K Number
K120675
Date Cleared
2013-02-14

(346 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Digital Automatic Blood Pressure Monitor BPM11 Series is for use by medical professional or at home. The BPM11 Series is intended to measure the systolic and diastolic blood pressure, and pulse rate (heartbeat rate) of an individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the upper arm of an individual. The inflatable cuff circumference is limited to 17cm - 44cm via 3 different size of cuff. 3 different cuff sizes are 17-22cm, 22-32cm and 32-44cm.

Device Description

Digital Automatic Blood Pressure Monitor BPM11 Series is a non-invasive blood pressure measurement system for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate (heartbeat rate) of an individual in each measurement and then displays the readings on a digital panel. The BPM11 Series utilizes the oscillometric methodology, in which an inflatable cuff is wrapped around the upper arm of an individual, for the blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movements into a digital reading.

AI/ML Overview

Here's a summary of the acceptance criteria and study information for the Digital Automatic Blood Pressure Monitor BPM11 Series, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
ANSI/AAMI SP-10-2002 requirements"The results demonstrated that BPM11 Series meets the requirement of ANSI/AAMI SP-10-2002."

2. Sample Size for Test Set and Data Provenance:

The document does not explicitly state the sample size of the test set or the country of origin of the data. It only mentions that "A systematic & independent clinical test was conducted." It also doesn't specify if the study was retrospective or prospective.

3. Number of Experts and Qualifications for Ground Truth of Test Set:

The document does not specify the number of experts used or their qualifications for establishing the ground truth for the clinical test.

4. Adjudication Method for the Test Set:

The document does not describe the adjudication method used for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool that would typically involve human readers.

6. Standalone (Algorithm Only) Performance:

Yes, a standalone performance study was conducted. The clinical test evaluated the performance of the device itself (the algorithm/hardware combined) in measuring blood pressure and pulse rate, independent of human interpretation or assistance. The device operates automatically.

7. Type of Ground Truth Used:

The type of ground truth used is not explicitly stated, but for blood pressure monitor validation against a standard like ANSI/AAMI SP-10, the ground truth is typically established by simultaneous measurements using a reference method (e.g., a mercury sphygmomanometer with auscultation by trained observers), compared against the device's readings.

8. Sample Size for the Training Set:

The document does not mention a training set or any details about it. This suggests that the device's algorithm might be based on established physiological principles and calibrated, rather than being a machine learning model that requires a distinct training phase with labeled data in the way a diagnostic AI typically would.

9. How Ground Truth for the Training Set Was Established:

As no training set is mentioned, information on how its ground truth was established is not provided.

{0}------------------------------------------------

Section IV: 510k Summary

Applicant's Identification

  • Applicant Phone Number Fax Number Contact Person Date of Application
    Grandway Technology (Shenzhen) Limited (00852)-2851-6789 (00852)-2851-6278 Mr. Patrick Chow 24th Feb, 2012

Device's Identification

Device Proprietary Name

Digital Automatic Blood Pressure Monitor BPM11 Series

Model No .: MD11xy

x --- The first character (0, 1, 3, 5, 6 or 8) is for the minor change revision of device. The mentioned "minor change" refers to those device changes not to be affected the conformity test results of EMC & safety as well as device performance, i.e. IEC 60601-1 and EN 60601-1-2. y --- The second character (which is always 0) is for the

identification of cabinet (housing).

Common Name Classification Name

Non-invasive Blood Pressure Measurement System Non-invasive Blood Pressure Measurement System (Class II per 21 CFR 870.1130)

Marketed Devices to which Equivalence is Claimed

DEVICEMANUFACTURER510(k) Number
AVITA BPM6 Series Blood Pressure Meter(or Monitor)AVITA CorpK033397

Device Description

Digital Automatic Blood Pressure Monitor BPM11 Series is a non-invasive blood pressure measurement system for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate (heartbeat rate) of an individual in each measurement and then displays the readings on a digital panel.

