(126 days)
The Mini Patch, model: WL-2301A & WL-2301B Transcutaneous Electrical Nerve Stimulators are intended for temporary relief of pain associated with sore and aching muscles in the upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
The Mini Patch, models WL-2301B are transcutaneous electrical nerve stimulator used for pain relief by applying an electrical current to electrodes, which are attached on the skin. The output and waveform characteristic is fixed for every model, only the intensity is adjustable.
The Mini Patch, models WL-2301A & WL-2301AB, consist mainly of two parts: the stimulus generator, electrode. The stimulus generator generates the output current specified as the input of controller. The output port transmits the output current to the electrode, which is attached to the patient's skin so as to transmit this stimulus current to the patient for pain relief.
The stimulation mode for Mini Patch is burst mode with fixed pulse rate, frequency, and fixed timer, only amplitude is adjustable. This operation way is considered the simplification from the burst mode of comparison clear model, WL-2302(K020020). Every model of Mini-TENS has its individual stimulation operation cycle.
This 510(k) summary describes a Transcutaneous Electrical Nerve Stimulator (TENS) device, the "Mini Patch" (models WL-2301A & WL-2301B), intended for temporary pain relief. The submission focuses on demonstrating substantial equivalence to a predicate device, not on proving clinical efficacy through a new clinical effectiveness study. Therefore, many of the requested criteria (like ground truth, expert adjudication, MRMC studies, standalone performance, and training set information) are not applicable as this type of submission typically relies on engineering and performance testing to show the new device operates similarly and safely to an already cleared device.
Here's a breakdown based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
For this type of device and submission (510(k) for substantial equivalence), the "acceptance criteria" are primarily established through compliance with recognized standards and comparison to a predicate device's specifications. The study demonstrating this is a non-clinical comparison and verification/validation testing.
| Acceptance Criteria (Comparison to Predicate Device WL-2407 (K091757)) | Reported Device Performance (Mini Patch WL-2301A & WL-2301B) |
|---|---|
| Intended Use: Same as predicate | Same (temporary relief of pain in upper/lower extremities due to sore muscles) |
| Product Code: Same as predicate (NUH) | Same |
| Regulation Number: Same as predicate (21 CFR 882.5890) | Same |
| Prescription or OTC: Same as predicate (OTC) | Same |
| Manufacturer: Same as predicate (Well-Life) | Same |
| Power Source(s): Similar (Predicate: 1.5Vx3 AAA; New: 3Vx1) | 3Vx1 |
| Method of Line Current Isolation: Same as predicate (Type BF) | Same |
| Patient Leakage Current (Normal): Under 0.1 uA | Same (Under 0.1 uA) |
| Patient Leakage Current (Single Fault): Under 0.5 uA | Same (Under 0.5 uA) |
| Number of Output Modes: Similar (Predicate: 8; New: 3) | 3 |
| Waveform: Same as predicate (Biphasic, Rectangular) | Same |
| Max Output Voltage (@500Ω): Predicate: 40V; New: 30V | 30V |
| Max Output Current (@500Ω): Predicate: 80mA; New: 60mA | 60mA |
| Duration of Primary Phase: Predicate: 260 Max usec; New: 250 Max usec | 250 Max usec |
| Pulse Duration: Predicate: 500.26 ms/Max.; New: 33.58 ms/Max. | 33.58 ms/Max. |
| Frequency (Rate): Same as predicate (60 Max Hz) | Same (60 Max Hz) |
| Symmetrical phases? Same as predicate (No) | Same (No) |
| Net charge: Zero for symmetrical biphasic wave | Achieved (Zero for symmetrical biphasic wave) |
| Max. phase charge: Predicate: 20.8 uc; New: 15 uc | 15 uc |
| Max. current Density: Predicate: 0.050 mA/cm²; New: 0.055 mA/cm² | 0.055 mA/cm² |
| Max. Average Current (RMSA) (@500Ω): Predicate: 38.43mA; New: 29.394mA | 29.394 mA |
| Max. Power Density: Predicate: 0.0020Watts/cm²; New: 0.0017Watts/cm² | 0.0017Watts/cm² |
| Compliance with Voluntary Standards: ANSI/AAMI NS4, IEC 60601-1, IEC 60601-1-2 | Same |
| Compliance with 21 CFR 898: Yes | Same |
| Software Verification: Carried out according to FDA guidance | Confirmed as carried out |
The "study" that proves the device meets the acceptance criteria is the non-clinical testing, verification, and validation activities performed by the manufacturer, as summarized in sections 6, 8, and 9 of the 510(k) submission. These tests compare the new device's technical specifications and safety features against those of the predicate device and relevant voluntary standards to demonstrate that the small differences do not raise new questions of safety or effectiveness.
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: Not applicable in the context of a clinical test set. The submission refers to engineering and performance verification and validation tests for the device itself, not patient data. These tests would involve multiple units of the new device to confirm consistency and adherence to specifications.
