(147 days)
The EndoFLIP system is indicated for use in a clinical setting as a pressure and dimension measurement device and as an adjunct to other methods in the comprehensive evaluation of patients with symptoms consistent with esophageal sensory hypersensitivity.
The EndoFLIP® system and catheter are identical to the predicate K092850, except the addition of the new software which permits a programmed series of inflations of the balloon in steps of volume. This feature is equivalent to a modality offered by the predicate Barostat (K991288).
The device is not intended to perform a diagnostic test.
The provided text is a 510(k) summary for the EndoFLIP® Barostat Mode device. It describes the device, its indications for use, and a comparison to predicate devices, but it does not contain information about specific acceptance criteria or a study proving the device meets those criteria, nor any performance data resulting from such a study.
The document states:
- "No comparative bench testing was required as the change is a software change only."
- "The proposed software modification does not change the performance specification of the EndoFLIP®, which remains identical to the predicate K092850."
Therefore, I cannot fulfill your request for the following information based on the provided text:
- A table of acceptance criteria and the reported device performance
- Sample sized used for the test set and the data provenance
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Adjudication method for the test set
- If a multi reader multi case (MRMC) comparative effectiveness study was done, and the effect size
- If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- The sample size for the training set
- How the ground truth for the training set was established
The document focuses on demonstrating substantial equivalence to predicate devices based on indications for use, environment of use, patient population, technology/design, and materials, asserting that the software change does not alter performance specifications.
{0}------------------------------------------------
510(k) Summary Page I of 4 20-Aug-12
AUG 27 2012
| Company | Crospon Ltd.Galway Business ParkDanganGalway, Ireland |
|---|---|
| Official Contact: | John O'Dea PhD |
| Proprietary or Trade Name: | EndoFLIP® Barostat Mode |
| Common/Usual Name: | Gastrointestinal motility monitoring system |
| Classification Name: | FFX - system, gastrointestinal motility (electrical)CFR 876.1725Class 2 |
| Device: | EndoFLIP System and Catheter |
| Predicate Devices: | K991288 - G&J Electronics, Distender Series IIBarostatK092850 - EndoFLIP® systems - Crospon Ltd. |
Device Description:
The EndoFLIP® system and catheter are identical to the predicate K092850, except the addition of the new software which permits a programmed series of inflations of the balloon in steps of volume. This feature is equivalent to a modality offered by the predicate Barostat (K991288).
The device is not intended to perform a diagnostic test.
Indications for Use:
The EndoFLIP system is indicated for use in a clinical setting as a pressure and dimension measurement device and as an adjunct to other methods in the comprehensive evaluation of patients with symptoms consistent with esophageal sensory hypersensitivity.
Patient Population:
Patients with esophageal disorders.
Environment of Use:
Hospitals, Physician offices.
Contraindications:
The EndoFLIP® System is contraindicated where endoscopy is contraindicated.
The EndoFLIP System is contraindicated in patients with an increased risk of esophageal perforation or bleeding.
{1}------------------------------------------------
| 0(k) Summa Page 2 of 4 20-Aug-12lS |
|---|
| ------------------------------------------------ |
FLAVOR PROFILE
- COMPLEX, INTENSE, PRONOUNCED
- DARK CHOCOLATE, BROWN SUGAR
- FULL BODIED, SMOOTH
- DEVICA IS STANDARD
| dications for | |||
|---|---|---|---|
| ក្រសួរសង្គម ស្រុក ស្រុក ស្រុក និង ភូមិ | inders III MBAA The Times March As a Previous and Caroles Andresses Marchiller March Asia Previous and Career Andress Normalians West National Provinces Supersions Sons | ||
| natomical Si | Stomac | Esophagi | |
| ospitals. Physician offices tients with esophageal disord | |||
| wironments of tient Populatio | |||
| spitals. Physician offices Reflection Resolução Brand Lista em Rossing Brasilia Band Leathernia Single System Band Norther Threat System Transmille Hurai | |||
| ntraindicati | Not stated | he EndoFLIP® System is ontraindicated where endosed ontraindicated. | |
| EndoFLIP System is contraindi tients with an increased risk of | |||
| escriptio | scription use | ||
| inciple of Operat | Presseription use Pressure or isovolumetric volume within pressure or isovolumetric volume within balloon. | catheter is positioned in the economics for any within and within are within are and with bearling the bearling the best and the best and the best and the bested to be set to | |
| ecords volume and pressure charing the protocol. | |||
{2}------------------------------------------------
510(k) Summary
| Distender Series II Barostat(G&J Electronics) K991288 | EndoFLIP®Crospon (K092850) | EndoFLIP®Barostat Mode | |
|---|---|---|---|
