K Number
K121039
Date Cleared
2012-05-25

(50 days)

Product Code
Regulation Number
866.2900
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MSwab is a Collection, Transport and Preservation System intended for the collection and transport of clinical specimens containing Gram positive aerobic and facultative anaerobic bacteria, HSV 1 and HSV 2 from the collection site to the testing laboratory. In the laboratory, MSwab specimens are processed using standard clinical laboratory operating procedures for culture.

Device Description

Copan MSwab Collection, Transport and Preservation System is supplied in two different formats: a collection kit format and a tube only format. Each collection kit consists of a package containing a plastic screw-cap tube with conical shaped bottom filled with 1 ml or 1.6 ml of MSwab transport and preservation medium and a small sterile peel pouch containing one specimen collection swab that has a tip flocked with soft nylon fiber. The tube only format consists of a plastic screw-cap tube with conical shaped bottom filled with 1 ml or 1.6 ml of MSwab transport and preservation medium. The MSwab is intended for single use. MSwab transport and preservation medium is a maintenance medium comprising TRIS HCl. EDTA. TRIS Base, Dimethyl Sulfoxide (DMSO) and Bovine Serum Albumin. The medium is designed to maintain the viability of Gram positive aerobic and facultative anaerobic bacteria. HSV 1 and HSV 2 during transit to the testing laboratory. The nylon flocked specimen collection swab provided in each collection kit of the Copan MSwab Collection. Transport and Preservation System has a solid plastic shaft with a molded breakpoint site.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Copan MSwab Collection, Transport and Preservation System, as extracted from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
Bacterial Recovery (Swab Elution):No more than a 3 log₁₀ (1 x 10 +/- 10%) decline in Colony Forming Units (CFU) between the CFU count at 0 hours and the CFU count at 48 hours.All tested Gram-positive aerobic and facultative anaerobic bacteria (12 strains) showed an acceptable log₁₀ decline within the 3 log₁₀ limit at both 4-8°C and 20-25°C for 48 hours. (Refer to Tables 1 and 2 for specific figures, all "Interpretation" rows are "Acceptable Recovery".)
Bacterial Recovery (Roll-Plate):>5 CFU at 48 hours from the specific dilution that yielded 0 hour counts closest to 300 CFU.All tested Gram-positive aerobic and facultative anaerobic bacteria (12 strains) showed acceptable recovery (>5 CFU at 48 hours) at both 4-8°C and 20-25°C for 48 hours. (Refer to Tables 3 and 4 for specific figures, demonstrating counts well above 5 CFU).
Viral Recovery (4-8°C & 20-25°C):Any viral recovery at 48 hours (determined by fluorescing foci counts).All tested HSV-1 and HSV-2 strains showed acceptable recovery (detectable fluorescing foci) at 48 hours at both 4-8°C and 20-25°C. (Refer to Table 5 for 4-8°C results and Table 6 for 20-25°C results, all "Interpretation" rows are "Acceptable Recovery" and non-zero foci counts are reported).
Viral Recovery (-70°C):Any viral recovery at 14 days.The results demonstrated acceptable recovery for all samples tested (specifically HSV-1 and HSV-2) at -70°C for 14 days. (No specific table for this data in the provided text, but the conclusion states it).
Recovery Stability:Ability to maintain viability of specific bacterial (Streptococcus pyogenes, Streptococcus pneumoniae, Staphylococcus aureus (MRSA)) and viral (HSV 1, HSV 2) strains for at least 48 hours at 4-8°C and 20-25°C, when using aged MSwab System devices (up to 13 months following manufacture).The results from three representative lots demonstrated the ability of the MSwab System to maintain the viability of the bacterial and viral strains evaluated for at least 13 months following the date of manufacture.
pH Stability:Maintain pH within the target range at all time intervals tested (up to 13 months).The results from three representative lots demonstrated the ability to maintain pH within the target range at all time intervals tested.
Sterilization:Tubes and caps: Gamma irradiation in accordance with UNI EN ISO 11137:2006. Swabs: ETO treatment in accordance with UNI EN ISO 11135:2007. Aseptic filling process for medium.Compliant, stated that "Representative samples were tested to validate the medium filling process with respect to risk of microbial contamination." and that components were sterilized according to the specified ISO standards.
Biocompatibility:Swab component is non-cytotoxic, non-irritating, and non-sensitizing in accordance with ISO 10993.The results demonstrated that the swab component is non-cytotoxic, non-irritating and non-sensitizing.
Cytotoxicity:No alteration of MRC5 cell monolayers compared to a negative control, using MSwab System devices aged 1, 7, and 13 months.The results from three representative lots aged 1 month, 7 months and 13 months showed no alteration of the cell monolayers compared to the negative control.

2. Sample Size Used for the Test Set and Data Provenance

  • Bacterial Recovery Studies:
    • For each of the 12 bacterial strains, the test set involved inoculating swabs in triplicate for measurement at 0, 24, and 48 hours. This was done for the MSwab device and relevant predicate devices, across two temperature ranges (4-8°C and 20-25°C). The text also mentions "three representative lots" for stability and pH studies, implying that the recovery studies were conducted across these lots as well, but the specific number of test samples (swabs) per organism per lot for the primary recovery study is explicitly stated as triplicate.
  • Viral Recovery Studies:
    • For each of the 2 viral strains, the test set involved inoculating swabs in triplicate for measurement at 0, 24, and 48 hours. This was done for the MSwab device and relevant predicate devices, across two temperature ranges (4-8°C and 20-25°C).
  • Recovery Stability, pH Stability, Cytotoxicity: Mention "three representative lots."
  • Data Provenance: The document does not specify the country of origin of the data. The studies are described as prospective laboratory recovery studies, where organisms were purposefully inoculated onto the collection systems and then tested for viability over time.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • The ground truth in this context is the viability of microorganisms (CFU counts for bacteria, fluorescing foci counts for viruses) after storage in the transport medium.
  • The document does not report the number of experts or their qualifications used to establish this ground truth. This is a standard laboratory study where the measurement of CFU or foci counts is a direct, quantifiable outcome, typically performed by trained laboratory personnel following established protocols (like CLSI M40-A). It is not a subjective assessment requiring expert consensus in the same way an imaging study might.

