K Number
K121293
Date Cleared
2012-07-26

(87 days)

Product Code
Regulation Number
892.1650
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Omega Medical Imaging, LLC CS-series-FP with 3030+ Option systems are intended for use in radiographic/fluoroscopic application including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.

Device Description

The Omega Medical Imaging, LLC. CS-series-FP systems currently incorporate a 19.8cm x 19.8cm solid-state flat-panel detector (FPD) as an option. This 510(k) submission adds a larger format (29.8cm x 29.8cm) flat-panel detector as an additional option. The CS-series-FP fluoroscopy single and dual plane x-ray imaging systems are configured with a floor mounted C-arm and a patient table. The dual plane systems incorporate a ceiling suspended C-arm into the system. The flat-panel image detector utilizes a cesium iodide scintillator coupled to an amorphous silicon TFT panel. The captured digital image is processed by the acquisition system which includes image processing, viewing functions, local storage, and DICOM compatibility.

AI/ML Overview

The provided text is a 510(k) summary for the Omega Medical Imaging, LLC CS-series-FP with 3030+ Option radiographic/fluoroscopy system. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a traditional clinical study with defined acceptance criteria and performance metrics against a gold standard in the way an AI/ML device might.

Therefore, the requested information elements related to acceptance criteria, ground truth, expert review, and statistical analyses of clinical performance are largely not applicable in this context. The "study" described is a non-clinical performance testing to show the new device meets the same performance standards as its predicate.

Here's an attempt to interpret and answer your questions based on the provided text, highlighting where the information is not available or not relevant for this type of submission:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Implied from Substantial Equivalence Claim)Reported Device Performance (Summary)
Equivalence to Predicate Device:The CS-series-FP with 3030+ Option (larger FPD) utilizes
- Low-contrast object visibility comparable to predicateidentical technology and image processing to the predicate.
- Spatial and temporal resolution comparable to predicatePerformance data was compared with the existing system
- Dynamic range comparable to predicateutilizing the smaller FPD. These tests used commercially
- Compliance with safety and regulatory standardsavailable Test Objects to assess low-contrast objects,
spatial/temporal resolution, and dynamic range.
The manufacturer's opinion is that the larger FPD system
is essentially equivalent to the cleared smaller FPD system.
Regulatory Compliance:The device complies with applicable requirements of
- 21 CFR 1020.30, 1020.31, 1020.3221 CFR 1020.30, 1020.31, 1020.32.
- IEC 60601-1, -1-2, -1-3, -2-7, -2-28, -2-32, -2-43
- UL 60601-1, CAN/USA C22.2 No.601.1-M90

Explanation of "Acceptance Criteria" in this context: For a 510(k) submission, the primary acceptance criterion is substantial equivalence to a predicate device. This means demonstrating that the new device is as safe and effective as a legally marketed device. The "acceptance criteria" here are therefore the performance characteristics of the predicate device, which the new device aims to match or exceed.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not applicable in the traditional sense of patient data. The "test set" consisted of "commercially available Test Objects" (phantoms). The number and specific types of these test objects are not detailed, but they are not human or animal subjects.
  • Data Provenance: Not applicable. The "data" comes from non-clinical performance testing using phantoms, not patient data from a specific country or collected retrospectively/prospectively.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable. Ground truth as typically understood for clinical studies involving expert interpretation of images or diagnoses is not relevant here. The "ground truth" for non-clinical performance testing is inherent in the design specifications of the test objects themselves (e.g., a known number of low-contrast objects, a specific spatial resolution pattern). Performance is measured objectively by the system.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable. There was no human adjudication of results for the non-clinical performance tests. Measurements are objective and quantitative (e.g., measuring line pairs per millimeter, contrast-to-noise ratio).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. An MRMC study was not performed. This submission is for an imaging system (hardware), not an AI/ML diagnostic tool. There is no AI component or human-in-the-loop performance measurement described.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No. This is an imaging system, not an algorithm, so "standalone (algorithm only)" is not applicable. The performance tests evaluate the physical and technical capabilities of the X-ray system and its detector.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • The "ground truth" for the non-clinical testing was based on the known characteristics and specifications of the commercially available test objects (phantoms). These phantoms are designed to have specific, measurable properties (e.g., specific sizes of low-contrast objects, known spatial frequencies).

8. The sample size for the training set

  • Not applicable. As this is not an AI/ML device, there is no "training set" in the machine learning sense. The device is a conventional X-ray imaging system.

9. How the ground truth for the training set was established

  • Not applicable, as there is no training set for an AI/ML algorithm.

