
| Intravascular | ||||||||||||||
| Other (Specify) | ||||||||||||||
| Cardiac Adult | ||||||||||||||
| Cardiac Pediatric | ||||||||||||||
| Intravascular (Cardiac) | ||||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||||
| Intra-cardiac | ||||||||||||||
| Other (Specify) | ||||||||||||||
| Peripheral vessel | E | E | E | E | 2 | E | E | E | 5,7 | |||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645
Note 1 Small organ includes thyroid, breast and testicle. Note 1 Sman organ includes BlM; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only (Per 21 CRF
Prescription Use Only (Per 21 CFR 801.109)
(Division Sign-Off)
Sign-Off)
Radiological Devices
OUD
422
A-20
510(k): APLIO 500/400/300, v2.1 Page 208 of 3006
{25}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
Toshiba America Medical Systems, Inc.
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:_ PLT-704SBT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | P | P | P | P | P | 2 | P | P | P | 5,7 | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | 2 | P | P | P | 5,7 | ||
| Musculo-skeletal (Superficial) | P | P | P | P | P | 2 | P | P | P | 5,7 | ||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | P | P | P | P | 2 | P | P | P | 5,7 | ||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only
(Division Sign-Off)
Division of Radiological Devices
{26}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 PLT-70SBTF Transducer:_
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | B | M | PWD CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] | ||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | ||||||||||||||
| Fetal | ||||||||||||||
| Abdominal | E | E | E | E | 2 | E | E | E | 5,7 | |||||
| Intra-operative (Abdominal) | E | E | E | E | 2 | E | E | E | 5,7 | |||||
| Intra-operative (Neuro) | ||||||||||||||
| Laparoscopic | ||||||||||||||
| Pediatric | ||||||||||||||
| Small Organ (Specify) (1) | ||||||||||||||
| Neonatal Cephalic | ||||||||||||||
| Adult Cephalic | ||||||||||||||
| Trans-rectal | ||||||||||||||
| Trans-vaginal | ||||||||||||||
| Trans-urethral | ||||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||||
| Intravascular | ||||||||||||||
| Other (Specify) | ||||||||||||||
| Cardiac Adult | ||||||||||||||
| Cardiac Pediatric | ||||||||||||||
| Intravascular (Cardiac) | ||||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||||
| Intra-cardiac | ||||||||||||||
| Other (Specify) | ||||||||||||||
| Peripheral vessel | ||||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103645
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
al Devices
{27}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 PVT-705BTH Transducer:_
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD CWD] | ColorDoppler | Combined(Specify)* | THI DynamicFlow | Power | CHI2D | 4D | OtherNote | ||
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | E | E | E | E | 2 | E | । ਜ | E | 5,7 | |||
| Intra-operative (Abdominal) | E | E | E | E | 2 | E | E | E | 5,7 | |||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| (Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix
lological Devices
Previous 510(k) of the transducer: K103645
:
MAY 201
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion ` Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
(Division Sign-Off)
{28}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: _Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PLT-805AT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | P | P | P | P | P | 2 | P | P | P | 5,6,7,11 | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | P | P | P | P | P | 2 | P | P | P | 5,6,7,11 | ||
| Musculo-skeletal (Superficial) | P | P | P | P | P | 2 | P | P | P | 5,6,7,11 | ||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | P | P | P | P | 2 | P | P | P | 5,6,7,11 | ||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870 i
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M: B/PWD; BDF/PWD; BDF/MDF; 8DF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
(Division Sign-Off)
ion of Radiological Devices
MAY 2012
510k
{29}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
·
System: __ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PLT-1202S
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical ApplicationSpecific(Tracks 3) | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] | |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | P | P | P | P | 2 | P | P | 4,5,11 | ||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | P | P | P | P | 2 | P | P | 4,5,11 | ||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal (Conventional) | P | P | P | P | 2 | P | P | 4,5,11 | ||||
| Musculo-skeletal (Superficial) | P | P | P | P | 2 | P | P | 4,5,11 | ||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | N | N | N | N | 2 | N | N | 4,5,11 | ||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD
-
