(169 days)
The Fully Automatic Blood Pressure Monitors BP series and BM 40 are intended for used by adults with 12 years and older to measure the systolic blood pressure and pulse rate.
Measure blood pressure (systolic and diastolic) and pulse rate.
The arm-type uses an inflatable cuff which is wrapped around the patient's upper arm. The cuff is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minute. There is a maximum pressure safety setting at 300 mmHg. The device will not inflate the cuff higher than 300 mmHg. For BP-103H, BP-1304, BP-1305, BP-1307, BM 40, the blood pressure results are compared with WHO (World Health Organization) Blood Pressure classification, which are severe Hypertension, Moderate Hypertension, Mild Hypertension, High-normal, Normal, and Optimal. The corresponding LCD segment will be turned on along with the systolic, diastolic, and pulse rate information. BP-103H, BP-1304, BP-1307, BM 40 will display an irregular heartbeat symbol "( if an irregular heartbeat was detected during the measurement process. BM 40, BP-1307 can display average results in three ways: the average of all measurements, the average of all AM (5:00 AM-8:59AM) measurements, and the average of all PM (18:00 PM-19:59 PM) measurements. BP-103H, BP-1304, BP-1305 can calculates the average of the last three measurements. In addition, BP-1307 has LCD backlight. After three minutes without operation, the blood pressure monitors turn off automatically.
The provided document describes the Sejoy Electronics & Instruments Co., Ltd. Fully Automatic Digital Blood Pressure Monitor (BP series and BM 40 models) and its substantial equivalence to a predicate device.
Here's an analysis of the acceptance criteria and study information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The device performance is assessed against the requirements of the ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008 standards for manual, electronic or automated sphygmomanometers. While the specific numerical acceptance criteria (e.g., mean difference and standard deviation between the device and a reference measurement) are not explicitly stated in the document, it does state that the device meets or exceeds these requirements.
| Acceptance Criteria (Standard) | Reported Device Performance |
|---|---|
| Accuracy requirements of ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008 for "manual, electronic or automated sphygmomanometers". | "The test result all meet or exceed the requirement of the standards." (for Performance Test) and "Clinical tests were performed and complied the accuracy requirements of ANSI/AAMI SP10..." |
2. Sample Size Used for the Test Set and Data Provenance
The document states that "Clinical tests were performed and complied the accuracy requirements of ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008 National Standard for Manual, Electronic or Automated Sphygmomanometers."
However, the specific sample size for the test set is not provided in this document.
The data provenance (country of origin, retrospective or prospective) is also not specified.
3. Number of Experts Used to Establish Ground Truth and Their Qualifications
The document does not provide details on the number of experts used or their qualifications for establishing the ground truth. It simply states that performance tests were conducted according to the ANSI/AAMI SP10 standard, which typically involves comparison against a reference measurement performed by trained observers (oscillometric vs. auscultatory method).
4. Adjudication Method for the Test Set
The document does not explicitly describe an adjudication method for the test set. In the context of blood pressure monitor accuracy testing per AAMI SP10, the "ground truth" is typically established by trained observers taking auscultatory measurements simultaneously or in strict sequence with the automated device. The standard itself specifies the statistical comparison methods.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is more common for diagnostic imaging algorithms where human readers interpret medical images. Blood pressure monitors are standalone measurement devices, and their accuracy is typically evaluated by comparing their readings against a validated reference method (e.g., auscultation by trained professionals).
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, a standalone performance evaluation was done. The document refers to "Performance Test according to ANSI/AAMI SP10:2002/(R) 2008..." and "Clinical tests were performed and complied the accuracy requirements of ANSI/AAMI SP10...". These tests assess the device's accuracy in measuring blood pressure and pulse rate without necessarily involving human interpretation of the device's output in the loop, beyond reading the displayed values. The device itself is designed to provide automated measurements.
7. The Type of Ground Truth Used
The ground truth used for performance testing (as implied by adherence to ANSI/AAMI SP10) would be reference blood pressure measurements obtained by trained observers using a validated method, typically auscultation with a mercury or aneroid sphygmomanometer. The standard specifies the requirements for this reference method.
8. The Sample Size for the Training Set
This information is not provided in the document. Blood pressure monitors like these do not typically have an "AI algorithm" in the sense of requiring a large, separate training set for a machine learning model. Their design is based on established oscillometric principles.
9. How the Ground Truth for the Training Set Was Established
As noted above, the concept of a "training set" in the context of an AI algorithm is largely not applicable to this device. The device operates based on a physical measurement principle (oscillometry) and its internal algorithms process these physical signals. Any calibration data or internal parameters would be established through engineering design and validation, not through a separate "training set" with ground truth in the AI sense.
