(214 days)
P080018/S001
K/DEN: P080018/S001
No
The document describes a digital mammography system using a structured needle phosphor detector and image processing software. There is no mention of AI, ML, or any related concepts in the intended use, device description, or performance studies. The image processing described is traditional, rule-based adjustment of image parameters.
No.
The device is used to generate digital mammographic images for screening and diagnosis of breast cancer, which is for diagnostic purposes, not therapeutic.
Yes
The device generates digital mammographic images that can be used for screening and diagnosis of breast cancer, which directly indicates a diagnostic purpose.
No
The device description explicitly mentions a "Carestream CR Mammography Cassette with SNP-M1 Screen" which is a hardware component (a structured needle phosphor detector) used in conjunction with the software feature.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The description clearly states that this device is used in conjunction with a mammographic x-ray machine to generate digital mammographic images. These images are then interpreted by a qualified physician for screening and diagnosis of breast cancer.
- Input: The input is a mammographic X-ray, which is an imaging modality, not a biological sample from the patient.
- Output: The output is a digital image, not a result from a test performed on a biological sample.
The device is a component of a medical imaging system used for diagnostic purposes, but it does not perform in vitro testing.
N/A
Intended Use / Indications for Use
The KODAK DirectView CR Mammography Feature together with KODAK DirectView CR Mammography Cassette comprise a device which, when used in conjunction with a KODAK DirectView CR System and a mammographic x-ray machine, generates digital mammographic images that can be used for screening and diagnosis of breast cancer. It is intended for use in the same clinical applications as traditional screen-film based mammographic systems. The mammographic images can be interpreted by a qualified physician using either hardcopy film or softcopy display at a workstation.
Product codes
MUE
Device Description
The Carestream CR Mammography Cassette with SNP-M1 Screen is a structured needle phosphor detector used in conjunction with the Kodak DirectView CR Mammography Feature on Kodak DirectView CR Systems for generating digital mammographic images. The Carestream CR Mammography Cassette SNP-M1 Screen is used in the same manner as a traditional screen-film cassette when performing radiographic patient exposures and will be used in the DirectView CR Systems: CR850, readers problem on on and Elite CR readers using the Kodak DirectView CR Mammography Feature.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
X-ray
Anatomical Site
Breast
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Qualified physician, interpretation at a workstation.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Performance testing was conducted to verify the design output met the design input requirements and to validate the device conformed to the defined user needs and intended uses. Non-clinical physical laboratory testing was conducted to assess performance. Clinical image evaluation was also conducted. Predefined acceptance criteria was met and demonstrated that the device is substantially equivalent to and as safe and as effective as the predicate device.
Key Metrics
Not Found
Predicate Device(s)
P080018/S001
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.1715 Full-field digital mammography system.
(a)
Identification. A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”See § 892.1(e) for the availability of this guidance document.
0
Special 510(k): Carestream Health, Inc.
"510(k) Summary"
AUG 2 8 2012
510(k) Owner Name | Carestream Health, Inc. |
---|---|
510(k) Owner Address | 150 Verona Street |
Rochester, New York 14608 | |
510(k) Owner Phone | 585 627-6491 |
510(k) Owner Fax | 585 323-7643 |
Contact Name & Info | Katherine Honsinger |
Regulatory Affairs Manager | |
katherine.honsinger@carestreamhealth.com | |
Date Summary Prepared | September 29, 2011 |
Device Trade Name | Kodak DirectView CR Mammography System |
Device Common Name | CR Digital Mammography System |
Classification Name | Full Field Digital, System, Xray, Mammographic |
Regulation Name | Full-Field Digital Mammography System |
Device Class | Class II |
Device Code | MUE |
Regulation Number | 21 CFR 892.1715, Full-Field Digital Mammography |
System | |
Predicate Device | Kodak DirectView CR Mammography System |
(P080018/S001) |
Device Description
The Carestream CR Mammography Cassette with SNP-M1 Screen is a structured needle phosphor detector used in conjunction with the Kodak DirectView CR Mammography Feature on Kodak DirectView CR Systems for generating digital mammographic images. The Carestream CR Mammography Cassette SNP-M1 Screen is used in the same manner as a traditional screen-film cassette when performing radiographic patient exposures and will be used in the DirectView CR Systems: CR850, readers problem on on and Elite CR readers using the Kodak DirectView CR Mammography Feature.