4. Adjudication Method for the Test Set

  • No adjudication method is explicitly described. The determination of "acceptable recovery" is based on quantitative thresholds applied to the CFU counts (for bacteria) or the presence of viral recovery (for viruses), as defined in the acceptance criteria. These are objective measures rather than subjective interpretations requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not done. This device is a specimen collection, transport, and preservation system, not an AI-powered diagnostic imaging or interpretation system that would involve human "readers" or AI assistance. The study focuses on the device's ability to maintain microorganism viability, not on improving human diagnostic accuracy.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, in concept, this is a standalone performance study for the device. The "algorithm" here isn't a software algorithm but rather the physical and chemical properties of the MSwab system (medium, swab material, etc.) in preserving organisms. The studies directly measure the performance of the device itself (its ability to maintain viability) without human intervention in the preservation process or interpretation stage beyond the initial inoculation and final enumeration.

7. The Type of Ground Truth Used

  • The ground truth used is direct microbial enumeration:
    • Colony Forming Units (CFU) counts for bacteria.
    • Fluorescing foci counts for viruses.
    • For viral recovery at -70°C, the ground truth was simply "any viral recovery" (implying detectability).
  • This is a form of direct measurement of biological viability rather than expert consensus, pathology, or outcomes data.

8. The Sample Size for the Training Set

  • This information is not applicable to this type of study. The device is a physical collection and transport system, not a machine learning or AI algorithm that requires a "training set" in the computational sense. The reported studies are performance validation studies.

9. How the Ground Truth for the Training Set Was Established

  • This information is not applicable as there is no "training set" for this device.

{0}------------------------------------------------

510(k) Summary for

MAY 2 5 2012

K121039

Copan MSwab Collection, Transport and Preservation System

(per 21 CFR 807.92 and http://www.fda.gov/MedicalDeviceRegulationandGuidance/default.htm)

1. APPLICANT/SPONSOR

Copan Flock Technologies S.r.l. Via F. Perotti 16-18 25125 Brescia, Italy

Contact Person:Alberto Poli
Telephone:+39 030 2687290
FAX:+39 030 2687250

May 22, 2012 Date Prepared:

2. DEVICE NAME AND REGULATORY INFORMATION

Proprietary Name:Copan MSwab Collection, Transport and Preservation System
Common/Usual Name:Collection and Transport Device
Classification Name:Microbiological Specimen Collection and Transport Device
Regulation Section:21 CFR 866.2900
Classification:Class I
Product Codes:JTW (System, Transport, Aerobic)JTX (Transport Systems, Anaerobic)
Panel:Microbiology

3. PREDICATE DEVICES

  • · Copan Liquid Amies Elution Swab (ESwab) Collection and Transport System Copan Diagnostics Inc. K061301
  • · Copan Universal Transport Medium (UTM-RT) System Copan Diagnostics Inc. K042970

4. DEVICE DESCRIPTION

Copan MSwab Collection, Transport and Preservation System is supplied in two different formats: a collection kit format and a tube only format. Each collection kit consists of a

l

{1}------------------------------------------------

package containing a plastic screw-cap tube with conical shaped bottom filled with 1 ml or 1.6 ml of MSwab transport and preservation medium and a small sterile peel pouch containing one specimen collection swab that has a tip flocked with soft nylon fiber. The tube only format consists of a plastic screw-cap tube with conical shaped bottom filled with 1 ml or 1.6 ml of MSwab transport and preservation medium. The MSwab is intended for single use.

MSwab transport and preservation medium is a maintenance medium comprising TRIS HCl. EDTA. TRIS Base, Dimethyl Sulfoxide (DMSO) and Bovine Serum Albumin. The medium is designed to maintain the viability of Gram positive aerobic and facultative anaerobic bacteria. HSV 1 and HSV 2 during transit to the testing laboratory.

The nylon flocked specimen collection swab provided in each collection kit of the Copan MSwab Collection. Transport and Preservation System has a solid plastic shaft with a molded breakpoint site.

5. INTENDED USE/INDICATIONS FOR USE

MSwab is a Collection, Transport and Preservation System intended for the collection and transport of clinical specimens containing Gram positive aerobic and facultative anaerobic bacteria, HSV 1 and HSV 2 from the collection site to the testing laboratory. In the laboratory, MSwab specimens are processed using standard clinical laboratory operating procedures for culture.

6. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICES

The Copan MSwab products are substantially equivalent to the predicate specimen collection and transport devices. The Copan MSwab products and the predicate devices are similar in intended use and overall function.

The proposed and predicate devices are single-use products intended for the collection and transport of clinical specimens containing bacteria in the case of the Copan ESwab System and viruses in the case of the Copan UTM-RT System. Both the Copan MSwab and the predicate devices are offered in collection kit formats with specimen collection swab options.

2 .

{2}------------------------------------------------

- RECOMMENDATIONS FOR DEALING WITH "BAD" BEHAVIOR.
- KNOW YOURSELF. What "pushes your buttons"? What behaviors do you find most annoying?- KNOW THE CHILD. What are the child's strengths? What are the child's needs? What is the child trying to communicate through his/her behavior?- KNOW THE CONTEXT. What is happening in the classroom? What is happening at home? What is happening in the child's life?- KNOW THE LAW. What are the legal requirements for dealing with "bad" behavior? What are the ethical considerations?

