(53 days)
The Gauss Pixel App is indicated for use to aid current practices in recording the number of surgical sponges and for visibility for assessment of sponge images.
The Gauss Pixel App is a software program used on an iPad tablet to capture images of sponges to assist surgical personnel in the management of surgical sponges after surgical The App allows surgical personnel to categorize sponges by sponge type and use. provides an automated ongoing count of total sponge images and sponge images by tag. It also provides a visual record of images for further evaluation. This program is not intended to replace existing sponge counting practices and sponges should be retained per the user's standard sponge management practice until the case is complete and sponge counting has been finalized.
Here's a summary of the acceptance criteria and study findings for the Gauss Pixel App, based on the provided document:
1. Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance (Gauss Pixel App) |
|---|---|
| Record Images as Indicated by User | Confirmed: The application provided instructions for use and recorded images as indicated by the user. |
| Accurately Tag Images as Indicated by User | Confirmed: The application accurately tagged images as indicated by the user. |
| Accurately Provide Automated Counting (Total) | Confirmed: The application accurately provided automated counting both in total. |
| Accurately Provide Automated Counting (by Type) | Confirmed: The application accurately provided automated counting by type. |
| Allow Visual Review and Management of Images | Confirmed: The application allowed visual review and management (re-tagging, deletion) of all images as appropriate. |
| Function as Intended | Demonstrated: Results of performance testing through the software verification and validation process demonstrate that the Gauss Pixel App functions as intended. |
| Substantial Equivalence to Predicate Device | Demonstrated: The Gauss Pixel App is as safe and effective as the predicate, has the same intended uses and indications, and utilizes a new technological method (software) which complements current clinical practices and does not raise new issues of safety or effectiveness. Software verification and validation demonstrate it functions as intended. |
2. Sample Size and Data Provenance
The document does not explicitly state the sample size used for a test set in the conventional sense of a clinical trial or independent validation. The performance claims are based on "software verification and validation testing." This typically involves internal testing by the developer to ensure the software meets its specified functional and non-functional requirements.
There is no information provided regarding the country of origin of data or whether it was retrospective or prospective. Given the nature of a 510(k) submission for a Class I device and the focus described (software verification and validation), it's highly probable that this involved internal testing rather than a large-scale clinical study with external patient data.
3. Number of Experts and Qualifications
The document does not specify the number of experts, their qualifications, or their role in establishing a "ground truth" for a test set. The validation described is primarily around the software's functional performance (e.g., counting, tagging, image display) rather than assessing clinical accuracy against expert opinion in a diagnostic context.
4. Adjudication Method
No adjudication method (e.g., 2+1, 3+1) is mentioned, as the described validation focuses on software functionality rather than interpreting findings against a 'ground truth' that would require expert consensus.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC comparative effectiveness study is mentioned. The device is described as an adjunctive tool to aid current practices, not as a replacement or independent diagnostic tool. The focus is on verifying its own functionality (counting, image display) rather than its impact on human reader performance. Therefore, an effect size of human readers improving with/without AI assistance is not reported.
6. Standalone Performance
Yes, a standalone (algorithm only without human-in-the-loop performance) assessment was done for the core functionalities of the application. The document states: "Results of performance testing confirmed that the application provided instructions for use, recorded images as indicated by the user, accurately tagged images as indicated by the user, accurately provided automated counting both in total and by type and allowed visual review and management (re-tagging, deletion) of all images as appropriate." This describes the independent functional performance of the software.
7. Type of Ground Truth Used
For the described performance testing, the "ground truth" appears to be the expected functional behavior of the software as defined by its specifications and design. For example:
- For "accurately provided automated counting," the ground truth would be the actual number of images taken and the actual type assigned by a user, against which the software's automated count is compared.
- For "accurately tagged images," the ground truth is the tag the user intended to apply, against which the software's recorded tag is compared.
It does not refer to medical ground truth like pathology, outcomes data, or expert consensus on a diagnostic interpretation.
