K Number
K120244
Manufacturer
Date Cleared
2012-07-05

(160 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SpectraShield 9500 Surgical N95 Respirator is a single use, disposable, surgical N95 respirator, tested for continuous use up to 8 hours, embedded with a zeolite carrier containing a silver-copper agent on the outer layer and is not an antimicrobial drug. SpectraShield 9500 kills 99.99% of test bacteria after one hour of contact with the surface of the respirator. In vitro (laboratory) tests have demonstrated 99.99% kill on the surface of the outer layer of the respirator when tested in vitro against single isolates of the following test bacteria: Streptococcus pyogenes, MRSA (Methicillin Resistant Staphylococcus aureus), and Haemophilus influenzae, under tested contact conditions.

No clinical studies have been conducted comparing the ability of an untreated surgical N95 respirator and the SpectraShield model 9500 surgical N95 respirator to protect the wearer from infection, and the antibacterial treatment cannot effect pathogens that are inhaled around the edges of the respirator.

The SpectraShield 9500 Surgical N95 respirator is a single use device, intended for occupational use to protect against microorganisms, body fluids, and particulate material.

Device Description

The SpectraShield model 9500 Surgical Mask is a molded shape surgical mask composed of 4 layers of material, molded to form the mask. A 2-ply meitblown polypropylene middle layer is sandwiched by inner and outer layers of 100% polyester nonwoven fabric. The inner and outside layers of polyester nonwoven fabric include fibers that have been embedded with an antibacterial agent to provide antibacterial performance. The mask has 2 latex-free non-allergenic elastic straps and an aluminum nose strip.

AI/ML Overview

The SpectraShield 9500 Surgical N95 Respirator, as described in K120244, demonstrates its performance through a series of laboratory-based tests comparing it to a predicate device (K090414). The primary goal of this 510(k) submission was to amend the Intended Use Statement of the SpectraShield 9500 Surgical N95 Respirator to include "tested for continuous use up to 8 hours," based on existing bioefficacy data showing sustained performance.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally established by meeting or exceeding the performance of the predicate device (K090414) in various standardized tests. The new device is essentially the same as the predicate, with the requested change being an update to the Intended Use Statement.

TestAcceptance Criteria (Predicate Device K090414)Reported Device Performance (New Device)
Fluid Resistance (ASTM F1862)Pass: Fluid Resistant @ 160mm HgPass: Fluid Resistant @ 160mm Hg
Particulate Filtration Efficiency (ASTM F2299)Pass: 99.87% at 0.1 micronsPass: 99.87% at 0.1 microns
Differential Pressure (Mil M36954C)Pass: 4.3mm H2O/cm2Pass: 4.3mm H2O/cm2
Bacterial Filtration Efficiency (ASTM F2101)Pass: 99.9%Pass: 99.9%
Flammability (Class 16CFR 1610)Class 1Class 1
Cytotoxicity (ISO 10993-10)Pass: USP reactivity score = < 2Pass: USP reactivity score = < 2
Primary Skin Irritation (ISO 10993-10)Pass: PSI Score = 0, Non-irritantPass: PSI Score = 0, Non-irritant
Repeated Patch Dermal Sensitization (ISO 10993-10)Pass: 0% incidence sensitization response "0" severity at each evaluated time point.Pass: 0% incidence sensitization response "0" severity at each evaluated time point.
Systemic Toxicity (ISO 10993-11)Pass: No mortality or evidence of systemic toxicity from the extracts was observed.Pass: No mortality or evidence of systemic toxicity from the extracts was observed.
Physico-chemical (USP Physico-chemical Test-Plastics)Pass: Test results met the USP limits.Pass: Test results met the USP limits.
Gas off TestingTotal antibacterial particles released from the device were verified to be within safe inhalation levels.Total antibacterial particles released from the device were verified to be within safe inhalation levels.
Leach off testingTotal leachable antibacterial particles released from the device were verified to be within safe orally ingestible levels.Total leachable antibacterial particles released from the device were verified to be within safe orally ingestible levels.
BioEfficacy (initial)S. pyogenes: > 4.40 log10 reduction - 1 hour, H. influenzae: > 6.20 log10 reduction - 1 hour, MRSA: > 4.83 log10 reduction - 1 hour (To Inoculums measured, >10^6)S. pyogenes: > 4.40 log10 reduction - 1 hour, H. influenzae: > 6.20 log10 reduction - 1 hour, MRSA: > 4.83 log10 reduction - 1 hour (To Inoculums measured, >10^6)
BioEfficacy (after repeated exposures to perspiration over 12 hours)S. pyogenes: > 4.25 log10 reduction - 1 hour, H. influenzae: > 4.18 log10 reduction - 1 hour, MRSA: > 4.11 log10 reduction - 1 hour (To Inoculums measured, >10^6)S. pyogenes: > 4.25 log10 reduction - 1 hour, H. influenzae: > 4.18 log10 reduction - 1 hour, MRSA: > 4.11 log10 reduction - 1 hour (To Inoculums measured, >10^6)