{0}------------------------------------------------

K121293

Omega Medical Imaging, LLC

JUL 26 2012

510(k) SUMMARY

Company Name: Omega Medical Imaging, LLC

675 Hickman Circle Address: Sanford, FL 32771

407-323-9400 Telephone No:

Registration No .: 1052701

Contact person: Brian J. Fleming

Date Prepared: 20 April 2012

CS-series-FP with 3030+ Option radiographic/fluoroscopy Device (trade) name: svstem

Common/usual name: Image Intensified Fluoroscopic X-ray System

Classification Name: Solid State X-ray Imager, Class II, 90 MQB Angiographic X-ray System, Class II, 90 IZI

Classification Panel: Radiology

CFR Section: 892.1650 and 892.1600

Device Class: Class II

90 MQB, 90 IZI Device Code:

Predicate device(s):

Omega Medical Imaging CS-series-FP (K100102) .

Device description:

The Omega Medical Imaging, LLC. CS-series-FP systems currently incorporate a 19.8cm x 19.8cm solid-state flat-panel detector (FPD) as an option. This 510(k) submission adds a larger format (29.8cm x 29.8cm) flat-panel detector as an additional option. The CS-series-FP fluoroscopy single and dual plane x-ray imaging systems are configured with a floor mounted C-arm and a patient table. The dual plane systems incorporate a ceiling suspended C-arm into the system. The flat-panel image detector utilizes a cesium iodide scintillator coupled to an amorphous silicon TFT panel. The captured digital image is processed by the acquisition system which includes image processing, viewing functions, local storage, and DICOM compatibility.

Materials: The solid-state x-ray imager (FPD) construction and materials comply with UL2601-1, IEC60601-1, CSA 22.2 No. 601.1-M90, and is CE Marked. The imager is classified by Underwriters Laboratories.

{1}------------------------------------------------

Omega Medical Imaging, LLC

Intended use:

  • The Omega Medical Imaging, LLC. CS-series-FP with 3030+ Option systems are . intended for use in radiographic/fluoroscopic applications including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.

Comparison with Predicate Devices:

  • The Omega Medical Imaging large format option to the CS-series-FP system utilizes . identical technology of the above mentioned predicate device. The input scintillator is Cesium lodide coupled to a Thin-Film-Transistor (TFT) panel. The image processing is achieved with conventional computer based image processing systems. It is the opinion of Omega Medical Imaging that the CS-series-FP with 3030+ Option utilizing the larger format FPD detector is essentially equivalent to the existing Omega Medical Imaging cleared CS-series-FP system utilizing the smaller format FPD.

Substantial Equivalence:

  • SE was determined on non-clinical performance testing. Included in this report is detailed . data comparing performance with the existing Omega Medical Imaging CS-series-FP system utilizing the 19.8cm x 19.8cm format FPD image acquisition system. The tests that were performed utilized commercially available Test Objects that include low-contrast objects with varying absorbers simulating different patient sizes, spatial and temporal resolution test objects, and dynamic range test objects. Also included is the "Non-clinical Considerations" of the flat-panel detector.
    It is the opinion of Omega Medical Imaging that the CS-series-FP with 3030+ Option with the larger format FPD (29.8cm x 29.8cm) is essentially equivalent to the cleared Omega Medical Imaging CS-series-FP (K100102) smaller format detector systems.

Safety information:

  • The Omega CS-series-FP with 3030+ Option systems comply with the applicable ● requirements of 21 CFR 1020.30, 21 CFR 1020.31, and 21 CFR 1020.32.
  • The Omega CS-series-FP with 3030+ Option systems comply with the international safety . standards IEC 60601-1, IEC 60101-1-2, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, and IEC 60601-2-43.
  • The Omega CS-series-FP with 3030+ Option systems comply with UL 60601-1 and . CAN/USA C22.2 No.601.1-M90

Conclusion:

The Omeqa CS-series-FP with 3030+ Option systems with the larger format (29.9cm x 29.9cm) FPD do not introduce any new indications for use, nor does the use of the device result in any new potential hazard. Omega considers the CS-series-FP with 3030+ Option systems with this larger format flat-panel detector to be substantially equivalent with the predicate device.

5-2

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread. The eagle is facing to the right and has three lines emanating from its body.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

JUL 26 2012

Mr. Brian J. Fleming President & CEO Omega Medical Imaging, LLC 675 Hickman Circle SANFORD FL 32771

Re: - K121293

Trade/Device Name: CS-series-FP with 3030+ Option Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA and OWB Dated: April 20, 2012 Received: May 1, 2012

Dear Mr. Fleming:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirences as begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nonication. The results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

CS-series-FP with 3030+ Option Device Name:

Indications for Use:

The Omega Medical Imaging, LLC CS-series-FP with 3030+ Option systems are intended for use in radiographic/fluoroscopic application including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

image.png

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k)

Page 1 of 1.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.