8. Sample Size for the Training Set
The document does not specify a sample size for a training set. This is consistent with the nature of the device as a functional tool for managing sponges, not a machine learning model that requires a large training dataset to develop its core functionality (like image recognition for diagnostic purposes). The "training" described implicitly refers to the software development and testing cycles designed to ensure correct functionality.
9. How Ground Truth for the Training Set Was Established
Given the information provided, the concept of a "ground truth for the training set" as it pertains to typical AI/ML development (e.g., labeled data for model training) is not applicable here. The ground truth for the development and verification of the software's core functions (counting, tagging) would have been established by the software's design specifications and the expected, correct outputs for given inputs during testing. This is standard software engineering practice rather than a data-driven ground truth for a machine learning algorithm.
{0}------------------------------------------------
Attachment 2
16 Mar 2012 revised section for K120473
5.0 510(k) Summary
Submitters Name and Address Gauss Surgical, Inc. 22700 Alcalde Road Cupertino CA 95014
Contact Person
Peggy McLaughlin Consulting Vice President, Clinical & Regulatory Affairs Gauss Surgical, Inc. 22700 Alcalde Road Cupertino CA 95014 Fax: 650 941-8222 Telephone: 650 504-8501 Email: MPMcLaughlin@sbcglobal.net
Date Prepared 16 Mar 2012
Name of Medical Device
Device Classification Name: Counter, Surgical Sponge Device Classification Number: 21 CFR 880.2740 Device Class: Class 1 Proprietary Name: Gauss Pixel App
Predicate Device
Bag-It™ Sponge Counter Bags (Tyco/Covidien) (K912824)
Device Description
The Gauss Pixel App is a software program used on an iPad tablet to capture images of sponges to assist surgical personnel in the management of surgical sponges after surgical The App allows surgical personnel to categorize sponges by sponge type and use. provides an automated ongoing count of total sponge images and sponge images by tag. It also provides a visual record of images for further evaluation. This program is not intended to replace existing sponge counting practices and sponges should be retained per the user's standard sponge management practice until the case is complete and sponge counting has been finalized.
Intended Use
The Gauss Pixel App is intended to be used to aid current clinical practices in recording the number of surgical sponges and for visibility for assessment of sponge images.
The App is intended to be used in the Operating Theatre to assist users with sponge management practices by providing another method to visually capture and display images of used sponges using software running on an iPad mobile platform. Running the
Gauss Surgical, Inc. Traditional 510(k) Submission
K 120473 16 Mar 2012
•
APR - 9 2012
{1}------------------------------------------------
. Attachment 2
16 Mar 2012 revised section for K120473
software, the user captures images of each sponge prior to storing per their standard sponge management practices. The user may categorize the sponge by type. The user may review each image to re-assign sponge type, confirm image quality or delete the image if appropriate. The software counts the images and displays an ongoing count of total sponges and sponges by sponge type as categorized by the user. The software allows users to re-review each sponge image and a display panel of all used sponges to assist in sponge management and review. The sponge image is captured in a fully expanded state, providing a full view of each sponge prior to storage per current clinical practices.
The indications for use for the predicate and proposed device differ in that the proposed device is intended to 'aid current practices' which means it is intended to be used in conjunction with current practices. The software automates the counting procedure and this counting has been verified through verification and validation testing. This difference does not affect the safety or efficacy of the device as it is labeled as an adjunctive product.
Technology
Both the Pixel App and the predicate device utilize visual assessment of sponges to count the number of surgical sponges used in a surgical procedure. The predicate device uses a plastic bag with partitions to separate each sponge for counting and review purposes while the Pixel App uses software running on an iPad mobile platform for counting and review purposes. The Pixel App and the predicate device are intended to be used together so that either method of visual assessment may be used during the surgical procedure and a final sponge count may be made by the user. This software technology has been verified to provide an accurate count of all images and by image type as categorized by the user and does not introduce new questions of safety as the user is advised to retain all sponges using standard sponge management practices and can rereview sponges physically at the end of the procedure if needed.