2. Sample Size and Data Provenance for Test Set

The document does not explicitly state the sample sizes used for each specific test in the "test set" in terms of number of masks. Instead, it references standardized ASTM, Mil, ISO, and USP test methods, which typically define their own sample size requirements. The data provenance is described as "in vitro (laboratory) tests," indicating that the tests were conducted in a controlled lab environment. There is no information about country of origin of the data or whether it was retrospective or prospective.

3. Number of Experts and Qualifications for Ground Truth

This type of submission (510(k) for a surgical mask with an amended intended use) relies on objective, repeatable physical and biological performance tests, not human expert interpretation of results for ground truth. Therefore, there were no "experts" in the traditional sense (e.g., radiologists) used to establish ground truth for a test set. The ground truth for performance measures (e.g., filtration efficiency, bacterial kill) is based on the results of the specified laboratory tests performed according to recognized standards.

4. Adjudication Method for the Test Set

No adjudication method (e.g., 2+1, 3+1) is described or applicable, as the evaluation of this device is based on quantifiable laboratory test results against predefined standards, not on subjective human assessments that would require adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There was no MRMC comparative effectiveness study done. The device is a surgical mask, and its performance is evaluated through material science and microbiology testing rather than clinical interpretation by human readers. Therefore, the concept of "how much human readers improve with AI vs. without AI assistance" is not applicable.

6. Standalone Performance (Algorithm Only)

A standalone performance evaluation was conducted for various physical and biological parameters of the mask. The "algorithm only" concept does not apply here as the device is a physical product (a mask), not an AI algorithm. The performance metrics presented in the table above represent the standalone performance of the device without human-in-the-loop for the core functionalities tested.

7. Type of Ground Truth Used

The ground truth used for evaluating the device's performance is based on objective laboratory test results according to established national and international standards (e.g., ASTM, Mil, ISO, USP). For bioefficacy, the ground truth is the measured bacterial reduction percentage based on controlled in vitro bacterial challenges. No pathology or outcomes data is mentioned in this summary for establishing ground truth.

8. Sample Size for the Training Set

The document does not describe a "training set" in the context of machine learning or AI, as this device submission is for a physical medical device. The tests performed are part of a product validation rather than an AI model development.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" in the context of this traditional medical device submission, the concept of establishing ground truth for a training set is not applicable.

{0}------------------------------------------------

K120244

Image /page/0/Picture/1 description: The image shows the logo for "NEXERA MEDICAL". The logo is in black and white, with the word "NEXERA" in large, bold letters. Above the X is a diamond shape. Below the ERA is the word "MEDICAL" in smaller letters.