| Principle of Operation | Pressure sensor in the device is connected tothe balloon via a measurement line. | Pressure sensor in the catheterballoon. Dest electrodes are located at16 points along the catheter inside aballoon. The balloon is placed suchthat its midpoint is located in a stomawithin a stomach fitted with a gastricband | Pressure sensor in the catheterballoon. Dest electrodes are located at16 points along the catheter inside aballoon. The balloon is placed suchthat its midpoint is located in theesophagus. |
| Data Recording | 4 channels of data can be recorded on a chartrecorder | Data is saved internally and can beexported to a USB mass storagedevice or printed to an attached USBprinter | Data is saved internally and can beexported to a USB mass storagedevice or printed to an attached USBprinter |
| Electrical SafetyBiocompatibility | Not stated | IEC60601-1 2nd Ed. + Am.1 + Am.2All materials have passedbiocompatibility tests in accordancewith ISO 10993-1 | IEC60601-1 2nd Ed. + Am.1 + Am.2Identical to K092850 |
| Compatibility With TheEnvironment And OtherDevices | User sources balloons and tubing. | User sources balloons and tubing. | EndoFLIP operates with customcatheters only. |
| Performance | Volume range and resolution is dependent onthe cylinder size used. Range available is 25mlto 1200ml. Resolution ranges from a pistonstep size of 0.181ml to 0.804ml | EndoFLIP operates with customcatheters only. | Balloon volume is controlled andvolume is displayed:Range: 0 to 50 ml.Resolution: 1 mL |
| Balloon volume is controlled andvolume is displayed:Range: 0 to 50 ml.Resolution: 1 mL |
{3}------------------------------------------------
510(k) Summarv Page 4 of 4 20-Aug-12
Substantial Equivalence
The EndoFLIP® system in Barostat.mode is viewed as substantially equivalent to the predicate devices presented above. In summary we have found that the following key elements support a determination of substantially equivalent:
Indications -
The proposed indications for use are nearly identical to the predicate, Barostat K991288.
The EndoFLIP system is indicated for use in a clinical setting as a pressure and dimension measurement device and as an adjunct to other methods in the comprehensive evaluation of patients with symptoms consistent with esophageal sensory hypersensitivity vs. the predicate Distender Series II Dual Drive Barostat device which is used for volume/pressure measurement in the alimentary tract as an adjunct to other diagnostic methods as part of a comprehensive evaluation of patients with symptoms consistent with gastrointestinal motility disorders.
Note: EndoFLIP is a measurement system. It is not intended to perform a diagnostic test.
Environment of Use -
The proposed environments of use are identical to the predicates. EndoFLIP® K092850 and Barostat (K991288), which are for Hospitals and Physician offices.
Patient Population -
The proposed patient population is patients with esophageal disorders whereas the predicate Barostat (K991288) lists patients with symptoms consistent with gastrointestinal motility disorders, which are equivalent.
Technology / Design / Features -
The technology is identical to the predicate EndoFLIP® systems (K092850) and we have only added a new software feature which is the Barostat mode, which allows programmed volume distentions, each volume distention being identical in nature to the single volume distention permitted using the existing software.
Materials -
The materials are identical to the predicate EndoFLIP® K092850 and thus not applicable as the modification is software only.
Performance Specifications -
The proposed software modification does not change the performance specification of the EndoFLIP®, which remains identical to the predicate K092850.
Performance Testing - Bench.
No comparative bench testing was required as the change is a software change only.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the department's name around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that suggest movement and progress.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Crospon, Ltd. % Mr. Paul Dryden President ProMedic, Inc. 24301 Woodsage Drive BONITA SPRINGS FL 34134
AUG 27 2012
Re: K120997
Trade/Device Name: EndoFLIP® Barostat Regulation Number: 21 CFR§ 876.1725 Regulation Name: Gastrointestinal motility monitoring system Regulatory Class: II Product Code: FFX Dated: August 20, 2012 Received: August 21, 2012
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set de rior thator and to systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part now are regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerély yours,
Sincerely yours,
Benjamin K. Fisher
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use Statement
Page 1 of 1
510(k) Number: K120997
Device Name:
EndoFLIP® Barostat
Indications for Use:
The EndoFLIP system is indicated for use in a clinical setting as a pressure and dimension measurement device and as an adjunct to other methods in the comprehensive evaluation of patients with symptoms consistent with esophageal sensory hypersensitivity.
Prescription Use XX (Part 21 CFR 801 Subpart D) or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Halur Rem
ision Sign-Off) of Reproductive, Gastro-Renal, xical De (k) Numbe
§ 876.1725 Gastrointestinal motility monitoring system.
(a)
Identification. A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.(b)
Classification. Class II (performance standards).