(70 days)
Not Found
No
The description details a standard enzymatic assay for glucose measurement based on absorbance readings and chemical reactions. There is no mention of AI/ML algorithms for data analysis, interpretation, or decision-making.
No.
This device is for in vitro diagnostic use to measure glucose levels, which aids in diagnosis and treatment monitoring, but it does not directly provide therapy.
Yes
This device is for "in vitro diagnostic use in the quantitative determination of glucose" and its measurements are "used in the diagnosis and treatment of carbohydrate metabolism disorders," explicitly indicating its diagnostic purpose.
No
The device description clearly outlines a two-component reagent system and absorbance readings, indicating a chemical assay that requires physical reagents and a chemistry analyzer, not solely software.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use/Indications for Use: The very first section explicitly states "For in vitro diagnostic use". It also describes the purpose of the device as the "quantitative determination of glucose in human serum, plasma, urine and CSF", which are all biological samples analyzed in vitro (outside the body). The stated use in the "diagnosis and treatment of carbohydrate metabolism disorders" further confirms its diagnostic purpose.
- Device Description: The description details a chemical method using reagents to analyze a sample, which is characteristic of an in vitro diagnostic test.
The information provided clearly aligns with the definition of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
For in vitro diagnostic use in the quantitative determination of glucose in human serum, plasma, urine and CSF on the ADVIA Chemistry systems. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and insulin overdose.
Product codes
CFR
Device Description
The ADVIA Chemistry Glucose Hexokinase 3 (GLUH 3) method uses a two-component reagent. Sample is added to Reagent 1, which contains the buffer, ATP, and NAD. Absorbance readings of the sample in Reagent 1 are taken and are used to correct for interfering substances in the sample. Reagent 2 is added, which initiates the conversion of glucose and the development of absorbance at 340/410 nm. The difference between the absorbance in Reagent 1 and Reagent 2 is proportional to the glucose concentration.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Matrix comparison: 56 matched serum and plasma (Sodium Fluoride/Potassium Oxalate) samples were evaluated. Some of these samples were spiked and/or diluted. There were a total of 82 data-points included in calculations. Matrix-Comparison between serum and Na FI plasma tubes using the ADVIA 1650 Chemistry GLUH_3 assay gave the following correlation statistics using the linear regression calculations: x - serum, y -Sodium Fluoride/Potassium Oxalate plasma. Regression Equation: y=1.011 * x + 0.8. Slope 95% CI: 1.003 to 1.019. r=0.999. Sample Range: 2-601. Intercept 95% Cl: -1.18 to 2.73.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
0
510(k) Summary
MAY 1 5 2012
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| Submitter's Name: | Kira Gordon
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave,
Tarrytown, NY 10591 | | | | |
|--------------------------|------------------------------------------------------------------------------------------------|--|--|--|--|
| Date of Preparation: | May 10, 2012 | | | | |
| Name of Product: | ADVIA® Chemistry Glucose Hexokinase_3 (GLUH_3) reagen | | | | |
| FDA Classification Name: | Glucose test system. | | | | |
Predicate Device:
The following table describes the predicate devices, device classification, regulation and product code associated with this pre-market notification:
New Product | Predicate Device | 510(k) number | Device Class | Regulation | Product Code |
---|---|---|---|---|---|
ADVIA® | |||||
Chemistry | |||||
GLUH_3 Reagent | |||||
(additional | |||||
plasma type) | ADVIA® | ||||
Chemistry | |||||
GLUH_3 Reagent | K101854 | Class II | 862.1345 | CFR |
Device Description: 1
The ADVIA Chemistry Glucose Hexokinase 3 (GLUH 3) method uses a two-component reagent. Sample is added to Reagent 1, which contains the buffer, ATP, and NAD. Absorbance readings of the sample in Reagent 1 are taken and are used to correct for interfering substances in the sample. Reagent 2 is added, which initiates the conversion of glucose and the development of absorbance at 340/410 nm. The difference between the absorbance in Reagent 1 and Reagent 2 is proportional to the glucose concentration.
Intended Use:
For in vitro diagnostic use in the quantitative determination of glucose in human serum, plasma, urine and CSF on the ADVIA Chemistry systems. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and insulin overdose.
