(15 days)
No
The document describes standard image processing techniques (attenuation correction, registration, fusion) and does not mention AI, ML, or related concepts.
No
Explanation: The device is a medical tool for detecting, localizing, and diagnosing diseases and for assessing organ function. It is used for planning, guiding, and monitoring therapy, but it does not directly administer or perform the therapy itself.
Yes
The intended use explicitly states the device is for "detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders."
No
The device description explicitly states it is a "combination of the Discovery NM630 gamma camera... and a 4 slice CT subsystem containing previously certified components." This indicates the device includes significant hardware components (gamma camera, CT subsystem, gantries, table, etc.) in addition to software.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The GE Optima NM/CT 640 system is an imaging system that uses nuclear medicine (SPECT) and computed tomography (CT) to visualize the inside of the patient's body. It detects the uptake of radioisotope tracers within the patient and uses X-rays for anatomical localization and attenuation correction.
- No Sample Analysis: The description clearly states the system detects radioisotope tracer uptake in the patient body. It does not mention analyzing samples taken from the patient's body.
Therefore, the device's function falls under the category of medical imaging, not in vitro diagnostics.
N/A
Intended Use / Indications for Use
The GE Optima NM/CT 640 system is intended for General Medicine imaging procedures for detection of Nuclear radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scanning modes supported by various acquisition types, and a CT component which is intended specifically for enabling attenuation correction and anatomical localization on SPECT studies.
The GE Optima NM/CT 640 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy.
The GE Optima NM/CT 640 system is a Nuclear Medicine (NM) system, which is intended to yield General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multiisotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include an assortment of collimators, gating by physiological signals, and real-time automatic body contouring.
The GE Optima NM/CT 640 system includes an integrated CT. which is intended specifically for attenuation correction and anatomical localization purposes only, a processing and review workstation, and may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may be used for patients of all ages.
Utilizing the associated CT portion and post processing, the device can produce attenuation corrected SPECT images using CT-based attenuation maps, as well as functional and anatomical mapping images registered with the SPECT images for purposes of localization and fusion.
The device does not support stand-alone CT operation.
Product codes
90 KPS, 90 JAK
Device Description
The Optima NM/CT 640 system is a combination of the Discovery NM630 gamma camera (K111445), the certified table from the Discovery NM/CT 670 (K093514) and a 4 slice CT subsystem containing previously certified components. It consists of two back-to-back gantries, a single table, a single power distribution unit, a console with single acquisition system and associated accessories (e.g. ECG gating. Table Extender. Head Holder, etc.). The system is delivered with a processing and review workstation (currently Xeleris 3-K093982). The system is intended to be in clinics or hospitals having normal HVAC controls.
The GE Optima NM/CT 640 system includes an integrated CT. which is intended specifically for attenuation correction and anatomical localization purposes only, a processing and review workstation, and may include signal analysis and display equipment, patient and equipment supports, components and accessories.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
SPECT, CT
Anatomical Site
Not Found
Indicated Patient Age Range
The system may be used for patients of all ages.
Intended User / Care Setting
appropriately trained healthcare professionals / clinics or hospitals
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The Optima NM/CT 640 does not require clinical studies to support substantial equivalence.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
K093514, K093982, K993645, K021491
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.1200 Emission computed tomography system.
(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.
0
K12109
APR 1 9 2012
510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
March 8, 2012 Date: Submitter: GE Healthcare, GE Medical Systems Israel, Functional Imaging 4 HAYOZMA St TIRAT HACARMEL, 30200, ISRAEL Primary Contact Person: Eli Werner Regulatory Affairs Leader GE Healthcare, GE Medical Systems Israel, Functional Imaging +972-4-8563666 +972-4-8577664 Secondary Contact Person: John Jaeckle Chief Regulatory Affairs Strategist GE Healthcare 262-424-9547 Device: Trade Name: Optima NM/CT 640 Common/Usual Name: Single Photon Emission Computed Tomography (SPECT ) & Computed Tomography X-Ray (CT) Classification Names: 21CFR 892.1200 & 21CFR 892.1750 90 KPS & 90 JAK Product Code: Predicate Device(s): K111445- Discovery NM 630 and K052434- Hawkeve 4 Option for Dual-Head Variable Angle Gamma Camera Device Description: The Optima NM/CT 640 system is a combination of the Discovery NM630 gamma camera (K111445), the certified table from the Discovery NM/CT 670 (K093514) and a 4 slice CT subsystem containing previously certified components. It consists of two back-to-back gantries, a single table, a single power distribution unit, a console with single acquisition system and associated accessories (e.g. ECG gating. Table Extender. Head Holder, etc.). The system is delivered with a processing and review workstation (currently Xeleris 3-K093982). The system is intended to be in clinics or hospitals having normal HVAC controls.
