K Number
K121243
Manufacturer
Date Cleared
2013-05-03

(373 days)

Product Code
Regulation Number
892.1560
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MD-1000A/P is used for ophthalmology to measure axial length (AL), anterior chamber depth (ACD), lens thickness (LT), and corneal thickness (CT).

The MD-1000A is used for ophthalmology to measure axial length (AL), anterior chamber depth (ACD) and lens thickness (LT).

The MD-1000P is used for ophthalmology to measure corneal thickness (CT).

The device should be operated by trained doctors. It should be used cautiously to patients without independent behavior abilities. Cornea trauma or inflammation patients are prohibited to use the device.

Device Description

The MD-1000 Series of Ultrasonic Biometer/Pachymeter for Ophthalmology consists of three models of products: the MD-1000A/P Ultrasonic Biometer for Ophthalmology, the MD-1000A Ultrasonic Biometer for Ophthalmology and the MD -1000P Ultrasonic Pachymeter.

The MD-1000A/P Ultrasonic Biometer for Ophthalmology is an ultrasonic measuring instrument based on pulse reflection. The MD-1000A/P contains two function units: A Mode Eye Axis Biometric Parameter Measuring Unit (A Biometer) and Corneal Thickness Measuring Unit (Pachymeter).

The A Biometer consists of a 10MHz A-Probe (probe model: Prb1000A/10-C) and an axis biometric parameter measuring unit. The axis is usually divided into three segments: anterior chamber, lens and vitreous body. Since the tissue within the eye varies in different areas, the acoustic velocity through these areas is also different. The summation of these three segments (ACD + LENS + VITR) provides the axial length (AL). Based on the interface reflections of the three different tissues, the ultrasonic A-biometry measures the transmitting time of ultrasound through each tissue and calculates the length of each segment to finally get the axial length.

The Pachymeter consists of a 20MHz P-Probe (probe model: Prb1000P) and the measuring unit. It is on the basis of the measurement of the time interval between the anterior and posterior interface reflection waves of cornea to get the corneal thickness (CT).

The MD-1000 Series has a built-in Thermal Printer, used to print out patient information, A-Scan measuring waveform, IOL calculating parameter and result as corneal thickness measuring result and corneal thickness distribution map.

The built-in memory of the MD-1000 Series can store up to180 patients' records. After examination, it can be connected with a computer to upload the stored measuring data and information, thus realizing mass storage.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information derived from the provided text for the MD-1000 Series Ultrasonic Biometer/Pachymeter:

1. Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Predicate Device)Reported Device Performance (MD-1000 Series)
AL Measuring Range15~39mm15~40mm
AL Measuring Accuracy≤±0.06mm≤±0.05mm
Corneal Thickness ResolutionN/A (implied by predicate, not explicitly stated as criteria)1um
Corneal Thickness Measuring RangeN/A (implied by predicate, not explicitly stated as criteria)0.23mm~1.2mm
Corneal Thickness AccuracyN/A (implied by predicate, not explicitly stated as criteria)Error≤±5µm

Note: The acceptance criteria for the MD-1000 series are compared directly to the specified performance of its predicate device, the ODM-1000 series, as the document states that the MD-1000 Series has "similar performance characteristics...as its predicate device" and highlights improvements (like AL accuracy and range). For corneal thickness, the predicate's specific criteria aren't detailed, so the MD-1000's reported performance is listed.


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size:
    • Corneal thickness measurement: 60 eyes
    • Axial length measurement: 47 eyes
    • Total subjects: 107
  • Data Provenance: The study was conducted by "a hospital" in China (implied by the submitter's location and the mention of 93/42/EEC MEDDEV. 2.7.1 Rev.3, which is a European medical device directive, but the specific country of the hospital is not explicitly stated). The study appears to be prospective as it involved selecting subjects to "implement axial length measurement or corneal thickness measurement" with the device and a comparator.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly state the number of experts or their qualifications used to establish ground truth. The study compares the MD-1000A/P to a "similar product (NIDEX-US-500)" which serves as a reference or "ground truth" for comparison. The assumption is that measurements from the NIDEX-US-500 are considered reliable.