JUL 5 5 2012

Section 5.0: 510(k) SUMMARY

510(k) Owner: NexEra Medical, Inc. 3343 West Commercial Blvd, Suite 103 Ft. Lauderdale, FL 33309

Contact:Paul Sallarulo, President CEO
Phone:954-495-2020, x 2031
Fax:954-491-7281

Establishment TBD Registration Number: Date Summarv July 2, 2012 Prepared: Device: Trade Name: SpectraShield model 9500 Surgical Mask Common /Classification Name: Surgical mask Classification Product Code: ONT Regulation Number: 21CFR 878.4040 Predicate K090414 SpectraShield 9500 Surgical N95 Respirator Device Information: The SpectraShield model 9500 Surgical Mask is a molded shape surgical mask composed Device of 4 layers of material, molded to form the mask. A 2-ply meitblown polypropylene Description: middle layer is sandwiched by inner and outer layers of 100% polyester nonwoven fabric. The inner and outside layers of polyester nonwoven fabric include fibers that have been embedded with an antibacterial agent to provide antibacterial performance. The mask has 2 latex-free non-allergenic elastic straps and an aluminum nose strip. Intended Use: The SpectraShield 9500 Surgical N95 Respirator is a single use, disposable surgical N95 respirator, tested for continuous use up to 8 hours, embedded with a zeolite carrier containing a silver-copper agent on the outer layer and is not an antimicrobial drug. SpectraShield 9500 kills 99.99% of test bacteria after one hour of contact with the surface of the respirator. In vitro (laboratory) tests have demonstrated 99.99% kill on the 3343 West Commercial Blvd., Suite 103 - Ft. Lauderdale, FL 33309 USA Telephone: 954-495-2020 ~ Fax: 954-495-7281 ~ www.nexeramed.com

Section 5.0: 510(K) Summary Page | 1

{1}------------------------------------------------

surface of the outer layer of the respirator when tested in vitro against single isolates of the following test bacteria: Streptococcus pyogenes, MRSA (Methicillin Resistant Staphylococcus aureus), and Haemophilus influenzae under tested contact conditions.

No clinical studies have been conducted comparing the ability of the untreated surgical N95 respirator and the SpectraShield model 9500 surgical N95 respirator to protect the wearer from infection and the antibacterial treatment cannot effect pathogens that are inhaled around the edges of the respirator.

The SpectraShield 9500 Surgical N95 respirator is a single use device intended for occupational use to protect against microorganisms, body fluids and particulate material.

