K Number
K120569
Date Cleared
2012-07-20

(144 days)

Product Code
Regulation Number
882.5890
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GEMORE TENS System, model GM310TP/GM320TP/GM330TP/GM340TP Transcutaneous Electrical Nerve Stimulators are indicated for:

  • symptomatic relief of chronic, intractable pain .
  • adjunctive treatment for post-surgical and post-traumatic acute pain
  • relief of pain associated with osteoarthritis of the knee and rheumatoid arthritis of the hand
Device Description

The Gemore TENS System / Model GM310TP, GM320TP, GM330TP, and GM340TP transcutaneous electrical nerve stimulators for pain relief, are the electrically power devices used to apply an electrical current to electrode on a patient's skin to relieve pain. This series of TENS devices are indicated for:

  • symptomatic relief of chronic, intractable pain
  • . adjunctive treatment for post-surgical and post-traumatic acute pain
    . relief of pain associated with osteoarthritis of the knee and rheumatoid arthritis of the hand
    For this definition of intended use, the Gemore TENS system is classified with the "NYN" & "GZJ" product code under 21CFR/Part 882.5890 FDA regulation classification number.
    The Gemore TENS system consists mainly of two parts: the stimulus generator and electrode. The stimulus generator generates the output current specified as the input of controller. The output of controller is able to be adjusted for B/C/M/P1/P2/P3/P4/P5 eight operation modes which bear different combination for the output parameters of pulse width, pulse rate, and pulse cycle. For every operation mode, the operator could adjust output intensity, pulse rate and operation time. The output parts transmit the output current to the electrodes which is attached to the patient's skin so as to .transmit the generated stimulus to patient for pain relief treatment.
    GM310TP/GM320TP/GM330TP/GM340TP is completely identical to that of model as mentioned in our previous submission of GEM STIM TENS, model GM310TE/ GM320TE/GM330TE/GM340TE (K032720). To change the indication for use as that of the chosen predicate devices, Gemore changed the software for these four models.
    For the device included in this submission, we use the following of our 510(K) legally marketed predicate garment electrodes:
  • <1>K062675, "Gemore Reuseable Self Adhesive Electrode" ; Wire Series/ Model FA5050 ( 5x5 CM or bigger size) self adhesive electrode.
    GM310TP/GM320TP/GM330TP/GM340TP is a selectable dual channel, 9V battery operated TENS device with the following features:
  • <1> The operation function is dual channels completely identical to the model being modified, GM310TE/GM320TE/GM330TE/GM340TE (K032720).
  • <2> The output of controller is able to be adjusted for B/C/M/P 1/P2/P3/P4/P5 eight operation modes which bear different combination for the output parameters of pulse width, pulse rate, and pulse cvcle.
  • <3> The output strength is adjustable at 0~80 mA, with maximum setting time 60 minutes counting from switching ON.
  • <4> The LCD display is provided for the indication of operation status including CH1/CH2, output program mode, pulse width, pulse rate, timer, and battery low warning
AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the GEMORE TENS System:

Summary of Device Performance and Acceptance Criteria

The submission for the GEMORE TENS System (models GM310TP/GM320TP/GM330TP/GM340TP) is a 510(k) premarket notification, which seeks to demonstrate substantial equivalence to a legally marketed predicate device (Empi - SELECT TENS System, K061650). This type of submission generally does not involve a clinical study to directly prove efficacy against acceptance criteria in the same way a de novo or PMA submission might. Instead, it focuses on demonstrating that the new device is as safe and effective as the predicate device by meeting similar technological characteristics and performance specifications.

The "acceptance criteria" here are implicitly defined by the performance characteristics of the predicate device and relevant voluntary standards. The "study" proving the device meets these criteria is a series of non-clinical tests and comparisons.

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a substantial equivalence claim based on non-clinical testing and comparison to a predicate, the "acceptance criteria" are the performance specifications and safety standards met by the predicate device and relevant industry standards. The "reported device performance" is how the new device compares to these.