- Data Provenance: Not applicable for clinical data. The data provenance is from the manufacturer's internal engineering and testing labs in Taiwan, ROC, where the device was developed and tested. This is a retrospective comparison to an existing predicate device's specifications and prospective engineering testing of the new device's physical and electrical properties.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. This submission is for substantial equivalence of a TENS device based on engineering and performance testing, not a diagnostic or clinical study requiring expert-established ground truth from patient data. The "ground truth" here is the adherence to electrical engineering specifications and safety standards as outlined by regulatory bodies and voluntary standards.
4. Adjudication method for the test set
- Not applicable. No expert adjudication of clinical cases was performed, as this was not a clinical study involving interpretation of patient data.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is a TENS unit, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a physical medical device (TENS unit), not an algorithm. Its "standalone" performance refers to its operational characteristics (output voltage, current, etc.) which were assessed through engineering tests as described in the comparison tables.
7. The type of ground truth used
- Engineering Specifications and Safety Standards (e.g., ANSI/AAMI, IEC 60601-1, IEC 60601-1-2): The "ground truth" for this submission is adherence to established electrical and mechanical safety standards and conformity to the operational parameters of a legally marketed predicate device (Well-Life OTC TENS, K091757), which itself was cleared based on similar principles.
8. The sample size for the training set
- Not applicable. This device does not use machine learning or AI, so there is no training set in the conventional sense.
9. How the ground truth for the training set was established
- Not applicable. As there is no training set, this is irrelevant.
{0}------------------------------------------------
510(K) Summary
SEP 7 2012
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K121353
Submitter's Identifications: 1.
Well Life Healthcare Limited 1F. No. 16. Lane 454. Jungjeng Road. Yunghe City, Taipei County 234, Taiwan, ROC
Contact: Jenny Hsieh Telephone: + 886 2 2928 2112
Date of Summary Preparation: August 17, 2012
- Name of the Device:
Trade/Device Name: Mini Patch, model : WL-2301A & WL-2301B. Regulation Number: 21 CFR 882.5890 Regulation Name: Stimulator, Nerve, Transcutaneous For Pain Relief. Requlatory Class: II -Product Code: NUH
- Information of the 510(k) Cleared Device (Predicate Device): ന് Well-Life OTC TENS for Arm & Leg Pain Relief. (K091757)
- Device Description:
The Mini Patch, models WL-2301B are transcutaneous electrical nerve stimulator used for pain relief by applying an electrical current to electrodes, which are attached on the skin. The output and waveform characteristic is fixed for every model, only the intensity is adjustable.
The Mini Patch, models WL-2301A & WL-2301AB, consist mainly of two parts: the stimulus generator, electrode. The stimulus generator generates the output current specified as the input of controller. The output port transmits the output current to the electrode, which is attached to the patient's skin so as to transmit this stimulus current to the patient for pain relief.
The stimulation mode for Mini Patch is burst mode with fixed pulse rate, frequency, and fixed timer, only amplitude is adjustable. This operation way is considered the simplification from the burst mode of comparison clear model, WL-2302(K020020). Every model of Mini-TENS has its individual stimulation operation cycle.
{1}------------------------------------------------
5. Intended Use:
The Mini Patch, model: WL-2301A & WL-2301B Transcutaneous Electrical Nerve Stimulators are intended for temporary relief of pain associated with sore and aching muscles in the upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities.
Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are ଚ. as follows:
Compliance to applicable voluntary standards includes ANSI/AAMI, NS4-1985, as well as IEC 60601-1, and IEC 60601-1-2 requirement.
In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.
