K Number
K121108

Validate with FDA (Live)

Device Name
DOPPLEX ABILITY
Date Cleared
2013-01-03

(266 days)

Product Code
Regulation Number
870.2780
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Dopplex Ability is indicated for use on adult subjects at risk of having or developing peripheral arterial disease (PAD).

Dopplex Ability is intended for rapid measurement of ankle-brachial pressure index (ABPI) or ankle-brachial index (ABI) in adults and pulse volume recording (PVR) / volume plethysmography.

It is suitable for use in woundcare assessment, for assessing symptomatic PAD, and as a screening device for PAD. It may also be used on patients with venous or arterial ulcers prior to the application of compression therapy. Dopplex Ability can be used on patients with unilateral lower limb amputation.

Device Description

Dopplex Ability is intended to measure a patient's Ankle Brachial Index (ABI) and provide PVR. This is done through an automated process.

The operator places the four colour coded cuffs on each of the patient's limbs as described in the instruction for use and connect to the device.

When connected the operator pushes soft touch button to start measurement, the device then will automatically control the inflation / deflation of the cuffs on each limb and monitor minute variations in individual pressures to determine values to be used for the calculation of the ABI values for both the left and right hand of the patient.

Dopplex Ability uses pneumo-plethysmography uses obtain physiological measurements from patient's limbs. Measurements are conducted as a single occurrence on all four limbs, eliminating any requirement to rest the patient between measurements.

The test period takes approximately 3 minutes. The ABI indices are calculated using an in house algorithm.

ABI values are displayed on the LCD and can also be printed. In addition, records are generated of the Pulse Volume Waveforms (PVR), which may also be displayed on the LCD and can be printed.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the "Dopplex Ability" device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The provided text does not explicitly state numerical acceptance criteria in a dedicated table format. Instead, it describes the study design and concludes that the device performs as well as the predicate. The performance is demonstrated through a comparison with a predicate device.

Acceptance Criterion (Assumed)Reported Device Performance (from text)
Safety and Effectiveness"The non clinical and clinical data detailed within submission demonstrates that the device is safe and effective and performs as well as the legally marketed Predicate device, identified in this summary."
Measurement of ABPI / ABI and PVR (equivalent to predicate)"Clinical data submitted demonstrates the submission device to provide the performance as indicated for use by a comparison of measured indices using both devices." "The predicate device has been used in comparison bench testing and clinical assessment to evaluate and demonstrate the equal capability of both devices to measure ABI."
Improved Functionality (single occurrence measurement on all four limbs) (beyond predicate capability)"With respect to Dopplex Ability to function to measure all four limbs as a single occurrence, the submission device provides improved functionality above that of the legally marketed predicate device without addition concerns regarding safety and effectiveness." This implies the acceptance of this improved function without compromising safety/effectiveness compared to the predicate.

2. Sample Size Used for the Test Set and Data Provenance

The document states:

  • "The predicate device has been used in comparison bench testing and clinical assessment to evaluate and demonstrate the equal capability of both devices to measure ABI."
  • "Clinical data submitted demonstrates the submission device to provide the performance as indicated for use by a comparison of measured indices using both devices."

However, the specific sample size for the test set (number of patients or measurements) is not explicitly mentioned in the provided 510(k) summary.

Data Provenance: The document does not specify the country of origin for the clinical data. It also does not explicitly state if the study was retrospective or prospective. Given the nature of a 510(k) submission and the description of "clinical assessment" to demonstrate "equal capability," it is most likely a prospective clinical study designed for this comparison.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not provide information on the number of experts used or their qualifications for establishing ground truth for the test set. The study design described is a comparison of the new device (Dopplex Ability) to a predicate device (Vascular Assist), rather than a comparison to expert-derived ground truth as such. The predicate device's measurements are effectively treated as the "reference standard" for comparison.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method. Since the study focuses on quantitative measurement agreement between two devices rather than subjective interpretation, a formal adjudication process for "ground truth" might not have been applicable in the traditional sense. The comparison likely involved statistical methods to assess agreement (e.g., Bland-Altman analysis, correlation).

