K Number
K121229
Date Cleared
2012-07-23

(90 days)

Product Code
Regulation Number
870.4350
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

INSPIRE 8M: Hollow Fiber Oxygenator: The INSPIRE 8M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 8M is intended to be used for 6 hours or less. INSPIRE HVR DUAL: Hardshell Venous/Cardiotomy Reservoir: The INSPIRE HVR DUAL is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and suction blood. INSPIRE HVR DUAL can be used post-operatively for chest drainage. INSPIRE HVR DUAL is intended to be used for 6 hours or less.

Device Description

The INSPIRE 8 DUAL is consisting of an oxygenator, integrated with a heat exchanger (INSPIRE 8M), and a hardshell venous/cardiotomy reservoir (INSPIRE HVR DUAL). The reservoir is connected to the gas exchange module by means of a molded fitting joint. The INSPIRE 8 DUAL is a high efficiency microporous hollow fiber membrane oxygenator, integrated with a heat exchanger, connected to a hardshell venous/cardiotomy reservoir. The device can be operated at flow rates up to 8 liters per minute (I/min). The hollow fiber membrane oxygenator provides oxygenation and carbon dioxide removal from venous or suctioned blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The integrated hardshell reservoir collects, defoams, filters venous and suction blood, and can be used post-operatively for chest drainage. The INSPIRE 8 DUAL is a modified version of the currently marketed integrated oxygenator/hardshell venous cardiotomy reservoir system D903 AVANT 2 Ph.I.S.I.O (hereinafter referred to as D903 AVANT).

AI/ML Overview

This document describes the pre-market notification (510(k)) for the INSPIRE 8 DUAL device, which is an oxygenator with integrated heat exchanger and hardshell venous/cardiotomy reservoir. The submission focuses on demonstrating substantial equivalence to a predicate device (D903 AVANT 2 Ph.I.S.I.O) through non-clinical and in vitro testing.

There is no information in the provided text about an AI/ML-driven device or study involving human readers, therefore, sections 2, 3, 4, 5, 8, and 9 cannot be addressed.

Here's the information that can be extracted from the provided text regarding the device acceptance criteria and the study proving it meets these criteria:


1. A table of acceptance criteria and the reported device performance

The document does not provide explicit numerical "acceptance criteria" for each test. Instead, it states that the device "passed each test" and that the results "demonstrate that the INSPIRE 8 DUAL performs in a manner substantially equivalent to the D903 AVANT predicate device with respect to the relevant functional parameters." The acceptance criteria implicitly refer to meeting established performance specifications or demonstrating equivalence to the predicate device.

TESTTEST CLASSIFICATIONTEST TITLEReported Device Performance/Conclusion
1Physical/MechanicalOxygenating module structural integrityPassed
2Physical/MechanicalReservoir structural integrityPassed
3Physical/MechanicalOxygenating module blood, water, gas pathway integrityPassed
4Physical/MechanicalReservoir blood pathway integrityPassed
5Functional/PerformanceOxygenating module blood volume capacityPassed
6Functional/PerformanceReservoir blood rest volumePassed
7Functional/PerformanceOxygenating module gas transfer performance/blood side pressure dropPassed
8Functional/PerformanceOxygenating module heat exchange performance/water side pressure dropPassed
9Functional/PerformanceReservoir air handlingPassed
10Functional/PerformanceReservoir break-through time and volumePassed
11Functional/PerformanceReservoir defoaming efficiencyPassed
12Functional/PerformanceReservoir dynamic priming volume / Hold-upPassed
13Functional/PerformanceReservoir filtration efficiency - venous sectionPassed
14Functional/PerformanceReservoir filtration efficiency - cardiotomy sectionPassed
15Functional/PerformanceReservoir flow rate capacityPassed
16Functional/PerformanceReservoir pressure dropPassed
17Functional/PerformanceIntegrated device hemolysisPassed
18Functional/PerformanceIntegrated device blood compatibilityPassed
19Functional/PerformanceOxygenating module leaching of coatingPassed
20Functional/PerformanceReservoir leaching of coatingPassed
21Functional/PerformanceIntegrated device flaking of coatingPassed
22Functional/PerformanceOxygenating module uniformity of coatingPassed
23Functional/PerformanceReservoir uniformity of coatingPassed
BiocompatibilityN/AN/A (General Biocompatibility Testing)Applicable tests carried out in accordance with ISO 10993-1 and FDA memorandum.
Sterilization/PyrogenN/AN/A (Sterility and non-pyrogenicity)Effectiveness of production techniques demonstrated to assure device is sterile and non-pyrogenic.