4. Adjudication Method for the Test Set

The document does not mention an adjudication method for the test set. It states that both devices were used to measure the same eye, and the Bland-Altman statistical method was used to assess agreement between the two devices. This implies a direct comparison rather than an adjudication of discrepancies.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable to this device. The MD-1000 Series is an ultrasonic biometer/pachymeter, not an AI-assisted diagnostic imaging interpretation device. The study compared the device's measurements directly to another similar device, not human readers with or without AI assistance.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This section is not applicable in the AI sense. The device itself performs the measurements (axial length and corneal thickness) as a standalone instrument. The "algorithm" here refers to the device's internal measurement and calculation processes, which are inherently "standalone" in their function. There is no human interpretation of an algorithm's output to assess standalone performance in the context of AI. The clinical validation was a comparison of the device's measurements against another device, which is a form of standalone performance evaluation for a measurement device.


7. The Type of Ground Truth Used

The ground truth for the clinical validation was established by comparison to a legally marketed similar product (NIDEX-US-500). The measurements from the NIDEX-US-500 served as the reference for evaluating the MD-1000A/P's performance.


8. The Sample Size for the Training Set

The document does not provide information regarding a separate training set. As this device is a physical medical instrument (ultrasonic biometer/pachymeter) and not a machine learning algorithm that requires explicit "training" in the typical sense, a training set as understood in AI/ML is not applicable or detailed in this submission. Its improvement in accuracy is attributed to hardware changes like increased sampling frequency (24MHz vs 12MHz).


9. How the Ground Truth for the Training Set Was Established

As no training set is mentioned in the context of an AI/ML system, this question is not applicable. The device's operational principles are based on established ultrasound physics and algorithms, rather than being "trained" on a dataset in the way a modern AI would be.

{0}------------------------------------------------

K121243

510(k) SUMMARY

MAY 0 3 2013

  1. Submitter Information

Name: MEDA Co., Ltd.,

Address: F2C, F3D, F4C, F5, Building C2, Xinmao Science Skill Park, Huayuan Industry Development Area, Tianjin 300384, P. R. China +86-22-83713808 Phone:

+86-22-83713880 Fax:

Contact person: KAI CHEN

.

Medtech International, Inc. and United States Designated Agent of MEDA Co., Ltd. Address: 13505 Broadfield Drive, Potomac, MD 20854 Tel: 240 888 4001 Fax: 301 251 2881 Email: Drkaichen@gmail.com

Date Prepared: March 30, 2012

Device Information 2.

Trade Name:MD-1000A/P Ultrasonic Biometer for Ophthalmology
MD-1000A Ultrasonic Biometer for Ophthalmology
MD-1000P Ultrasonic Pachymeter
  • Common Name: Ultrasound A/P Scan System for Ophthalmology Ultrasound A Scan for Ophthalmology Ultrasound Pachymeter
  • Classification Name: Ultrasonic Pulsed Echo Imaging System Diagnostic Ultrasound Transducer

Regulation Number: 892.1560; 892.1570

Product Code: IYO; ITX

    1. Predicate Devices
      We selected the legally marketed similar produced by our company as the predicate devices. They have the same intended use and working principle, and are equivalent in structure (including accessories) and technical characteristics. The predicate devices are listed below:

Manufacturer: MEDA Co., Ltd

Device: ODM-1000A/P Biometer ODM-1000A Ultrasonic A-Biometer ODM-1000P Ultrasonic Pachymeter

510(k) Number: K063472

{1}------------------------------------------------

4. Device Description

The MD-1000 Series of Ultrasonic Biometer/Pachymeter for Ophthalmology consists of three models of products: the MD-1000A/P Ultrasonic Biometer for Ophthalmology, the MD-1000A Ultrasonic Biometer for Ophthalmology and the MD -1000P Ultrasonic Pachymeter.

The MD-1000A/P Ultrasonic Biometer for Ophthalmology is an ultrasonic measuring instrument based on pulse reflection. The MD-1000A/P contains two function units: A Mode Eye Axis Biometric Parameter Measuring Unit (A Biometer) and Corneal Thickness Measuring Unit (Pachymeter).