The GE Optima NM/CT 640 system is intended for General Intended Use: Medicine imaging procedures for detection of Nuclear radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scanning modes supported by various acquisition types, and a CT component which is intended specifically for enabling
DOC111487
1
attenuation correction and anatomical localization on SPECT studies.
The GE Optima NM/CT 640 system is a medical tool intended Indications for Use: for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy.
The GE Optima NM/CT 640 system is a Nuclear Medicine (NM) system, which is intended to yield General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multiisotope/multi peak frame/list mode single-photon imaging. The imaging-enhancement features include an assortment of collimators, gating by physiological signals, and real-time automatic body contouring.
The GE Optima NM/CT 640 system includes an integrated CT. which is intended specifically for attenuation correction and anatomical localization purposes only, a processing and review workstation, and may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may be used for patients of all ages.
Utilizing the associated CT portion and post processing, the device can produce attenuation corrected SPECT images using CT-based attenuation maps, as well as functional and anatomical mapping images registered with the SPECT images for purposes of localization and fusion.
The device does not support stand-alone CT operation.
The Optima NM/CT 640 utilizes the same fundamental scientific Technology: technology as its predicate devices: the Nuclear Medicine system Discovery NM630 (K111445) and the- Hawkeye 4 Option for Dual-Head Variable Angle Gamma Camera (K052434). The NM subsystem of Optima NM/CT 640 is a virtually identical NM
DOC111487
2
system to its predicate device Discovery NM630, except for the patient table, which is the same table as used on the Discovery NM/CT 670 (093514).
The Optima NM/CT 640 includes an integrated CT subsystem, which is intended specifically for attenuation correction and anatomical localization purposes only. This CT includes a previously certified tube and generator that is software limited to low mA (and hence low dose). The tube itself is the same as used on other GE CT systems such as CT/e. CT/e Dual (K993645. K021491) and Brivo CT 315/325. The CT gantry is capable of much faster rotation speeds than its Hawkeye 4 predicate and includes a new 4 row, optimized for low signal, Data Acquisition System (DAS) and detector. The CT gantry is independent of the NM gantry (however it has integrated covers). The NM- and CTsubsystems share a common patient table.
Determination of Substantial Equivalence:
The Optima NM/CT 640 and its applications are designed and independently tested to comply with recognized voluntary standards and will be certified to 21CFR performance standards as detailed in Section 9, 11 and 17 of this premarket submission. The system was design in accordance with the design controls of GE's quality system. The following quality assurance measures were applied to the development of the system:
- Risk Analysis
Requirements Reviews Design Reviews Module Testing on unit level System Integration testing Performance testing Safety testing
The Optima NM/CT 640 does not require clinical studies to support substantial equivalence.
Conclusion: GE Healthcare considers the Optima NM/CT 640 to be as safe, as effective, and with performance substantially equivalent to the predicate device(s).
DOC1111487
3
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three abstract shapes resembling human figures or birds in flight. The overall design is simple and conveys a sense of unity and progress.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
GE Medical System, Israel, Functional Imaging % Mr. Ned Devine Senior Staff Engineer Underwriters Laboratories Inc. 333 Pfingsten Road NORTBROOK IL 60062
APR 1 9 2012
Re: K121019
Trade/Device Name: Optima NM/CT 640 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS and JAK Dated: April 2, 2012 Received: April 4, 2012
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration:
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
4
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
5
INDICATIONS FOR USE FORM
510(k) Number (if known):
Device Name:
Optima NM/CT 640
Indications for Use:
The GE Optima NM/CT 640 system is a medical tool intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and in assessment of organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of turnors, planning, guiding, and monitoring therapy.
4121019
The GE Optima NM/CT 640 system is a Nuclear Medicine (NM) system, which is intended to yield General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The scanning modes include planar mode (Static, Multi-gated, Dynamic and Whole body scanning) and tomographic mode (SPECT, Gated SPECT, Whole body SPECT). The acquisition types include single and multi-isotope/multi peak frame/list mode single-photon imaging-enhancement features include an assortment of collimators, gating by physiological signals, and real-time automatic body contouring.
The GE Optima NM/CT 640 system includes an integrated CT, which is intended specifically for attenuation correction and anatomical localization purposes only, a processing and review. workstation, and may include signal analysis and display equipment, patient and equipment supports, components and accessories. The system may be used for patients of all ages.
Utilizing the associated CT portion and post processing, the device can produce attenuation corrected SPECT images using CT-based attenuation maps, as well as functional and anatomical mapping images registered with the SPECT images for purposes of localization and fusion.
The device does not support stand-alone CT operation.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence/of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign Off
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
page 1 of 1
.
DOC111487
210(K