{0}------------------------------------------------
PS lot 5
Image /page/0/Picture/1 description: The image shows the word "SEJOY" in bold, black letters. There is a small mark above the "J" that looks like an accent mark. The letters are closely spaced together, and the word appears to be a logo or brand name. The background is white.
世佳电子有限公司 Sejoy Electronics & Instruments Co., Ltd.
510(k) Summary
AUG 1 0 2012
The assigned 510(k) number is: K120554
1. Date Prepared:
2011.11.25
2. Submitter's Identification:
Name: Sejoy Electronics & Instruments Co., Ltd.
Add .: Building 2, No.202, Zhenzhong Road, West Lake Economy & Technology Zone, Hangzhou, China 310030 Contact Person: Yunhua Ren Phone: +86-571-81957767 Fax: +86-571-81957750 Email: info@sejoy.com
3. Name of the Device:
Trade Name: Fully Automatic Digital Blood Pressure Monitor Including the following models:
-
BP series
-
BP-102, BP-103, BP-103H, BP-1209, BP-1303A, BP-1304, BP-1305, BP-1307 for Arm-type Fully Automatic Digital Blood Pressure Monitor and
-
BM 40 for Arm-type Fully Automatic Digital Blood Pressure Monitor
Common Name: Blood Pressure Monitor
Classification name: Non-invasive blood pressure measurement System 21 CFR 870-1130, Class II, 74-DXN.
Classification Information: 4.
Regulation Number: 870.1130 Product Code: DXN Device Class: II Panel: 74 Cardiovascular
Predicate Device Information: న.
7
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "SEJOY" in bold, black letters. There is a small, triangle-like shape above the letter "J". The letters are closely spaced together, and the word appears to be a logo or brand name.
世佳电子有限公司 Sejoy Electronics & Instruments Co., Ltd.
The Arm-type Fully Automatic Digital Blood Pressure Monitors are substantially equivalent to the following device:
Premier Blood Pressure Monitor (Model: UA-1020), FDA 510(k) K111686, manufactured by A&D Company, Limited.
6. Device Description:
The arm-type uses an inflatable cuff which is wrapped around the patient's upper arm. The cuff is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minute. There is a maximum pressure safety setting at 300 mmHg. The device will not inflate the cuff higher than 300 mmHg. For BP-103H, BP-1304, BP-1305, BP-1307, BM 40, the blood pressure results are compared with WHO (World Health Organization) Blood Pressure classification, which are severe Hypertension, Moderate Hypertension, Mild Hypertension, High-normal, Normal, and Optimal. The corresponding LCD segment will be turned on along with the systolic, diastolic, and pulse rate information. BP-103H, BP-1304, BP-1307, BM 40 will display an irregular heartbeat symbol "( if an irregular heartbeat was detected during the measurement process. BM 40, BP-1307 can display average results in three ways: the average of all measurements, the average of all AM (5:00 AM-8:59AM) measurements, and the average of all PM (18:00 PM-19:59 PM) measurements. BP-103H, BP-1304, BP-1305 can calculates the average of the last three measurements. In addition, BP-1307 has LCD backlight. After three minutes without operation, the blood pressure monitors turn off automatically.
| FeaturesModels | A | B | C | D | E | F | G | H | I | J | K | L |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| BP-102 | ● | ● | 120 Memories | ○ | ○ | ○ | ○ | ○ | 150×112×56 | 22-36cm | 83×35 | ○ |
| BP-103 | ● | ● | 120 Memories infour groups | ● | ○ | ○ | ○ | ○ | 110×155×70 | 22-36cm | 84.1×55.1 | ○ |
| BP-103H | ● | ● | 120 Memories infour groups | ● | ○ | ● | ● | ● | 110×99×66 | 22-36cm | 84.1×55.1 | ○ |
| BP-1209 | ● | ● | 120 Memories | ● | ○ | ○ | ○ | ○ | 134×99×66 | 22-36cm | 62.7×46.4 | ○ |
| BP-1303A | ● | ● | 120 Memories infour groups | ● | ○ | ○ | ○ | ○ | 110×155×70 | 22-36cm | 62.7×46.4 | ○ |
| BP-1304 | ● | ○ | 120 Memories infour groups | ● | ○ | ● | ● | ● | 110×155×70 | 22-36cm | 62.7×46.4 | ○ |
The detail comparisons among devices are listed below:
{2}------------------------------------------------
Pg 3 of 5
| BP-1305 | • | • | 120 Memories intwo groups | • | o | • | • | • | 166×114×72 | 22-36cm | 84.1×55.1 | o |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| BP-1307 | • | • | 120 Memories intwo groups | • | • | o | • | • | 166×114×72 | 22-36cm | 102.1×68.9 | • |
| BM 40 | • | • | 120 Memories intwo groups | • | • | o | • | • | 119×109×60 | 22-35cm or30-42cm | 83.4×54.41 | o |
世佳电子有限公司
A = Powered by 4 AA size batteries
B= Powered by AC adaptor
C = Memory Size
D = Time & Date
E = Results Average in Three way
F = WHO (World Health Organization) Classification Indicator
G = Last 3 Results Average
H = Irregular Heartbeat Detection
I = Outside Demission (L x W x H in mm)
J = Cuff Size
K = LCD Size (Viewing Area in mm)
L = LCD Backlight
• = Yes
o = No
The devices are all designed and manufactured according to ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008, manual, electronic or automated sphygmanometers.