The A Biometer:

The A Biometer consists of a 10MHz A-Probe (probe model: Prb1000A/10-C) and an axis biometric parameter measuring unit. The axis is usually divided into three segments: anterior chamber, lens and vitreous body. Since the tissue within the eye varies in different areas, the acoustic velocity through these areas is also different. The summation of these three segments (ACD + LENS + VITR) provides the axial length (AL). Based on the interface reflections of the three different tissues, the ultrasonic A-biometry measures the transmitting time of ultrasound through each tissue and calculates the length of each segment to finally get the axial length.

The Pachymeter:

The Pachymeter consists of a 20MHz P-Probe (probe model: Prb1000P) and the measuring unit. It is on the basis of the measurement of the time interval between the anterior and posterior interface reflection waves of cornea to get the corneal thickness (CT).

The MD-1000 Series has a built-in Thermal Printer, used to print out patient information, A-Scan measuring waveform, IOL calculating parameter and result as corneal thickness measuring result and corneal thickness distribution map.

The built-in memory of the MD-1000 Series can store up to180 patients' records. After examination, it can be connected with a computer to upload the stored measuring data and information, thus realizing mass storage.

5. Intended Use

The MD-1000A/P is used for ophthalmology to measure axial length (AL), anterior chamber depth (ACD), lens thickness (LT), and corneal thickness (CT),

The MD-1000A is used for ophthalmology to measure axial length (AL), anterior chamber depth (ACD) and lens thickness (LT).

The MD-1000P is used for ophthalmology to measure corneal thickness (CT).

The device should be operated by trained doctors. It should be used cautiously on babies or patients without independent behavior abilities. It is prohibited to use the device on patients with eye trauma, inflammation or infection.

Technological characteristics 6.

a) Safety

{2}------------------------------------------------

The electrical, mechanical, environmental safety testing was performed according to standard IEC60601-1: 2005, Deviations USA/Canada to IEC 60601-1: 2005 and IEC 60601-2-37: 2007.

The EMC testing was conducted in accordance with standard IEC 60601-1-2: 2007. The safety standards performed by the MD-1000 Series are identical with those of the predicate product.

b) Characteristics

A Biometer:

Ultrasonic frequency: 10MHz; Display resolution: 0.01mm; Adjustable gain scope: 050dB; Measuring scope (AL): 15 mm40mm; Measuring accuracy: Error≤±0.05mm; Measuring Parameter: ACD, LENS, VITR and AL; Measuring mode: automatic mode or manual mode can be selectable. Automatic mode for Normal, Cataract and Aphakic mode; Measuring method: contact and immersion, can be selectable; IOL calculation: SRK/T, SRK-11, BINK-11, HOLLADAY, HOFFER-Q, HAIGIS.

Pachymeter:

Ultrasonic frequency: 20MHz; Display resolution: 1um; Measuring scope: 0.23mm~1.2mm; Measuring accuracy: Error≤±5µm.

The MD-1000 Series has similar performance characteristics, controls and indicators, and specifications as its predicate device. It incorporates the same operational principles and technology as the ODM-1000 Series.

The main differences between the MD-1000 Series and its predicate device are: they use different monitors, housing, AC-DC adapters and have different specifications of A-Probe. The details are listed below:

ModelSubject DevicePredicate Device
AccessoryMD-1000 series(MD-1000A/P, MD-1000A, MD-1000P)ODM-1000 series(ODM -1000A/P, ODM -1000A, ODM -1000P)
GeneralSafety andEffectivenessMeasuring Range (AL): 15~40mmMeasuring Accuracy (AL): ≤±0.05mmMeasuring Range (AL): 15~39mmMeasuring Accuracy (AL): ≤±0.06mm
A-probeModel: Prb1000A/10-CTransducer Diameter: 4.8 mmFocal Length: $30\pm3$ mmModel: Prb1000A/10Transducer Diameter: 4 mmFocal Length: $23\pm3$ mm
Monitor6.4" True Color TFT LCD (640×480 pixels,18Bit)5.1" LCD (320×240 pixels)
PowerAdapterModel: HES49-12040 12V/4A(UL E164433)Model: MW128RA0703F01 7.5V/3A(UL E145177)
Foot SwitchMD-1000.FS-2 (for MD-1000A/P andMD-1000P)MD-1000.FS-1 (for MD-1000A)Foot Switch for ODM-1000A/P andODM-1000PFoot Switch for ODM-1000A

{3}------------------------------------------------

It can be found from the above comparison that the AL measurement range and accuracy of the two devices are a little different.