5. If an MRMC Comparative Effectiveness Study Was Done, and Effect Size

The provided text does not indicate a Multi-Reader Multi-Case (MRMC) comparative effectiveness study. The study described is a comparison of two devices (Dopplex Ability vs. Vascular Assist) rather than comparing human readers with and without AI assistance. Therefore, there is no mention of an effect size for human reader improvement with AI.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, the study primarily describes standalone performance of the Dopplex Ability device itself, comparing its measurements to those of the predicate device. The device is described as an "automated process" where "the device then will automatically control the inflation / deflation of the cuffs on each limb and monitor minute variations in individual pressures to determine values to be used for the calculation of the ABI values." The "in house algorithm" calculates the ABI values. While a human initiates the test and places cuffs, the core measurement and calculation are automated.

7. The Type of Ground Truth Used

The "ground truth" for the test was the measurements obtained by the legally marketed predicate device (Vascular Assist). The study's objective was to demonstrate that the Dopplex Ability device's measurements of ABI and PVR were "equal" or provided "the performance as indicated for use by a comparison of measured indices using both devices."

8. The Sample Size for the Training Set

The document does not specify a sample size for a training set. The device performs quantitative measurements based on pneumo-plethysmography and an "in house algorithm" for ABI calculation. While such algorithms might be developed and "trained," the summary only describes the clinical validation against the predicate, not the internal development or training process.

9. How the Ground Truth for the Training Set Was Established

Since a training set sample size is not provided, the method for establishing its ground truth is also not described in this document.

{0}------------------------------------------------

Huntleigh Healthcare Limited | - | | | | T L E | G | | | T L E | G | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |

510(k) Summary Dopplex Ability
Name & Address:Huntleigh Healthcare Limited - Diagnostic Products Division Unit 35, Portmanmoor Road Cardiff CF24 5HN United KingdomJAN 0 3 2013
Telephone:+(44) 2920 485885
Fax:+(44) 2920 492520
Prepared:8 December 2012
Contact:David Moynham - Regulatory Affairs Engineer
Device Name:Dopplex Ability
Common NameHydraulic, pneumatic, or photoelectric plethysmographs
ClassificationClassProduct CodeClassification Regulation
IIJOM870.2780
Classification Name:Hydraulic, pneumatic, or photoelectric plethysmographs
Predicate Device:Vascular Assist (K002186) manufactured by Huntleigh Healthcare Limited, 35 Portmanmoor Road, Cardiff, CF24 5HN, UK
Indications for Use:Dopplex Ability is indicated for use on adult subjects at risk of having or developing peripheral arterial disease (PAD).
Dopplex Ability is intended for rapid measurement of ankle-brachial pressure index (ABPI) or ankle-brachial index (ABI) in adults and pulse volume recording (PVR) / volume plethysmography.
It is suitable for use in woundcare assessment, for assessing symptomatic PAD, and as a screening device for PAD. It may also be used on patients with venous or arterial ulcers prior to the application of compression therapy. Dopplex Ability can be used on patients with unilateral lower limb amputation.
The Indications for Use of Dopplex Ability are specific to the measurement of ankle-brachial pressure index (ABPI) or ankle-brachial index (ABI) in adults and pulse volume recording (PVR) for those considered at risk of having or developing peripheral arterial disease (PAD). The Predicate device indication for use, is a broader indication "The device is used for the assessment of blood flow in veins and arteries to assist in the identification of vascular disease". Peripheral arterial disease (PAD) is a sub group of the broader indication "vascular disease". The Predicate device includes the capability of measuring ABPI / ABI and PVR. Clinical data submitted demonstrates the submission

Dopplex Ability FDA 510(k) Premarket Notification

3

{1}------------------------------------------------

Huntleigh Healthcare Limited = Limited = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =

device to provide the performance as indicated for use by a comparison of measured indices using both devices. This supports the demonstration of its safety and effectiveness.

Description :

Dopplex Ability is intended to measure a patient's Ankle Brachial Index (ABI) and provide PVR. This is done through an automated process.