. .

ItemCopan MSwab System(Bacteriology and Virology Claims)Copan ESwab System(Predicate for Bacteriology Claim)Copan UTM-RT System(Predicate for Virology Claim)
Intended UseMSwab is a Collection, Transport and Preservation System intended for the collection and transport of clinical specimens containing Gram positive aerobic and facultative anaerobic bacteria, HSV 1 and HSV 2 from the collection site to the testing laboratory. In the laboratory, MSwab specimens are processed using standard clinical laboratory operating procedures for culture.Copan Liquid Amies Elution Swab (ESwab) Collection and Transport System is intended for the collection and transport of clinical specimens containing aerobes, anaerobes and fastidious bacteria from the collection site to the testing laboratory. In the laboratory, ESwab specimens are processed using standard clinical laboratory operating procedures for bacterial culture.Copan Universal Transport Medium (UTM-RT) System is intended for the collection and transport of clinical specimens containing viruses, chlamydiae, mycoplasma or ureaplasma from the collection site to the testing laboratory. UTM-RT can be processed using standard clinical laboratory operating procedures for viral, chlamydial, mycoplasma and ureaplasma culture.
MicroorganismsSupportedGram positive aerobic and facultative anaerobic bacteria, HSV 1 and HSV 2Aerobic, anaerobic and fastidious bacteriaViruses, chlamydiae, mycoplasma and ureaplasma
Medium FormulationTRIS HCLEDTATRIS BaseDMSOBovine Serum AlbuminDistilled waterSodium chloridePotassium chlorideCalcium chlorideMagnesium chlorideMonopotassium phosphateDisodium phosphateSodium thioglycollateDistilled waterHank's Balanced SaltsBovine Serum AlbuminL-CysteineGelatinSucroseL-Glutamic AcidHEPES BufferVancomycinAmphotericin BColistinPhenol Red
Storage Temperature5-25°C5-25°C2-25°C
Container for MediumTube; Plastic, conical bottomTube; Plastic, conical bottomTube; Plastic, conical bottom
Product ConfigurationMedium in Tube with Cap;Kit with Medium and Swab in Peel PouchOptionsKit with Medium and Swab in Peel PouchOptionsMedium in Tube with Cap;Kit with Medium and Swab in Peel PouchOptions
Swab TipFlocked nylonFlocked nylonPolyester

{3}------------------------------------------------

7. STANDARD/GUIDANCE DOCUMENTS REFERENCED

  • · Clinical and Laboratory Standards Institute (CLSI) document M40-A, Vol. 23, No. 34, "Ouality Control of Microbiological Transport Systems: Approved Standard
  • · FDA/CDRH General Program Memorandum G95-1. Use of International Standard ISO-10993. "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". (May 1. 1995)

8. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE

Recovery Studies: Recovery studies were performed using the Copan MSwab and the predicate devices to determine the ability of the products to maintain viability of various strains of Gram positive aerobic and facultative anaerobic bacteria. HSV 1 and HSV 2.

The Gram positive aerobic and facultative anaerobic bacteria included in the recovery studies were:

Streptococcus pyogenesATCC 19615
Streptococcus pneumoniaeATCC 6305
Streptococcus pneumoniaeATCC 49136
Enterococcus faecalisATCC 29212
Staphylococcus epidermidisATCC 12228
Staphylococcus aureusATCC 29213
Streptococcus agalactiae (Group B Strep)ATCC 13813
Kocuria rhizophilaATCC 9341
Listeria monocytogenesATCC 19114
Bacillus cereusATCC 10876
Staphylococcus aureus (Methicillin resistant)ATCC 43300
Staphylococcus aureusATCC 6538
Staphylococcus aureus (Methicillin resistant)ATCC 700698

The viruses included in the recovery studies were: ATCC VR-539 Hernes Simnlex Virus Type 1 (HSV 1)

Herpes Simplex Virus Type 1 (HSV-1)ATCC VR-733
Herpes Simplex Virus Type 2 (HSV 2)ATCC VR-734

The recovery studies were performed using the Copan MSwab and the predicate devices for all organisms at two different temperature ranges, 4-8℃ and 20-25℃, corresponding to refrigerator and room temperature, respectively. Swabs accompanying each transport system were inoculated in triplicate with 100 ul of specific concentrations of organism suspension. Swabs were then placed in their respective transport medium tubes and were held for 0, 24

{4}------------------------------------------------

and 48 hours. Viability performance was determined for each.test organism at the 48 hours time point compared to the 0 hour time point.

For the bacterial recovery studies, each swab was processed according to the swab elution or roll-plate method at the appropriate time intervals. Acceptable recovery for the swab elution method was defined as no more than a 3 log10 (1 x 10 +/- 10%) decline in CFU between the CFU count at 0 hour and the CFU count at 48 hours. Acceptable recovery for the roll-plate method was defined as >5 CFU at 48 hours from the specific dilution that yielded 0 hour counts closest to 300 CFU. The results of the bacterial recovery studies for the Copan MSwab System are presented in Tables 1 through 4.

For the viral recovery studies, each swab was vortexed and removed from its transport medium tube at the appropriate time interval. 200ul aliquots of the suspension in the medium tube were inoculated into shell vials. All cultures were processed by standard laboratory culture technique and examined after a specified incubation time. Organism viability was determined by fluorescing foci counts. Acceptable recovery was defined as any viral recovery at 48 hours. The results of the viral recovery studies for the Copan MSwab System are presented in Tables 5 and 6.

