K Number
K121340

Validate with FDA (Live)

Device Name
SOMNODENT G2
Manufacturer
Date Cleared
2012-05-30

(27 days)

Product Code
Regulation Number
872.5570
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SomnoDent G2 is intended for the treatment of nighttime snoring and mild to moderate obstructive sleep apnea in patients 18 years of age or older.

Device Description

The SomnoDent G2 device is an intraoral device intended for the treatment of nighttime snoring and mild to moderate obstructive sleep apnea in patients 18 years of age or older. The device functions as a mandibular repositioner, which acts to increase the patient's pharyngeal space during sleep. The increase in the patient's pharyngeal space improves their ability to exchange air during sleep. The device is customized for each patient and has an adjustable coupling mechanism enabling the amount of mandibular advancement to be set by the dentist or physician at the time of fitting the device. The SomnoDent G2 is a modification to the SomnoDent Flex (K073004). The SomnoDent G2 is identical to SomnoDent Flex, except for differences in the adjustment mechanism (which is an addition of a material of construction, which is biocompatible), and a change (which to an addition of a differences introduced by these modifications, when compared to the predicate product, do not introduce new safety issues.

AI/ML Overview

The SomnoDent G2 is an intraoral device for treating snoring and mild to moderate obstructive sleep apnea. The provided text, a 510(k) summary, focuses on demonstrating substantial equivalence to a predicate device (SomnoDent Flex, K073004) rather than presenting a study demonstrating its performance against specific acceptance criteria for efficacy in patients. Therefore, the device performance is primarily assessed through comparison to the predicate.

Here's an analysis of the information, acknowledging that the primary "study" is a comparative bench testing study to establish substantial equivalence:

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a 510(k) submission focused on substantial equivalence through bench testing rather than clinical efficacy, the "acceptance criteria" discussed are largely related to mechanical performance and safety, and the "reported device performance" is framed in terms of equivalence to the predicate.

Acceptance Criteria (Implied from testing)Reported Device Performance
Device functions as intended (mandibular repositioner, increases pharyngeal space)The SomnoDent G2 and predicate device function as a mandibular repositioner, which acts to increase the patient's pharyngeal space during sleep.
Withstands mechanical forces (simulated bruxism) and functions as designedBench testing included a mechanical force test to simulate bruxism, which was effective and determined worst-case use. The results indicated that the subject device performs as well or better than the predicate device.
Winglet bond withstands applicable pressuresMechanical testing was conducted on the winglet bond of the subject device. Testing results indicate that the bonds withstand applicable pressures, ensuring the device is safe and effective.
Mechanical performance of the advancement mechanism is equivalent to predicateComparative testing around the coupling mechanism was conducted. The testing concluded that the advancement of the subject device is substantially equivalent to the predicate device with regards to mechanical performance.
Modifications do not introduce new safety issuesThe document explicitly states that "Any differences introduced by these modifications, when compared to the predicate product, do not introduce new safety issues." This is the overarching acceptance criterion addressed by the mechanical testing.
Biocompatibility of new material (polycarbonate)The new material for the advancement mechanism is described as "biocompatible, medical grade polycarbonate." This implies that it meets acceptance criteria for biocompatibility, though no specific test results are detailed in this summary beyond the statement.
Adjustable range functionalityThe device has an adjustable range from 6 to 10mm (compared to the predicate's 6mm, though the text states "a change in adjustable range from 6 to 10mm" where the predicate device might not have been 10mm, leaving slight ambiguity, but the G2 confirms 6-10mm capability). This functionality is stated as present.

2. Sample Size Used for the Test Set and the Data Provenance

  • Sample Size: The document does not specify a "sample size" in terms of number of devices or units tested for the bench tests. It refers generally to "the subject device" and "the predicate device," suggesting testing was performed on representative units.
  • Data Provenance: Not applicable in the traditional sense of patient data. The data is from laboratory bench testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

  • This is not applicable to this type of submission. The ground truth for mechanical and safety tests is established by engineering standards and validated testing protocols, not expert clinical consensus on patient outcomes.

4. Adjudication Method for the Test Set

  • Not applicable. Bench testing results are typically objectively measured and evaluated against predefined engineering specifications, not adjudicated by a panel.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

  • Not applicable. This is not a study involving human readers or AI.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not applicable. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

  • The ground truth for the bench testing was engineering specifications and mechanical performance standards. For comparison purposes, the performance of the legally marketed predicate device (SomnoDent Flex) served as a benchmark for equivalence. Biocompatibility standards would also serve as ground truth for material selection.

8. The Sample Size for the Training Set

  • Not applicable. This is not an AI/machine learning device that would require a training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for SomnoMed. The logo consists of a stylized graphic above the company name. Below the company name, in smaller font, is the text "Dental * Sleep * Medical".