Feature / Acceptance Criteria (Derived from Predicate/Standards)GEMORE TENS System (Reported Performance)Notes on Acceptance
Intended UseSymptomatic relief of chronic, intractable pain; Adjunctive treatment for post-surgical and post-traumatic acute pain; Relief of pain associated with osteoarthritis of the knee and rheumatoid arthritis of the hand.Meets: Identical to the predicate device's Indications for Use.
Product CodeGZJ & NYNMeets: Identical to the predicate device.
Electrical Safety (IEC 60601-1)CompliantMeets: Declared compliance.
Electromagnetic Compatibility (IEC 60601-1-2)CompliantMeets: Declared compliance.
Transcutaneous Electrical Nerve Stimulators (ANSI/AAMI NS4)CompliantMeets: Declared compliance (predicate also complies with IEC standards).
Software VerificationCarried out in accordance with FDA software guidance.Meets: Declared compliance for software.
Power Source9V batteryDifferent: Predicate uses 1.5Vx2 (AA Size). Deemed not to affect safety/effectiveness.
Method of Line Current IsolationType BFMeets: Same as predicate.
Patient Leakage Current (Normal Condition)Under 0.1 uAMeets: Same as predicate.
Patient Leakage Current (Single Fault Condition)Under 0.5 uAMeets: Same as predicate.
Number of Output Modes2Meets: Same as predicate.
Channel SynchronizationSynchronousMeets: Same as predicate.
Method of Channel IsolationOutput transformerMeets: Same as predicate.
Output RegulationCurrentMeets: Same as predicate.
Software/Firmware/Microprocessor controlYesMeets: Same as predicate.
Automatic Shut Off?YesMeets: Same as predicate.
Display IndicatorsOn/Off Status, Low BatteryMeets: Same as predicate. (Note: Predicate says "Same" for Voltage/Current Level, but new device says "No"). Substantial equivalence maintained.
Timer Range (Minutes)15, 30, and 60Different: Predicate has 1~60. Deemed not to affect safety/effectiveness.
Weight140 gSimilar: Predicate is 138.9 g.
Dimensions (W x H x D)108x61.5x25 mmSimilar: Predicate is 109.5x60x34.9 mm.
Housing MaterialsABSMeets: Same as predicate.
WaveformMonophasicMeets: Same as predicate.
ShapeRectangularMeets: Same as predicate.
Maximum Output Voltage (V) (+/- 20%)40V @500Ω, 96V @2kΩ, 110V @10kΩDifferent: Predicate is lower. Deemed not to affect safety/effectiveness, possibly offering more robust output at higher impedances.
Maximum Output Current (mA) (+/- 20%)80mA @500Ω, 48mA @2kΩ, 11mA @10kΩDifferent: Predicate is lower. Deemed not to affect safety/effectiveness.
Duration of primary phase (usec)260 (usec) MaxDifferent: Predicate is 400 (usec) Max. Deemed not to affect safety/effectiveness, but shows a difference.
Pulse Duration (usec)260 MaxDifferent: Predicate is 400 Max.
Frequency (Hz) / Rate (pps)160 MaxDifferent: Predicate is 150 Max.
Symmetrical phases (for multiphasic waveforms)No (monophasic)Meets: Same as predicate (both are monophasic).
Max Phase Charge (uQ)20.8 uQ (for all modes)Meets / Similar: Same as predicate (20.8uQ).
Max Current Density (mA/cm²)Ranges from 0.083 to 0.133Similar: Predicate value is 0.1248. Differences are within acceptable limits for substantial equivalence.
Max Avg Current (mA)Ranges from 2.08 to 3.328Similar: Predicate value is 3.12 (in most cases), 13.12 (P1). Differences are within acceptable limits for substantial equivalence.
Max Power Density (w/cm²)Ranges from 0.0033 to 0.0053Similar: Predicate value is 0.0037. Differences are within acceptable limits for substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not applicable in the context of this 510(k) submission. This is a technical comparison and non-clinical performance verification, not a clinical trial with a "test set" of patients. The "test set" for performance characteristics would refer to the TENS devices themselves being tested in a lab environment.
  • Data Provenance: The data comes from the manufacturer's (Gemore Technology Co., Ltd., Taiwan) internal testing and comparison against the specifications of the predicate device (Empi - SELECT TENS System). It is entirely non-clinical.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not applicable. Ground truth for a device's technical specifications and compliance with standards is established through engineering and electrical testing, not typically by clinical experts in the sense of reviewing patient data.
  • Qualifications of Experts: N/A.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. There is no ambiguous patient data requiring expert adjudication in a 510(k) submission focused on technical equivalence.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is specifically for evaluating the diagnostic performance of a device (often imaging AI) by multiple human readers across multiple cases, and comparing performance with and without AI assistance. The GEMORE TENS System is a pain relief device, and this specific submission is for technical equivalence, not clinical efficacy through such a study design.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Standalone Performance: While the TENS device itself operates "standalone" in delivering electrical stimulation, the concept of "standalone performance" and "human-in-the-loop" typically applies to AI diagnostic algorithms. For a TENS device, the "performance" is its ability to deliver specified electrical outputs accurately. This was verified through engineering tests. There's no AI algorithm here that performs a diagnostic task independently or with human input.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The "ground truth" here is the technical specifications and performance characteristics of the predicate device, as well as the relevant voluntary standards (ANSI/AAMI NS4-1985, IEC 60601-1, and IEC 60601-1-2) for electrical safety and electromagnetic compatibility. The new device's performance was measured against these established engineering benchmarks.