Viral recovery studies also were performed using the Copan MSwab and the predicate Copan UTM-RT for samples stored at -70℃ for 14 days. Acceptable recovery was defined as any viral recovery at 14 days. The results demonstrated acceptable recovery for all samples tested.

The results of the recovery studies demonstrate the ability of the Copan MSwab Collection, Transport and Preservation System to maintain the viability of all bacterial strains evaluated for at least 48 hours at temperatures of 4-8°C and 20-25°C, and the viability of all viral strains evaluated for at least 48 hours at temperatures of 4-8℃ and 20-25℃, and for at least 14 days at a temperature of -70°C.

{5}------------------------------------------------

TABLE 1. SUMMARY OF RESULTS FOR BACTERIAL RECOVERY STUDIES

SWAB ELUTION METHOD, 4-8°C

.

.

·

.

.

.

.

·

:

.

OrganismDilution:0.5 McFarlandbacterialsuspension withsalineProductLot NumberAverage ofCFUs recoveredat time0 hrsAverage ofCFUs recoveredat time24 hrsAverage ofCFUs recoveredat time48 hrsLogdeclineInterpretation
StreptococcuspyogenesATCC 19615diluted1:10MSwab2045$3.9 x 10^5$$2.7 x 10^5$$2.3 x 10^5$-0.23AcceptableRecovery
2045/1$4.8 x 10^5$$2.5 x 10^5$$2.1 x 10^5$-0.36AcceptableRecovery
2045/2$4.2 x 10^5$$2.3 x 10^5$$2.0 x 10^5$-0.32AcceptableRecovery
StreptococcuspneumoniaeATCC 6305diluted1:10MSwab2045$1.2 x 10^5$$1.8 x 10^4$$2.1 x 10^3$-1.76AcceptableRecovery
2045/1$1.2 x 10^5$$1.9 x 10^4$$1.7 x 10^3$-1.85AcceptableRecovery
2045/2$1.2 x 10^5$$2.0 x 10^4$$1.9 x 10^3$-1.80AcceptableRecovery
StreptococcuspneumoniaeATCC 49136diluted1:10MSwab2045$2.0 x 10^5$$2.0 x 10^3$$2.1 x 10^3$-1.98AcceptableRecovery
2045/1$1.8 x 10^5$$2.1 x 10^4$$1.8 x 10^3$-2.00AcceptableRecovery
2045/2$1.9 x 10^5$$1.8 x 10^4$$1.7 x 10^3$-2.05AcceptableRecovery
Enterococcus faecalisATCC 29212diluted1:10MSwab2045$1.1 x 10^6$$1.0 x 10^6$$9.4 x 10^5$-0.07AcceptableRecovery
2045/1$1.1 x 10^6$$1.0 x 10^6$$9.2 x 10^5$-0.08AcceptableRecovery
2045/2$1.1 x 10^6$$1.0 x 10^6$$9.5 x 10^5$-0.06AcceptableRecovery
StaphylococcusepidermidisATCC 12228diluted1:10MSwab2045$1.6 x 10^6$$1.2 x 10^6$$9.2 x 10^5$-0.24AcceptableRecovery
2045/1$1.7 x 10^6$$1.3 x 10^6$$1.1 x 10^6$-0.19AcceptableRecovery
2045/2$1.4 x 10^6$$1.2 x 10^6$$9.5 x 10^5$-0.17AcceptableRecovery
Staphylococcus aureusATCC 29213diluted1:10MSwab2045$1.3 x 10^6$$1.0 x 10^6$$8.8 x 10^5$-0.17AcceptableRecovery
2045/1$1.0 x 10^6$$9.5 x 10^5$$7.4 x 10^5$-0.13AcceptableRecovery
2045/2$1.0 x 10^6$$9.5 x 10^5$$8.0 x 10^5$-0.10AcceptableRecovery
Streptococcus agalactiaeATCC 13813diluted1:10MSwab2045$2.1 x 10^6$$1.8 x 10^6$$1.5 x 10^6$-0.15AcceptableRecovery
2045/1$1.9 x 10^6$$1.8 x 10^6$$1.5 x 10^6$-0.10AcceptableRecovery
2045/2$2.2 x 10^6$$1.4 x 10^6$$1.3 x 10^6$-0.23AcceptableRecovery
Kocuria rhizophilaATCC 9341diluted1:10MSwab2045$2.0 x 10^5$$1.5 x 10^5$$1.4 x 10^5$-0.15AcceptableRecovery
2045/1$2.1 x 10^5$$1.4 x 10^5$$1.4 x 10^5$-0.18AcceptableRecovery
2045/2$2.2 x 10^5$$1.6 x 10^5$$1.4 x 10^5$-0.19AcceptableRecovery
Listeria monocytogenesATCC 19114diluted1:10MSwab2045$1.4 x 10^6$$8.1 x 10^5$$3.0 x 10^5$-0.67AcceptableRecovery
2045/1$1.3 x 10^6$$9.0 x 10^5$$3.4 x 10^5$-0.58AcceptableRecovery
2045/2$1.5 x 10^6$$8.8 x 10^5$$4.2 x 10^5$-0.55AcceptableRecovery
Bacillus cereusATCC 10876diluted1:10MSwab2045$1.5 x 10^5$$6.9 x 10^4$$1.6 x 10^4$-0.97AcceptableRecovery
2045/1$1.6 x 10^5$$7.6 x 10^4$$1.8 x 10^4$-0.95AcceptableRecovery
2045/2$1.6 x 10^5$$8.3 x 10^4$$1.4 x 10^4$-1.06AcceptableRecovery
Staphylococcus aureus(MRSA)ATCC 43300diluted1:10MSwab2045$2.1 x 10^6$$1.2 x 10^6$$8.0 x 10^5$-0.42AcceptableRecovery
2045/1$1.9 x 10^6$$1.3 x 10^6$$8.9 x 10^5$-0.33AcceptableRecovery
2045/2$1.9 x 10^6$$1.4 x 10^6$$9.4 x 10^5$-0.30AcceptableRecovery
Staphylococcus aureusATCC 6538diluted1:10MSwab2045$1.5 x 10^6$$9.2 x 10^5$$6.9 x 10^5$-0.34AcceptableRecovery
2045/1$1.6 x 10^6$$1.0 x 10^6$$8.3 x 10^5$-0.29AcceptableRecovery
2045/2$1.7 x 10^6$$1.3 x 10^6$$9.5 x 10^5$-0.25AcceptableRecovery
Staphylococcus aureus(MRSA)ATCC 700698diluted1:10MSwab2045$2.0 x 10^6$$1.6 x 10^6$$1.4 x 10^6$-0.15AcceptableRecovery
2045/1$2.1 x 10^6$$1.6 x 10^6$$1.4 x 10^6$-0.18AcceptableRecovery
2045/2$1.9 x 10^6$$1.5 x 10^6$$9.5 x 10^5$-0.30AcceptableRecovery
OrganismDilution:0.5 McFarland bacterialsuspension with salineProductLot NumberAverage of CFUs recovered at time 0 hrsAverage of CFUs recovered at time 24 hrsAverage of CFUs recovered at time 48 hrsLog10 declineInterpretation
