K Number
K120192
Date Cleared
2012-10-16

(267 days)

Product Code
Regulation Number
878.4040
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Disposable Surgical Gowns, which are blue colored and EO sterilized, are indicated to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. It could achieve Level 3 Barrier Performance as per AAMI PB70:2003. It is for single use only.

Device Description

Disposable Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Disposable Surgical Gowns are manufactured from three layers nonwoven fabric that is comprised of Spunbond (SMS) materials. Disposable Surgical Gowns are reinforced with an additional layer of SPP (Syndiotactic Polyprogylene) laminating film in the sleeve and body areas for higher barrier protection. The proposed Disposable Surgical Gowns include four sizes, which are presented in Table V-1 General Description.

AI/ML Overview

The provided text describes a 510(k) summary for disposable surgical gowns, focusing on their physical and biological properties. It does not contain any information about an AI-powered device, a test set, expert readers, or ground truth establishment in the context of medical imaging or diagnostic algorithms.

Therefore, I cannot fulfill your request to describe acceptance criteria and study proving device meets acceptance criteria for an AI device based on the provided text.

The document discusses:

  • Device: Disposable Surgical Gowns
  • Intended Use: Protection during surgical procedures.
  • Acceptance Criteria (for this product): Primarily physical barrier performance (e.g., Level 3 per AAMI PB70), tear strength, fire protection, and biocompatibility, along with manufacturing standards. These are verified through bench tests against established standards (e.g., ASTM, AAMI, ISO).
  • Study Proving Acceptance: "Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards..."

The text provided does not contain the information requested for an AI-powered device, such as:

  1. A table of acceptance criteria and reported device performance (for an AI).
  2. Sample size for a test set or data provenance for an AI.
  3. Number of experts or their qualifications for ground truth establishment for an AI.
  4. Adjudication method for a test set for an AI.
  5. MRMC comparative effectiveness study or effect size for AI assistance.
  6. Standalone performance of an algorithm.
  7. Type of ground truth (expert consensus, pathology, outcomes data) for an AI.
  8. Sample size for a training set for an AI.
  9. How ground truth for the training set was established for an AI.

The information is related to the regulatory approval process for a physical medical device (surgical gowns), not an AI/software as a medical device.

{0}------------------------------------------------

OCT
16
2012

Exhibit #2 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number: K120192

Date of Submission: September 25, 2012 1.

. मं Sponsor

Weihai Hongyu Nonwoven Fabric Products Co., Ltd. 9-1 Pudong Road, Area of Economy and Technique, Weihai. China, 264205 Establishment Registration Number: 3008489813 Contact Person: Yanfei Liu Position: Registration Manager Tel: + (86)-631-3636943 Fax: + (86)-631-3636910 Email: liuyanfei340@163.com

  1. Submission Correspondent

Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net

  1. Proposed Device Identification

Proposed Device Name: Disposable Surgical Gowns Proposed Device Model: M, L, XL, XXL Classification: II Product Code: FYA Regulation Number: 21 CFR 878.4040 Review Panel: General& Plastic Surgery

{1}------------------------------------------------

Intended Use Statement:

Disposable Surgical Gowns, which are blue colored and EO sterilized, are indicated to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. It could achieve Level 3 Barrier Performance as per AAMI PB70:2003. It is for single use only.

  • Predicate Device Identification 5.
    Predicate device for Disposable Surgical Gowns 510(k) Number: K100971 Product Name: Hangzhou ATek Medical and Textile Surgical Gowns Manufacturer: Hangzhou ATek Medical and Textile Co. Ltd.

6. Device Description

Disposable Surgical Gowns are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Disposable Surgical Gowns are manufactured from three layers nonwoven fabric that is comprised of Spunbond (SMS) materials. Disposable Surgical Gowns are reinforced with an additional layer of SPP (Syndiotactic Polyprogylene) laminating film in the sleeve and body areas for higher barrier protection.

The proposed Disposable Surgical Gowns include four sizes, which are presented in Table V-1 General Description

Table v-1 General Description of Proposed Device
Proposed DeviceMaterialSizeColorStyleSterility
DisposableSurgical GownsSMSMBlueReinforcedEO Sterilized
LBlueReinforcedEO Sterilized
XLBlueReinforcedEO Sterilized
XXLBlueReinforcedEO Sterilized

1 Description of Proposed Device

Non-Clinical Test Conclusion 7.

Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

{2}------------------------------------------------

  • ASTM F2407-06 Standard Specification for Surgical Gowns Intended for Use in Healthcare > Facilities
  • AAMI PB70-2003 Liquid barrier performance and classification of protective apparel and drapes ﺮ intended for use in health care facilities
  • AATCC 42-2000 Water Resistance: Impact Penetration Test; >
  • AATCC 127-2003 Water Resistance: Hydrostatic Pressure Test; V
  • 16 CFR Part 1610:2008 Standard for the Flammability of Clothing Textiles ア
  • ASTM D 5034 09 Standard Test Method for Breaking Strength and Elongation of Textile Fabrics ﺮ (Grab Test)
  • ASTM D 5733 99 Standard Test Method for Tearing Strength of Nonwoven Fabrics by the ア Trapezoid Procedure
  • v. A STM D 1683 - 07 Standard Test Method for Failure in Sewn Seams of Woven Apparel Fabrics
  • ISO 9073-10:2003 Textiles Test Methods for Nonwovens Part 10: Lint and Other Particles ﺮ Generation in the Dry State
  • ASTM F 1868 02 Standard Test Method for Thermal and Evaporative Resistance of Clothing ア Materials Using a Sweating Hot Plate - Part B - Isothermal Evaporative Resistance
  • ISO 10993-5:2009: Biological Evaluation of Medical Devices Part 5: Tests for in vitro ﺮ Cytotoxicity
  • ISO 10993-10:2010 Biological Evaluation of Medical Devices Part 10: Tests for irritation and skin ア sensitization
  • ISO 11135-1:2007 Sterilization of health care products- Ethylene oxide- Part 1: Requirements for ﺮ development, validation and routing control of a sterilization process for medical devices
ItemProposed DevicePredicate Device
Product CodeFYAFYA
Regulation No.21 CFR 878.404021 CFR 878.4040
ClassIIII
Intended UseDisposable Surgical Gowns are intended tobe worn by operating room personnel duringsurgical procedures to protect both thesurgical patient and the operating roompersonnel from transfer of microorganisms,body fluids, and particulate material.Disposable gowns are worn by operating roompersonnel during surgical procedures to protectboth the surgical patient and OR personnelfrom the transfer of body fluids and particulatematerial.
StyleReinforcedReinforced and Non-reinforced
SizeM, L, XL, XXLL, XL, XXL
ColorBlueUnknown
Weight per square45 g/m²Unknown
Thickness0.16mmUnknown
Physical SpecificationsResistance to Blood and Liquid PenetrationLevel 3 per AAMI PB 70Level 3 per AAMI PB 70
Resistance to Tear≥20 N≥20 N
Fire ProtectionClass IClass I
Features for Safe UseFree of toxic ingredientsthe average value of coefficient of variationof linting is 6.2%Free of Nonfast dyesFree of toxic ingredientsUnknownFree of Nonfast dyes
Mechanical specificationsBurst strength≥40 Kpa≥40 Kpa
Tensile strength≥20 N≥20 N
DurabilityDisposableDisposable
Performance TestingConform toAATCC 42-2000AATCC 127-200316 CFR Part 1610:2008ASTM D5034 - 09ASTM D5733 - 99ASTM D1683 - 07ISO 9073 Part 10:2003ASTM F1868 - 02Conform toAATCC 42-2000AATCC 127-200316 CFR Part 1610:2008ASTM D5034 - 09ASTM D5733 - 99ASTM D1683 - 07ISO 9073 Part 10:2003ASTM F1868 - 02
MaterialSMS, SPP, Polypropylene, PE, Nylon, PolyesterSMS, Hydro-Entangled (Spunlace), SFT
BiocompatibilityConform to the requirement of ISO 10993 series StandardsConform to the requirement of ISO 10993 series Standards
Cytotoxicity
IrritationNo intracutaneous reactivityNo intracutaneous reactivity
SensitizationNo delayed dermal contact sensitizationNo delayed dermal contact sensitization
SterilizationSAL106106
MethodEO SterilizationEO Sterilization
ValidationConforms to ISO 11135Conforms to ISO 11135
Package IntegrityConforms to ISO 11607Conforms to ISO 11607
EO ResidualConforms to ISO 10993-7Conforms to ISO 10993-7
Label and LabelingConforms to FDA RequirementsConforms to FDA Requirements

8. Summary of Comparison

E2-3

{3}------------------------------------------------

Exhibit #2 510(k) Summary

·

.

·

. .

,

·

.

·

:

.

:

E2-4

.


:

{4}------------------------------------------------

1:2-5

9. Substantially Equivalent Conclusion

·

The proposed devices, Disposable Surgical Gowns, are determined to be Substantially Equivalent (SE) to the predicate device, Hangzhou ATek Medical and Textile Surgical Gowns (K100971), in respect of safety and effectiveness. 1 ・

י

י

:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

OCT 1 6 2012

Weihai Hongyu Nonwoven Fabric Products Company, Limited C/O Ms. Diana Hong General Manager Mid-Link Consulting Company, Limited P.O. Box 237-023 Shanghai, 200237 China

Re: K120192

Trade/Device Name: Disposable Surgical Gowns Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FYA Dated: September 25, 2012 Received: October 1, 2012

· Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2- Ms. Hong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Antion V. Van

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number: Device Name: Disposable Surgical Gowns Proposed device model: M, L, XL and XXL

Indications for Use:

Disposable Surgical Gowns, which are blue colored and EO sterilized, are indicated to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. It could achieve Level 3 Barrier Performance as per AAMI PB70:2003. It is for single use only.

ZPRESCRIPTION USE (Part 21 CFR 801 Subpart D)

OVER-THE-COUNTER USE (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F. Clamie. William

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

K120192 510(k) Number:

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.