8. The Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. This device does not use machine learning or AI that requires a "training set" of data.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable. As there is no training set, there is no ground truth establishment for it.

Conclusion of the Submission:

The study presented here is a pre-market submission (510(k)) demonstrating substantial equivalence between the new GEMORE TENS System and a previously cleared predicate device. This is achieved through:

  • Detailed technical comparison of features, intended use, operating modes, and physical characteristics.
  • Performance comparison of electrical output parameters (voltage, current, pulse duration, frequency, phase charge, current density, average current, power density).
  • Declaration of compliance with recognized voluntary standards (ANSI/AAMI NS4-1985, IEC 60601-1, and IEC 60601-1-2).
  • Software verification according to FDA guidance.

The manufacturer asserts that despite some minor differences (e.g., power source, maximum output parameters, timer range), these differences do not affect the intended use or alter the fundamental scientific technology of the device, thus maintaining the same safety and effectiveness as the cleared predicate. The FDA agreed with this assessment, granting 510(k) clearance.

{0}------------------------------------------------

.. . .

: ! ! : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K120569 .

    1. Submitter's Identifications: Gemore Technology Co., Ltd. 11FL., NO. 29-5, Sec. 2, Chung Cheng E. RD., Tan Shui, Taipei Hsien, Taiwan
      Contact: Boden S.P. Lai / President & Official Correspondent Phone Number:886-2-8809-1799 Fax Number:886-2-8809-1781

Date of Summary Preparation: July 12, 2012.

    1. Name of the Device:
Trade name:GEMORE TENS System; models GM310TP/GM320TP/GM330TP/GM340TP
Common name:Stimulator, Nerve, Transcutaneous For Pain Relief.
Classification name:Stimulator, Nerve, Transcutaneous For Pain Relief.
Product Code:GZJ & NYN

3. Information of the 510(k) Cleared Device (Predicate Device):

For this 510(k) submission, we compared our models to the following FDA cleared TENS devices: - K061650: Empi - SELECT TENS System.

4. Device Description:

The Gemore TENS System / Model GM310TP, GM320TP, GM330TP, and GM340TP transcutaneous electrical nerve stimulators for pain relief, are the electrically power devices used to apply an electrical current to electrode on a patient's skin to relieve pain. This series of TENS devices are indicated for:

  • symptomatic relief of chronic, intractable pain
  • . adjunctive treatment for post-surgical and post-traumatic acute pain

. relief of pain associated with osteoarthritis of the knee and rheumatoid arthritis of the hand For this definition of intended use, the Gemore TENS system is classified with the "NYN" & "GZJ" product code under 21CFR/Part 882.5890 FDA regulation classification number.

{1}------------------------------------------------

Gemore Technology Co., Ltd.

The Gemore TENS system consists mainly of two parts: the stimulus generator and electrode. The stimulus generator generates the output current specified as the input of controller. The output of controller is able to be adjusted for B/C/M/P1/P2/P3/P4/P5 eight operation modes which bear different combination for the output parameters of pulse width, pulse rate, and pulse cycle. For every operation mode, the operator could adjust output intensity, pulse rate and operation time. The output parts transmit the output current to the electrodes which is attached to the patient's skin so as to .transmit the generated stimulus to patient for pain relief treatment.

GM310TP/GM320TP/GM330TP/GM340TP is completely identical to that of model as mentioned in our previous submission of GEM STIM TENS, model GM310TE/ GM320TE/GM330TE/GM340TE (K032720). To change the indication for use as that of the chosen predicate devices, Gemore changed the software for these four models.