2. Sample size used for the test set and the data provenance

The document does not specify the sample size for the in vitro tests. The data provenance is described as "In vitro testing was carried out" and "Applicable tests were carried out," but the location (country of origin) of these tests is not explicitly stated, although the submitter is Sorin Group Italia S.r.l. The studies are non-clinical/in vitro tests, not human data, so "retrospective or prospective" is not applicable in the human data sense.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is a submission for a medical device (oxygenator/reservoir) based on non-clinical and in vitro performance testing, not an AI/ML device relying on expert-established ground truth from medical images/data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As above, this does not involve human readers or adjudication of medical data.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This is not an AI/ML device and no MRMC study was performed or mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML algorithm. The performance described is the standalone performance of the physical device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The "ground truth" for the device's performance is established by validated in vitro test methods and established industry standards (ISO 7199, ISO 15675, ISO 15674, FDA Guidance documents) designed to measure the physical and functional parameters of cardiopulmonary bypass devices. The comparison is made against the performance of a predicate device (D903 AVANT 2 Ph.I.S.I.O).

8. The sample size for the training set

Not applicable. This is not an AI/ML device requiring a training set.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device requiring a training set.

{0}------------------------------------------------

K12/229

INSPIRE 8 DUAL Sorin Group Italia S.r.I. Traditional 510(k) April 24, 2012

510(k) SUMMARY
JUL 2 3 2012
SUBMITTER:Sorin Group Italia S.r.l.86, Via Statale 12 Nord41037 Mirandola (MO) Italy
CONTACT PERSON:Luigi VecchiPhone: 011 39 0535 29811Fax: 011 39 0535 25229
DATE PREPARED:April 24, 2012
DEVICE TRADE NAME:INSPIRE 8 DUAL hollow fiber oxygenator withintegrated hardshell venous/cardiotomyreservoir
COMMON NAMES:Hollow Fiber Oxygenator with integratedhardshell venous/cardiotomy reservoirHollow Fiber OxygenatorHardshell Venous/Cardiotomy Reservoir
CLASSIFICATION NAMES:Cardiopulmonary Bypass Oxygenator/Cardiopulmonary Bypass Heat Exchanger/Cardiopulmonary Bypass Blood Reservoir/Cardiopulmonary Bypass Defoamer/
PREDICATE DEVICE:D903 AVANT 2 Ph.I.S.I.O. Adult Hollow FiberOxygenator with Ph.I.S.I.O. coating (K033323)

DEVICE DESCRIPTION:

The INSPIRE 8 DUAL is consisting of an oxygenator, integrated with a heat exchanger (INSPIRE 8M), and a hardshell venous/cardiotomy reservoir (INSPIRE HVR DUAL). The reservoir is connected to the gas exchange module by means of a molded fitting joint.

The INSPIRE 8 DUAL is a high efficiency microporous hollow fiber membrane oxygenator, integrated with a heat exchanger, connected to a hardshell venous/cardiotomy reservoir.

The device can be operated at flow rates up to 8 liters per minute (I/min).

The hollow fiber membrane oxygenator provides oxygenation and carbon dioxide removal from venous or suctioned blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia or aids in the maintenance of

{1}------------------------------------------------

INSPIRE 8 DUALTraditional 510(k)
Sorin Group Italia S.r.I.April 24, 2012

normothermia during surgery. The integrated hardshell reservoir collects, defoams, filters venous and suction blood, and can be used post-operatively for chest drainage. The INSPIRE 8 DUAL is a modified version of the currently marketed integrated oxygenator/hardshell venous cardiotomy reservoir system D903 AVANT 2 Ph.I.S.I.O (hereinafter referred to as D903 AVANT).

INDICATION FOR USE: ·

The intended use for the two elements that constitute the integrated device are:

INSPIRE 8M: Hollow Fiber Oxygenator

The INSPIRE 8M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 8M is intended to be used for 6 hours or less.

INSPIRE HVR DUAL: Hardshell Venous/Cardiotomy Reservoir

The INSPIRE HVR DUAL is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and suction blood. INSPIRE HVR DUAL can be used post-operatively for chest drainage. INSPIRE HVR DUAL is intended to be used for 6 hours or less.

TECHNOLOGICAL CHARACTERISTICS:

The INSPIRE 8 DUAL has the same fundamental technological characteristics, principles of operation and control mechanisms as the predicate device.

Sorin believes that the INSPIRE 8 DUAL is substantially equivalent to the D903 AVANT on the basis of operating principles and basic function.

The INSPIRE 8 DUAL and the D903 AVANT share the same fundamental technological characteristics except for some modifications that do not affect the basic device function. Any differences do not raise any new issues of safety and effectiveness.

The INSPIRE 8 DUAL is ethylene oxide sterilized and has a non-pyrogenic fluid path. It is for sinale use only.

NON CLINICAL TEST RESULTS:

Applicable tests were carried out in accordance with the requirements of ISO 10993-1 and the FDA May 1st, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing of materials.