StreptococcuspyogenesATCC 19615diluted1:10MSwab2045$3.9 \times 10^5$$2.2 \times 10^5$$1.5 \times 10^5$-0.41AcceptableRecovery
2045/1$4.8 \times 10^5$$2.1 \times 10^5$$1.5 \times 10^5$-0.51AcceptableRecovery
2045/2$4.2 \times 10^5$$2.0 \times 10^5$$1.7 \times 10^5$-0.39AcceptableRecovery
StreptococcuspneumoniaeATCC 6305diluted1:10MSwab2045$1.2 \times 10^5$$1.0 \times 10^4$$9.4 \times 10^2$-2.11AcceptableRecovery
2045/1$1.2 \times 10^5$$1.2 \times 10^4$$1.1 \times 10^3$-2.04AcceptableRecovery
2045/2$1.2 \times 10^5$$1.2 \times 10^4$$1.1 \times 10^3$-2.04AcceptableRecovery
StreptococcuspneumoniaeATCC 49136diluted1:10MSwab2045$2.0 \times 10^5$$1.5 \times 10^4$$1.6 \times 10^3$-2.10AcceptableRecovery
2045/1$1.8 \times 10^5$$1.5 \times 10^4$$1.1 \times 10^3$-2.21AcceptableRecovery
2045/2$1.9 \times 10^5$$1.7 \times 10^4$$1.4 \times 10^3$-2.13AcceptableRecovery
Enterococcus faecalisATCC 29212diluted1:10MSwab2045$1.1 \times 10^6$$1.1 \times 10^6$$7.7 \times 10^5$-0.15AcceptableRecovery
2045/1$1.1 \times 10^6$$9.6 \times 10^5$$8.1 \times 10^5$-0.13AcceptableRecovery
2045/2$1.1 \times 10^6$$9.3 \times 10^5$$8.6 \times 10^5$-0.11AcceptableRecovery
StaphylococcusepidermidisATCC 12228diluted1:10MSwab2045$1.6 \times 10^6$$1.1 \times 10^6$$5.2 \times 10^5$-0.49AcceptableRecovery
2045/1$1.7 \times 10^6$$1.1 \times 10^6$$6.8 \times 10^5$-0.40AcceptableRecovery
2045/2$1.4 \times 10^6$$9.0 \times 10^5$$5.5 \times 10^5$-0.41AcceptableRecovery
Staphylococcus aureusATCC 29213diluted1:10MSwab2045$1.3 \times 10^6$$8.7 \times 10^5$$5.2 \times 10^5$-0.40AcceptableRecovery
2045/1$1.0 \times 10^6$$8.7 \times 10^5$$5.8 \times 10^5$-0.24AcceptableRecovery
2045/2$1.0 \times 10^6$$9.9 \times 10^5$$6.0 \times 10^5$-0.22AcceptableRecovery
Streptococcus agalactiaeATCC 13813diluted1:10MSwab2045$2.1 \times 10^6$$1.5 \times 10^6$$1.3 \times 10^6$-0.21AcceptableRecovery
2045/1$1.9 \times 10^6$$1.7 \times 10^6$$1.3 \times 10^6$-0.16AcceptableRecovery
2045/2$2.2 \times 10^6$$1.4 \times 10^6$$1.0 \times 10^6$-0.34AcceptableRecovery
Kocuria rhizophilaATCC 9341diluted1:10MSwab2045$2.0 \times 10^5$$1.1 \times 10^5$$6.0 \times 10^4$-0.52AcceptableRecovery
2045/1$2.1 \times 10^5$$1.5 \times 10^5$$5.4 \times 10^4$-0.59AcceptableRecovery
2045/2$2.2 \times 10^5$$1.3 \times 10^5$$6.5 \times 10^4$-0.53AcceptableRecovery
Listeria monocytogenesATCC 19114diluted1:10MSwab2045$1.4 \times 10^6$$3.8 \times 10^5$$2.0 \times 10^5$-0.84AcceptableRecovery
2045/1$1.3 \times 10^6$$5.3 \times 10^5$$1.8 \times 10^5$-0.86AcceptableRecovery
2045/2$1.5 \times 10^6$$5.3 \times 10^5$$1.7 \times 10^5$-0.94AcceptableRecovery
Bacillus cereusATCC 10876diluted1:10MSwab2045$1.5 \times 10^5$$5.8 \times 10^4$$2.0 \times 10^3$-1.87AcceptableRecovery
2045/1$1.6 \times 10^5$$7.8 \times 10^4$$2.1 \times 10^3$-1.88AcceptableRecovery
2045/2$1.6 \times 10^5$$8.1 \times 10^4$$3.8 \times 10^3$-1.62AcceptableRecovery
Staphylococcus aureus(MRSA)ATCC 43300diluted1:10MSwab2045$2.1 \times 10^6$$1.0 \times 10^6$$5.0 \times 10^5$-0.62AcceptableRecovery
2045/1$1.9 \times 10^6$$1.4 \times 10^6$$4.2 \times 10^5$-0.65AcceptableRecovery
2045/2$1.9 \times 10^6$$9.1 \times 10^5$$4.3 \times 10^5$-0.65AcceptableRecovery
Staphylococcus aureusATCC 6538diluted1:10MSwab2045$1.5 \times 10^6$$7.6 \times 10^5$$5.1 \times 10^5$-0.47AcceptableRecovery
2045/1$1.6 \times 10^6$$1.1 \times 10^6$$6.4 \times 10^5$-0.40AcceptableRecovery
2045/2$1.7 \times 10^6$$1.0 \times 10^6$$7.9 \times 10^5$-0.33AcceptableRecovery
Staphylococcus aureus(MRSA)ATCC 700698diluted1:10MSwab2045$2.0 \times 10^6$$1.6 \times 10^6$$1.1 \times 10^6$-0.26AcceptableRecovery
2045/1$2.1 \times 10^6$$1.5 \times 10^6$$1.3 \times 10^6$-0.21AcceptableRecovery
2045/2$1.9 \times 10^6$$1.4 \times 10^6$$8.2 \times 10^5$-0.36AcceptableRecovery
OrganismDilution:0.5 McFarland bacterialsuspension with salineProductLot NumberAverage of CFUsrecovered at time0 hrsAverage of CFUsrecovered at time24 hrsAverage of CFUsrecovered at time48 hrsInterpretation
StreptococcuspyogenesATCC 19615diluted$10^{-3}$MSwab2045283.3172.3153.0AcceptableRecovery
2045/1266.7220.0169.0AcceptableRecovery
2045/2275.7240.7160.3AcceptableRecovery
StreptococcuspneumoniaeATCC 6305diluted$10^{-1.5}$MSwab2045227.3153.0111.7AcceptableRecovery
2045/1215.7171.0109.0AcceptableRecovery