For the AL measuring accuracy, the sampling frequency of the MD-1000A/P to the ultrasonic echo is 24MHz, but the sampling frequency of the ODM-1000A/P is 12MHz. The improvement of the sampling frequency can improve the precision of time measurement, thus improving the AL measuring accuracy.

For the AL measurement range, the original design specifications of the ODM-1000A/P is 15 ~ 40mm, but in actual test, due to the limitation of the LCD display resolution (320 × 240 pixels), the display range reaches 39mm only; the MD-1000A/P utilizes LCD monitor with the resolution of 640 × 480 pixels, thus satisfying the requirements of 15 ~ 40mm AL measurement range.

The ODM-1000 Series uses a 5.1" LCD monitor and membrane keyboard. The MD-1000 Series uses a 6.4" True Color TFT LCD monitor and touch screen. The Safety Test Report showed that it met the requirements of IEC 60601-1 and did not affect the safety of the MD-1000 Series.

The Prb1000A/10-C configured with the MD-1000A/P uses the transducer of φ4.8mm and Focal Length of 30mm. Although there is difference from the characteristics of the A-Probe configured with the ODM-1000A/P, based on the working principle of ultrasonic distance measurement, it is not required to change algorithm and processing hardware. Its algorithm is the same with that of the ODM-1000A/P; and for the processing hardware, except that the sampling frequency increases to 24MHz for the purpose of improving measuring accuracy, the structure of the hardware is not changed. The PEMS Validation Report, Acoustic Output Test Report and Biological Evaluation Test Report proved that, the changes of the A-Probe size and focal length did not affect the safety and AL measuring accuracy.

The MD-1000 Series and ODM-1000 Series use different power adapters and foot switches, but both of them meet the requirements of IEC 60601-1.

While there are some differences between the MD-1000 Series and its predicate device, they do not affect safety or effectiveness.

7. Brief Discussion of Non-clinical Tests

The safety tests based on FDA-recognized standards were conducted by TÜV SÜD Laboratories, and all results comply with the requirements of relevant standards.

The software and essential performance have passed verification and validation, and the results comply with the requirements.

8. Brief Discussion of Clinical Evaluation and Validation

The clinical data of the MD-1000A/P Ultrasonic Biometer for Ophthalmology have been collected and evaluated according to 93/42/EEC MEDDEV. 2.7.1 Rev.3.

The medical literature related to measuring accuracy and scope mainly comes from the Journal of Research in Medical Sciences, J Cataract Refract Surg, Int Ophthalmol and etc.

As part of the product design validation, we have entrusted a hospital to make clinical effectiveness validation. Below is a summary of the details about the clinical validation study:

{4}------------------------------------------------

a) 107 subjects were selected to implement axial length measurement or corneal thickness measurement, among which 60 eyes were for corneal thickness measurement and 47 eyes were for axial length measurement. The age and sex of subjects were not limited and all subjects participated in the clinical tests voluntarily.

b) The MD-1000A/P and similar product (NIDEX-US-500) were used repeatedly to measure corneal thickness of the same eye. One measurement was taken per eye with the two products respectively and a total of 60 subjects (eyes) were measured.

The axial length measurements were conducted in the same way, a total of 47 subjects (eyes) were measured.

c) The Bland-Altman statistical method was used to process the measuring data and assess the agreement of the measurements by the two devices. Take 95% as the limit of agreement.

The statistical processing to the clinical test data showed that there is acceptable agreement for the test results of corneal thickness and axial length between the MD-1000A/P and its similar product; therefore, the MD-1000A/P can satisfy the requirements of the clinical applications.

The clinical test results showed that the MD-1000A/P met the requirements of the intended clinical applications.

No adverse effect was found during the collection and evaluation of clinical data and clinical validation.

The clinical test prompted that the device should be used cautiously on patients without independent behavior abilities or who are highly sensitive to contacting measurements, whose ineffective cooperation may result in inaccurate measurements.