New DevicePredicate Device
510(k) #To be determinedK090414
CompanyNexEra Medical, Inc.NexEra Medical, Inc.
Name/ModelSpectraShield 9500 Surgical N95 Respirator *(*with amended Intended use Statement)SpectraShield 9500 Surgical N95 Respirator
FabricsNonwoven polyester containing a silver-copper zeolite (antibacterial agent) and ameltblown polypropylene substrate.Nonwoven polyester containing a silver-copper zeolite (antibacterial agent) and ameltblown polypropylene substrate.
Nosepiece100% Aluminum100% Aluminum
Straps(2) Polyamide fiber and elastic straps, latex free(2) Polyamide fiber and elastic straps, latex free
Mask StyleMolded shapeMolded shape
Fluid ResistanceASTM F1862Pass: Fluid Resistant @ 160mm HgPass: Fluid Resistant @ 160mm Hg
New DevicePredicate Device
ParticulateFiltration EfficiencyASTM F2299Pass: 99.87% at 0.1 micronsPass: 99.87% at 0.1 microns
DifferentialPressureMil M36954CPass: 4.3mm H20/cm2Pass: 4.3mm H20/cm2
BacterialFiltrationEfficiencyASTM F2101Pass: 99.9%Pass: 99.9%
FlammabilityClass 16CFR 1610Class 1Class 1
Cytotoxicity10993-10Pass: USP reactivity score = < 2Pass: USP reactivity score = < 2
Primary skinirritationISO10993-10Pass: PSI Score = 0, Non-irritantPass: PSI Score = 0, Non-irritant
Repeated PatchDermalSensitizationISO 10993-10Pass: 0% incidence sensitization response"0" severity at each evaluated time point.Pass: 0% incidence sensitization response"0" severity at each evaluated time point.
Systemic ToxicityISO 10993-11Pass: No mortality or evidence of systemictoxicity from the extracts was observed.Pass: No mortality or evidence of systemictoxicity from the extracts was observed.
Physico-chemicalUSPPhysico-chemical Test-PlasticsPass: Test results met the USP limits.Pass: Test results met the USP limits.
Gas off TestingTotal antibacterial particles released fromthe device were verified to be within safeinhalation levels.Total antibacterial particles released fromthe device were verified to be within safeinhalation levels.
Leach off testingTotal leachable antibacterial particlesreleased from the device were verified to beTotal leachable antibacterial particlesreleased from the device were verified to be
New DevicePredicate Device
within safe orally ingestible levels.within safe orally ingestible levels.
BioEfficacy :To Inoculums measured, >106S.pyogenes: > 4.40log10reduction - 1 hourH.influenzae: > 6.20log10reduction - 1 hourMRSA: > 4.83log10reduction - 1 hourTo Inoculums measured, >106S.pyogenes: > 4.40log10reduction - 1 hourH.influenzae: > 6.20log10reduction - 1 hourMRSA: > 4.83log10reduction - 1 hour
BioEfficacy :after repeatedexposures toperspiration over12 hoursTo Inoculums measured, >106S.pyogenes: > 4.25log10reduction - 1 hourH.influenzae: > 4.18log10reduction - 1 hourMRSA: > 4.11log10reduction - 1 hourTo Inoculums measured, >106S.pyogenes: > 4.25log10reduction - 1 hourH.influenzae: > 4.18log10reduction - 1 hourMRSA: > 4.11log10reduction - 1 hour
Intended UseStatementThe SpectraShield 9500 Surgical N95Respirator is a single use, disposable surgicalN95 respirator, tested for continuous use upto 8 hours, embedded with a zeolite carriercontaining a silver-copper agent on the outerlayer and is not an antimicrobial drug.SpectraShield 9500 kills 99.99% of testbacteria after one hour of contact with thesurface of the respirator. In vitro (laboratory)tests have demonstrated 99.99% kill on thesurface of the outer layer of the respiratorwhen tested in vitro against single isolates ofthe following test bacteria: Streptococcuspyogenes, MRSA (Methicillin ResistantStaphylococcus aureus), and Haemophilusinfluenzae under tested contact conditions.No clinical studies have been conductedcomparing the ability of the untreatedsurgical N95 respirator and the SpectraShieldmodel 9500 surgical N95 respirator toprotect the wearer from infection and theantibacterial treatment cannot effectpathogens that are inhaled around the edgesof the respirator.The SpectraShield 9500 Surgical N95Respirator is a single use, disposable surgicalN95 respirator, embedded with a zeolitecarrier containing a silver-copper agent onthe outer layer and is not an antimicrobialdrug. SpectraShield 9500 kills 99.99% of testbacteria after one hour of contact with thesurface of the respirator. In vitro (laboratory)tests have demonstrated 99.99% kill on thesurface of the outer layer of the respiratorwhen tested in vitro against single isolates ofthe following test bacteria: Streptococcuspyogenes, MRSA (Methicillin ResistantStaphylococcus aureus), and Haemophilusinfluenzae under tested contact conditions.No clinical studies have been conductedcomparing the ability of the untreatedsurgical N95 respirator and the SpectraShieldmodel 9500 surgical N95 respirator toprotect the wearer from infection and theantibacterial treatment cannot effectpathogens that are inhaled around the edgesof the respirator.
New DevicePredicate Device
The SpectraShield 9500 Surgical N95respirator is a single use device intended foroccupational use to protect againstmicroorganisms, body fluids and particulatematerial.respirator is a single use device intended foroccupational use to protect againstmicroorganisms, body fluids and particulatematerial.

510(k) Summary Device Comparison Table

3343 West Commercial Blvd., Suite 103 - Ft. Lauderdale, FL 33309 USA Telephone: 954-495-2020 ~ Fax: 954-495-7281 ~ www.nexeramed.com

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for "NEXERA MEDICAL". The logo is in black and features a stylized "X" with a diamond shape above it. The word "MEDICAL" is written in smaller letters below the "ERA" portion of "NEXERA".