Note 4 TDI
Note 5 ApliPure -
Note 6 MicroPure Note 7 Precision Imaging
Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color
Note 11 Elastography
Note 12 Fusion Note 13 2D WMT
MAY 2012
Prescription Use Only (Per 21 CRF801.109)
Manfred D. Krahn
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Division C Divisi un of Radiolo
A-25
510(k): APLIO 500/400/300, v2.1 Page 213 of 3006
{30}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: _ Aplio 300, Aplio 400, Aplio 500 v2, 1 Transducer:___PLT-1204BT
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | OtherNote |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,1 | ||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,1 | ||
| Musculo-skeletal (Superficial) | P | P | P | P | P | 2 | P | P | P | 4,5,6,7,1 | ||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | N | N | N | N | N | 2 | N | N | N | 4,5,6,7,1 | ||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
-
Note 4 TDI
Note 5 ApliPure -
Note 6 MicroPure
Note 7 Precision Imaging -
Note 8 STIC
Note 9 3D Color (Volume Color) -
Note 10 STIC Color
Note 11 Elastography
Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
(Division Sign
of Radiolo
- 2313 401 401 10.
{31}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
Aplio 500 v2.1 System: Transducer:_ PLT-1204BX
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation Clinical Application Specific PWD CWD Color Combined THI Dynamic Power CHI 4D Other B M Doppler (Specify) Flow 2D [Note] (Tracks 3) Ophthalmic Fetal Abdominal Intra-operative (Abdominal) Intra-operative (Neuro) Laparoscopic Pediatric P P P P 5,7 Small Organ (Specify) (1) P P P 2 Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Musculo-skeletal P P P P 5,7 b P P 2 (Conventional) Musculo-skeletal (Superficial) P P P P 2 P P P 5,7 Intravascular Other (Specify) Cardiac Adult Cardiac Pediatric Intravascular (Cardiac) Trans-esoph. (Cardiac) Intra-cardiac Other (Specify) Peripheral vessel P P P P P P P 5,7 2 Other (Specify)
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note I Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M: B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109)
Division G
{32}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
A M SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE SE FELLER
System: _Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PLT-1204MV
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | P | P | P | P | P | 2 | P | P | P | P | 5,7,8,9,1 | |
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | P | P | P | P | P | 2 | P | P | P | P | 5,7,8,9,1 | |
| Musculo-skeletal (Superficial) | P | P | P | P | P | 2 | P | P | P | P | 5,7,8,9,1 | |
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | P | P | P | P | 2 | P | P | P | P | 5,7,8,9,1 | |
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
Note 5 ApliPure
Note 6 MicroPure
Note 7 Precision Imaging
Note 8 STIC Note 9 3D Color (Volume Color)
Note 10 STIC Color
Note 11 Elastography Note 12 Fusion
Note 13 2D WMT
Prescription Use Only (Per 21 CFR 801.109)
{33}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: __ Aplio 300, Aplio 400. Aplio 500 v2.1 Transducer:_ PET-508MA
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | E | E | E | E | E | 3 | E | 4,13 | ||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: None
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
ical Devices
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD
-
Note 4 TDI
Note 5 ApliPure -
Note 6 MicroPure
Note 7 Precision Imaging -
Note 8 STIC
Note 9 3D Color (Volume Color) -
Note 10 STIC Color
Note 11 Elastography
Note 12 Fusion
Note 13 2D WMT
MAY
Prescription Use Only (Per 21 CRF801.109)
A-29
510(k): APLIO 500/400/300, v2.1 Page 217 of 3006
{34}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PET-510MB
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] | |
| Ophthalmic | |||||||||||||
| Fetal | |||||||||||||
| Abdominal | |||||||||||||
| Intra-operative (Abdominal) | |||||||||||||
| Intra-operative (Neuro) | |||||||||||||
| Laparoscopic | |||||||||||||
| Pediatric | |||||||||||||
| Small Organ (Specify) (1) | |||||||||||||
| Neonatal Cephalic | |||||||||||||
| Adult Cephalic | |||||||||||||
| Trans-rectal | |||||||||||||
| Trans-vaginal | |||||||||||||
| Trans-urethral | |||||||||||||
| Trans-esoph. (non-Card.) | |||||||||||||
| Musculo-skeletal(Conventional) | |||||||||||||
| Musculo-skeletal (Superficial) | |||||||||||||
| Intravascular | |||||||||||||
| Other (Specify) | |||||||||||||
| Cardiac Adult | |||||||||||||
| Cardiac Pediatric | |||||||||||||
| Intravascular (Cardiac) | |||||||||||||
| Trans-esoph. (Cardiac) | P | P | P | P | P | 3 | P | 4,13 | |||||
| Intra-cardiac | |||||||||||||
| Other (Specify) | |||||||||||||
| Peripheral vessel | |||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: KI 10870
Note 1 Small organ includes thyroid, breast and testicle.
Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD
Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD
- Note 4 TDI
Note 5 ApliPure
Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color
Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRF801.109) (Division Sign-Off)
vision of Radiological Devices
510k MAY 2012
{35}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: _ Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer: PET-512MC
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | ||||||||||||
| Cardiac Pediatric | ||||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | E | E | E | E | E | 3 | E | 4,13 | ||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | ||||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K103629
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21 CRE801 109 Division Sign-Off
vision of Radiological Devices
MAY 2012
A-3 I
{36}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
System: Aplio 300, Aplio 400, Aplio 500 v2.1 Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other(Note) |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | P | |||||||||||
| Cardiac Pediatric | P | |||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | |||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT Prescription Use Only (Per 21 CRF801.109)
Division Sign-Off
ogical Devices
.
{37}------------------------------------------------
510(k) Premarket Notification Aplio™500(v2.1) TUS-A500 Ultrasound System
Aplio 300, Aplio 400, Aplio 500 v2.1 System: Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify)* | THI | DynamicFlow | Power | CHI2D | 4D | Other[Note] |
| Ophthalmic | ||||||||||||
| Fetal | ||||||||||||
| Abdominal | ||||||||||||
| Intra-operative (Abdominal) | ||||||||||||
| Intra-operative (Neuro) | ||||||||||||
| Laparoscopic | ||||||||||||
| Pediatric | ||||||||||||
| Small Organ (Specify) (1) | ||||||||||||
| Neonatal Cephalic | ||||||||||||
| Adult Cephalic | ||||||||||||
| Trans-rectal | ||||||||||||
| Trans-vaginal | ||||||||||||
| Trans-urethral | ||||||||||||
| Trans-esoph. (non-Card.) | ||||||||||||
| Musculo-skeletal(Conventional) | ||||||||||||
| Musculo-skeletal (Superficial) | ||||||||||||
| Intravascular | ||||||||||||
| Other (Specify) | ||||||||||||
| Cardiac Adult | P | |||||||||||
| Cardiac Pediatric | P | |||||||||||
| Intravascular (Cardiac) | ||||||||||||
| Trans-esoph. (Cardiac) | ||||||||||||
| Intra-cardiac | ||||||||||||
| Other (Specify) | ||||||||||||
| Peripheral vessel | P | |||||||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K110870
Note 1 Small organ includes thyroid, breast and testicle. Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 5 ApliPure Note 6 MicroPure Note 7 Precision Imaging Note 8 STIC Note 9 3D Color (Volume Color) Note 10 STIC Color Note 11 Elastography Note 12 Fusion Note 13 2D WMT
Prescription Use Only (Per 21, GRF80 109 (Division Sign-Off
vision of Radiological Devices
- 4 - 8 - 4 - 4 - 4 - 4 - 4 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -
§ 892.1560 Ultrasonic pulsed echo imaging system.
(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.