The operator places the four colour coded cuffs on each of the patient's limbs as described in the instruction for use and connect to the device.

When connected the operator pushes soft touch button to start measurement, the device then will automatically control the inflation / deflation of the cuffs on each limb and monitor minute variations in individual pressures to determine values to be used for the calculation of the ABI values for both the left and right hand of the patient.

Dopplex Ability uses pneumo-plethysmography uses obtain physiological measurements from patient's limbs. Measurements are conducted as a single occurrence on all four limbs, eliminating any requirement to rest the patient between measurements.

The test period takes approximately 3 minutes. The ABI indices are calculated using an in house algorithm.

ABI values are displayed on the LCD and can also be printed. In addition, records are generated of the Pulse Volume Waveforms (PVR), which may also be displayed on the LCD and can be printed.

Model REFDeviceFeatures
DA100PBDopplex AbilityAC powered with integral printer and battery
DA100PDopplex AbilityAC powered with integral printer
DA100BDopplex AbilityAC powered with battery
DA100Dopplex AbilityAC powered

Substantial Equivalence:

Dopplex Ability is substantially equivalent to the cleared device: Vascular Assist (K002186), cleared 05/02/2001

Technologies Summary:

The predicate device utilises Doppler technology to measure physiological parameters used in calculating into ABI indices. Whereas Dopplex Ability uses pneumo-plethysmography to obtain physiological parameters from patient's limbs

The predicate device has been used in comparison bench testing and clinical assessment to evaluate and demonstrate the equal capability of both devices to measure ABI.

Page 2 of 3

Dopplex Ability FDA 510(k) Premarket Notification

{2}------------------------------------------------

Huntleigh Healthcare Limited |------------------------------------------------------------------------------------------------------------------------------------------------

Conclusion:

The non clinical and clinical data detailed within submission demonstrates that the device is safe and effective and performs as well as the legally marketed Predicate device, identified in this summary. With respect to Dopplex Ability to function to measure all four limbs as a single occurrence, the submission device provides improved functionality above that of the legally marketed predicate device without addition concerns regarding safety and effectiveness.

Page 3 of 3

Dopplex Ability FDA 510(k) Premarket Notification

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract shapes resembling birds in flight, rendered in black.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

JAN 0 3 2013

Huntleigh Healthcare Limited c/o Mr. David Moynham Senior Regulatory Affairs Engineer 35 Portmanmoor Road Cardiff, CF24 5HN (UK)

Re: K121108

Trade/Device Name: Dopplex ability Regulation Number: 21 CFR 870.2780 Regulation Name: Hydraulic, pneumatic, or photoelectric plethysmographs Regulatory Class: Class II (two) Product Code: JOM Dated: December 10, 2012 Received: December 12, 2012

Dear Mr. David Moynham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. David Moynham

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerelv vours.

Bram D. Zuckerman

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications For Use

Huntleigh Healthcare Limited Huntleigh Healthcare Limited

510(k) Number:

Device Name:

Dopplex Ability

Indications for Use:

Dopplex Ability is indicated for use on adult subjects at risk of having or developing peripheral arterial disease (PAD).

Dopplex Ability is intended for rapid measurement of ankle-brachial pressure index (ABPI) or anklebrachial index (ABI) in adults and pulse volume recording (PVR) / volume plethysmography.

It is suitable for use in woundcare assessment, for assessing symptomatic PAD, and as a screening device for PAD. It may also be used on patients with venous or arterial ulcers prior to the application of compression therapy. Dopplex Ability can be used on patients with unilateral lower limb amputation.

Prescription Use
YESAND/OR
(Part 21 CFR 801 Subpart D)

Over-The-Counter Use NO (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K121108

Page 1 of 1

Huntleigh Healthcare Ltd.

§ 870.2780 Hydraulic, pneumatic, or photoelectric plethysmographs.

(a)
Identification. A hydraulic, pneumatic, or photoelectric plethysmograph is a device used to estimate blood flow in a region of the body using hydraulic, pneumatic, or photoelectric measurement techniques.(b)
Classification. Class II (performance standards).