2045/2217.3159.0104.3AcceptableRecovery
StreptococcuspneumoniaeATCC 49136diluted$10^{-1.5}$MSwab2045279.7206.3123.7AcceptableRecovery
2045/1256.7207.7113.0AcceptableRecovery
2045/2270.0195.7123.0AcceptableRecovery
Enterococcus faecalisATCC 29212diluted$10^{-3.5}$MSwab2045207.3172.3158.3AcceptableRecovery
2045/1214.0167.7158.3AcceptableRecovery
2045/2223.3183.3140.0AcceptableRecovery
StaphylococcusepidermidisATCC 12228diluted$10^{-3.5}$MSwab2045280.7238.7123.0AcceptableRecovery
2045/1277.0238.0113.0AcceptableRecovery
2045/2273.3151.3117.0AcceptableRecovery
Staphylococcus aureusATCC 29213diluted$10^{-3.5}$MSwab2045252.7209.3179.0AcceptableRecovery
2045/1212.7188.3160.0AcceptableRecovery
2045/2217.0188.7157.3AcceptableRecovery
Streptococcus agalactiaeATCC 13813diluted$10^{-3}$MSwab2045174.0121.0117.7AcceptableRecovery
2045/1162.3127.0100.3AcceptableRecovery
2045/2186.0148.3124.7AcceptableRecovery
Kocuria rhizophilaATCC 9341diluted$10^{-2}$MSwab2045271.7237.0212.0AcceptableRecovery
2045/1267.7223.3202.0AcceptableRecovery
2045/2269.3235.0179.0AcceptableRecovery
Listeria monocytogenesATCC 19114diluted$10^{-3}$MSwab2045267.3243.3209.7AcceptableRecovery
2045/1275.0242.0192.7AcceptableRecovery
2045/2274.3226.7198.3AcceptableRecovery
Bacillus cereusATCC 10876diluted$10^{-1.5}$MSwab2045234.7185.3167.0AcceptableRecovery
2045/1246.0182.7151.7AcceptableRecovery
2045/2231.0179.0144.7AcceptableRecovery
Staphylococcus aureus(MRSA)ATCC 43300diluted$10^{-3.5}$MSwab2045204.7185.0172.7AcceptableRecovery
2045/1211.7185.7177.3AcceptableRecovery
2045/2196.7183.7161.0AcceptableRecovery
Staphylococcus aureusATCC 6538diluted$10^{-3.5}$MSwab2045149.0123.098.7AcceptableRecovery
2045/1172.0138.0108.3AcceptableRecovery
2045/2167.3142.398.3AcceptableRecovery
Staphylococcus aureus(MRSA)ATCC 700698diluted$10^{-3.5}$MSwab2045257.0223.3204.0AcceptableRecovery
2045/1254.3229.0203.3AcceptableRecovery
2045/2257.7238.0202.7AcceptableRecovery
OrganismDilution:0.5 McFarland bacterialsuspension with salineProductLot NumberAverage of CFUs recovered at time 0 hrsAverage of CFUs recovered at time 24 hrsAverage of CFUs recovered at time 48 hrsInterpretation
Streptococcus pyogenesATCC 19615diluted$10^{-3}$MSwab2045283.3184.0113.3AcceptableRecovery
2045/1266.7196.0123.0AcceptableRecovery
2045/2275.7211.0130.0AcceptableRecovery
Streptococcus pneumoniaeATCC 6305diluted$10^{-1.5}$MSwab2045227.3131.783.0AcceptableRecovery
2045/1215.7149.373.0AcceptableRecovery
2045/2217.3137.380.0AcceptableRecovery
Streptococcus pneumoniaeATCC 49136diluted$10^{-1.5}$MSwab2045279.7148.371.0AcceptableRecovery
2045/1256.7171.081.7AcceptableRecovery
2045/2270.0171.071.0AcceptableRecovery
Enterococcus faecalisATCC 29212diluted$10^{-3.5}$MSwab2045207.3140.0117.7AcceptableRecovery
2045/1214.0140.0117.7AcceptableRecovery
2045/2223.3134.096.0AcceptableRecovery
Staphylococcus epidermidisATCC 12228diluted$10^{-3.5}$MSwab2045280.7147.089.7AcceptableRecovery
2045/1277.0154.093.3AcceptableRecovery
2045/2273.3167.788.0AcceptableRecovery
Staphylococcus aureusATCC 29213diluted$10^{-3.5}$MSwab2045252.7176.7112.0AcceptableRecovery
2045/1212.7178.3126.7AcceptableRecovery
2045/2217.0131.7107.3AcceptableRecovery
Streptococcus agalactiaeATCC 13813diluted$10^{-3}$MSwab2045174.0112.780.3AcceptableRecovery
2045/1162.3119.788.0AcceptableRecovery
2045/2186.0112.784.0AcceptableRecovery
Kocuria rhizophilaATCC 9341diluted$10^{-2}$MSwab2045271.7198.3165.3AcceptableRecovery
2045/1267.7196.3145.7AcceptableRecovery
2045/2269.3204.7142.3AcceptableRecovery
Listeria monocytogenesATCC 19114diluted$10^{-3}$MSwab2045267.3221.3167.0AcceptableRecovery
2045/1275.0221.0162.0AcceptableRecovery
2045/2274.3211.3172.0AcceptableRecovery
Bacillus cereusATCC 10876diluted$10^{-1.5}$MSwab2045234.7156.0116.0AcceptableRecovery
2045/1246.0167.0120.3AcceptableRecovery
2045/2231.0161.7117.3AcceptableRecovery
Staphylococcus aureus(MRSA)ATCC 43300diluted$10^{-3.5}$MSwab2045204.7173.7153.0AcceptableRecovery
2045/1211.7174.3142.3AcceptableRecovery
2045/2196.7174.3137.3AcceptableRecovery
Staphylococcus aureusATCC 6538diluted$10^{-3.5}$MSwab2045149.0117.072.7AcceptableRecovery
2045/1172.0113.382.0AcceptableRecovery
2045/2167.3133.386.3AcceptableRecovery
Staphylococcus aureus(MRSA)ATCC 700698diluted$10^{-3.5}$MSwab2045257.0205.0171.7AcceptableRecovery
2045/1254.3203.0176.3AcceptableRecovery
2045/2257.7205.3170.3AcceptableRecovery

