K Number
K121227
Device Name
MANUKA FILL
Date Cleared
2012-05-23

(30 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MANUKA FILL wound dressings are sterile, single-use, wound care dressings for use in moist wound management. MANUKA FILL wound dressings may be used Over-The-Counter for:

  • Minor Abrasions
  • Lacerations
  • Minor Cuts
  • Minor Scalds and Burns

Under the supervision of a healthcare professional, MANUKA FILL wound dressings may be used for:

  • Leg Ulcers
  • Pressure Ulcers
  • 1st and 2nd Degree Burns (Superficial and Partial Thickness)
  • Diabetic Foot Ulcers
  • Surgical Wounds
  • Traumatic Wounds
Device Description

MANUKA FILL wound dressings are sterile, single-use wound care dressings for use in moist wound management. The primary device is Leptospermum scoparium honey from New Zealand that is harvested and processed under controlled conditions. This honey is sealed into low density polyethylene tubes (LDPE) before sterilization using gamma irradiation.

AI/ML Overview

This is a 510(k) premarket notification for the MANUKA FILL wound dressings. The submission focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study for a novel device. Therefore, many of the requested categories (like MRMC study, sample sizes for test/training sets, number of experts for ground truth, adjudication methods) are not applicable or not provided in this type of submission.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Biocompatibility (Cytotoxicity, Skin Irritation, Sensitization, Wound Healing)Adherence to US FDA General Program Memorandum #G95-1 Part-10993-1 of the International Standard Organization (ISO) Standard (Biological Evaluation of Medical Devices)Met the acceptance criteria for all tests conducted and is considered biocompatible under the conditions tested.
Sterilization ValidationAcceptance criteria not explicitly detailed but implied by 'All acceptance criteria were met'.All acceptance criteria were met.
Shelf-life (Accelerated and Real-time)Acceptance criteria not explicitly detailed but implied by 'All acceptance criteria were met'.All acceptance criteria were met.
Packaging ValidationAcceptance criteria not explicitly detailed but implied by 'All acceptance criteria were met'.All acceptance criteria were met.

2. Sample Size Used for the Test Set and Data Provenance

This document primarily describes nonclinical testing for biocompatibility, sterility, shelf-life, and packaging. It does not refer to a "test set" in the context of clinical performance data like a diagnostic algorithm. Therefore, information on sample size for a test set and data provenance (country of origin, retrospective/prospective) is not provided. The biocompatibility tests were likely conducted on laboratory animals and/or in vitro, but specific details are not given.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This submission focuses on the safety and effectiveness of a wound dressing through nonclinical testing, not on the performance of a diagnostic algorithm requiring expert-established ground truth.

4. Adjudication Method for the Test Set

Not applicable, as there is no "test set" in the context of a clinical performance study for an algorithm.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-powered device or a diagnostic algorithm, so an MRMC study is irrelevant.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an algorithm.

7. The Type of Ground Truth Used

For the nonclinical tests:

  • Biocompatibility: The "ground truth" would be the physiological and cellular responses observed in the lab, compared against established biological safety standards (ISO Standard, US FDA General Program Memorandum #G95-1).
  • Sterilization, Shelf-life, Packaging: The "ground truth" would be objective measurements and observations confirming sterility, product integrity over time, and packaging effectiveness, compared against defined specifications and regulatory requirements.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/ML device; there is no "training set."

9. How the Ground Truth for the Training Set was Established

Not applicable.

{0}------------------------------------------------

K121227
page 1 of 3

MAY 2 3 2012

Image /page/0/Picture/2 description: The image shows a logo with the letters "lmp" in a stylized, interconnected design. The letters are bold and outlined, with rounded corners and a retro aesthetic. The background is black, and the letters are white, creating a high-contrast visual. The logo has a slightly distressed or aged appearance, adding to its vintage feel.

Links Medical Products Inc

510(k) SUMMARY

Submitted by:

ું મું

Owner's Name: Address:

Contact: Telephone: Fax: E-mail:

Links Medical Products, Inc. 9247 Research Drive Irvine, CA 92618 Tom Buckley, Chief Executive Officer 949-753-0001 949-753-7412 tbuckley@linksmed.com

Contact Person:

Company Address

Contact: Telephone Fax: E-mail:

Date Prepared: Trade Name: Common Name: Classification Name: Device Class: Product Code: Predicate Device: Predicate 510(k) #: Device Description:

NanoSmart, Inc. 29442 Pointe Royale Laguna Niguel, CA 92677 James Smith, Ph.D. 949-340-7261 949-340-7141 irsmith00@cox.net

April 17, 2012 MANUKA FILL wound dressings Wound Dressing Dressing, Wound, Drug Unclassified FRO Manukapli wound dressings (Manuka Medical, Ltd.) K092689 MANUKA FILL wound dressings are sterile, single-use wound care dressings for use in moist wound management. The primary device is Leptospermum scoparium honey from New Zealand that is harvested and processed under

controlled conditions. This honey is sealed into low density polyethylene tubes (LDPE) before sterilization using gamma irradiation.