1
Comparison to Predicate Device:
A comparison of the important features of the devices are provided in the following table:
Item | New Device | Predicate Device |
---|---|---|
Analyte | Glucose | Same |
Intended Use | For in vitro diagnostic use in the | |
quantitative determination of glucose in | ||
human serum, plasma, urine and CSF on | ||
the ADVIA Chemistry systems. Such | ||
measurements are used in the diagnosis | ||
and treatment of carbohydrate | ||
metabolism disorders including diabetes | ||
mellitus, neonatal hypoglycemia, | ||
idiopathic hypoglycemia, and insulin | ||
overdose | Same | |
Sample type | human serum, plasma, urine and CSF | Same |
Plasma type | Li-Heparin, K-EDTA and | |
Na-Fluoride/K-Oxalate | Li-Heparin, K-EDTA | |
Instrument to | ||
be used | ADVIA Chemistry | Same |
Method | ||
Principle | based on the method by Slein using | |
hexokinase and glucose-6-phosphate | ||
dehydrogenase enzymes. | Same | |
Calibrators | Siemens Healthcare Diagnostics | |
Chemistry Calibrator REF 09784096 | Same | |
Assay range | 4 -700 mg/dL | Same |
Accuracy / | ||
Correlation | Serum: | |
$Y = 1.001x + 0.3$ ; N=99 r=1.000 | ||
Plasma (Li-Heparin): | ||
$Y = 1.001x + 0.2$ , N=88; r=1.000 | ||
Plasma (K-EDTA) | ||
$Y = 1.002x - 0.0$ , N=87; r=1.000 | ||
Plasma (Na Fluoride/Potassium Oxalate) | ||
$Y = 1.011x + 0.8$ , N=82; r=0.999 | ||
CSF: | ||
$Y = 1.005x - 0.1$ , N=113; r=1.000 | ||
Urine: | Serum: | |
Same | ||
Plasma (Li-Heparin): | ||
same | ||
Plasma (K-EDTA) | ||
Same | ||
Plasma (Na Fluoride/Potassium Oxalate) | ||
none | ||
CSF: | ||
Same | ||
Urine: | ||
Same |
Comments on Substantial Equivalence:
- Analytical performance:
- a. Precision/Reproducibility:
Not applicable for this modification.
b. Linearity/assay reportable range:
Not applicable for this modification.
- c. Traceability (controls, calibrators, or method): Not applicable for this modification.
2
d. Detection limit (functional sensitivity):
Not applicable for this modification.
-
e. Analytical specificity:
Not applicable for this modification. -
f. Assay cut-off:
Not applicable.
2. Comparison studies: ·
a. Method comparison with predicate device:
Not applicable for this modification.
b. Matrix comparison: .
56 matched serum and plasma (Sodium Fluoride/Potassium Oxalate) samples were evaluated. Some of these samples were spiked and/or diluted. There were a total of 82 data-points included in calculations. Matrix
- Comparison between serum and Na FI plasma tubes using the ADVIA 1650 Chemistry GLUH_3 assay gave the following correlation statistics using the
linear regression calculations:
x - serum, y -Sodium Fluoride/Potassium Oxalate plasma
Regression Equation | Sample Range | ||
---|---|---|---|
v=1.011 * x + 0.8 | |||
Slope 95% CI: 1.003 to 1.019 | 6.13 | 0.999 | ર-601 |
Intercept 95% Cl: -1.18 to 2.73 |
- Clinical studies:
a. Clinical sensitivity:
Not applicable for this submission.
b. Clinical specificity:
Not applicable for this submission.
-
- Clinical cut-off:
Not applicable for this submission.
- Clinical cut-off:
- Expected values/Reference range:
Not applicable for this modification.
Conclusion:
Comparative testing of the ADVIA 1650 Chemistry Glucose Hexokinase 3 reagent using Sodium Fluoride/Potassium Oxalate plasma type demonstrates substantially equivalent performance to the GLUH_3 reagent using serum as a plasma type cleared under K101854.
Kira Gordon Regulatory Affairs & Compliance May 10, 2012
3
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces left. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in all capital letters.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration
10903 New Hampshire Avenue Silver Spring, MD 20993
Siemens Healthcare Diagnostics Inc. c/o Kira Gordon 511 Benedict Ave Tarrytown, N.Y. 10509
MAY 1 5 2012
Re: K120681
Trade Name: Advia® Chemistry GLUH 3 Reagent Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Codes: CFR Dated: March 2, 2012 Received: March 6, 2012
Dear Ms Gordon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
4
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical
Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm
Sincerely yours,
N. Kim, Ph.D.
Couriney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
5
Indication for Use
510(k) Number (if known):
Device Name:
ADVIA® Chemistry GLUH 3 Reagent
Indication For Use:
For in vitro diagnostic use in the quantitative determination of glucose in human serum, plasma, urine and CSF on the ADVIA Chemistry systems. Such measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, idiopathic hypoglycemia, and insulin overdose.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use _ (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) R 120681
Page 1 of 1