·

·

· ·

:

{6}------------------------------------------------

TABLE 2. SUMMARY OF RESULTS FOR BACTERIAL RECOVERY STUDIES

SWAB ELUTION METHOD, 20-25°C

,

.

:

:

'

·

.

·

.

'

7

.

:

:

.

,

{7}------------------------------------------------

TABLE 3. SUMMARY OF RESULTS FOR BACTERIAL RECOVERY STUDIES

:

·

·

:

.

:

.

:

:

:

:

8

:

.

.

:

.

.

{8}------------------------------------------------

TABLE 4. SUMMARY OF RESULTS FOR BACTERIAL RECOVERY STUDIES

.

.

·

:

{9}------------------------------------------------

TABLE S. SUMMARY OF RESULTS FOR VIRAL RECOVERY STUDIE

OrganismDilutionProductLot NumberAverage of foci of infected cells at time 0 hrsAverage of foci of infected cells at time 24 hrsAverage of foci of infected cells at time 48 hrsLog₁₀ declineInterpretation
HSV 1ATCC VR539diluted10⁻²MSwab2045535.3302.6175.6-0.48AcceptableRecovery
HSV 1ATCC VR539diluted10⁻²MSwab2045/ 1505.7354.7195.0-0.41AcceptableRecovery
HSV 1ATCC VR539diluted10⁻²MSwab2045/ 2501.7369.3213.7-0.37AcceptableRecovery
HSV 2ATCC VR734diluted10⁻¹MSwab2045902.0526.0209.7-0.63AcceptableRecovery
HSV 2ATCC VR734diluted10⁻¹MSwab2045/ 1889.7419.0245.7-0.56AcceptableRecovery
HSV 2ATCC VR734diluted10⁻¹MSwab2045/ 2954.0486.0275.7-0.54AcceptableRecovery

TARLE 6. SUMMARY OF RESULTS FOR VIRAL RECOVERY STILDIES

OrganismDilutionProductLot NumberAverage of foci ofinfected cells attime0 hrsAverage of foci ofinfected calls attime24 hrsAverage of foci ofinfected cells attime48 hrsLogi, declineInterpretation
HSV 1dilutedATCC VR53910-52045535.3199.3142.6-0.57AcceptableRecovery
MSwab2045/ 1505.7195.7. 140.7ન્વ સ્ક્રAcceptableRecovery
2045/ 2501.7152.0103.3-0.69AcceptableRecovery
HSV 2dilutedATCC VR73410MSwab2045902.0387.7184.7-0.69AcceptableRecovery
2045/ 1889.7414.7203,7-0.64AcceptableRecovery
954.0375.3187.0-0.71AcceptableRecovery