FEB 1 4 2013

1/3

{1}------------------------------------------------

The BPM11 Series utilizes the oscillometric methodology, in which an inflatable cuff is wrapped around the upper arm of an individual, for the blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movements into a digital reading.

Intended Use (Indication for Use)

Digital Automatic Blood Pressure Monitor BPM11 Series is for use by medical professional or at home. The BPM11 Series is intended to measure the systolic and diastolic blood pressure, and pulse rate (heartbeat rate) of an individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the upper arm of an individual. The inflatable cuff circumference is limited to 17cm - 44cm via 3 different size of cuff. 3 different cuff sizes are 17-22cm, 22-32cm and 32-44cm.

Comparison of Technological Characteristics between New Device and Predicate Devices

The Digital Automatic Blood Pressure Monitor BPM11 Series is a non-invasive measuring device and utilizes the oscilliometric methodology to measure the blood pressure reading. The key components of device are a pressure sensor, electric valve and an electronic control module together with an electric pump, which inflate (and deflate) the inflatable cuff automatically according to our designed architecture. The predicate device adopts exactly same methodology and key components for measuring the blood pressure reading.

Clinical & Non-clinical Tests

A systematic & independent clinical test was conducted to validate the performance of the Digital Automatic Blood Pressure Monitor BPM11 Series. The results demonstrated that BPM11 Series meets the requirement of ANSI/AAMI SP-10-2002.

Comprehensive safety and EMC tests were performed and compiled to demonstrate BPM11 Series is safe for use. Tests include EN1060-1:1995/A2:2009 EN1060-3:1997/A2:2009 IEC60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007) EN60601-1-2:2007 FCC Part 15 ISO10993-5:2009 ISO10993-10:2002 + A1:2006 EN60601-1-4:2007

Conclusion

{2}------------------------------------------------

Digital Automatic Blood Pressure Monitor BPM11 Series has the same intended use and similar technological characteristics as predicate device (K033397). Moreover, bench testing contained in this submission and clinical testing supplied demonstrated that no differences in the technological characteristics and questioning on safety or effectiveness to be raised. Thus, the Digital Automatic Blood Pressure Monitor BPM11 Series is substantially equivalent to the predicate device.

13

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.

Public Health Service

February 14, 2013

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

Grandway Technology (Shenzhen) Limited c/o Mr. Patrick Chow, General Manager Block 6 and 7, Zhu Keng Industrial Zone, Ping Shan Long Gang District Shenzhen, Guang Dong People's Republic of China

Re: K120675

Trade/Device Names: Digital Automatic Blood Pressure Monitor with models MD1100, MD1110, MD1130, MD1150, MD1160 and MD1180 Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: February 6, 2013 Received: February 8, 2013

Dear Mr. Chow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Patrick Chow

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Owen P. Faris -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Section V: Statement of Indications for Use

510(k) Number (if known): K120675

Device Name:

Digital Automatic Blood Pressure Monitor BPM11Series

Model No.: MD11xy

x --- The first character (0, 1, 3, 5, 6 or 8) is for the minor change revision of device. The mentioned " minor change" refers to those device changes not to be affected the conformity test results of EMC & safety as well as device performance, i.e. IEC 60601-1 and EN 60601-1-2.

v --- The second character (which is always 0) is for the identification of cabinet (housing).

Indications For Use:

Digital Automatic Blood Pressure Monitor BPM11 Series is for use by medical professional or at home. The BPM11 Series is intended to measure the systelic and diastolic blood pressure, and pulse rate (heartbeat rate) of an individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the upper arm of an individual. The inflatable cuff circumference is limited to 17cm-44cm via 3 different size of cuff. 3 different cuff sizes are 17-22cm, 22-32cm and 32-44cm.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21CFR 801Subpart D) AND/OR

Over-The-Counter Use X (21CFR 801Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/5/Picture/15 description: The image shows the text "Owen P. Faris -S" in a bold, sans-serif font. The letters "DA" are stylized with a geometric, three-dimensional design. The text is horizontally aligned and appears to be a name or title.

Page 1 of 1

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).