| Technologic method for: | Predicate | Gauss Pixel App |
|---|---|---|
| Sponge management | Sponges are collected andstored in partitions within theplastic bag. | Sponges are collected andstored in partitions within theplastic bag or other spongemanagement method selectedby user. |
| Sponge counting | User manually countssponges as seen in partitionsand calculates a total. | User captures image of eachsponge prior to storing percurrent sponge managementpractices and software countseach image, providing anongoing total on the iPadmobile platform. |
| Sponge typing | Not available unless user optsto use separate containers foreach sponge type. | User assigns sponge type foreach image captured. |
Tachmologio Co
Gauss Surgical, Inc. Traditional 510(k) Submission
K 120473 16 Mar 2012
2/3
{2}------------------------------------------------
Attachment 2 -
16 Mar 2012 revised section for K120473
| Technologic method for: | Predicate | Gauss Pixel App |
|---|---|---|
| Sponge counting by type | Not available unless user optsto use separate containers foreach sponge type and totalseach type separately. | Software counts each imageby assigned type, providing anongoing total. |
| Visual assessment of eachsponge | User can view and assesseach sponge in a wadded upstate in partitions. | User can view and assesseach sponge in anopen/unfolded state by re-reviewing any selected image. |
| Visual assessment of allsponges used in a clinicalprocedure | User can view and assess theentire 'sheet' of sponges (orseries of sheets) as stored inthe individual partitions. | User can view and assess allsponges in an open/unfoldedstate in a grid-like display ofall sponges captured during aprocedure. |
Performance
Results of performance testing confirmed that the application provided instructions for use, recorded images as indicated by the user, accurately tagged images as indicated by the user, accurately provided automated counting both in total and by type and allowed visual review and management (re-tagging, deletion) of all images as appropriate. Results of performance testing through the software verification and validation process demonstrate that the Gauss Pixel App functions as intended and is substantially equivalent to the predicate device.
Substantial Equivalence
The Gauss Pixel App is as safe and effective as the predicate. The Gauss Pixel App has the same intended uses and indications and utilizes a new technological method (software) which complements current clinical practices and does not raise new issues of safety or effectiveness. Software verification and validation demonstrate that the Gauss Pixel App functions as intended. Thus, the Gauss Pixel App has been shown to be substantially equivalent to the predicate device.
3/3
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three overlapping profiles facing left. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circle around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Gauss Surgical, Inc. % Ms. Peggy McLaughlin Consulting VP, Clinical and Regulatory Affairs 22700 Alcalde Road Cupertino, California 95014
Re: K120473
Trade/Device Name: Gauss Pixel App Regulation Number: 21 CFR 880.2740 Regulation Name: Surgical sponge scale Regulatory Class: I Product Code: LWH Dated: March 16, 2012 Received: March 19, 2012
Dear Ms. McLaughlin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any 1 vith all the Act's requirements, including, but not limited to: registration and listing (21
APR - 9 2012
{4}------------------------------------------------
Page 2 - Ms. Peggy McLaughlin
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Attachment 1
16 Mar 2012 revised section for K120473
4.0 Indications for Use Statement
Indications for Use Form
510(k) Number: K120473
Device Name: Gauss Pixel App
Indications for Use:
The Gauss Pixel App is indicated for use to aid current practices in recording the number of surgical sponges and for visibility for assessment of sponge images.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Richard P. Bell for Ned Ogden
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
ic,
510(k) Number_k 120473
Gauss Surgical, Inc. Traditional 510(k) Submission
Page / of /
K 120473 16 Mar 2012
§ 880.2740 Surgical sponge scale.
(a)
Identification. A surgical sponge scale is a nonelectrically powered device used to weigh surgical sponges that have been used to absorb blood during surgery so that, by comparison with the known dry weight of the sponges, an estimate may be made of the blood lost by the patient during surgery.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 880.9. The device also is exempt from the current good manufacturing practice requirements of the quality system regulation in part 820 of this chapter, with the exception of § 820.180, with respect to general requirements concerning records, and § 820.198, with respect to complaint files.