For the device included in this submission, we use the following of our 510(K) legally marketed predicate garment electrodes:

  • <1>K062675, "Gemore Reuseable Self Adhesive Electrode" ; Wire Series/ Model FA5050 ( 5x5 CM or bigger size) self adhesive electrode.
    GM310TP/GM320TP/GM330TP/GM340TP is a selectable dual channel, 9V battery operated TENS device with the following features:

  • <1> The operation function is dual channels completely identical to the model being modified, GM310TE/GM320TE/GM330TE/GM340TE (K032720).

  • <2> The output of controller is able to be adjusted for B/C/M/P 1/P2/P3/P4/P5 eight operation modes which bear different combination for the output parameters of pulse width, pulse rate, and pulse cvcle.

  • <3> The output strength is adjustable at 0~80 mA, with maximum setting time 60 minutes counting from switching ON.

  • <4> The LCD display is provided for the indication of operation status including CH1/CH2, output program mode, pulse width, pulse rate, timer, and battery low warning

5. Intended Use:

The indication for use is defined by the following statement:

The GEMORE TENS System, model GM310TP/GM320TP/GM330TP/GM340TP Transcutaneous Electrical Nerve Stimulators are indicated for:

  • symptomatic relief of chronic, intractable pain
  • . adjunctive treatment for post-surgical and post-traumatic acute pain
  • relief of pain associated with osteoarthritis of the knee and rheumatoid arthritis of the . hand
    1. Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:

Compliance to applicable voluntary standards includes ANSI/AAMI. NS4-1985, as well as IEC 60601-1, and IEC 60601-1-2 requirement. In addition to compliance of the voluntary standards, the software verification has been carried out in accordance to the FDA software guidance.

{2}------------------------------------------------

K120569

7. Similarities and differences comparison:

For the similarities and differences comparison between the application model GM310TP/ GM320TP/GM340TP and predicate 510(K) cleared model, the Emip-select TENS System, please see the details of the following table:

Comparison featuresGemore ModelPredicate 510(K) cleared Model
Model NameGEMORE TENS System, modelGM310TP/ GM320TP/ GM330TP/GM340TPEmip-Select TENS System
510(K) numberK120569K061650
Indication for useindicated for:• symptomatic relief of chronic,intractable pain• adjunctive treatment forpost-surgical and post-traumaticacute pain• relief of pain associated withosteoarthritis of the knee andrheumatoid arthritis of the handused for the symptomatic relief andmanagement of chronic, intractablepain and pain associated witharthritis. They are also used as anadjunctive treatment for post-surgicaland post-trauma acute pain
Product CodeGZJ & NYNSame
Program modeB/C/M and P1~P5 pre-program mode(Note: The program mode is similarto predicate device)B/C/M & ARB/SMP/MA/SPMpreprogram mode
Conformance standardsIEC 60601-1IEC 60601-1-2ANSI/AAMI NS4IEC 60601-1IEC 60601-1-2
Outlook & Dimensions108x61.5x25 mm109.5x60x34.9 mm
Weight140 g138.9 g

8. Comparision for the basic unit characteristics : As the following table

ParameterNew DevicePredicate Device
510(K) NumberK120569K061650
Device Name and ModelGEMORE TENSSystemSELECT TENSSystem
ManufacturerGemoreEmpi
Power Source(s)9V battery1.5Vx2 (AA Size)
- Method of Line current IsolationType BFSame
- Patient Leakage Current---
- Normal condition (uA)Under 0.1Same
- single Fault condition (uA)Under 0.5Same
Average DC current through electrodes whendevice is on but no pulses are being appliedNot applicableSame
(UA)
Number of Output Modes2Same
Number ofSynchronous or Alternating ?SynchronousSame
OutputMethod of Channel IsolationOutput transformerSame
Channels:
Regulated Current or Regulated Voltage?CurrentSame
Software/Firmware/Microprocessor control?YesSame
Automatic Overload Trip?NoSame
Automatic No-Load Trip?NoSame
Automatic Shut Off?YesSame
User Override control?NoSame
IndicatorOn/Off Status?YesSame
Display:Low Battery?YesSame
Voltage/Current Level?NoSame
Timer Range (Minutes)15,30, and 601~60
Compliance with Voluntary Standards?IEC 60601-1IEC 60601-1-2ANSI/AAMI NS4.IEC 60601-1IEC 60601-1-2
Compliance with 21 CFR 898?YesSame
Weight (g)140138.9
Dimensions (mm.) [W x H x D]108x61.5x25109.5x60.3x34.9
Housing Materials and constructionABSSame

{3}------------------------------------------------

K120569

9. Comparison of Output Specification : As the following table

Maximum output specification:

ParameterYour DevicePredicate Device
Mode or Program NameGM320TP as representativeSELECT TENS System
Waveform (e.g., pulsed monophasic, biphasic)MonophasicSame
Shape (e.g., rectangular, spike, rectified sinusoidal)RectangularSame
Maximum Output Voltage (volts) (+/- 20 %)40V @500Ω96V @2k Ω110V @10k Ω26V @500Ω40V @2k Ω40V @10k Ω
Maximum Output Current (specify units) (+/- 20 %)80mA @500Ω48mA @2k Ω11mA @10k Ω52mA @500Ω20mA @2k Ω4mA @10k Ω
Duration of primary (depolarizing) phase (usec)260 (usec) Max400 (usec) Max
Pulse Duration (usec)260 Max400 Max
Frequency (Hz) [or Rate (pps)]160 Max150 Max
For multiphasicwaveforms only:Symmetrical phases?NoSame
Phase Duration (include units),(Stage range, if applicable), (bothphases, if asymmetrical)Not ApplicableSame

Detailed comparison for each operation mode:

Max Phase ChargeMax Current DensityMax Avg CurrentMax Power DensityOperation
GemoreEmipGemoreEmipGemoreEmipGemoreEmipMode
20.8uQ20.8uQ0.0830.12482.08 mA3.12 mA0.00330.0037B
mA/cm²mA/cm²w/ cm²w/ cm²
20.8uQ20.8uQ0.1330.12483.328 mA3.12 mA0.00530.0037C
mA/cm²mA/cm²w/ cm²w/ cm
20.8uQ20.8uQ0.1330.12483.328 mA3.12 mA0.00530.0037M
mA/cm²mA/cm²w/ cm²w/ cm²
20.8uQ20.8uQ0.1330.12483.328 mA13.12 mA0.00530.0037Pl
mA/cm2mA/cm²w/ cm-w/ cm²
20.8uQ20.8uQ0.1330.12483.328 mA3.12 mA0.00530.0037P2
mA/cm²mA/cm²w/ cm2w/ cm²
20.8uQ20.8uQ0.1330.12483.328 mA3.12 mA0.00530.0037p3
mA/cm²mA/cm²w/ cmw/ cm-
20.8uQ20.8uQ0.1330.1248B.328 mA3.12 mA0.00530.0037P4
mA/cm²mA/cm²w/ cm²w/ cm
20.8uQ20.8uQ0.1330.12483.328 mA3.12 mA0.00530.0037િર
mA/cm2mA/cm²w/ cm²w/ cm

10. Conclusions

.

The GEMORE TENS System, model GM310TP/GM320TP/GM330TP/GM340TP have the same technological characteristics as the of the 510(K) cleared device, the Empi - SELECT TENS System. (K061650). Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted model could maintain the same safety and effectiveness as that of cleared device.

In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device. Therefore, the GEMORE TENS System, model GM310TP/GM320TP/GM330TP/GM340TP are substantial equivalent to the chosen predicate model.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around the perimeter. In the center of the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans and provide essential human services.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

GEMORE TECHNOLOGY CO., LTD. c/o Mr. Boden S.P. Lai General Manager 11 FL. No. 29-5, Sec. 2 Chung Cheng E. Rd. Tan Shui, Taipei Hsien Taiwan 251

JUL 2 0 2012

Re: K120569

Trade/Device Name: GEMORE TENS System Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: II Product Code: GZJ, NYN Dated: July 13, 2012 Received: July 16, 2012

Dear Mr. Lai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Spècial Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{5}------------------------------------------------

Page 2 - Mr. Boden S.P. Lai --

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Kesia Alexander

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications For Use

510(k) Number (if known): K120569

Device Name: GEMORE TENS System, model GM310TP/GM320TP/ GM330TP/GM340TP

Indications For Use:

The GEMORE TENS System, model GM310TP/GM320TP/GM330TP/GM340TP Transcutaneous Electrical Nerve Stimulators are indicated for:

  • symptomatic relief of chronic, intractable pain .
  • adjunctive treatment for post-surgical and post-traumatic acute pain
  • relief of pain associated with osteoarthritis of the knee and rheumatoid arthritis of the hand

Prescription Use V (Part 21 CFR 801 Subpart D)

OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sra-Offi

(Division Siza-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

Page 1 of

510(k) Nomb

§ 882.5890 Transcutaneous electrical nerve stimulator for pain relief.

(a)
Identification. A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.(b)
Classification. Class II (performance standards).