7. Intended Use:
The Fully Automatic Blood Pressure Monitors BP series and BM 40 are intended for used by adults with 12 years and older to measure the systolic blood pressure and pulse rate.
The intended user and the indication for use of the Fully Automatic Blood Pressure Monitors BP series and BM 40 as described in the labeling are nearly the same as their predicated devices, Premier Blood Pressure Monitor (Model: UA-1020).
8. Summary comparing technological characteristics with predicate device:
| Technological Characteristics | Comparison result |
|---|---|
| Design principle | Identical |
| Appearance | Similar |
| Patients contact Materials | Identical |
| Performance | Similar |
{3}------------------------------------------------
Pg 4 of 5
世佳电子有限公司 Sejoy Electronics & Instruments Co., Ltd.
| Biocompatibility | Identical |
|---|---|
| Mechanical Safety | Identical |
| Energy Source | Similar |
| Standards Met | Identical |
| Electrical Safety | Identical |
| EMC | Identical |
| Function | Similar |
The difference of technological characteristic between the predicate device and the submit arm-type fully automatic blood pressure monitors is the appearance, the performance, energy source and the function.
9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards including ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008 as well as IEC/EN 60601-1: 2005+CORR.1 (2006) +CORR.2 (2007) / EN 60601-1:2006 /AC: 2010, IEC/EN 60601-1-2:2007, ISO 10993-5: 2009 Biological evaluation of medical devices -Part 5:Tests for in vitro cytotoxicity and ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
Guidance documents include the "FDA Non-invasive Blood Pressure (NIBP) Monitor Guidance" and "FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "FDA Bluebook Memorandum G95-1Use of International Standard ISO 10993."
Non-clinical Tests:
- Electromagnetic Compatibility Test according to IEC/EN 60601-1-2:2007 .
- . General Safety Provisions Test according to IEC/EN 60601-1: 2005+CORR.1 (2006) +CORR.2 (2007) / EN 60601-1:2006 /AC: 2010
- . Performance Test according to ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008. The test result all meet or exceed the requirement of the standards.
- Biocompatibility Test according to FDA Bluebook Memorandum G95-1Use of International Standard ISO 10993, ISO 10993-5: 2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity and ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
10. Discussion of Clinical Tests Performed:
Clinical tests were performed and complied the accuracy requirements of ANSI/AAMI
1 0
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "SEJOY" in bold, black letters. The letters are slightly spaced apart, and the font appears to be sans-serif. There is a small, indistinct mark above the "J" in the word. The overall impression is a simple, clear logo or brand name.
世佳电子有限公司 Sejoy Electronics & Instruments Co., Ltd.
SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008 "National Standard for Manual, Electronic or Automated Sphygmomanometers".
11. Conclusions:
Our Arm-type Fully Automatic Blood Pressure Monitors BP-102, BP-103, BP-103H, BP-1209, BP-1303A, BP-1304, BP-1305, BP-1307 and BM 40 have the same intended use and similar technological characteristics as the A&D Premier Blood Pressure Monitor (Model: UA-1020) whose 510 (k) number is K111686.
Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared devices.
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device, and the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's emblem, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the emblem in a circular fashion.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
AUG 1 0 2012
Sejoy Electronics & Instruments Co., Ltd. c/o Mr. Edmondo Roark 1259 Monten Street #4 Cincinnati, OH 45208
Re: K120554
Trade/Device Name: Arm-type Fully Automatic Digital Blood Pressure Monitor - (Nine Models: BP-102, BP-103, BP-103H, BP-1209, BP-1303A, BP-1304, BP-1305, BP-1307 and BM 40)
Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: II (two) Product Code: DXN Dated: July 23, 2012 Received: July 23, 2012
Dear Mr. Roark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may) are so revisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{6}------------------------------------------------
Page 2 – Mr. Roark
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Sincerely yours,
C. Bram D. Zuckerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
EJO
世佳电子有限公司 Sejoy Electronics & Instruments Co., Ltd.
Indications for Use
510(k) Number (if known):
Blood Pressure Monitors BP series and BM 40 Device Name:
Indications For Use:
Measure blood pressure (systolic and diastolic) and pulse rate.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use V (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of _ 1
5
sion Sign-Off) sion of Cardiovascular Devices
510(k) Number
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).