3343 West Commercial Blvd., Suite 103 – Ft. Lauderdale, FL 33309 USA Telephone: 954-495-2020 ~ Fax: 954-495-7281 ~ www.nexeramed.com

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "NEXERA MEDICAL". The logo is in black and white. The word "NEXERA" is in large, bold letters, with a diamond shape above the "X". Below "NEXERA" is the word "MEDICAL" in smaller letters.

3343 West Commercial Blvd., Suite 103 – Ft. Lauderdale, FL 33309 USA Telephone: 954-495-2020 ~ Fax: 954-495-7281 ~ www.nexeramed.com

{4}------------------------------------------------

The subject device (SpectraShield model 9500 Surgical mask with the revised IFU Conclusion: referencing "tested for continuous use up to 8 hours"), and the predicate device (K090414) are the same device. The intention of this 510k submittal is to change the IFU to include the statement "tested for continuous use up to 8 hours".

The predicate device (K090414) was tested for bio-efficacy after repeated exposures to perspiration over a 12 hour period (see K090414 Repeat Challenge Protocol and Testing). The repeat challenge testing required the predicate device be repeatedly exposed to perspiration over a 12 hour period. Following the 12 hour exposure the predicate device was tested and demonstrated 99.99% kill on the surface of the outer layer of the respirator when tested in vitro against Streptococcus pyogenes, MRSA (Methicillin Resistant Staphylococcus aureus), and Haemophilus influenzae under tested contact conditions. The intention of the repeated challenge and sustained exposure was to demonstrate that the device would still function as intended (99.99% kill) after wearing the device for 12 hours.

The IFU for the predicate device references "single use, disposable device". The proposed change to the IFU would read "single use, disposable device, tested for continuous use up to 8 hours."

It is our conclusion that the proposed change to the IFU does not change the intended use of the device, and we believe the change to the IFU further clarifies the intended use of the device. Additionally, we note the proposed change to the IFU demonstrate the device is as safe and as effective as the predicate device and performs equally as well.

3343 West Commercial Blvd., Suite 103 - Ft. Lauderdale, FL 33309 USA Telephone: 954-495-2020 ~ Fax: 954-495-7281 ~ www.nexeramed.com

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, which is a common symbol of the United States. The logo is black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Paul Sallarulo Nexera Medical Incorporated 3343 West Commercial Boulevard Suite 103 Fort Lauderdale, Florida 33309

5 2012 JUL

Re: K120244 Trade/Device Name: SpectraShield Model 9500 Surgical Respirator Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: ONT Dated: June 21, 2012 Received: June 25, 2012

Dear Mr. Sallarulo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading:

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2- Mr. Sallarulo

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

For

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use Statement Section 4.0:

510(k) Number: 510(k) submission K12 0244

SpectraShield model 9500 Surgical Respirator Device Name:

Indications for Use:

The SpectraShield 9500 Surgical N95 Respirator is a single use, disposable, surgical N95 respirator, tested for continuous use up to 8 hours, embedded with a zeolite carrier containing a silver-copper agent on the outer layer and is not an antimicrobial drug. SpectraShield 9500 kills 99.99% of test bacteria after one hour of contact with the surface of the respirator. In vitro (laboratory) tests have demonstrated 99.99% kill on the surface of the outer layer of the respirator when tested in vitro against single isolates of the following test bacteria: Streptococcus pyogenes, MRSA (Methicillin Resistant Staphylococcus aureus), and Haemophilus influenzae, under tested contact conditions.

No clinical studies have been conducted comparing the ability of an untreated surgical N95 respirator and the SpectraShield model 9500 surgical N95 respirator to protect the wearer from infection, and the antibacterial treatment cannot effect pathogens that are inhaled around the edges of the respirator.

The SpectraShield 9500 Surgical N95 respirator is a single use device, intended for occupational use to protect against microorganisms, body fluids, and particulate material.

Prescription Use

AND/OR

Over -the-counter Use X (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)

Nalip. hanyhim. 1

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.