Recovery Stability: Recovery studies were performed using aged MSwab System devices at specified time intervals up to 13 months following the date of manufacture. The bacteria and viruses selected for the stability testing were:

Streptococcus pyogenesATCC 19615
Streptococcus pneumoniaeATCC 6305
Staphylococcus aureus (Methicillin resistant)ATCC 43300
Herpes Simplex Virus Type 1 (HSV 1)ATCC VR-539
Herpes Simplex Virus Type 2 (HSV 2)ATCC VR-734

The recovery stability studies were performed at two different temperature ranges, 4-8°C and 20-25°C. Viability performance was determined for each test organism at the 48 hours time point compared to the 0 hour time point. The results from three representative lots demonstrated the ability of the MSwab System to maintain the viability of the bacterial and viral strains evaluated for at least 13 months following the date of manufacture.

{10}------------------------------------------------

pH Stability: The pH of MSwab medium was measured at specified time intervals up to 13 months following the date of manufacture. The results from three representative lots demonstrated the ability to maintain pH within the target range at all time intervals tested.

Sterilization: Copan MSwab tubes and caps are treated by gamma irradiation in accordance with UNI EN ISO 11137:2006, "Sterilization of health care products - Radiation." The tubes are filled with MSwab liquid medium aseptically under controlled conditions. Representative samples were tested to validate the medium filling process with respect to risk of microbial contamination.

The nylon flocked specimen collection swabs provided with the MSwab collection kits are individually wrapped inside a peel pouch and sterilized by ETO treatment in accordance with UNI EN ISO 11135:2007, "Sterilization of health care products - Ethylene oxide."

Biocompatibility: The nylon flocked swab component of the MSwab System was tested in accordance with ISO 10993, "Biological Evaluation of Medical Devices." The results demonstrated that the swab component is non-cytotoxic, non-irritating and non-sensitizing.

Cytotoxicity: Testing was performed to evaluate the cytotoxicity of the MSwab System using an MRC5 cell line. The results from three representative lots aged 1 month, 7 months and 13 months showed no alteration of the cell monolayers compared to the negative control.

9. CONCLUSIONS DRAWN FROM NON-CLINICAL TESTS

The non-clinical testing conducted to evaluate the Copan MSwab Collection, Transport and Preservation System demonstrated the successful performance of the device for its intended use and the substantial equivalence of the MSwab System to the cited predicate devices.

{11}------------------------------------------------

All study protocols, reports, data, and criteria were reviewed and found acceptable. The device was found to demonstrate expected performance, comparable to the predicate devices (k 061301) and (k 042970)

Conclusion

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

Reviewer Signature: Jason Heflyer Date: 05.24.2012

Management Concurrence: Saguntar Date: 5-25-12

Management Concurrence:

PILOT TRIAGE PROGRAM: OIVD, 4/11/12, v1.1

{12}------------------------------------------------

Appendix B

Statement for the Record, K 121039

This 510(k) was reviewed under OIVD's Pilot Triage Program. This program represents an internal workload management tool intended to reduce internal FDA review resources for 510(k) applications that are of good quality upon receipt by FDA.

The information in the 510(k) is complete and supports a substantial equivalence (SE) determination. Please refer to the applicant's 510(k) summary for a summary of the information that supports this SE determination.

OIVD, 4/2/12, v1.1

.
.
.
.
.

{13}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/13/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", indicating the department's name and national affiliation.

Food and Drug Administration

10903 New Hampshire Avenue Silver Spring, MD 20993

COPAN Flock Technologies c/o Cynthia Sinclair, RAC Aptiv Soutions 49 Plain Street North Attleboro, MA 02760

MAY 2 5 2012

Re: K121039

Trade/Device Name: Copan MSwab Collection, Transport and Preservation System Regulation Number: 21 CFR 866.2900

Regulation Name: Microbiological specimen collection and transport devices Regulatory Class: Class I

Product Code: JTW Dated: April 3, 2012 Received: April 5, 2012

Dear Ms. Sinclair:

We have reviewed your Section 510(k) premarket notification of intent to market the device i referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket which in the FDA finding of substantial equivalence of your device to a legally marketed

{14}------------------------------------------------

Page 2 - Cynthia Sinclair, RAC

predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Saggitgus.

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{15}------------------------------------------------

Indications for Use

510(k) Number (if known): K12 < 0 3 9

Device Name: Copan MSwab Collection, Transport and Preservation System

Indications for Use:

MSwab is a Collection, Transport and Preservation System intended for the collection and transport of clinical specimens containing Gram positive aerobic and facultative anaerobic bacteria, HSV 1 and HSV 2 from the collection site to the testing laboratory. In the laboratory, MSwab specimens are processed using standard clinical laboratory operating procedures for culture.

Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Freddie Lee Poole

vision Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K12 10.39

§ 866.2900 Microbiological specimen collection and transport device.

(a)
Identification. A microbiological specimen collection and transport device is a specimen collecting chamber intended for medical purposes to preserve the viability or integrity of microorganisms in specimens during storage of specimens after their collection and during their transport from the collecting area to the laboratory. The device may be labeled or otherwise represented as sterile. The device aids in the diagnosis of disease caused by pathogenic microorganisms.(b)
Classification. Class I (general controls). The device, when solely intended for use in the collection of concentrated parasites from specimens and transport, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9.