IN VITRO TEST RESULTS:

In vitro testing was carried out in accordance with the relevant requirements of "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions; Final Guidance for Industry and FDA Staff" issued on November 13, 2000, "Guidance for Extracorporeal Blood Circuit Defoamer 510(k) Submissions; Final Guidance for Industry and FDA" issued on November 29, 2000 ISO 7199 "Cardiovascular implants and artificial organs - Blood-gas exchangers (oxygenators)"; ISO 15675 "Cardiovascular implants and artificial organs — Cardiopulmonary bypass systems — Arterial blood line

{2}------------------------------------------------

INSPIRE 8 DUALTraditional 510(k)
Sorin Group Italia S.r.I.April 24, 2012

filters" and ISO 15674, "Cardiovascular implants and artificial organs - Hard-shell cardiotomy/venous reservoir systems (with/without filter) and soft venous reservoir bags".

In vitro testing was carried out to demonstrate both the substantial equivalence with the predicate device and also to comply with safety and effectiveness requirements.

Testing supplied in the 510(k) premarket notification includes performance tests and physical/mechanical integrity tests that demonstrate compliance with performance specifications.

The tests that were performed are listed in the following summary table. The INSPIRE 8 DUAL passed each test mentioned in the table below.

TESTTESTCLASSIFICATIONTEST TITLE
1Physical/MechanicalOxygenating module structural integrity
2Physical/MechanicalReservoir structural integrity
3Physical/MechanicalOxygenating module blood, water, gaspathway integrity
4Physical/MechanicalReservoir blood pathway integrity
5Functional/PerformanceOxygenating module blood volume capacity
6Functional/PerformanceReservoir blood rest volume
7Functional/PerformanceOxygenating module gas transferperformance/blood side pressure drop
8Functional/PerformanceOxygenating module heat exchangeperformance/water side pressure drop
9Functional/PerformanceReservoir air handling
10Functional/PerformanceReservoir break-through time and volume
11Functional/PerformanceReservoir defoaming efficiency
12Functional/PerformanceReservoir dynamic priming volume / Hold-up
13Functional/PerformanceReservoir filtration efficiency - venoussection
14Functional/PerformanceReservoir filtration efficiency - cardiotomysection
15Functional/PerformanceReservoir flow rate capacity
16Functional/PerformanceReservoir pressure drop
17Functional/PerformanceIntegrated device hemolysis
18Functional/PerformanceIntegrated device blood compatibility
19Functional/PerformanceOxygenating module leaching of coating
20Functional/PerformanceReservoir leaching of coating

{3}------------------------------------------------

TESTTHESTCLASSIFICATIONTEST TITLE
21Functional/PerformanceIntegrated device flaking of coating
22Functional/PerformanceOxygenating module uniformity of coating
73Functional/PerformanceReservoir uniformity of coating

CONCLUSIONS:

The results of in vitro studies demonstrate that the INSPIRE 8 DUAL performs in a manner substantially equivalent to the D903 AVANT predicate device with respect to the relevant functional parameters. Test results of this study demonstrate the INSPIRE 8 DUAL is equivalent to the predicate device with respect to device function.

Additional testing has also demonstrated the effectiveness of production techniques to assure that the device is sterile and non-pyrogenic.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUL 2 3 2012

Sorin Group Italia s.r.1. c/o Mr. Barry Sall Principal Consultant Parexel International Consulting 195 West Street Waltham, WA 02451

Re: K121229

Trade/Device Name: Inspire 8 DUAL Hollow Fiber Oxygenator with Integrated Hardshell Venous/Cardiotomy Reservoir Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: April 24, 2012 Received: April 24, 2012

Dear Mr. Sall:

We have reviewed your Section 510(k) premarket notification of intent to market the device we neve rollowed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 101 ass stated in the Med 28, 1976, the enactment date of the Medical Device Amendments, or to conninered print to trial 2011-17, in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic (110.) that the novice, subject to the general controls provisions of the Act. The r ou may, merelore, manot of the Act include requirements for annual registration, listing of general controls provisions of practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is elabilited been of . Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Barry Sall

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be har nou a determination that your device complies with other requirements of the Act that I Drinas intact and regulations administered by other Federal agencies. You must or uny I eath all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Or K F at 607), hoo!ing (21 CFR 803); good manufacturing practice requirements as set do rece related daty systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the qualify of oversions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire speethe ad 100 ros JVA /CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regulations regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Tou may of amall Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

INSPIRE 8 DUAL Sorin Group Italia S.r.I.

Indications for Use:

INSPIRE 8M: Hollow Fiber Oxygenator

The INSPIRE 8M is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It provides gas exchange support and blood temperature control. INSPIRE 8M is intended to be used for 6 hours or less.

INSPIRE HVR DUAL: Hardshell Venous/Cardiotomy Reservoir

The INSPIRE HVR DUAL is intended for use in adult and small adult surgical procedures requiring cardiopulmonary bypass. It collects, defoams and filters venous blood and suction blood. INSPIRE HVR DUAL can be used post-operatively for chest drainage. INSPIRE HVR DUAL is intended to be used for 6 hours or less.

× Prescription Use _ (Part 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 807 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_ 220

68

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”