K Number
K033397

Validate with FDA (Live)

Manufacturer
Date Cleared
2004-04-09

(168 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is noninvasive and provides systolic, diastolic blood pressure and pulse rate measurements by using an cuff which is wrapped around the arm. All values can be read out in one LCD panel. Measurement is for adult only.

Device Description

AVITA BPM6 series Blood Pressure Meter(or Monitor) is designed to measure the systolic and diastolic blood pressure, and pulse rate(heart of an individual).

AI/ML Overview

The provided document is a 510(k) summary for the AVITA BPM6 series Blood Pressure Meter. It lacks detailed information on specific acceptance criteria and a dedicated study report proving those criteria were met, as would be expected for a comprehensive clinical study. However, some pieces of information can be inferred or extracted.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Clinical Repeatability (compared to auscultatory method)"statistically and clinically acceptable"
Conformance to EN-1060-1Conforms
Conformance to EN-1060-3Conforms
Conformance to ANSI/AAMI SP-10Conforms
Conformance to IEC 60601-1Conforms
Conformance to IEC 60601-1-2Conforms

Note: The document only provides a high-level statement about "statistically and clinically acceptable" repeatability without specific numerical thresholds or results. ANSI/AAMI SP-10 defines specific accuracy requirements for automated sphygmomanometers, which would constitute the numerical acceptance criteria for blood pressure accuracy. While the document states conformance, it does not explicitly list the detailed results demonstrating this conformance.

2. Sample Size Used for the Test Set and Data Provenance

The document mentions a "comparison study with device that use auscultatory method" but does not provide any details regarding the sample size used for this clinical comparison or the provenance (country of origin, retrospective/prospective nature) of the data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document states a comparison study was performed using the "auscultatory method." This method typically involves trained healthcare professionals (e.g., clinicians, nurses) using a stethoscope and sphygmomanometer to measure blood pressure. However, the document does not specify the number or qualifications of these experts who established the ground truth.

4. Adjudication Method for the Test Set

The document does not provide any information about an adjudication method. For a comparison study based on the auscultatory method, the ground truth is often established by a single trained observer. If multiple observers were used, a method like averaging or consensus might be employed, but this is not mentioned.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No. The document describes a "comparison study" with the auscultatory method, which is a method for generating ground truth for device performance, not an MRMC study comparing human readers with and without AI assistance. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Yes, in essence. The "comparison study with device that use auscultatory method" evaluated the performance of the AVITA BPM6 series Blood Pressure Meter itself (the algorithm/device) against a accepted reference standard (auscultatory method). This constitutes a standalone performance evaluation of the device's accuracy in measuring blood pressure.

7. The Type of Ground Truth Used

The ground truth used was based on the auscultatory method, which is considered a clinical reference standard for blood pressure measurement. This method involves direct observation by a trained human using a stethoscope and sphygmomanometer.

8. The Sample Size for the Training Set

The document does not specify a training set sample size. Given that this is a 510(k) for a blood pressure monitor, the focus is on demonstrating equivalence to an existing device and conformance to standards, rather than the development of a complex machine learning algorithm that would require a distinct training set. The device likely relies on a well-established oscillometric algorithm, not one trained on a large, novel dataset in the context of this submission.

9. How the Ground Truth for the Training Set Was Established

As no specific training set is mentioned in the context of this 510(k) submission, there is no information provided on how ground truth for such a set would have been established. Any underlying algorithm development would have occurred prior to this submission and is not detailed here.

{0}------------------------------------------------

APR - 9 2004

510(K) SUMMARY

K033397 p112-

This summary of 510(k) safety and effectiveness information is being submitted in accordance in with the requirements of SMDA and 21 CFR §807.92

1.0Submitter's Name:AVITA Corp
Address:9F, No. 78, Sec. 1, Kwang-Fu Rd., San-Chung, Taipei County,Taiwan, R.O.C.
Phone:001-886-2-85121568
Fax:001-886-2-85121347
Contact:Mr. Geo Lin, General Manager
2.0Device Name:AVITA BPM6 series Blood Pressure Meter(or Monitor)
Model No.:BPM6XY (X= 0 or 1 , Y=1,2 or 3)The first Character (X= 0 or 1 ) represents for type of case housing.The second Character (Y=1,2 or 3 ) represents for type of product specification.
3.0Classification:Class II
4.0Predicate Device:AVITA BPM6 series Blood Pressure Meter(or Monitor) has similar general design with OMRON HEM-757 Blood PressureMonitor(K001670) marketed by Omron Healthcare, Inc..
5.0Device Description:AVITA BPM6 series Blood Pressure Meter(or Monitor) is designed to measure the systolic and diastolic blood pressure, and pulse rate(heart of an individual).
6.0Intended Use:AVITA BPM6 series Blood Pressure Meter is intended to measure human begins Systolic, Diastolic blood pressure and heart rate using the oscillometric method. All values can be read out in one LCD DISPLAY. Measurement position is on adult arm only.
7.0Performance Summary:In terms of operating specification, Safety & EMC requirements, the device conforms to applicable standards included EN-1060-1, EN-1060-3, ANSI/AAMI SP-10, IEC 60601-1 and IEC 60601-1-2 requirements. A comparison study with device that use auscultatory method was performed to validate the performance of the AVITA BPM6 series Blood Pressure Meter. The comparison study demonstrated that the clinical repeatability of AVITA BPM6 series Blood Pressure Meter is statistically and clinically acceptable.

$\frac{cv}{\Pi}$

JK2

{1}------------------------------------------------

Conclusions: 8.

o. "Conclusione:
The AVITA BPM6 series Blood Pressure Meter have the same intended use and similar mle AVTA Blance 21000 Bress 21000
technological characteristics as OMRON HEM-757 Blood Pressure Monitor(K001670) technological characteristions as Chitter Contained in this submission and mailkeled by Official realthours, mon more connological characteristics do
clinical testing supplied demonstrate that any differences in the same of the st cifical testing supplied demonstrate that any and the AVITA BPM6 series Blood not raiso and not is substantially equivalent to the predicate devices.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized depiction of an eagle or bird-like figure with three wing-like strokes.

Public Health Service

APR - 9 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AVITA Corporation c/o Ms. Jennifer Reich Harvest Consulting Corp. 3892 South America West Trail Flagstaff, AZ 86001

Re: K033397

Trade Name: AViTA BPM6 Series Blood Pressure Meter (Monitor) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Received: March 02, 2004

Dear Ms. Reich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Ms. Jennifer Reich

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Dana R. Lo Jones

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 4 - Ms. Jennifer Reich

Indications for Use

510(k) Number: K033397

Device Name: Avita BPM6 Series Blood Pressure Meter (or Monitor)

Indications For Use:

The device is noninvasive and provides systolic, diastolic blood pressure and pulse rate measurements by using an cuff which is wrapped around the arm. All values can be read out in one LCD panel. Measurement is for adult only.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use V . (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Duna R. Lochner

(Division 3) - Jil) Division of Cardiovascular Devices

510(k) Number_K033399

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).