1
Intended Use
Indications for Use and Intended Use of the modified device remain unchanged from the predicate device.
The Kodak DirectView CR Mammography Feature together with the CR Mammography Cassette comprise a device which, when used in conjunction with a Kodak DirectView CR System and a mammographic X-ray machine, generates digital mammographic images that can be used for screening and diagnosis of breast cancer. It is intended for use in the same clinical applications as traditional screen-film based mammooraphic systems. The mammographic images can be interpreted by a qualified physician using either hardcopy film or softcopy display at a workstation.
Comparison of Technological Characteristics
The SNP-M1 storage phosphor screen consists of a structured needle phosphor layer. The predicate device EHR-M3 storage phosphor screen consists of a powdered-base phosphor laver. The SNP-M1 needle-shaped phosphor crystals, which are oriented at 90 degrees to the screen surface, reduce the spread of light within the screen, and produce a more uniform screen structure compared to the EHR-M3 powder-based phosphors. The SNP-M1 structured needle phosphor screen with reader optimization kit is designed to provide improved image quality, increased sharpness, and reduced noise, delivered at a lower dose compared to the EHR-M3 powdered-based phosphor screen. Either screen type is adhered to a rigid aluminum cassette plate assembly which is inserted within the cassette housing. There are minor modifications to the cassette housing to protect the SNP-M1 screen/plate assembly from scratching or damage.
As with the EHR-M3 screen, the SNP-M1 screen is used in the same manner as a traditional screen-film cassette when performing radiographic patient exposures and will be used in all DirectView CR Systems: CR850, CR975, Classic CR and Elite CR readers that utilize the Kodak DirectView CR Mammography Feature. The CR readers are modified to optimize image quality of images captured with the SNP-M1 Screen.
There is no difference in image acquisition. Use and functional operation of the CR Mammography Cassette with SNP-M1 Screen in the CR850, CR975, Classic CR and Elite CR Readers is the same as that for the predicate device CR Mammography Cassettes with EHR-M3 Screen.
There are no significant changes in the image processing software used with the CR Mammography Cassette with SNP-M1 Screen. The Kodak DirectView CR Mammography Feature utilizing Kodak ECLIPSE Image Processing Software v5.4 is the same as that used with the CR Mammography Cassette with EHR-M3 Screen. A modified looks database enhances the image quality by adjusting noise, contrast, brightness, sharpness and detail of the images captured with the SNP-M1 Screen.
Discussion of Testing
محمد مع مع مع مع مع مع قوم مع
Performance testing was conducted to verify the design output met the design input requirements and to validate the device conformed to the defined user needs and intended uses. Non-clinical physical laboratory testing was conducted to assess performance. Clinical image evaluation was also conducted. Predefined acceptance criteria was met and demonstrated that the device is substantially equivalent to and as safe and as effective as the predicate device.
2
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
AUG 28 2012
Ms. Katherine Honsinger Regulatory Affairs Manager Carestream Health, Inc. 150 Verona Street ROCHESTER NY 14608
Re: K120246
Trade/Device Name: KODAK DirectView CR Mammography System Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: August 23, 2012 Received: August 24, 2012
Dear Ms. Honsinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your because of the device is substantially equivalent (for the indications ferenced above and have vire) to legally marketed predicate devices marketed in interstate for use stated in the encreating to 10gail) to the Medical Device Amendments, or to conniered print to May 20, 1978, are occordance with the provisions of the Federal Food, Drug, de rices that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The T ou may, distore, mailer of the Act include requirements for annual registration, listing of general controls provisions of mactice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such II your device is classified (see above) into enas ffecting your device can be found in Title 21, additional controls. Existing major regulations and to 895. In addition, FDA may publish further Sour of Fouring your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination cates of the ot that FDA has made a determination that your device complies with other requirements of the Act that FDA has made a determination ana younder Federal agencies. You must of any Federal statutes and regulations annimities out not limited to: registration and listing (21 Comply with an the Act 3 requirements, artical device reporting (reporting (reporting of
3
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirements as set forum in the quality systems (