{1}------------------------------------------------

Image /page/1/Figure/0 description: The image contains handwritten text. The top line reads 'K121227'. The second line reads 'posi 2 of 3'. The text appears to be written in black ink on a white background.

Image /page/1/Figure/1 description: The image shows a logo with the letters "LMP" in a stylized, blocky font. The letters are outlined in white and filled in with black. The logo has a retro or vintage feel due to the font style and the slightly distressed appearance of the image.

Links Medical Products Inc.

Intended Use:

MANUKA FILL wound dressings are sterile, single-use wound care dressings for use in moist wound management. MANUKA FILL Wound Dressings may be used Over-The-Counter for:

  • s Minor Abrasions
  • Lacerations

  • Minor Cuts

  • Minor Scalds and Burns A

Under the supervision of a healthcare professional, MANUKA FILL wound dressings may be used for:

  • Leg Ulcers

  • Pressure Ulcers s
  • 1st and 2nd Degree Burns (Superficial and Partial A Thickness)
  • A Diabetic Foot Ulcers
  • a Surgical Wounds
  • Traumatic Wounds A

The technical characteristics of MANUKA FILL wound dressing are substantially equivalent to the predicate device. The devices are similar in function, composition, and intended use. Leptospermun scoparium honey is the primary ingredient for MANUKA FILL and the predicate device. Both MANUKA FILL and the predicate device are provided as a single-use device in individually-sterilized packaging.

Standard biocompatibility tests were performed on the MANUKA FILL wound dressings; including cytotoxicity, skin irritation, sensitization, and wound healing studies. All tests were performed in accordance with US FDA General Program Memorandum #G95-1 Part-10993-1 of the International Standard Organization (ISO) Standard (Biological Evaluation of Medical Devices) by North American Science Associates, Inc. (NAMSA). The MANUKA FILL wound dressing met the acceptance criteria for all tests conducted and is considered biocompatible under the conditions tested.

LINKS MEDICAL PRODUCTS, INC. 9247 Research Drive Irvine, CA 92618

Technology Comparison:

Nonclinical Testing:

ર-2

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo with the letters 'LMP' in a stylized, interconnected design. The letters are bold and outlined, with the 'L' and 'M' sharing a common base. The 'P' is a rounded shape connected to the 'M'. The logo has a retro or vintage feel due to its design and the image quality.

K 12 12 27
Page 3 of 3

Links Medical Products Inc.

Additional testing included sterilization validation, shelf-life under accelerated and real-time conditions, and packaging validation. All acceptance criteria were met for all tests conducted.

Conclusion of Comparison: MANUKA FILL and the predicate device were both demonstrated to be biocompatible and meet performance requirements for sterility, shelf-life, and packaging. Based upon technological characteristics and nonclinical performance data, MANUKA FILL wound dressings are substantially equivalent and as safe and effective as the currently-marketed predicate device.

Links Medical Products, Inc. 9247 Research Drive Irvine, CA 92618

5-3

{3}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Links Medical Products, Incorporated % Nano Smart Incorporated James Smith, Ph.D. Consultant 29442 Pointe Royale Laguna Niguel, California 92677

WAY 2 3 2012

Re: K121227 Trade/Device Name: MANUKA FILL wound dressings Regulatory Class: Unclassified Product Code: FRO Dated: April 16 , 2012 Received: April 23, 2012

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

Page 2 - Dr. Smith

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

INDICATIONS FOR USE

510(k) Number (if known):

Device Name: MANUKA FILL wound dressings

Indications for Use:

MANUKA FILL wound dressings are sterile, single-use, wound care dressings for use in moist wound management. MANUKA FILL wound dressings may be used Over-The-Counter for:

  • Minor Abrasions

  • A Lacerations
  • A Minor Cuts
  • Minor Scalds and Burns

Under the supervision of a healthcare professional, MANUKA FILL wound dressings may be used for:

  • Leg Ulcers

  • Pressure Ulcers

  • A 1st and 2nd Degree Burns (Superficial and Partial Thickness)
  • Diabetic Foot Ulcers

  • Surgical Wounds A
  • Traumatic Wounds

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Kroneberger MM

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K121227

N/A