MAY 3 0 2012

510(k) Summary (21 CFR 807.92) Page 1 of 3

510(k) Number: K121340

Submission Owner:SomnoMed, Inc.7460 Warren ParkwaySuite 190Frisco, TX 75034Phone: 888-447-6673
Fax: 972-377-3404
Email: kjayne@somnomed.com
Official Correspondent:Kathryn A. Jayne
Date Prepared:February 15, 2012
Trade Name:SomnoDent® G2
Common Name:Intraoral device for snoring and mild to moderate obstructivesleep apnea (OSA)
Classification Name:Device, Anti-Snoring
Regulation Number:21 CFR 872.5570
Product Code:LRK
Class:II
Panel:Dental
Predicate Devices:K073004, SomnoDent Flex

Description of the device: ~

The SomnoDent G2 device is an intraoral device intended for the treatment of nighttime snoring and mild to moderate obstructive sleep apnea in patients 18 years of age or older. The device functions as a mandibular repositioner, which acts to increase the patient's pharyngeal space during sleep. The increase in the patient's pharyngeal space improves their ability to exchange air during sleep. The device is customized for each intient and has an adjustable coupling mechanism enabling the amount of mandibular advancement to be set by the dentist or physician at the time of fitting the device. The SomnoDent G2 is a modification to the SomnoDent Flex (K073004). The SomnoDent G2 is identical to SomnoDent Flex, except for differences in the adjustment mechanism (which is an addition of a material of construction, which is biocompatible), and a change (which to an dadition of a fifterences introduced by these modifications, when compared to the predicate product, do not introduce new safety issues.

Indications for Use:

The SomnoDent G2 is intended for the treatment of nighttime snoring and mild to moderate obstructive sleep apnea in patients 18 years of age or older.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for SomnoMed. The logo consists of a stylized graphic above the company name. Below the name, in smaller font, are the words "Dental - Sleep - Medical".

510(k) Summary (21 CFR 807.92)

Page 2 of 3

Substantial Equivalence:

Substantial Equivalence Table

Substantial Equivalence Table
SomnoDent G2SomnoDent FlexK073004
Intended Use
Intended as an intraoral deviceYesYes
Intended to reduce snoring or helpalleviate snoringYesYes
Treatment of mild to moderateobstructive sleep apneaYesYes
Intended for nighttime useYesYes
Indicated for single patient multiuseYesYes
Indicated for use at home or sleeplaboratoriesYesYes
Target population: adultsYesYes
Prescription deviceYesYes
Design
Customized fit for each patientYesYes
Separate upper and lower tray piecesYesYes
Works by mandibular advancementYesYes
Can be adjusted or refitYesYes
Lower jaw adjustment using suppliedcomponentsYesYes
Permits patient to breathe throughmouthYesYes
Upper and lower trays disengage foreasy removalYesYes
Cleaned and inspected daily by patientYesYes
Material
Trays constructed from a soft liningmaterial adhered to a hard surfaceacrylicYesYes
Advancement mechanism constructedof surgical grade stainless steelNoYes
Advancement mechanism constructedof biocompatible, medical gradepolycarbonateYesNo

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for SomnoMed. The logo consists of a black abstract shape above the company name. Below the name, it says "Dental - Sleep - Medical" in a smaller font.

510(k) Summary (21 CFR 807.92) Page 3 of 3

Substantial equivalence discussion:

The SomnoDent G2 is considered to be substantially equivalent to the SomnoDent Flex device. The SomnoDent G2 and predicate device function as a mandibular repositioner, which acts to increase the patient's pharyngeal space during sleep. The increase in the patient's pharyngeal space improves their ability to exchange air during sleep. The devices are customized for each patient and have an adjustable coupling mechanism enabling the amount of mandibular advancement to be set by the dentist or physician at the time of fitting the device. The SomnoDent G2 is identical to SomnoDent Flex, except for differences in the adjustment mechanism (which is an addition of a material of construction, which is biocompatible), and a change in adjustable range from 6 to 10mm. Any differences introduced by these modifications, when compared to the predicate product, do not introduce new safety issues.

Summary of Testing:

To demonstrate substantial equivalence, bench testing was conducted on the advancement mechanism to ensure the device performed as intended and is safe and Testing included a mechanical force test to simulate bruxism, which was effective. determined worst case use, to ensure the wings functioned as designed and withstood appropriate pressures. The results indicated that the subject device performs as well or better than the predicate device. Mechanical testing was conducted on the winglet bond of the subject device. Testing results indicate that the bonds withstand applicable pressures, ensuring the device is safe and effective. In addition, comparative testing around the coupling mechanism was conducted between the subject and predicate The testing concluded that the advancement of the subject device is devices. substantially equivalent to the predicate device with regards to mechanical performance. Testing demonstrates that the subject device, the SomnoDent G2, is as safe, as effective, and performs as well as or better than the predicate device, the SomnoDent Flex.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

MAY 3 0 2012

Somnomed Inc. C/O Mr. Jeff D. Rongero Responsible Third Party Official Underwriters Laboratories 12 Laboratory Drive Research Triangle, North Carolina 27709

Re: K121340

Trade/Device Name: SomnoDent G2 Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea

Regulatory Class: II Product Code: LRK Dated: May 21, 2012 Received: May 25, 2012

Dear Mr. Rongero:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Rongero

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing proportive requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

h for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for SomnoMed, a company that specializes in dental, sleep, and medical solutions. The logo features a stylized graphic element above the company name, "SomnoMed," which is written in a bold, sans-serif font. Below the company name, the words "Dental - Sleep - Medical" are written in a smaller, sans-serif font.

Indications for Use

510(k) Number (if known):

Device Name: SomnoDent G2

Indications for Use:

The SomnoDent G2 is intended for the treatment of nighttime snoring and mild to moderate obstructive sleep apnea in patients 18 years of age or older.

Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C) ·

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Swan Runner

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

CONFIDENTIAL

12334

§ 872.5570 Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea.

(a)
Identification. Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.(b)
Classification. Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”