9-Conclusions

The results of non-clinical tests as well as clinical evaluation demonstrate that the MD-1000A/P Ultrasonic Biometer for Ophthalmology, the MD-1000A Ultrasonic Biometer for Ophthalmology and the MD-1000P Ultrasonic Pachymeter are equivalent in safety, effectiveness and performance to the legally marketed predicate devices.

    1. MEDA CO., LTD will update and include in this summary any other information deemed reasonably necessary by the FDA.
      5

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.

Public Health Service

May 3, 2013

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-002

MEDA Co. Ltd. % Mr. Kai Chen U.S. Designated Agent Medtech International, Inc. 13505 Broadfield Drive Potomac, MD 20854

Re: K121243

Trade/Device Name: MD-1000 Series Ultrasonic Biometer for Ophthalmology Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: IYO, ITX Dated: April 19, 2013 Received: April 29, 2013

Dear Mr. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{6}------------------------------------------------

Page 2 - Mr. Kai Chen

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia YAlexander -S

for Malvina B. Eydelman, M.D. Director Division of Opthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

510(k) Submission of MEDA. CO., LTD

MD-1000 Ultrasonic Biometer/Pachymeter For Ophthalmology

Indications for Use Statement

510(k) Number (if known): K121243

Device Name: MD-1000A/P Ultrasonic Biometer for Ophthalmology

MD-1000A Ultrasonic Biometer for Ophthalmology

MD-1000P Ultrasonic Pachymeter

Indications for Use:

The MD-1000A/P is used for ophthalmology to measure axial length (AL), anterior chamber depth (ACD), lens thickness (LT), and corneal thickness (CT).

The MD-1000A is used for ophthalmology to measure axial length (AL), anterior chamber depth (ACD) and lens thickness (LT).

The MD-1000P is used for ophthalmology to measure corneal thickness (CT).

The device should be operated by trained doctors. It should be used cautiously to patients without independent behavior abilities. Cornea trauma or inflammation patients are prohibited to use the device.

Prescription Use _ V

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

Over-The-Counter Use_

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR ...

Concurrence of CDRH, Office of Device Evaluation (ODE)

1

Ka N. To
2013.05.02 17:34:16 -04'00'

(Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices 510(k) Number:______ K121243

{8}------------------------------------------------

Al-4: Indications for Use Statement

MD-1000 Ultrasonic Biometer/Pachymeter For Ophthalmology

Diagnostic Ultrasound Indications for Use Format

MD-1000A/P&MD-1000A System:

Transducer: Prb1000A/10-C A-Probe

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track1 Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other*(A-mode)
OphthalmicOphthalmicP
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Bladder)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix

  • Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Tissue Motion Doppler, and Color Velocity Imaging

Image /page/8/Picture/10 description: The image shows the text "Ka N. To" on the first line and "2013.05.02 17:34:33 -04'00'" on the second line. The date is May 2nd, 2013. The time is 5:34 PM and 33 seconds. The time zone is -04'00'.

(Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices 510(k) Number:_ K121243

2

{9}------------------------------------------------

510(k) Submission of MEDA. CO., LTD

Al-4: Indications for Use Statement

MD-1000 Ultrasonic Biometer/Pachymeter For Ophthalmology

MD-1000A/P&MD-1000P System:

Transducer: Prb1000P P-Probe

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation
General(Track1 Only)Specific(Tracks 1 & 3)BMPWDCWDColorDopplerCombined(Specify)Other*(A-mode)
OphthalmicOphthalmicP
Fetal Imaging& OtherFetal
Abdominal
Intra-operative (Specify)
Intra-operative (Neuro)
Laparoscopic
Pediatric
Small Organ (Specify)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Card.)
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Intravascular
Other (Bladder)
CardiacCardiac Adult
Cardiac Pediatric
Intravascular (Cardiac)
Trans-esoph. (Cardiac)
Intra-cardiac
Other (Specify)
PeripheralVesselPeripheral vessel
Other (Specify)

N = new indication; P = previously cleared by FDA; E = added under this appendix

  • Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging

Ka N. To
2013.05.02 FDA 17:34:51-04'00'"

(Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices 510(k) Number:_ K121243

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.