7. Comparison of Significant device features
| ComparisonFeatures | Models to be Compared | |
|---|---|---|
| Model | OTC TENS for Arm & Leg | Mini Patch |
| 510(K) No. | WL-2407 | WL-2301A & WL-2301B |
| Prescription orOTC | K091757 | K121353 |
| Indication for use | OTC | Same |
| FDA product code | intended for temporary relief of painassociated with sore and achingmuscles in the upper and lowerextremities (arm and/or leg) due tostrain from exercise or normalhousehold and work activities | Same |
| Electrode Used | NUH | Same |
| Manufacturer | K082065, Snap Type self adhesiveelectrode (5 X 5 cm) | K082065, Snap Type self adhesiveelectrode (2.925 x 5.6 cm) |
| Well-Life | Same |
{2}------------------------------------------------
8. Comparison of Unit Characteristics & Output Specification
| Parameter | Predicate Device | New Device | ||
|---|---|---|---|---|
| 510(K) Number | K091757 | K121353 | K121353 | |
| Device Name and Model | WL-2407 | WL-2301A | WL-2301B | |
| Manufacturer | Well-Life | Same | Same | |
| Power Source(s) | 1.5Vx3(AAA Size) | 3Vx1 | 3Vx1 | |
| - Method of Line current Isolation | Type BF | Same | Same | |
| - Patient Leakage Current | --- | --- | --- | |
| - Normal condition (uA) | Under 0.1 | Same | Same | |
| - single Fault condition (uA) | Under 0.5 | Same | Same | |
| Average DC current through electrodes whendevice is on but no pulses are being applied(UA) | Not applicable | Same | Same | |
| Number of Output Modes | 8 | 3 | 3 | |
| Number of OutputChannels: | Synchronous orAlternating? | Synchronous | Same | Same |
| Method of ChannelIsolation | Output Coil | Same | Same | |
| Regulated Current or Regulated Voltage? | Voltage | Same | Same | |
| Software/Firmware/Microprocessor control? | Yes | Same | Same | |
| Automatic Overload Trip? | No | Same | Same | |
| Automatic No-Load Trip? | Yes | Same | Same | |
| Automatic Shut Off? | Yes | Same | Same | |
| User Override control? | No | Same | Same | |
| Indicator Display: | On/Off Status? | Yes | Same | Same |
| Low Battery? | Yes | Same | Same | |
| Voltage/CurrentLevel? | Yes | Same | Same | |
| Timer Range (Minutes) | 5~60 | 20 | 20 | |
| Compliance with Voluntary Standards? | ANSI/AAMI NS4,IEC 60601-1 &IEC 60601-1-2. | Same | Same | |
| Compliance with 21 CFR 898? | Yes | Same | Same | |
| Weight (g) | 43 | 20 | 20 | |
| Dimensions (mm.)[W x H x D] | 90×52.5×19.4 | 59×30×12.15 | 60.2×31.9×14.4 | |
| Housing Materials and construction | ABS | Same | Same |
15121353
——————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————————
| Parameter | Predicate Device | New Device |
|---|---|---|
| Mode or Program Name | WL-2407 | WL-2301A WL-2301B |
| Waveform (e.g., pulsed monophasic, biphasic) | Biphasic | Same Same |
| Shape (e.g., rectangular, spike, rectified sinusoidal) | Rectangular (with deformation on the top of each pulse) | Same Same |
| Maximum Output Voltage (volts) (+/- 20 %) | 40V@500Ω62V@2kΩ104V@10kΩ | 30V@500Ω50V@2kΩ80V@10kΩ 30V@500Ω50V@2kΩ80V@10kΩ |
| Maximum Output Current (specify units) (+/- 20 %) | 80mA@500Ω31mA@2kΩ10.4mA@10kΩ | 60mA@500Ω25mA@2kΩ8.0mA@10kΩ 60mA@500Ω25mA@2kΩ8.0mA@10kΩ |
| Duration of primary (desplarizing) phase (usec) | 260 Max | 250 Max 250 Max |
| Pulse Duration (usec) | 500.26 ms/Max. | 33.58 ms/Max. 33.58 ms/Max. |
| Frequency (Hz) [or Rate (pps)] | 60 Max | 60 Max 60 Max |
| Formultiphasicwaveformsonly: | Symmetrical phases?Phase Duration (include units), (Stage range, if applicable), (both phases, if asymmetrical) | NoNot applicable SameSame SameSame |
{3}------------------------------------------------
9. Significant output characteristics comparison table:
| Comparison feature | Model | Remarks | |
|---|---|---|---|
| Net charge | WL-2407 | WL-2301/2301B | Note 1 |
| Max. phase charge | 20.8 uc | 15uc | |
| Max. current Density | 0.050 mA/cm² | 0.055 mA/cm² | |
| Max. Averagecurrent(RMSA) | 500 Ω | 38.43mA | 29.394 mA |
| 2K Ω | 14.892mA | 12.248 mA | |
| 10 Ω | 4.996mA | 3.9192 mA | |
| Max. Power Density | 0.0020Watts/ cm² | 0.0017Watts/ cm² | |
| Burst Mode | No | 6Sec/On & 1Sec/Off |
Note: Net Charge is Zero for the symmetrical stimulation biphasic wave
10. Conclusions
The Well-Life Mini Patch, model: WL-2301A & WL-2301B Transcutaneous Electrical Nerve Stimulators has the same intended use and the similar technological characteristics as the cleared device- Well-Life OTC TENS for Arm & Leg Pain Relief. (K091757) . Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device.
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 7 2012
Well Life Healthcare, Ltd % Ms. Jenny Hsieh 1F1, No.16, Lane 454, Jungjeng Rd. Yunghe City, Taipei County Taiwan, R.O.C.
Re: K121353
Trade/Device Name: Mini Patch, Model WL-2301A & WL-2301B Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: II Product Code: NUH Dated: August 17, 2012 Received: August 21, 2012
Dear Ms. Hsieh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications For Use
510(k) Number (if known): K121353
Device Name: Mini Patch, model : WL-2301A & WL-2301B
Indications For Use:
The Mini Patch, model: WL-2301A & WL-2301B Transcutaneous Electrical Nerve Stimulators are intended for temporary relief of pain associated with sore and aching muscles in the upper and lower extremities (arm and/or leg) due to strain from exercise or normal household and work activities
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Kunz
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K121353
§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.
(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).